Ayse Rezzan
Kose
a and
Hur
Koser
*b
aSchool of Engineering and Applied Science, Yale University, 15 Prospect St., Becton Center, Room 608, New Haven, CT 06520-8267, USA. E-mail: ayse.kose@yale.edu; Fax: +1 (203) 432-6420; Tel: +1 (203) 432-8573
bSchool of Engineering and Applied Science, Yale University, 15 Prospect St., Becton Center, Room 507, New Haven, CT 06520-8267, USA. E-mail: hur.koser@yale.edu; Fax: +1 (203) 432-6420; Tel: +1 (203) 432-9629
First published on 11th November 2011
We present a low-cost, flow-through nanocytometer that utilizes a colloidal suspension of non-functionalized magnetic nanoparticles for label-free manipulation and separation of microparticles. Our size-based separation is mediated by angular momentum transfer from magnetically excited ferrofluid particles to microparticles. The nanocytometer is capable of rapidly sorting and focusing two or more species, with up to 99% separation efficiency and a throughput of 3 × 104 particles/s per mm2 of channel cross-section. The device is readily scalable and applicable to live cell sorting with biocompatible ferrofluids, offering competitive cytometer performance in a simple and inexpensive package.
The pressing need for the development of alternative separation techniques has led to a proliferation of new platforms.12–21 These new separation schemes generally fall into two broad categories: techniques that utilize alternative labels, including immunomagnetic separation,14 and label-free techniques such as dielectrophoresis (DEP) and hydrodynamic separation that rely on physical differences between cells and particles (e.g., size, density, shape, morphology or dielectric properties). Some of the currently available cell separation technologies and their performance metrics are summarized in Table 1. Fluorescence activated cellular sorting (FACS) is one of the most widely employed systems due to its high accuracy and versatility.12,22–28 However, the combination of precision flow systems with delicate laser optics makes FACS a complex, expensive technology to acquire (∼$250,000). Additionally, the need to rely on trained personnel to run assays on FACS renders it expensive to use and maintain, as well. Immunomagnetic separation, on the other hand, allows for the use of cheap permanent magnets to separate target cellsviaantibody-functionalized magnetic microbead labels.14,16 This approach offers high throughput and reduced cost, but still relies on manual labor steps.
Separation Mode | Purity(%) | Throughtput | Label | Cost ($) setup/consumables | Typical assay run-time | |
---|---|---|---|---|---|---|
a Reports from company directly. | ||||||
FACS 11,12 | Flow/electrostatic actuation | >88 | ∼104cells s−1 | Yes,fluo | >250,000/1000a | ∼1 h |
Immunomagnetic, batch14,15 | Batch/magnetic actuation of label | >70 | 106cells s−1 | Yes,magnetic | 390/320a | ∼1 h |
Immunomagnetic, flow14,15 | Flow/magnetic actuation of label | >99 | 103cells s−1 | Yes,magnetic | ∼1000/300 (estimated) | ∼1 h |
Hydrodynamic17 | Flow/hydrodynamic forces | 99 | ∼3 cells s−1 | No | ∼100/100 (estimated) | ∼1 h (for 1μl blood) |
Dielectrophoresis 19 | Batch or flow/electrodynamic actuation | 95 | ∼3 × 103cells s−1 | No | ∼1000/1 (estimated) | 10 min |
Ferro-microfluidic21 | Batch or flow; ferrohydrodynamic actuation | 98 | ∼3 × 104 particles/s mm−2 | No | ∼100/1 | ∼1 min |
Advances in micro- and nano-fabrication technologies have recently enabled the realization of miniaturized sorters that achieve particle manipulation with reduced manual labor using fluidic, electric, and magnetic forces.29–35 Replacing conventional cell sorting devices with micro systems also increases analytic sensitivity,36 though typically at the expense of throughput. For instance, Davis et al. employ a matrix of silicon microposts to hydrodynamically fractionate blood cells.17 While highly accurate in the analysis of small samples (∼ 1 μl), the device requires high-resolution microfabrication and is prone to clogging. DEP is a widely explored, alternative technique that exploits electric fields for cellular manipulation and sorting; the downside of this technique is the sensitivity of the system to flow and the buffer's electrical properties.19,37 Moreover, the use of large electric fields and their gradients in DEP may polarize cells and potentially damage or otherwise interfere with cellular functions.
