Transportation of magnetic beads between different reagents plays a crucial role in many biological assays e.g. for purification of biomolecules or cells where the beads act as a mobile solid support. Therefore, usually a complex set-up either for fluidic processing or for manipulation of magnetic beads is required. To circumvent these drawbacks, we present a facile and automated method for the transportation of magnetic beads between multiple microfluidic chambers on a centrifugal microfluidic cartridge “LabDisk”. The method excels by requiring only one stack of stationary permanent magnets, a specific microfluidic layout without actively controlled valves and a predefined frequency protocol for rotation of the LabDisk. Magnetic beads were transported through three fluidically separated chambers with a yield of 82.6% ± 3.6%. Bead based DNA purification from a dilution series of a Listeria innocua lysate and from a lambda phage DNA standard was demonstrated where the three chambers were used for binding, washing and elution of DNA. Recovery of L. innocua DNA was up to 68% ± 24% and for lambda phage DNA 43% ± 10% compared to manual reference purification in test tubes. Complete purification was conducted automatically within 12.5 min. Since all reagents can be preloaded onto the LabDisk prior to purification, no further hands-on steps are required during processing. Due to its modular and generic character, the presented method could also be adapted to other magnetic bead based assays e.g. to immunoassays or protein affinity purification, solely requiring the adjustment of number and volumes of the fluidic chambers.
In general, methods for handling magnetic beads can be divided into three groups: (1) Beads are (temporarily) immobilized by a magnetic field and sequentially flushed by different reagents.9,13 However, for resuspension and efficient washing of the beads, the magnetic field needs to be removed or switched off and the reagent flow needs to be stopped in order not to drag away any beads. (2) Beads are actively transported between different reagents e.g. by actuated permanent magnets8,14–16 or coil arrays17,18 requiring rather complex and costly set-ups for magnet actuation or fabrication of the coil arrays. (3) Beads are attracted over interfaces of adjacent laminar reagent streams by magnetophoresis19 using permanent magnets3,7,12 thereby facilitating continuous processing. However, successful bead transport is highly sensitive to the interplay of magnetic field strength, orientation and the flow rate of the reagents.
Although automated manipulation of magnetic beads is state of the art in pipetting workstations and various microfluidic platforms3,7,11–13,20,21 smart solutions for bead handling in centrifugal microfluidic platforms are very limited and need to be further explored, especially because centrifugal microfluidic platforms22–25 are expected to have a crucial impact on future automated point-of-care testing.20,26–29
So far, only one centrifugal microfluidic system has been published, which exploited the handling of magnetic beads and sequentially exposed them to different reagents.30 Here, a primary magnet mounted on a linear geared stepping motor underneath the rotating cartridge was used to control the position of a secondary magnet which was integrated in the bottom layer of a double layered cartridge. Depending on the position of the primary magnet, spinning frequency and direction of rotation, the secondary magnet could be moved to four different positions in the bottom layer and thereby manipulate magnetic beads within the microfluidic structure that was integrated into the top layer of the cartridge. During processing, the beads were captured by the magnet in a first chamber and mixed with the sample for immunomagnetic capturing of pathogens. The beads were then flushed with the washing buffer to a second position where they were captured for washing. The washing buffer was transferred to a downstream waste reservoir while the beads remained fixed for subsequent thermal pathogen lysis. The purified cell lysate containing the released DNA could then be used for further analysis. The required fluidic control of the buffer streams was achieved by a complex actuation scheme involving a series of integrated ferrowax valves (LIFM) that could be opened or closed by infra-red laser irradiation.
In this work, we aim at an improved and generic bead handling method on a centrifugal microfluidic LabDisk that excels by requiring solely a stationary permanent magnet, a spinning drive, and a single layered microfluidic LabDisk. The need for costly magnet actuation systems is circumvented by defined positioning of the LabDisk with respect to a stationary permanent magnet. The presented method of bead handling in combination with the modular fluidic layout allows stand-alone processing since all reagents can be loaded onto the LabDisk simultaneously prior to starting the automated frequency protocol. The novel method is demonstrated by magnetic bead based DNA purification, one of the most prevalent applications involving transportation of magnetic beads. Aiming at significant reduction of complexity and costs, the method is meant to facilitate entry of centrifugal microfluidic platforms into the point-of-care market.
