Raquel Beatriz Ribeiro
Mesquita
ab,
Maria Teresa Soares Oliveira Barbosa
Ferreira
a,
Ricardo Luiz Alves
Segundo
a,
Catarina Fernanda Carvalho Pinheiro
Teixeira
b,
Adriano Agostinho
Bordalo
b and
António Osmaro Santos Silva
Rangel
*a
aCBQF/Escola Superior de Biotecnologia, Universidade Católica Portuguesa, R. Dr. António Bernardino de Almeida, 4200 - 072, Porto, Portugal. E-mail: aorangel@esb.ucp.pt
bLaboratory of Hydrobiology, Institute of Biomedical Sciences Abel Salazar (ICBAS) and Institute of Marine Research (CIIMAR), Universidade do Porto, Lg. Abel Salazar 2, 4099 - 003, Porto, Portugal
First published on 29th October 2009
In this work, a sequential injection methodology for monitoring nitrite and nitrate in estuarine waters without any previous treatment is described. The developed system was applied to the measurement of nitrite and nitrate in estuarine waters of three rivers in the NW Portugal, allowing an automatic, fast (ca 60 h−1) and precise method (relative standard deviation lower than 2%). The procedure was based on the reaction between nitrite, sulfanilamide and N-(1-naphthyl)-ethylenediamine dihydrochloride (N1NED), whereas the determination of nitrate resulted from its reduction to nitrite, using an in-line cadmium column, followed by the same reaction. The samples were collected in three locations for each river (Douro, Cávado and Ave) covering the lower, middle and upper section of the estuaries. Despite the presence of a salinity gradient, this parameter showed no interference in the accuracy of the determinations. The results obtained for the described method for nitrite were statistically comparable to those obtained by the reference procedure. For the determination of nitrate, recovery tests confirmed that the sequential injection methodology provided good quality results.
The increase of nitrate in water bodies such as river, lakes, estuaries and the coastal zone, can lead to eutrophication, meaning excess of nutrients followed by algal blooms, oxygen depletion and fish deaths. An estuary being by definition an area where “sea water is measurably diluted with freshwater derived from land drainage”,4 serves as a boundary system between the fluvial environment and the coastal zone. Indeed, estuaries may exhibit typical environmental gradients with its own habitats and species making them an irreplaceable natural resource.
The monitoring of nitrites and nitrate in estuarine waters is an effective way to trace possible contamination sources. But estuarine waters are rather complex matrices as they change significantly with the proximity or distance from the sea end, both in terms of concentration values for many analytes and physicochemical properties such as temperature, pH and conductivity.
Sequential injection analysis has proved to be an effective tool for meeting the requirements of routine water analysis5 adding robustness and versatility to the other flow methodologies characteristics of automation, high throughput and reagent economy. The application of SI methodologies to the determination of inorganic nitrogen ionic forms has been quite successful. Most authors describe the spectrophotometric determination of nitrite based on a colorimetric reaction and then present either the concentration of nitrite,6–8 or the concentration of both nitrate and nitrite after reduction of nitrate.9,10
Another spectrophotometric approach was described by Mikuska,11 replacing the traditional colorimetric reaction for chemiluminescent detection. There were also a few works using electrochemical detection systems, such as potentiometry12 and voltammetry.13
Although most of the above mentioned works were applied to waters, only three were applied to saline waters, one applied to coastal waters6 and two applied to sea waters.8,10 There is no reference to any work applied to estuarine water samples.
In this research, a sequential injection methodology was developed for the determination of both forms of anionic nitrogen, nitrite and nitrate. This method is based on the Griess reaction between nitrite, sulfanilamide and N1NED which results in a coloured product with maximum absorbance at 543 nm.14 Therefore, the concentration of nitrite is obtained directly from its reaction with the colour reagent. As for nitrate, it is reduced to nitrite in a copperised cadmium column before the colorimetric reaction; the obtained result corresponds to the sum of nitrite and nitrate in the sample. The nitrate concentration in the sample is then obtained by the difference between the value obtained for nitrite plus reduced nitrate and the one obtained for nitrite.
