Sara Melo
Oliveira
,
Teresa Isabel
Marques da Silva Lopes
,
Ildikó Vargáné
Tóth
and
António Osmaro
Santos Silva Rangel
*
Escola Superior de Biotecnologia, Universidade Católica Portuguesa, Rua Dr. António Bernardino de Almeida, 4200-072, Porto, Portugal. E-mail: aorangel@esb.ucp.pt; Fax: +351 225090351; Tel: +351 225580064
First published on 20th October 2008
A multi-commuted flow system coupled to a gas diffusion device was developed for the spectrophotometric determination of ammonium nitrogen in sea and estuarine waters. The efficiency of complexing agents to prevent precipitation of metallic hydroxides, due to the high pH value of the carrier solution, was studied. Under the optimised conditions, no interference was observed from different expected interfering ions as well as volatile amines. The proposed method provided the determination of NH4+ in concentrations ranging from 50 to 1000 µg L−1, with detection and quantification limits of 18 and 35 µg L−1, respectively. A determination rate of 20 h−1 was achieved, with good repeatability for 10 consecutive injections of sea and estuarine samples (relative standard deviations lower than 2.0%). Accuracy of the methodology was assessed through recovery assays in 10 samples and also by analysis of certified reference material.
To cope with this increasing demand for ammonium determination in a large number of coastal water samples, several automatic methods based on flow procedures like segmented flow analysis (SFA) and flow injection analysis (FIA) were developed. The methods without a separation device employ spectrophotometric,2,5–8 fluorimetric9,10 or potentiometric4 detection. To overcome the problems inherent to the nature of saline samples, these methods required matrix matching of the standard solutions with NaCl, artificial seawater or low nutrient seawater.2,4–11 The level of interferences and matrix effects of saline waters can be significantly reduced by the introduction of a gas diffusion device in the flow system.3,11–14 Some of these works also involved ion chromatography separation to quantify ammonia and methylamines.13,14 Nevertheless, the inclusion of a chromatographic separation step compromises drastically the sample throughput, and contributes to a higher analysis cost.
In this work, an alternative method for the ammonium determination in marine waters exploiting the multicommutation concept,15 and using low toxicity reagents, is presented. The multicommuted approach offers a high degree of automation, and low reagent consumption and waste generation, since solutions are introduced in the flow network only when they are required for the determination, returning to the respective reservoirs during the rest of the time of the analytical cycle. The developed method was based on the spectrophotometric monitoring of the colour change caused by a pH variation of a bromothymol blue solution, after diffusion of ammonia to the acceptor channel of the gas diffusion cell. In order to apply the system to samples with a wide salinity range, a systematic study with different solutions to eliminate interferences from coastal water samples was carried out. To accomplish this objective, the use of several complexing agents was evaluated, and an extensive study of possible interfering species was performed.
The carrier solution was prepared weakly by dissolving 70 g of potassium sodium tartrate tetrahydrate and 20 g of sodium hydroxide in 1000 mL of water.
To prepare the bromothymol blue stock solution, 0.2002 g of the indicator was dissolved in 100.0 mL of ethanol. The 250.0 mL acceptor solution was prepared daily by appropriate dilution of the stock solution with water, resulting in a solution containing 0.06 mmol L−1 of bromothymol blue. The pH of this solution was adjusted to 6.8 with NaOH 0.25 mol L−1.
A 1000 mg L−1 stock standard solution of ammonium was prepared by dissolving in water 0.2967 g of NH4Cl, previously dried for 2 h at 105 °C. This solution was adjusted to pH 2 with H2SO4, in order to avoid loss of analyte by its conversion to NH3 and the final volume was adjusted to 100.0 mL. Working standard solutions were daily prepared from the above solution, by dilution in water, resulting in ammonium concentrations of 50.0, 100, 400, 700 and 1000 µg L−1.
For the study of the complexing agents, solutions were prepared from dissolution in water of the required amounts of sodium citrate dihydrate, ethylenediaminotetracetic acid disodium salt dihydrate (EDTA), potassium sodium tartrate tetrahydrate, and boric acid.
A certified reference material QC RW1 (VKI Reference Materials) was prepared by dilution of 1.00 mL of the concentrated reference solution in 100.0 mL of the sample, according with the manufacturer’s specifications.
