Ganeshram
Krishnamoorthy
*,
Edwin T.
Carlen
,
J. Bianca
Beusink
,
Richard B. M.
Schasfoort
and
Albert
van den Berg
BIOS Lab-on-a-Chip Group, MESA+ Institute for Nanotechnology, University of Twente, P.O. Box 217, 7500, AE Enschede, The Netherlands. E-mail: g.krishnamoorthy@utwente.nl; Fax: 0031 53 4893595; Tel: 0031 53 4892724
First published on 19th October 2009
Binding affinity of biomolecular interactions can be directly extracted from measured surface plasmon resonance biosensor sensorgrams by fitting the data to the appropriate model equations. The conventional method for affinity estimation uses a series of analytes and buffers that are injected serially to a single immobilized ligand on the sensing surface, including a regeneration step between each injection, to generate information about the binding behavior. We present an alternative method to estimate the affinity using a single analyte concentration injected to multiple ligand densities in a microarray format. This parameter estimation method eliminates the need for multiple analyte injections and surface regeneration steps, which can be important for applications where there is limited analyte serum, fragile ligand-surface attachment, or the detection of multiple biomolecule interactions. The single analyte injection approach for binding affinity estimation has been demonstrated for two different interactant pairs, β2 microglobulin/anti-β2 microglobulin (β2M) and human IgG/Fab fragments of anti-human IgG (hIgG), where the ligands are printed in a microarray format. Quantitative comparisons between the estimated binding affinities measured with the conventional method are β2M: KD = 1.48 ± 0.28 nM and hIgG: KD = 12.6 ± 0.2 nM and for the single injection method are β2M: KD = 1.52 ± 0.22 nM and hIgG: KD = 12.5 ± 0.6 nM, which are in good agreement in both cases.
Since the first report of the surface plasmon resonance imaging (iSPR) biosensor system in the late 1980's,11 the iSPR technique is now commonly accepted as being suitable for the label-free measurement of multiple simultaneous biomolecular interactions, compared to other established methods, such as the quartz crystal microbalance,12 Suprex MALDI mass spectroscopy13 and kinetic capillary electrophoresis,14 due to its compatibility with microarray substrates. Currently, there are many commercially available iSPR systems, including the Biacore Flexchip,15GWC SPRimagerII,16IBIS-iSPR,17,53 Genoptics SPRi-Lab+,18 and also many custom-made iSPR systems.19,20
Biomolecular interactions measured with iSPR,21–23 can provide binding information, such as binding quantitation and specificity, and rate and affinity constants24 of two or more interactants, which can lead to further understanding of interactant binding. For example, in the development process of new drugs, kinetic experiments provide insights into potential drug candidates as well as the definition of lead targets.25 Kinetic parameter estimation of biomolecular interactions from integrated microarray-iSPR is increasingly being utilized for multi-analyte or multi-ligand studies.24–36Table 1 shows a compilation of SPR-microarray studies from literature for various application areas, including protein–DNA interactions,26,27epitope mapping,28,32antibody screening,29,31 demonstration of new imaging systems,30,34,36protein–carbohydrate interactions,33 allergen–antibody interactions,35 affinity ranking,37 and small molecules analysis.38
Ref. | No. spots | Spot conc. | No. ligands | No. analytes | Analyte conc. | Application |
---|---|---|---|---|---|---|
26 | 20 | Same | 5 · 4 | 1 | Same | Protein–DNA interaction |
27 | 36 | Same | 6 · 6 | 1 | Same | Protein–DNA interaction |
28 | 54 | Same | 18 · 3 | 18 (1/ligand type) | Different | Epitope mapping |
29 | 96 | Same | 2 · 36; 1 · 4; 1 · 20 | 4 | Multiple | Antibody screening |
30 | 900 | Same | 30 · 30 | 30 (1/ligand type) | Different | New iSPR system demonstration |
31 | 8 | — | 1 · 8 | 386 (1/ligand type) | Different | Antibody screening |
31 | 16 | — | 16 · 1 | 80 (5/ligand type) | Different and multiple | Antibody screening |
32 | 12 | — | 12 · 1 | 1 | Same | Epitope mapping |
32 | 9 | — | 9 · 1 | 1 | Same | Epitope mapping |