The described methodology presents the novelty of using two aligned flow cells in the cell compartment (with different optical paths) in order to meet the requirements for the determination of each analyte. In fact, nitrite concentrations in estuarine waters are much lower than nitrate values, and with a narrower concentration range. So, in order to maximize sensitivity for nitrite, a 2 cm optical path flow cell was used. In order to obtain a wider concentration range for nitrate, and not requiring such a low limit of detection, a 1 cm optical path flow cell was used. With this arrangement, the two methodologies could be comprised in the same manifold and use the same spectrophotometer.
The overall conditions of the multiparametric methodology were studied to be applicable to estuarine waters. To assess if the methodology could cope with salinity gradients and other possible interfering agents, we analysed waters from estuaries of three rivers in NW Portugal with different constrains. River Ave watershed (1,400 km2) drains mainly agriculture and industrial areas, River Cávado (1,600 km2) crosses through disperse rural and urban areas and River Douro, which drains the largest watershed (98,000 km2) and presents a much higher freshwater discharge (>500 m3s−1), flowing into an estuary that suffers the pressure of over one million inhabitants. As a result, the samples collected from these rivers are thought to present different matrices.
A stock solution of ortho-phosphoric acid 5 mol L−1 was obtained from the concentrated acid (d = 1.71, 85%).
The colour reagent was monthly prepared by dissolving 5 g of sulfanilamide in 25 mL of 5 mol L−1ortho-phosphoric acid and by mixing with 0.5 g of N-(1-naphthyl)-ethylenediamine dihydrochloride (N1NED) dissolved in water; after homogenizing the mixture the volume was completed to 250 mL. A concentration of 20 g L−1 sulfanilamide and 2 g L−1 N1NED in 0.5 M of ortho-phosphoric acid was obtained in these conditions.
The conditioner of the cadmium column was monthly prepared by dissolving 0.2 g of di-sodium ethylenediamine tetraacetic acid (EDTA) and 10 g of ammonium chloride in 500 mL of water, to a final concentration of 0.4 g L−1EDTA and 20 g L−1 ammonium chloride. The pH of this solution was adjusted between 9–9.5 with commercial ammonia solution (d = 0.91, 25%).
A standard stock solution of sodium nitrite of 100 mmol L−1 was obtained by dissolving 700 mg of the dried solid in water to a final volume of 100 mL. This stock solution was diluted 100 times (1 mmol L−1) (monthly prepared) and from this solution, another dilution (1:
50) was carried out and this solution, prepared weekly, was 20 μmol L−1.
A standard stock solution of sodium nitrate of 100 mmol L−1 was obtained by dissolving 850 mg of the dried solid in 100 mL. From this stock solution, a dilution of 1:
100 was carried out and this solution (1 mmol L−1) was used for a month.
From the nitrite standard solution of 20 μmol L−1 and the nitrate standard solution of 1 mmol L−1, working standards were weekly prepared in the range of 0.50–8.00 μmol L−1 of nitrite and 12.5–300 μmol L−1nitrate.
The column tube, made of PTFE with 90 mm long and 3.2 mm inner diameter, was filled with the prepared granules. Ordinary domestic sponge was placed at both ends of the column to entrap the cadmium granules.
After preparing the column, it was washed with conditioner (NH4Cl-EDTA) and activated by passing a mixture of 25% nitrate solution and 75% conditioner. This mixture resulted from adding 25 mL of a 20 μmol L−1 of nitrate (obtained by proper dilution of the 1 mmol L−1 solution) to 75 mL of the conditioner solution.
For each river the water was collected from three different locations, one close to the ocean (location 1) and two others upstream (locations 2 and 3), the number 3 being the location with less influence from the ocean.
The estuarine water samples were introduced in the system either directly or after filtration. For the filtration, a hydrophilic membrane (Whatman, cellulose acetate) of pore size 0.45 μm was used.
![]() | ||
Fig. 1 Manifold for the spectrophotometric determination of nitrite and nitrate in estuarine waters: CR, colour reagent: sulfanilamide 20 g L−1 and N1NED 2 g L−1 in 0.5 M H3PO4; S, water sample or standard; Col., copperised cadmium column of 90 mm long and 3.2 mm width; EDTA, conditioner solution: EDTA 0.4 g L−1 and 20 g L−1 NH4Cl; W, waste; P, peristaltic pump; HC, holding coil 400 cm long; SV, eight port selection valve; Ri, reaction coils of 140 cm; λ, spectrophotometer (543 nm) with the two aligned cells. |
Solutions were propelled by a Gilson Minipuls 3 peristaltic pump with a PVC pumping tube. The pump was connected to the central channel of an eight port selection valve (Valco VICI 51652-E8). All tubing connecting the different components was made of Teflon from Omnifit, with 0.8 mm i.d.