Magnetophoresis is an alternative to navigate around the challenges of using high electric fields within conductive buffer solutions. Magnetic fields are typically not hazardous to cells and are not significantly affected by the ionic content or the pH levels of the solution. They can be applied externally, eliminating the need for physical contact between the field source and the analyte.38,39 Besides their common use in immunomagnetic separation, magnetic fields can manipulate cells and microparticles via label-free, negative magnetophoresis, as well.21,40 In this approach, magnetic nanoparticles within a ferrofluid medium are attracted towards higher magnetic field regions, thereby effectively displacing and guiding non-magnetic objects away from field sources.41 Displaced objects experience magnetic repulsion forces in proportion to their volumes, making size-based separation possible in flow-through devices. The source of magnetic fields in such separators has traditionally been permanent magnets.34,35 To obtain high separation accuracy, these platforms require a very tight focusing of the input sample stream, resulting in low throughput.
In a recent study, we reported a new micro-separation and sorting platform that utilizes a low-cost electrode substrate to create locally programmable magnetic fields within a nano-colloidal suspension medium.21 We demonstrated the proof of principle for the sorting mechanism in a closed microfluidic channel with no external flow, and elucidated the physical mechanism of cell manipulation based on size, shape and elasticity.
In this paper, we demonstrate a flow-through, ferrofluid-mediated nano-cytometer that continuously separates, sorts and focuses micro-particles based on size. Our technique utilizes inexpensive materials and achieves label-free and labor-free separation. The nanocytometer device delivers over 98% sample purity and high throughput per unit cross-section (>3 × 104 particles/s per mm2), rendering it a practical candidate for clinical cytometry.
At this early stage, we used microspheres to demonstrate the capability of our technology and study the separation and sorting mechanism in the presence of flow. We elucidate in detail the flow effects, such as the flow-induced hydrodynamic shear on the effectiveness of the ferrofluid-based separation. We believe that once these issues are understood and devices are characterized, separating and sorting live cells is possible, as we demonstrated in a prior publication using our initial prototype platform with no flow.21 Our technology is still clinically relevant even if the separation is confined to microspheres only, as microspheres of different sizes can be functionalized (i.e., labeled) to separate and sort protein, DNA, virus, bacterial or cellular targets from complex mixtures.
![]() | ||
Fig. 1 Ferro-microfluidic particle sorting platform. (a) The device consists of a poly-dimethylsiloxane (PDMS) channel over copper electrodes. Microspheres, suspended in ferrofluid, are introduced from the center inlet and hydrodynamically focused by the ferrofluid flow from the side inlets. (b) Top view of the beginning of the separation channel. The electrodes are connected in quadrature, with a 90° phase difference to generate a magnetic field pattern that travels along +y-axis within the separation chamber, rolling the microspheres in the opposite direction. With the appropriate excitation frequency, the stream of larger particles become trapped and focused between the center electrodes, while smaller particles continuously move away. At the end of the separation channel, particles with different sizes reach different outlet reservoirs. |
![]() | ||
Fig. 2 Rotational interactions between nanoparticles and microspheres. (a) Under the influence of a magnetic field rotating within the y–z plane, most magnetite nanoparticles in the EMG700 ferrofluid get aligned with the field and begin whirling. Nanoparticles proximal to the surface of a nonmagnetic microsphere experience increased rotational drag, which results in partial angular momentum transfer and causes the microsphere to roll in the opposite direction. (b) On the other hand, presence of shear flow near the channel ceiling causes both nanoparticles and microspheres to spin counter-clockwise within the x–z plane with similar angular velocities. The surface velocity of the microsphere is much faster and opposite to that of the proximal nanoparticles, and continuously disrupts their rotation. At high enough shear rates, this disruption also begins to interfere with the angular momentum transfer due to the rotating field in the orthogonal plane. |
With uniform electrode spacing, particle manipulation in bio-ferrofluidic devices is limited to the separation of one microparticle or cell species from others. With varying electrode gaps, the balance between magnetic force and torque effects can be altered locally, and simultaneous sorting of more than two species becomes possible at a given excitation frequency. Specifically, larger electrode gaps can progressively trap smaller microparticles at a fixed frequency; hence, a gradient of electrode spacing may sort microparticles in descending size order. We proposed this theoretical possibility in our prior work;21 here, we experimentally demonstrate this sorting functionality in a simple, low-cost and disposable nanocytometer. One significant advantage that our approach offers is the potential for discrete sorting, with each species sorted and tightly focused onto a separate particle stream. This capability renders our bio-ferrofluidic nanocytometer highly practical simultaneously as a potential separator, sorter and counter of multiple microparticle species in parallel.