33 | 8 | Multiple | 2 · 4 | 2 (1/ligand type) | Different | Protein–carbohydrate interaction |
34 | 156 | Same | 12 · 13 Matrix + 156 (ref.) | 5 | Multiple | New iSPR system demonstration |
35 | 24 | — | 24 · 1 | 24 × (8/10/12) | Multiple | Allergen–antibody interactions |
36 | 110 | Same | 1 sample shown | 6 | Multiple | New iSPR system demonstration |
37 | 302 | Same | 2 · 151 | 191 | Different | Affinity ranking |
37 | 288 | Same | 3 · 96 | 191 | Different | Affinity ranking |
38 | 36 | Same | 1 · 6 | 6 | Multiple | Small molecules analysis |
38 | 36 | Same | 1 · 6 (8 comp.) | 40 (5/ligand type) | Different and multiple | Small molecules analysis |
Conventional kinetic parameter estimation is done by measuring sensor responses in a serial way by introducing a range of analyte concentrations to a fixed ligand density on the sensing surface where between each analyte injection step, dissociation and surface regeneration steps are performed each requiring a different buffer solution. Although the conventional procedure is important for general ligand–ligate binding parameter estimation, there are certain applications that can benefit from alternative parameter estimation methods. We present an alternative binding affinity estimation method that uses a single analyte injection and multiple ligand densities in a microarray format; different than the conventional estimation method used to extract the binding affinity. This parameter estimation method eliminates the need for multiple analyte injections and surface regeneration steps, which can be important for applications where there is limited analyte serum, fragile ligand–surface attachment, or the detection of multiple biomolecule interactions. This approach was briefly described in our previous article, which describes a new lab-on-a-chip device integrated with iSPR system.39 In this article we demonstrate the effectiveness of the new method by comparing the estimated binding affinity of two well-known interactant pairs along with the conventional method. Differences between various measurement scenarios in this combination (microarray-iSPR) are discussed, as well as advantages and disadvantages of the single injection method.
![]() | ||
Fig. 1 Measurement scenarios (a) Conventional method with single ligand [B] and multiple analyte concentrations [Ap] (b) Single ligand density microarray [Bm·n] and multiple analyte concentrations [Ap] (c) Multiple ligand densities and multiple analyte concentrations (d) Multiple ligand density and single analyte concentration (e) Multiple ligand types [B1]-[Bq] and multiple analyte concentrations [Ap]1-[Ap]q (f) Multiple ligand types [B1]-[Bq] and multiple analyte concentrations of [A]1-[A]q (g) Multiple ligand densities and types [Bm·n]1-[Bm·n]q and multiple analyte concentrations [Ap]1-[Ap]q (h) Multiple ligand densities and types [Bm·n]1-[Bm·n]q and multiple analyte concentrations [A]1-[A]q. (m · n is multiplication of the number of row m and the number of columns n of the microarray). |
The most commonly reported SPR binding kinetics measurement, or conventional measurement, uses panalytes [A]p serially injected to a single ligand density [B], where each of the p injections is separated by a surface regeneration step (Fig. 1a) and measurement responses are recorded over the entire experimental time. The responses are recorded as t = [
1t,
2t,…,
pt], where
t is a matrix of p response vectors
it for each analyte concentration i sampled over the experimental time te. Kinetic model parameters are extracted from the measured responses by finding the global-minimum of an error function, such as the “chi-square” function
, where σit is the standard deviation of each data point, such that the chosen model function ft(aj), best matches the measured responses
t with an optimal set of model parameters aj. This type of parameter extraction is referred to as a “global” fitting procedure since the model parameters aj are estimated over all measured responses, which is considered a more accurate method for parameter estimation.40 The scenario where a constant ligand density is printed in a microarray (Fig. 1b) is similar to the single spot ligand density (Fig. 1a).