All samples were collected in previously washed and dried polyethylene bottles and filtered through a 0.45 µm cellulose acetate membrane filter (Whatman). Samples were introduced in the flow system without the need to carry out any further treatment.
To control the direction of the solutions, five three-way solenoid valves (NResearch, 161 T031, New Jersey, USA) were used. The solenoid valves were operated by means of a power drive (CoolDriveTM, NResearch). A 386 personal computer (SD700, Samsung, Seoul, South Korea) equipped with an interface card (PCL-818L, Advantech) running a lab-made software written in QuickBasic 4.5 (Microsoft, USA) controlled the switching of the solenoid valves.
The gas diffusion device consisted of two separate Perspex blocks, pressed against each other by 6 screws.16 A hydrophobic membrane (HVHP09050, Millipore Durapore®, Madrid, Spain) with a porosity of 0.45 µm was placed between the two blocks, being replaced weekly.
Absorbance measurements were carried out by a UV/Vis spectrophotometer (Unicam 8625, Cambridge, United Kingdom) set at 620 nm, equipped with a flow-through cell with 18 µL of internal volume and 1-cm flow path (Hellma 178.712-QS, Mullheim/Baden, Germany). A chart recorder (Kipp & Zonen BD111, Delft, Holland) connected to the spectrophotometer was used to register the analytical signals.
Step | Description | Position of the commutation valves | Time/s | ||||
---|---|---|---|---|---|---|---|
1 | 2 | 3 | 4 | 5 | |||
a The letters N and F correspond to positions “on” and “off” of the commutation valves, respectively. | |||||||
1 | Wash connection between valves V3 and V5 | N | F | N | F | N | 15 |
2 | Wash connection between confluence and valve V5 | N | F | F | N | N | 15 |
3 | Wash acceptor and donor channels | F | N | F | F | F | 20 |
4 | Sample introduction | F | N | N | F | F | 18 |
5 | Sample introduction and stop BTB flow | N | N | N | F | F | 12 |
6 | Stop BTB flow | N | N | F | F | F | 48 |
7 | Propel BTB toward the detector. Signal registration | F | N | F | F | F | 90 |
![]() | ||
Fig. 1 Schematic configuration of the multicommuted flow system for the spectrophotometric determination of ammonium in marine waters. BTB: bromothymol blue 0.06 mmol L−1, 1.7 mL min−1; CS: potassium and sodium tartrate 70 g L−1 + NaOH 0.5 mol L−1, 0.84 mL min−1; S: sample or standard, 0.85 mL min−1; H2O, 0.82 mL min−1; P: peristaltic pump; Vi: solenoid valve; C: confluence; R: reaction coil (100 cm); W: waste; GDU: gas diffusion unit; D: detector (620 nm). In the valves, the position “on” is represented by a continuous line and the position “off” is represented by a dotted line. |
The analytical cycle started with the manifold washing steps. First, the connection between valves V3 and V5 was washed and filled with the new sample (step 1). Then, the sample remaining in the tubing between the confluence and valve V5 was removed with the carrier solution (step 2). Finally, donor and acceptor channels were washed with the respective solutions (step 3). The above-mentioned steps were only necessary when a new sample was introduced in the flow system. Afterwards, the sample was introduced in the system with the carrier solution, over 30 s (steps 4 and 5). To increase the efficiency of the diffusion process, the acceptor stream was stopped during 60 s, as soon as the donor solution reached the gas diffusion area (steps 5 and 6). In the last step, the acceptor solution with the diffused analyte was sent towards the detector, and then recirculated to the acceptor solution flask. This recycling was considered as this solution did not suffer any significant alteration during the analytical cycle, as previously shown in a systematic study.16 To ensure baseline stability the volume of the acceptor solution was relatively large (250 mL) and was maintained under constant stirring.
Applying the previous flow system to saline samples, recovery percentages close to 120% were obtained (Table 2), and after several injections, a poor precision of the results was noticed. Moreover, the formation of a white precipitate inside the tubing was observed, leading to the deterioration of the flow system performance. This fact is probably due to the precipitation of some metal ions in alkaline medium, present in high levels in saline waters, as metal hydroxides, such as Ca(OH)2 and Mg(OH)2. Despite this precipitation occurs in the donor channel and so the colour detection is not affected, the decrease in repeatability affects the overall method performance.