Fig. 1 Schematic microfluidic structure for characterisation of GTM. (a) Microfluidic structure with magnetic bead pellet in chamber 1 and reagents in inlet chambers i1–i5 (b) After starting the centrifugal frequency protocol, reagents are transferred from inlet chambers into the associated chambers 1–3. The bead pellet is rehydrated and resuspended. (c) Beads are centrifuged to the rim of chamber 1 by spinning the LabDisk. (d) The LabDisk is stopped at a defined azimuthal position of the liquid phase in chamber 1 (blue) with respect to the external magnet. The magnet attracts the beads, which form clusters and are moved across the liquid-gas interface into the air-gap that connects chamber 1 and chamber 2. (e) The magnet holds the beads fixed while the LabDisk is rotated in 0.5° increments. Thereby, the beads are transported to an azimuthal position radially inwards of chamber 2 (dashed arrow). (f) The LabDisk is reaccelerated and the beads are centrifuged into the liquid phase in chamber 2 (red). (C)–(F) show photographs of the bead transport. Steps (c)–(f) are repeated a second time to transport the beads from chamber 2 into chamber 3 that is connected to a chamber representing downstream fluidics for further assay steps. Grey shaded area represents a hydrophobic Teflon coating (see section: fabrication and preparation of the LabDisk). |
The only required components to perform GTM are a LabDisk with one single microfluidic layer comprising the bead-transfer structure and a spinning device equipped with a permanent magnet fixed above the disk. Due to the modular layout of the GTM unit operation, the number of bead-transfer steps can be adjusted to the number of liquids that the beads have to transit. Additionally, the single volume of the chambers can be adjusted to the respective volumes of the liquids. For characterisation of GTM, a LabDisk with a microfluidic structure for two serial bead-transfers was designed and fabricated (Fig. 1). The structure consists of chamber 1 (volume 250 μL), chamber 2 (vol. 130 μL) and chamber 3 (vol. 50 μL), separated by two air-gaps between adjacent chambers. Furthermore, chambers 1–3 are connected to upstream inlet chambers i1–i5. Chamber 1 is connected to three inlet chambers to allow preloading of several liquids that are merged after a first centrifugation step. Magnetic beads are prestored in chamber 1 (described in chapter 1 ESI†).
Initially, the LabDisk is fixed to the rotor of the processing device and a stack of three neodymium iron boron (NdFeB) magnets (#W-07-N and #S 03 01-N, Supermagnete.de, Germany) is positioned above the microfluidic chambers (Fig. S1, ESI†). The magnet stack has a radial position of r = 36 mm, a radial distance of i = 2.5 mm to the liquid–gas interface and was positioned with a distance of d = 0.5 mm above the microfluidic chambers (Fig. 2a–b). The position of the magnet stack remains constant during the entire operation. To initiate bead transfer by GTM, the LabDisk is stopped at a defined azimuthal position of chamber 1 with respect to the magnet. The magnet attracts the beads which form clusters and are moved across the liquid–gas interface into the air-gap that connects chamber 1 and chamber 2. The LabDisk is then rotated in increments of 0.5° while the beads are held by the magnet. Thereby, the beads are transported to the azimuthal position of chamber 2. Subsequent acceleration centrifuges the beads into the liquid phase of chamber 2. The complete sequence of GTM is depicted in Fig. 1 and in a movie file provided in the ESI.† After incubation of the beads in the liquid of processing chamber 2, GTM is conducted a second time to transfer the beads into the liquid phase of chamber 3.