The second experimental scenario consists of multiple ligand densities of the same type printed in a microarray format (Fig. 1c and 1d). For the case where both the analyte and ligand densities are varied (Fig. 1c), the measured responses form a three-dimensional matrix consisting of p response vectors for all m · n ligands t = [
1t,
2t,…,
m·nt]p. The model parameters are extracted with a global fitting procedure as previously described. An interesting alternative, and the topic of this article, is the injection of a single analyte concentration, referred to as “single injection”, and using multiple ligand densities (Fig. 1d) to extract the model parameters. In this case, the measured response matrix contains m · n response vectors
t = [R1t,R2t,…,Rm·nt]. Compared to the conventional method (Fig. 1a and 1b), this method offers some clear advantages such as a reduction in analyte sample and the elimination of regeneration steps. For example, five different analyte concentrations used in a conventional kinetics experiment, each 100 µl, requires about 500 µl of sample; a five-fold reduction in sample volume. We expect a reduction in ligand consumption using a microarray-based approach. The main disadvantage with the single analyte injection approach is that ligand surface density cannot be well controlled with conventional spotting techniques, which can be problematic if each spot requires calibration, which is usually the case when screening for binding interaction parameters, however, this limitation will diminish with improvements in spotting techniques. From a model parameter extraction perspective, the accuracy of the fitting procedure favors the former method (Fig. 1c) where m · n response vectors are typically larger than p response vectors from multiple analyte concentrations, however, this requires the greatest experimental effort and requires surface regeneration.
The third experimental scenario consists of four possibilities, shown in Fig. 1e–h, consisting of multi-analyte and multi-ligand measurement scenarios. In this case, multiple ligand types (Fig. 1e and 1f) with duplicates as well as multiple ligand densities (Fig. 1g and 1h) are immobilized on the surface. This approach is mainly useful in the screening of drug targets where hundreds to thousands of molecules are to be screened. These scenarios are not described in detail in this article and have been included for sake of completeness.
In this article we evaluate the single injection kinetics method (Fig. 1d) and compare results from two different ligand–analyte systems with measurements from the conventional method (Fig. 1a/1b) and assess its use for actual experimental applications described in Fig. 1c and 1d.
From the rate equation, assuming [A]≫[B] and initial conditions f (t = 0) = 0, the model response function is f = α(1 − e−βt) = f(ka,kd,Rm), where α = ka[A]Rm/ka[A] + kd and β = −ka[A] + kd.40 The dissociation rate constant can be evaluated using the initial condition fd(t = td) = R0 resulting in fd(t) = R0e−kdt, where R0 is the response at time td, the onset of the dissociation phase.40 The simple 1:1 interaction model is used for parameter extraction in this article.
Different analyte concentrations: [A1] = 72, [A2] = 36, [A3] = 18, [A4] = 9, [A5] = 4.5, [A6] = 2.25, [A7] = 1.1 and [A8] = 0.5 nM, were prepared in the binding buffer. An iSPR image of the fabricated microarray is shown in Fig. 2a. The SPR dips were measured for all the 24 spots (Fig. 2b). Regeneration was done with 10 mM Glycine-HCl (pH 1.6) between each analyte injections in the case of multiple analyte injections. In this case, the association and dissociation profiles were measured for 900 and 600 seconds respectively.