To overcome these problems, preliminary off-line studies were carried out to evaluate the efficiency of different complexing agents. However, addition of this complexing agents to the NaOH solution caused a significant decrease on pH of the carrier solution. In order to maintain pH values above 12 and ensure total conversion of ammonium to ammonia, NaOH concentration was increased to 0.5 mol L−1. Evaluation of complexing agents was carried out by adding 2.00 mL of saline samples to 2.00 mL of a solution containing NaOH 0.5 mol L−1 plus the complexing agent, followed by the measurement of the absorbance of the formed mixture at 420 nm. In the first set of experiments, several concentrations of sodium citrate, EDTA and potassium sodium tartrate were tested separately.
The results, presented in Fig. 2 revealed that the use of EDTA or potassium sodium tartrate in concentrations of 70 g L−1 prevented occurrence of precipitation. However, citrate did not prevent precipitation, even when a concentration of 180 g L−1 was employed. Citrate is typically applied in ammonia determination by the indophenol blue method at a pH value of 10.5. However, in this method, pH values higher than 12 are required to promote the maximum conversion of ammonium to ammonia, and consequently increase the diffusion efficiency. According to others,13 the ability of citrate to chelate Mg2+ and Ca2+ and thereby inhibit the precipitation of these metals as hydroxides, was only effective at pH values below 11. This was confirmed by measuring the pH of the solutions containing sample, hydroxide and citrate. For solutions containing all citrate concentrations tested, pH values > 12.7 were obtained.
![]() | ||
Fig. 2 Evaluation of different complexing agents on precipitation prevention in sea and estuarine waters, in alkaline medium. |
In a second stage, studies to evaluate the efficiency of the combination of EDTA with citrate were also carried out, using fixed EDTA concentrations of 30 and 40 g L−1, while the citrate concentration was varied up to 180 g L−1. Results revealed (Fig. 3) that the use of citrate combined with EDTA does not prevent precipitation completely, although a higher efficiency of citrate on precipitation impediment was observed when this complexing agent was combined with a concentration of 40 g L−1 of EDTA.
![]() | ||
Fig. 3 Study of the efficiency of different concentrations of citrate on precipitation prevention of coastal waters in alkaline medium, using EDTA concentrations of 30 (A) and 40 g L−1 (B). |
Considering an International Standard,17 the efficiency of a solution composed by EDTA plus boric acid was studied. In a first approach, fixing EDTA concentration to 30 g L−1, the concentration of boric acid was studied in a range of 0–20 g L−1. Afterwards, using a concentration of 16 g L−1 of boric acid, the EDTA concentration was evaluated from 0 to 40 g L−1. Results, depicted in Fig. 4, revealed that the addition of boric acid allows the reduction of EDTA concentration. Taking in consideration the overall results, three efficient complexing agents were found to be successful to prevent precipitation of metal hydroxides in saline waters: EDTA, tartrate and EDTA in combination with boric acid. However, tartrate is a more environmental friendly option and was referred by the U.S. Environmental Protection Agency18 as an advantageous alternative to EDTA. Therefore, tartrate was selected as the complexing agent and its concentration was studied in the flow system. This evaluation was carried out by adding different concentrations of tartrate to the carrier solution containing NaOH 0.5 mol L−1. Tartrate concentrations ranging from 70 to 120 g L−1 were studied. Baseline instability was observed with the tartrate concentration increase. Thus, a tartrate concentration of 70 g L−1 was chosen for further work.