Fig. 3, top row depicts a series of three fluorescent microscopy image snapshots obtained during the course of the experiment from the inlet, center and outlet regions of the device, respectively. These images demonstrate the progression of microparticle separation as they are dragged downstream. The experiments were run multiple times with comparable results. Fig. 3, bottom row displays the corresponding particle counts (as normalized probability density functions) across the width of the separation channel, as determined from images accumulated over the three regions for at least 10 s. At the end of a given run, samples from both active outlets (a total of 10 samples from each outlet, 10 μl each) were pipetted onto microscope slides, and the number of microparticles in each droplet was counted using ImageJ.42 Particle counts were then averaged. Fig. 4 depicts the images taken from the inlet and the two active outlets, together with the corresponding sample purities. As expected, we did not find any microparticles in outlet 3 (control). We found that 98.2 ± 0.4% of the particles collected in outlet 1 were 2.2 μm and 86.3 ± 3.8% of particles collected in outlet 2 were 4.8 μm. The error bars represent the ± standard deviation in the independent sampling of the output wells (n = 10) at different times during the experiment. The lower purity value obtained in outlet 2 resulted in part from the electrostatic attraction of 2.2 μm microspheres to the larger particles. We define separation efficiency as the ratio of the number of target microparticles collected in their corresponding outlet to their total number in all outlets. Under this definition, the separation efficiency for 2.2 μm and 4.8 μm particles was 99% and 77%, respectively.
![]() | ||
Fig. 3 Binary separation of 2.2 μm and 4.8 μm microspheres (7 App, 1500 Hz). The microfluidic device features a PDMS channel 20 μm high, 2.7 cm long and 1200 μm wide. The electrodes are 210 μm wide and 220 μm apart. (Top) Fluorescence microscopy snapshots of the channel inlet, center and outlet regions from above, depicting 2.2 μm particles deviating from the center line and 4.8 μm particles being focused at the center spacing. Microspheres of different sizes eventually reach different outlets. (Bottom) Probability density functions (PDF's) associated with the y-location of each particle size as the microspheres crossed the depicted green dashed line at the beginning, center and end of the separation channel. 4.8 μm microspheres are tightly focused in single-file at the center stream, while 2.2 μm microspheres are continuously cleared. Scale bar: 200 μm. |
![]() | ||
Fig. 4 Microscope images of samples obtained from the initial particle mixture (a), as well as the two active outlets of the binary sorter (b) and (c). (d) Purity of the separated samples. 98.2 ± 0.4% of the particles collected from outlet 1 are 2.2 μm, while 86.3 ± 3.8% of the particles collected from outlet 2 are 4.8 μm particles. Scale bar: 100 μm. |
![]() | ||
Fig. 5 Triple size separation at 8 App and 1200 Hz. The microfluidic channel used in triple separation features electrodes with gaps that gradually increase (from 100 to 300 μm) to trap and focus subsequently smaller microspheres. The microsphere mixture is hydrodynamically focused over the smallest (i.e., 100 μm wide) electrode gap at the inlet of the separation channel. (Top) Fluorescence microscopy snapshots of the separation channel's inlet, center and outlet from above. The largest (9.9 μm) microspheres get trapped and focused over the narrowest electrode gap, as the smaller particles (4.8 μm and 2.2 μm microspheres) are directed to the neighboring spacings. Eventually, all three sizes get trapped and focused along neighboring spacings. (Bottom) PDF's of particle counts along the channel width, as sampled at the corresponding dashed green lines. Scale bar: 400 μm. |
One point worth mentioning is that the 2.2 μm diameter microspheres appear less focused within their own output stream compared to the other size groups. One reason for this phenomenon is that the electrode gap that traps the 2.2 μm microspheres is large, resulting in reduced magnetic forces locally and an increase in the time that is necessary for tight focusing. Longer devices could solve this issue if tighter focusing is absolutely necessary, without sacrificing throughput.