A ThermoSpectromic Helios γ UV-Vis spectrophotometer set at the wavelength of 543 nm used as detection system. For the nitrite determination, the flow-cell was a Starna Brand 75.3 Q (20 mm light path, 40 μL inner volume), and for nitrate determination a Hellma 178.711-QS flow-cell (10 mm light path, 30 μL inner volume). The two flow cells were aligned in the same cell support of the spectrophotometer.
Analytical signals were recorded in a Metrohm E 586 Labograph strip recorder. A personal computer (Samsung SD 700) equipped with a PCL818L interface card, running with a homemade software written in QuickBasic 4.5, controlled the selection valve (SV) position and the pump sense and speed.
The sequence of steps with the respective time and volumes for the determination of nitrite and nitrate is shown in Table 1.
Step | SV position | Time/s | Pump speed | Pump direction | Volume/μL | Description |
---|---|---|---|---|---|---|
A | 4 | 4.5 | 30 | a | 214 | Aspiration of conditioner (EDTA) |
B | 2 | 2.5 | 30 | a | 119 | Aspiration of sample/standard |
C | 4 | 10 | 10 | b | 148 | Propelling to column for reduction |
D | 8 | 4 | 40 | b | 243 | Propelling to waste to wash holding coil |
E | 1 | 9 | 40 | a | 548 | Aspiration of colour reagent |
F | 2 | 9.2 | 40 | a | 560 | Aspiration of sample/standard |
G | 3 | 55 | 40 | b | 3348 | Propelling to detector, mixture of reagent and sample and measurement (nitrite determination) |
H | 4 | 3 | 10 | a | 44 | Aspiration for preparation of the port |
I | 8 | 2 | 40 | b | 122 | Propelling to waste to wash holding coil |
J | 1 | 5 | 40 | a | 304 | Aspiration of colour reagent |
L | 4 | 3.4 | 20 | a | 102 | Aspiration of reduced sample/standard |
M | 5 | 45 | 40 | b | 2738 | Propelling to detector, mixture of reagent and sample and measurement (nitrite + nitrate determination) |
N | 4 | 5 | 30 | a | 238 | Aspiration of conditioner (EDTA) to wash the column |
O | 8 | 5 | 40 | b | 304 | Propelling to waste to wash holding coil |
The first step (step A) is the aspiration of conditioner, through the cadmium column, followed by the aspiration of sample (step B). This way the sample is dispersed in the conditioner. The mixing is promoted by reversing the flow while sending the stacked plugs through the column, for nitrate reduction (step C).
After washing the holding coil, the determination of nitrite is carried out during the nitrate reduction process (steps D to G). For the nitrite determination, the sample is aspirated after the colour reagent, and then the flow is reversed, promoting the mixture on the way to the detector.
After the determination of nitrite and the preparation of the port of the cadmium column (steps H and I), the determination of nitrite + nitrate is carried out (steps J to M). The colour reagent is aspirated, followed by the sample from the column. Then this mixture is sent to the detector.
The coloured product formed corresponds to the sum of the nitrite and the nitrate in the sample. The concentration of nitrate is obtained by the difference between both determinations. The last two steps (N and O) of the protocol sequence aim to prepare the column for a new cycle.
![]() | ||
Fig. 2 Study of the aspiration volumes of the sample (◊) and of the reagent (□); the points in black represent the chosen values. |
The initial preparation of the colour reagent was done accordingly to the reference procedure,14 Reagent 1 in Table 2. The colour reagent was composed by different components: sulfanilamide, N1NED and phosphoric acid, the first two being directly involved in the colorimetric reaction.