![]() | ||
Fig. 4 Study of the efficiency of different concentrations of boric acid on precipitation prevention of coastal waters in alkaline medium, using 30 g L−1 of EDTA (A) and different concentrations of EDTA, using 16 g L−1 of H3BO3 (B). |
Species studied | Concentration tested | Relative deviation (%) | Expected concentration in seawater |
---|---|---|---|
a N/A - not available. | |||
Methylamine (nmol L−1) | 800 | 3.4 | 58720 |
Dimethylamine (nmol L−1) | 2000 | 2.7 | 36012 |
Trimethylamine (nmol L−1) | 1000 | 2.4 | 51420 |
Ethylamine (nmol L−1) | 1500 | 3.0 | 140020 |
Diethylamine (nmol L−1) | 3000 | 2.1 | 7120 |
Triethylamine (nmol L−1) | 3000 | 4.3 | N/A |
Triethanolamine (nmol L−1) | 8000 | −3.1 | N/A |
Urea (µmol L−1) | 10 | −0.5 | 320 |
HCO3− (mg L−1) | 200 | 4.1 | 14021 |
CO32− (mg L−1) | 100 | 4.4 | N/A |
Cu2+ (mg L−1) | 1 | −2.2 | 0.00122 |
Al3+ (mg L−1) | 1 | 0.6 | 0.00122 |
Fe3+ (mg L−1) | 1 | 1.1 | 0.00422 |
Hg2+ (mg L−1) | 1 | 2.2 | 0.000222 |
S2− (mg L−1) | 10 | 1.6 | 0.09623 |
Ca2+ (mg L−1) | 500 | 0.6 | 41221 |
Mg2+ (mg L−1) | 1500 | 1.9 | 128421 |
K+ (mg L−1) | 500 | 4.0 | 39921 |
Sr2+ (mg L−1) | 20 | 3.7 | 7.9421 |
SO42− (mg L−1) | 3000 | 4.5 | 271221 |
Br − (mg L−1) | 100 | 0.5 | 6721 |
H3BO3 (mg L−1) | 50 | 2.0 | 2521 |
F − (mg L−1) | 10 | −0.5 | 1.321 |
![]() | ||
Fig. 5 Recorder output obtained in the spectrophotometric determination of ammonium, corresponding to the injection of a set of standard solutions (a = 0, b = 50, c = 100, d = 400, e = 700, and f = 1000 µg L−1), and recovery assays with addition of 50, 200, 500 and 800 µg NH4+ L−1, using a seawater sample. |
The repeatability of the procedure was assessed from the relative standard deviation (RSD) calculated from ten consecutive injections of the certified reference material prepared in four different samples, providing RSD values less than 2.0%.
The detection and quantification limits were calculated as recommended by the IUPAC.24 Detection and quantification limits of 18 and 35 µg NH4+ L−1 were obtained, respectively.
The determination throughput was calculated considering the time spent in all steps of the analytical cycle. A time of 168 s was needed for each determination, and a 50 s washing time was required between different samples, resulting in a determination frequency of 20 h−1.
Conductivity and pH measurements were carried out for each sample. Salinity values of the samples were calculated based on the practical salinity scale,25 using the conductivity values and the measurement temperature (21 °C). These values are presented in Table 4.
Sample | Sample characteristics | [NH4+] (µg L−1) | Recoverya (%) | |
---|---|---|---|---|
Added | Founda | |||
a n = 9. b Salinity. | ||||
1 | pH = 7.71 S = 11.2b | 0 | 61.4 ± 1.9 | — |
50 | 111 ± 3 | 98.7 ± 5.5 | ||
200 | 250 ± 5 | 94.5 ± 2.3 | ||
500 | 560 ± 2 | 99.8 ± 0.5 | ||
800 | 858 ± 4 | 99.6 ± 0.5 | ||
2 | pH = 7.77 S = 9.6b | 0 | 49.7 ± 5.6 | — |
50 | 98.0 ± 2.9 | 96.5 ± 5.8 | ||
200 | 247 ± 9 | 98.8 ± 4.4 | ||
500 | 562 ± 7 | 102 ± 1 | ||
800 | 853 ± 11 | 100 ± 1 | ||
3 | pH = 7.93 S = 10.0b | 0 | 75.6 ± 1.5 | — |
50 | 125 ± 1 | 99.0 ± 2.2 | ||
200 | 267 ± 2 | 95.7 ± 1.1 | ||
500 | 570 ± 3 | 98.9 ± 0.6 | ||