![]() | ||
Fig. 2 (a) Real-time image of the microarray with multiple ligand densities of β2M (b) SPR measurement dips for twenty-four ligand spots (c) Conventional kinetics estimation for ligand density 125 µg/ml and 250 µg/ml and different analyte concentrations [A1]-[A8]. The point plots represents the residual plots obtained from the results of 1:1 model fit function to the recorded iSPR data. |
Multi-analyte/Single-ligand (Fig. 1a/1b) | Multi-analyte/Multi-ligand (Fig. 1c) | Single-analyte/Multi-ligand (Fig. 1d) |
---|---|---|
[B2]: KD = 1.48 ± 0.28 | ||
R m = 66.4 ± 5.4 m° | [A1]: KD = 1.50 ± 0.60 | |
[B3]: KD = 1.21 ± 0.18 | [A2]: KD = 1.57 ± 0.67 | |
R m = 52.6 ± 3.7 m° | [A3]: KD = 1.50 ± 0.66 | |
K D = 1.48 ± 0.28 | [B4]: KD = 1.37 ± 0.29 | [A4]: KD = 1.61 ± 0.58 |
R m = 66.4 m° | R m = 12.8 ± 3.6 m° | [A5]: KD = 1.57 ± 0.53 |
[B5]: KD = 1.43 ± 0.28 | [A6]: KD = 1.29 ± 0.55 | |
R m = 5.9 ± 1.0 m° | [A7]: KD = 1.60 ± 0.64 | |
[B6]: KD = 1.49 ± 0.27 | [A8]: KD = 1.57 ± 0.82 | |
R m = 1.6 ± 0.9 m° | ||
Weighted average | K D = 1.36 ± 0.11 | K D = 1.52 ± 0.22 |
![]() | ||
Fig. 3 Single injection kinetics estimation of β2M/a-β2M: Sensorgram obtained for the injection single analyte concentration over five different ligand densities for creating the spots ((a) a-β2M: [A] = 36 nM (b) a-β2M: [A] = 72 nM). Residual plots are shown for the respective analyte concentrations. |
The single injection measurement (Fig. 1d) was performed and example images and sensorgrams are shown in Fig. 3a and 3b, respectively. The affinity for each analyte concentration is listed in the third column of Table 2 (see Supporting information for more details†). The Rm values are not listed in the tables for the single injection kinetics approach as the different analyte concentrations have five different Rm values due to the five different ligand densities. It is also due to the well known fact that Rm is directly proportional to ligand densities and is an indirect way to quantify the ligand spots. The varying spacing between measured responses from ligand densities [B2] and [B3] in Fig. 3 is due to the lack of precise control of the spot concentration as recently described.53 A certain ligand density is achieved by a timed exposure of the ligand with the surface; we use the term density rather than concentration to describe the final amount of ligand. The ligand densities are directly proportional to the extracted Rm values (see Supporting information for more details†). In Fig. 2 the values of the ligand density for creating the spots are given. The extracted affinities KD are in good agreement among all three measurements shown here, as well as another recent report.50
For the single injection measurement, it is important that the analyte concentration is large enough to ensure that the response signal-to-noise ratio is large enough to avoid measurement errors. Another important consideration for the single injection approach is that at least one injection of a known analyte concentration has to be done to calibrate the analyte/ligand behavior. If the affinity constant is known, then one can easily prepare the analyte close to the affinity constant. The obtained rate constants are shown in Table 4.