Reagent | [Sulfanilamide]/g L−1 | [N1NED]/g L−1 | [H3PO4]/M | Calibration curve |
---|---|---|---|---|
1 | 10.0 | 1.0 | 1.5 | A = 0.0281 [NO2−] + 0.0461; R2 = 0.9933 |
2 | 20.0 | 2.0 | 1.5 | A = 0.0307 [NO2−] + 0.0732; R2 = 0.9999 |
3 | 30.0 | 3.0 | 1.5 | A = 0.0282 [NO2−] + 0.1012; R2 = 0.9752 |
So, the study of the concentration was carried out maintaining the proportion between sulfanilamide and N1NED of the reference procedure,14 Reagent 2 and 3 in Table 2.
As the Reagent 2 showed, not only a better sensitivity, but also a better linearity, those were the concentrations chosen for sulfanilamide and N1NED.
The ortho-phosphoric acid not only ensures an acidic pH for the reaction but also contributes to dissolve the sulfanilamide; so, in order to study its minimisation, a batch procedure was used. To assess the minimal quantity of H3PO4 required for the dissolution, 0.5 g of sulfanilamide were weighed (for a final volume of 25 mL of colour reagent) and increments of 500 μL of acid were made until complete dissolution. In the end, 2.5 mL of acid were added resulting in an acid concentration of 0.5 M. A calibration curve was made to ensure that the sensitivity had not been affected. Actually, no significant change was observed, the slope had even increased a little (A = 0.0350 [NO2−] + 0.0181). Aiming to further increase the sensitivity and to lower the quantification limit, the flow cell of 1 cm light path was replaced by a flow cell of 2 cm light path and, as it could be expected, the sensitivity doubled.
Before incorporating the nitrate determination in the developed system, water samples were analysed, in order to evaluate if there was salinity interference in these conditions. The water samples were analysed by the developed sequential injection methodology and the results compared to the reference procedure.14
To evaluate accuracy, a linear relationship between the concentrations obtained with the developed sequential injection methodology (CSIA) and with the reference procedure (CRef.Met.) was established. The equation found was: CSIA = 1.05 (±0.021) × CRef.Met. − 3.5 × 10−2 (±0.052); R2 = 9.99 × 10−1 where the values in parenthesis represent 95% confidence limits. These figures showed that the estimated slope and intercept do not differ statistically from values 1 and 0 respectively. Therefore, there was no evidence for systematic differences between the two set of results.15
The linear regression was assessed by residuals analysis. Independence, randomness and normality was validated. The quality of the regression was also confirmed by the coefficient of determination above 0.999.
After these studies, two calibration curves were traced, one with nitrite standards and one with nitrate standards with the same concentrations. The slopes were quite similar (slope difference < 2%) as well as the intercepts (difference < 7%). An average of 100.4% (standard deviation 3.2) conversion percentage was obtained for the calibration standards.
Having established the conditions for the reduction process, the study of aspiration volumes for both the reduced sample and colour reagent was carried out. The volume of colour reagent was set to obtain the maximum sensitivity within the expected dynamic range. The study of the volume of the sample to be aspirated from the column was made together with the study of the volume of sample sent to the column. A range from 60 to 115 μL was studied and, as the sensitivity increased up to the volume of 105 μL, this was the chosen value.
Concerning the determination of nitrate, two extra steps were included aiming for the preparation of the column for the next cycle and wash of the system. Firstly, the conditioner is aspirated through the column to remove residues of the previous sample (and leaving it filled with conditioner), and then (final step) the holding coil was washed preparing the system for the next cycle.
This way, the column was never left either empty or filled with water, therefore enabling to maintain a good percentage of conversion. Nevertheless, after some time that percentage decreased. When a percentage of conversion below 95% occurred, the cadmium was removed from the column and the procedures of washing, copperising and conditioning repeated before repacking into the column again.
The possibility of having mixed standards of nitrite and nitrate was tested. As the results showed that there was no significant difference in the slopes (<2.8%) between the calibrations with single standards and mixed standards, mixed standards were used.
As two different flow cells were used in the combined arrangement, two ports of the selection valve were connected to the spectrophotometer. The optical path of these two cells were aligned in the cell compartment. As the two measurements, corresponding to each determination, are sequential there was no mutual signal interference. Firstly, the coloured product formed by nitrite reaction flows through the 2 cm flow cell. When the product of the nitrite + nitrate reaction is sent to the 1 cm flow cell, the 2 cm flow cell is filled with carrier (deionised water).