800 | 872 ± 5 | 99.5 ± 0.6 | ||
4 | pH = 7.69 S = 34.2b | 0 | 81.5 ± 2.2 | — |
50 | 132 ± 1 | 101 ± 2 | ||
200 | 283 ± 9 | 101 ± 4 | ||
500 | 579 ± 9 | 99.5 ± 1.7 | ||
800 | 884 ± 6 | 100 ± 1 | ||
5 | pH = 7.82 S = 32.4b | 0 | 70.0 ± 2.4 | — |
50 | 120 ± 2 | 101 ± 5 | ||
200 | 274 ± 3 | 102 ± 2 | ||
500 | 587 ± 8 | 103 ± 2 | ||
800 | 889 ± 8 | 102 ± 1 | ||
6 | pH = 8.00 S = 34.8b | 0 | 38.5 ± 1.3 | — |
50 | 88.1 ± 1.4 | 99.2 ± 2.8 | ||
200 | 228 ± 4 | 94.9 ± 1.8 | ||
500 | 520 ± 2 | 96.4 ± 0.5 | ||
800 | 830 ± 5 | 98.9 ± 0.6 | ||
7 | pH = 8.00 S = 33.0b | 0 | 119 ± 3 | — |
50 | 169 ± 2 | 99.8 ± 3.4 | ||
200 | 313 ± 4 | 96.9 ± 2.0 | ||
500 | 602 ± 5 | 96.7 ± 1.0 | ||
800 | 909 ± 6 | 98.9 ± 0.8 | ||
8 | pH = 7.96 S = 33.6b | 0 | 88.7 ± 2.2 | — |
50 | 141 ± 2 | 105 ± 5 | ||
200 | 281 ± 5 | 96.3 ± 2.5 | ||
500 | 604 ± 10 | 103 ± 2 | ||
800 | 891 ± 8 | 100 ± 1 | ||
9 | pH = 7.97 S = 31.9b | 0 | < LOQ (21.9 ± 2.5) | — |
50 | 51.5 ± 2.1 | 103 ± 4 | ||
200 | 196 ± 6 | 98.1 ± 3.0 | ||
500 | 481 ± 9 | 96.2 ± 1.9 | ||
800 | 776 ± 12 | 96.9 ± 1.5 | ||
10 | pH = 8.00 S = 32.4b | 0 | 61.1 ± 4.3 | — |
50 | 112 ± 2 | 102 ± 3 | ||
200 | 256 ± 5 | 97.6 ± 2.5 | ||
500 | 564 ± 7 | 100 ± 1 | ||
800 | 866 ± 6 | 100 ± 1 |
The accuracy of the proposed method was assessed by recovery tests and analysis of certified reference material.
Recovery studies consisted in the addition of 1.00 mL of standard solutions of ammonium to volumetric flasks of 25.00 mL, adjusting the volume with the respective sample. Ammonium concentrations of 50.0, 200, 500 and 800 g L−1 were added to all samples. Each concentration level was prepared in triplicate, and all assays were analysed also in triplicate, resulting in 9 peaks for each concentration level.
Recoveries between 94.5 and 105% were obtained (Table 4). Statistical test (t-test) was used to evaluate if the mean recovery value did not significantly differ from 100%.19 The results demonstrated that the recovery values were not statistical different from 100% at a 95% confidence level, since the calculated t-value (1.81) was lower than the correspondent t-critical value (2.02) (n = 40), thus indicating the absence of systematic errors.
Concerning the certified reference material QC RW1, the certified value is 100.9 ± 1.3 μg N–NH4+ L−1 and the acceptance limit is 100.2–101.5 μg N-NH4+ L−1. Each sample was injected in the flow system ten times. Analysis of the certified sample material prepared using samples 2, 3, 5 and 6 provided concentrations of 100.7 ± 1.3, 101.2 ± 2.0, 101.1 ± 1.6, and 100.5 ± 1.6 µg L−1 of N–NH4+, respectively. For all samples tested, the values obtained were within the acceptance limit specified by the reference material.
The majority of the flow methods for ammonium determination in sea or estuarine waters require the preparation of the working standard solutions in NaCl, artificial seawater or LNSW. In the herein described method, the standards are prepared in deionized water. Despite the absence of salinity in the standard solutions used for calibration, accurate results were attained when samples with salinities ranging from 9.6 to 34.8 were analysed.
This journal is © The Royal Society of Chemistry 2009 |