Fig. 2 Forces acting on the magnetic beads. (a) Scheme of microfluidic structure with liquid phase in chamber 1 (blue) and a second liquid phase in chamber 2 (red). The cluster of magnetic beads is attracted in the x direction. Cross-section A–A is magnified in (b) showing the opposing forces acting on the beads. The magnetic field was calculated using COMSOL. (c) Magnetic force in the x direction (calculated using eqn (3) and values from Table 1) versus opposing surface force (calculated with eqn (4) and values from Table 1). For radial distances between liquid–gas interface and magnet ≲5 mm, the magnetic force overcomes the surface force and the beads are attracted over the interface. For the experiments, the distance was set to i = 2.5 mm. |
The prerequisite for the beads to cross the liquid–gas interface is that the magnetic force Fmag overcomes the counteracting force of the surface tension Fsurface (eqn (1)):
Fmag > Fsurface | (1) |
The magnetic force can be calculated from the volume of the bead cluster Vcluster, the difference in magnetic susceptibilities of the magnetic beads χmag and surrounding liquid χH2O, the strength and gradient of the magnetic field (grad(B))·B and the magnetic vacuum permeability μ0 (eqn (2)).4
(2) |
For the calculations, only the x-component of the magnetic force is considered (Fig. 2a–b) (3).1
(3) |
The counteracting force from the surface tension is given in (eqn (4)).
Fsurface = 61/3 · π2/3 · σliquid · V1/3cluster | (4) |
a Measured on analytical balance (Ohaus Adventurer ARRV70, China). b Measured with vibrating sample magnetometer. Kindly provided by Dr A. Tschöpe (Universität des Saarlandes, Germany). c Locally dependent magnetic field components in the x, y and z direction were measured by a Koshava 4 teslameter (Wuntronic, Germany). The teslameter was mounted to an automatic positioning system. Measurements were conducted in 0.3 mm increments. From adjacent points of measurement, the local derivatives of the magnetic field components in the x, y and z direction were calculated. | ||||
---|---|---|---|---|
Volume of bead cluster | V cluster | ∼4 | mm³ | Measureda |
Volume susceptibility of mag. beads | χ mag | 0.29 | Measuredb | |
Magnetic field | B | B x, By, Bz | T | Measuredc |
Magnetic vacuum permeability | μ 0 | 1.25 × 10−6 | N A-2 | Constant |
Surface tension of water | σ H2O | 72.5 | mN m−1 | Matter constant |
For the experiments, a mixture of 200 μL of a Listeria innocua lysate and 5 μL of a DNA standard was used as sample. The sample was split into two aliquots of 102.5 μL that were then loaded into inlets i1 and i3 respectively. Furthermore, 50% (v/v) isopropanol, washing buffer (buffer AW2, Qiagen, Germany) and elution buffer (PCR grade water) were loaded into inlets i2, i4 and i5 as described in Table 2.
Inlet | Reagent | Volume [μL] |
---|---|---|
i1 | Sample lysate | 102.5 |
i2 | 50% (v/v) isopropanol | 50 |
i3 | Sample lysate | 102.5 |
i4 | Washing buffer AW2 | 130 |
i5 | PCR grade water | 50 |
The magnetic beads were prestored in chamber 1 as dried pellet. Finally, the inlet holes were sealed with adhesive tape and stand-alone DNA purification was started by the frequency protocol described in Table 3. The protocol automatically performs release of reagents into processing chambers 1–3 used for binding, washing and elution. Afterwards the eluted DNA was removed from the LabDisk by pipetting and stored for further analysis.