The possible salinity effect was assessed by comparing calibration curves using standard solutions with different salinities; these solutions were prepared by adding sodium chloride to the standards. The salinity values were adjusted to an intermediate and a maximum level of salinity in estuarine waters, 9 and 19 respectively. The calibration curve with pure standards was compared to the calibration curve with standards with salinity 9. The estimated slopes of both curves were not statistically different (assessed by the confidence intervals at 95%, which overlapped). The same was verified when the pure standards calibration curve was compared with the calibration curve with salinity 19, and for both nitrite and nitrate calibration curves.
A global linear regression was carried out using all data sets (a total of 116 experimental points) related to all calibration curves. The estimated linear model obtained was the following: A = 5.06 × 10−2 ± 1.08 × 10−3 [NO2−] −7.42 × 10−3 ± 4.94 × 10−3. The model parameters were similar to the ones estimated when only the pure standards data values were used (80 experimental points): A = 5.06 × 10−2 ± 1.02 × 10−3 [NO2−] −1.24 × 10−4 ± 4.68 × 10−3. This was also assessed by the magnitude of the confidence intervals of the slope and the intercept at 95% presented in the previous equations (i.e. ± half of the confidence interval at 95%). For the nitrate, the results obtained showed that calibration curves with different salinities were also identical. The estimated calibration curve was A = 3.15 × 10−3 ± 6.46 × 10−5 [NO3−] −7.24 × 10−2 ± 1.02 × 10−2 (global data with 116 data points) and A = 3.17 × 10−2 ± 7.57 × 10−5 [NO3−] −8.38 × 10−2 ± 1.19 × 10−2 (pure standards, 80 points).
As mentioned in Section 3.1, the quality of the regression was tested by residuals analysis (i.e. randomness and normality) and by the coefficient of determination (i.e.R2, which was above 0.987 in all cases). Overall it can be concluded that the salt addition did not affect the calibration curves.
Analyte | Dynamic concentration range | LOD | LOQ | % RSD, conc. ± standard deviation |
---|---|---|---|---|
Nitrite | 0.50–8.00 μM | 0.11 μM | 0.36 μM | 4.1%, 0.78 ± 0.04 μM (38.2 ± 1.6 μg L−1) |
(23–400 μg L−1) | (5.4 μg L−1) | (18 μg L−1) | 1.1%, 5.88 ± 0.06μM (284 ± 3 μg L−1) | |
Nitrate | 12.5–305 μM | 3.7 μM | 12.2 μM | 3.1%, 54.2 ± 1.7μM (3.5 ± 0.1 mg L−1) |
(0.80–20 mg L−1) | (0.2 mg L−1) | (0.8 mg L−1) | 0.8%, 304 ± 2μM (19.4 ± 0.1 mg L−1) |
The LOD and LOQ were calculated as the concentration corresponding to three and ten times the standard deviation of ten consecutive injection of deionised water, respectively, according to IUPAC recommendation.15
As mentioned in Section 3.1., linear regressions were assessed by residuals analysis (randomness, independence and normality verified) and the quality of the regression confirmed by the coefficient of determination.
The sample frequency was calculated based on the time spent per cycle. A complete analytical cycle took about 3.6 min for both nitrite and nitrate determination, including the reduction in the cadmium column, resulting in a sample throughput of 33 determinations per hour, including the two parameters. A complete analytical cycle takes into account not only the times in the protocol sequence but also the time required to change valve position and pump direction.
An analytical cycle corresponds to a sample consumption of 680 μL per determination of nitrite and nitrate, along with an overall reagent consumption of: 17 mg sulfanilamide; 1.7 mg N1NED; 41 mg phosphoric acid, 181 μg EDTA, 9.0 mg ammonium chloride and an effluent production of about 6.8 mL.
The conversion percentage was assessed daily for two weeks resulting in an average of 101% (standard deviation of 2.1). A statistical test (t-test) was used to evaluate if that value was not significantly different from 100%,15 which was proved as the calculated t-value was 1.346 with a correspondent critical value 3.163 for a 95% significance level.