However, simply increasing the particle concentration within the channel or using multiple channels in parallel can substantially increase the throughput of ferrofluid-mediated particle sorting. For instance, channel depth can be increased 5-fold (to 100 μm) without a significant loss in particle manipulation capability; this is because currents through the excitation electrodes still mimic the effects of an infinite current sheet for heights much smaller than the traveling field wavelength (∼1200 μm). A deeper channel would necessitate a slight increase in the overall channel length, as it would take several seconds longer for the microparticles to travel up to the channel ceiling in the worst-case scenario. The increase in channel depth could also alleviate shear limitations on angular momentum transfer from nanoparticles to microspheres, allowing for faster input flows. Combined with 25 separation chambers running in parallel (covering up to 4.5 cm in total channel width), the next generation device then should be able to achieve a throughput of 105 particles/s without any further modification to this separation technology.
The maximum throughput potential for future devices would then be limited in part by the dwell time (30 s), the minimum transverse separation distance (∼500 μm), and the minimum particle manipulation velocity (20 μm s−1). We can overcome these three limitations in part by increasing the transverse particle velocity through the use of higher excitation currents and a higher susceptibility ferrofluid. Another consideration regarding upper limits on throughput with such devices involves the availability of ceiling real-estate that can accommodate the increased number of input micro-moieties. A 5-fold increase in channel depth would ultimately result in a corresponding increase in surface concentration of microparticles. Based on our observations of ceiling coverage (supplementary videos†) over the electrode gap where microparticles are focused, we conclude that there exists plenty of ceiling real-estate available to accommodate this increase in surface concentration of microparticles.
The ferrofluid-mediated nanocytometer offers the potential for high sample purities and separation efficiencies that are comparable to those obtained with commercially available devices, but without the use of labels. Alternatively, the microspheres themselves may be used as labels to separate, isolate and concentrate protein, DNA, virus, bacterial or cellular targets from complex mixtures. Specifically, microspheres with different sizes may be simultaneously utilized to isolate multiple different targets from the same sample. Another advantage of our approach is that the microparticles that are separated can be focused into single-file streams within discrete particle tracks, which makes our devices compatible with integrated particle counting and analysis schemes. The disposable nature of the devices prevents cross-contamination between measurements and keeps the costs down. The nanocytometer is easy-to-use and virtually labor-free. With a bio-compatible ferrofluid, the platform is applicable for live cell separation, sorting and population characterization, as well. Integrated with sensors, the nanocytometer could offer rapid and practical new assays in the context of pathogen detection, blood counting and circulating tumor cell isolation, replacing expensive conventional techniques.
Footnote |
† Electronic supplementary information (ESI) available: Two supplementary videos that display the binary and ternary separation of different sizes of microspheres within the nanocytometer. See DOI: 10.1039/c1lc20864k |
This journal is © The Royal Society of Chemistry 2012 |