Frequency [rpm] | Duration [ms] | Notes | Action |
---|---|---|---|
600 | 15000 | All reagents are transferred from inlet chambers to binding, washing or elution chamber | |
480 | 2000 | Pellet resuspension | |
120 | 10000 | Alternation between 120 and 480 rpm repeated 30 times | Beads are magnetically attracted radially inwards at 120 rpm while centrifuged radially outwards at 480 rpm. DNA binds to the surface of the beads. |
480 | 10000 | ||
0 | 20000 | Defined positioning of LabDisk | Beads are magnetically attracted out of the lysate. |
0 | 1000 | Incremental rotation of 0.5°, repeated 30 times | Beads are moved from binding chamber to washing chamber through air-gap |
480 | 15000 | Beads are centrifuged into washing buffer | |
0 | 20000 | Defined positioning of LabDisk | Beads are magnetically attracted out of washing buffer. |
0 | 1000 | Incremental rotation of 0.5°, repeated 30 times | Beads are moved from washing chamber to elution chamber through air-gap |
480 | 15000 | Beads are centrifuged into elution buffer |
Fig. 3 LabDisk, replicated by the μTSL process featuring two independent GTM structures, each with three buffer chambers designed for a bind-wash-elute protocol used in magnetic bead based DNA purification. |
Pellet # | Before transport [mg] | After transport [mg] | Yield % |
---|---|---|---|
1 | 1.94 | 1.52 | 78.4 |
2 | 1.97 | 1.63 | 82.7 |
3 | 1.99 | 1.64 | 82.4 |
4 | 2.09 | 1.70 | 81.3 |
5 | 2.05 | 1.81 | 88.3 |
Mean yield | 82.6 | ||
Std. dev. | 3.6 |
Due to the small number of DNA copies, real-time PCR results at these dilutions are dominated by statistical effects. Therefore, PROBIT regression analysis using the statistical software SPSS was applied to estimate the concentration of recovered DNA of the two most diluted samples.38,39 By using the PROBIT method, the DNA concentration is determined by the ratio of positive PCR reactions to the overall number of PCR reactions performed per dilution. PROBIT analysis is described in detail in the ESI† (Fig. S2, Table S2). The LabDisk based purification of L. innocua DNA yielded between 29 ± 25% and 68 ± 24% compared to the manual reference purification in tubes (Fig. 4, Table 5). From the total number of 18 LabDisk and 18 reference purifications in tubes (6 dilutions, 3 replicates per dilution) integrated lambda phage DNA standard was measured by real time PCR (Table S3, ESI†). Compared to the theoretical maximum of 17500 inserted lambda phage DNA copies, 31 ± 10% could be recovered after the LabDisk based purification and 72 ± 21% by the manual reference in tube. Thus, the yield of recovered lambda phage DNA on the LabDisk versus manual reference is calculated to be 43 ± 10% (Table 6).
Fig. 4 Number of recovered L. innocua copies obtained by LabDisk based (diamond) or reference (triangle) purification from lysed L. innocua cultures. The investigated concentration range comprised six orders of magnitude. Quantification of DNA yield was performed by quantitative real time PCR. For concentrations ≤ 104 “lysed CFU” per sample, PROBIT regression analysis was used for calculation of recovery. |
Inserted number of lysed L. innocua | LabDisk recovered number of DNA copies | Reference recovered number of DNA copies | Ratio LabDisk/Reference (Errors calc. by Gaussian error propagation) |
---|---|---|---|
a According to PROBIT analysis, more DNA seems to be recovered at the highest dilution that was actually inserted into the reaction. This might be attributed firstly to low sample size for statistical PROBIT analysis and secondly to the comparison to plate counting which might be defective. | |||
(8.5 ± 2.5) × 107 | (1.7 ± 0.6) × 107 | (2.5 ± 0.1) × 107 | 68% ± 24% |
(8.5 ± 2.5) × 106 | (1.8 ± 0.3) × 106 | (4.4 ± 0.6) × 106 | 41% ± 9% |
(8.5 ± 2.5) × 105 | (2.2 ± 0.2) × 105 | (3.4 ± 0.9) × 105 | 65% ± 18% |
(8.5 ± 2.5) × 104 | (2.8 ± 0.1) × 104 | (7.5 ± 1.9) × 104 | 37% ± 10% |
(8.5 ± 2.5) × 103 | (2.5 ± 1.4) × 103 | (4.6 ± 2.8) × 103 | 54% ± 45% |
(8.5 ± 2.5) × 102 | (1.1 ± 0.6) × 103a | (3.7 ± 2.3) × 103a | 29% ± 25% |
Inserted number of lambda phage DNA copies | LabDisk recovered number of DNA copies | Reference recovered number of DNA copies | Ratio LabDisk/Reference (%) | Ratio on LabDisk/theor. maximum (%) |