![]() | ||
Fig. 4 (a) Real-time SPR image of multiple concentration hIgG ligand spots (b) SPR measurement dips for twenty-four ligand spots (c) Conventional kinetics estimation (1:1 model) of 500 µg/ml and 250 µg/ml ligand spots and multiple analyte concentrations. Residual plots shown for the respective ligand concentrations. |
Multi-analyte/Single-ligand (Fig. 1a/1b) | Multi-analyte/Multi-ligand (Fig. 1c) | Single-analyte/Multi-ligand (Fig. 1d) |
---|---|---|
[B1]: KD = 12.6 ± 0.2 | ||
R m = 43.8 m° | [A1]: - | |
[B2]: KD = 12.0 ± 1.3 | [A2]: - | |
R m = 41.0 m° | [A3]: KD = 11.1 ± 3.2 | |
K D = 12.0 ± 1.3 | [B3]: KD = 13.0 ± 1.7 | [A4]: KD = 10.0 ± 3.0 |
R m = 41.0 m° | R m = 2.6 m° | [A5]: KD = 12.6 ± 4.7 |
[B4]: KD = 12.8 ± 3.8 | [A6]: KD = 12.1 ± 1.7 | |
R m = 1.9 m° | [A7]: KD = 12.9 ± 0.8 | |
[B5]: KD = 12.3 ± 3.96 | [A8]: KD = 12.3 ± 1.6 | |
R m = 0.8 m° | ||
Weighted average | K D = 12.6 ± 0.2 | K D = 12.5 ± 0.6 |
Since the affinity is approximately 12 nM, the lower analyte concentrations can be neglected. Also mass transport affected the high concentration ligand spots when the analyte concentration is very low, as well as the low concentration spots with a high analyte concentration (results not shown). The single injection measurement was performed and example images and sensorgrams are shown in Fig. 5a and 5b, respectively.
![]() | ||
Fig. 5 Single injection kinetics estimation of hIgG/a-IgG interaction. Sensorgrams obtained for the injection single analyte concentration over five different ligand concentrations ((a) a-IgG: [A] = 100 nM (b) a-IgG: [A] = 200 nM). Residual plots shown for the respective analyte concentrations. |
The affinity for each analyte concentration is listed in the third column of Table 3. The extracted affinities KD are in good agreement among all three measurements. The estimated maximum response Rm shows a systematic decrease that is proportional to the reduced ligand densities (see Supporting information for more details†). The advantage of experimental time reduction and elimination of regeneration steps must be balanced with larger measurement errors when using the single injection approach as shown in Table 2 and 3 for the second scenario (Fig 1d). This is due to the fact that ligand immobilization is not well controlled with our present spotting technique, which is done offline and the immobilization is completely due to the diffusion of molecules from liquid droplets to the sensor surface over an incubation time. The obtained rate constants are shown in Table 4.
Interactant pairs | Conventional kinetics | Single injection kinetics | ||
---|---|---|---|---|
k a (M−1 s−1) | k d (s−1) | k a (M−1 s−1) | k d (s−1) | |
β2M/Anti-β2M | (1.31 ± 2.77) × 106 | (2.47 ± 5.32) × 10−3 | (2.77 ± 4.36) × 106 | (3.52 ± 5.64) × 10−3 |
HIgG/a-IgG | (1.42 ± 0.69) × 104 | (1.78 ± 0.88) × 10−5 | (1.88 ± 0.89) × 104 | (1.72 ± 1.31) × 10−4 |
For the conventional measurement, the analyte concentrations are highly controllable as demonstrated by the near equidistant sensor response curves (Fig. 2c). The number of sample points for each scenario is also different. In general, larger sample sizes (number of samples) result in higher accuracy parameter extraction compared to smaller sample sizes. Single injection kinetic estimations of multiple interactant pairs can reduce the time for kinetic experiments, as well as measurement costs as only a single sensor chip and single analyte injection are required.
The single injection approach can be extended to any number of target biological systems. As long as the ligand density is small and the analyte concentration is close to the real affinity of the biomolecules, this approach can be used. Considerations to mass transport limitations, rebinding effects, steric hindrance in hydrogels and other factors that may complicate the 1:1 binding model should be taken into account and low ligand densities are always the best to determine kinetic rate constants. The application of single injection kinetics over a microarray with multiple ligand densities can lead to fast kinetic parameter estimation of many samples from the same sensor surface. Certain aspects should be considered when designing such experiments, one of which is sample cross reactivity that can lead to inaccurate kinetic estimation.
Footnote |
† Electronic supplementary information (ESI) available: Supplementary material showing detailed data analysis including Fig. S1, Fig. S2, Fig. S3 and Table S1. See DOI: 10.1039/b9ay00176j |
This journal is © The Royal Society of Chemistry 2009 |