---|---|---|---|---|
17.5 × 103 | (5.5 ± 1.7) × 103 | (12.6 ± 3.7) × 103 | 43 ± 10 | 31 ± 10 |
In summary, DNA recovery of LabDisk based DNA purification versus manual reference was 43 ± 10% for lambda phage DNA and between 29 ± 25% and 68 ± 24% for a log10 dilution series of L. innocua lysate covering a range of six orders of magnitude in DNA concentration. Hence, the recovery should be further optimized for applications that require high sensitivity; for genotyping applications the recovery achieved would be already sufficient. The first and very obvious reason for the reduced DNA recovery of the LabDisk based purification is the loss of almost 20% of magnetic beads during the experiments reported, directly explaining 20% loss in DNA recovery. One second reason for the reduced DNA recovery of the LabDisk based purification might be attributed to a not fully optimized rehydration and resuspension of the dried magnetic bead pellet leading to a reduced silica surface available for DNA binding. Although in tube experiments, similar results for DNA purification using dry magnetic bead pellets or magnetic beads in suspension were obtained (chapter 6 in ESI; Fig. S3†), it has to be considered that for those experiments vortexing was used to resuspend the beads, which very likely might have been more efficient than resuspension of the pellet in the LabDisk. On the LabDisk, mixing was mainly conducted by cyclic variation of rotational speed resulting in a movement of the magnetic beads between a radial inward and a radial outward position (i.e. as a result of dominating magnetic or centrifugal forces). Insufficient resuspension results in less available bead surface for DNA binding. Hence, one way to improve resuspension of magnetic beads might be a more efficient mixing.
A more straightforward approach would replace the bead pelleting agent trehalose which was used in the experiments reported by an agent that better supports rehydration and resuspension of the dried magnetic bead pellet. For example polyvinylpyrrolidone has successfully been used for pelleting and rehydration of beads.40
This novel method excels by several features: First, the technical requirements are low: only one fluidic layer and no actively controlled valves integrated into the LabDisk facilitate production. Furthermore, the only requirement for the LabDisk spinning device is to have a permanent magnet installed at a fixed position and to allow azimuthal positioning of the disk. Second, the design of the microfluidic structure for GTM is modular and allows easy adjustment to the requirements of the assay, just by varying number and volume of the buffer chambers. Remaining shortcomings at the time of reporting comprise a limited bead transport efficiency of 82.6 ± 3.6% after two transportation steps.
As an application example for this novel method, DNA purification with commercially available DNA extraction reagents was demonstrated and extensively characterized. Yields of up to 68% compared to manual reference purifications were achieved. The yield may be further enhanced by improving resuspension of the prestored, pelleted beads or using different pelleting agents like water soluble polymers. The process of DNA purification was automated by a predefined frequency protocol and didn't require any human interaction during processing. In the future, lysis will also be included. For this operation mode, instead of the lysate, the original sample (bacteria stock, whole blood, or blood plasma) could be loaded into inlet i2 while the combined lysis and binding buffer could be loaded into inlet i3. The DNA extraction structure is composed in such a way that it may be concatenated with structures for multiplex real-time PCR41 or RPA42 analysis paving the way for automatic nucleic acid sample-to-answer analytics on a disposable foil LabDisk. In future, reagent prestorage in glass ampoules43 or stick-packs44 may be incorporated further enhancing the degree of integration.
Additional applications using magnetic bead based immunoassays or protein affinity purification may also be conducted with the presented method by simply adjusting number and volume of the isoradially arranged chambers.