The water samples were analysed by the developed sequential injection (SIA) system. For the nitrite determination, the samples were also analysed by the reference procedure14 (Ref. Met.) and the results obtained are presented in Table 4.
River Estuary | Sample ID | pH (25 °C) | G/μS cm−1 | Salinity | Ref.Met./μM | SIA/μM | RD (%) |
---|---|---|---|---|---|---|---|
Douro | RD1 | 8.08 | 23300 | 15.2 | 0.470 ± 0.014 | 0.403 ± 0.020 | −14.3 |
RD2N | 7.56 | 9230 | 5.8 | 0.791 ± 0.014 | 0.701 ± 0.084 | −11.4 | |
RD2F | 7.72 | 5480 | 3.3 | 0.446 ± 0.037 | 0.488 ± 0.042 | 9.3 | |
RD3F | 8.67 | 6540 | 4.0 | 1.096 ± 0.024 | 1.144 ± 0.020 | 4.3 | |
RD3 | 9.17 | 1710 | <2 | 0.739 ± 0.085 | 0.709 ± 0.030 | −4.0 | |
Cávado | RC1 | 7.23 | 7280 | 4.5 | 0.405 ± 0.145 | 0.403 ± 0.035 | −0.6 |
RC2F | 6.97 | 2330 | <2 | 1.353 ± 0.000 | 1.474 ± 0.012 | 9.0 | |
RC2N | 7.11 | 2250 | <2 | 1.072 ± 0.014 | 1.195 ± 0.012 | 11.5 | |
RC3F | 7.11 | 695 | <2 | 1.586 ± 0.028 | 1.601 ± 0.014 | 0.9 | |
RC3N | 6.92 | 765 | <2 | 1.707 ± 0.014 | 1.854 ± 0.023 | 8.6 | |
Ave | RA3 | 8.38 | 607 | <2 | 5.064 ± 0.000 | 4.908 ± 0.000 | −3.1 |
To evaluate accuracy, a linear relationship between CSIA (μM) and CRef. Met. (μM) was established; the equation found was: CSIA = 0.971 (±0.0504) × CMet.Ref. + 0.0614 (±0.0900), where the values in parenthesis are 95% confidence limits. These figures show that the estimated slope and intercept do not differ statistically from values 1 and 0, respectively. Therefore, there is no evidence for systematic differences between the two sets of results.15 As mentioned in Section 3.1., the linear regression was assessed by residuals analysis and the quality of the regression confirmed by the coefficient of determination.
The nitrate determination results from the reduction to nitrite prior to determination of total nitrite and subtraction of the nitrite value. For the nitrate determination, recovery studies were carried out using the estuarine waters, the results are presented in Table 5.
Sample ID | pH (25 °C) | G/μS cm−1 | Salinity | Initial | Added | Found | Recovery (%) | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
Conc./μM | SD | RSD % | Conc./μM | Conc./μM | SD | RSD % | |||||
RD1 | 8.16 | 24300 | 16.5 | 57.5 | 1.1 | 2.0 | 100 | 163 | 1.8 | 1.1 | 105.5 |
57.5 | 1.1 | 2.0 | 200 | 263 | 3.7 | 1.4 | 102.8 | ||||
RD1F | 8.18 | 24500 | 16.7 | 38.3 | 0.3 | 0.8 | 100 | 138 | 1 | 1.0 | 99.7 |
38.3 | 0.3 | 0.8 | 200 | 236 | 13 | 5.7 | 98.9 | ||||
RD2 | 8.37 | 23100 | 15.6 | 55.7 | 0.4 | 0.7 | 100 | 160 | 1.6 | 1.0 | 104.3 |
55.7 | 0.4 | 0.7 | 200 | 256 | 3.6 | 1.4 | 100.1 | ||||
RD2F | 8.34 | 21700 | 14.6 | 44.1 | 1.8 | 4.1 | 100 | 150 | 3.9 | 2.6 | 105.9 |
44.1 | 1.8 | 4.1 | 200 | 253 | 2.6 | 1.0 | 104.4 | ||||
RD3 | 8.03 | 1087 | < 2 | 64.7 | 0.1 | 0.2 | 100 | 158 | 2 | 1.1 | 93.9 |
64.7 | 0.1 | 0.2 | 200 | 253 | 8 | 3.1 | 94.1 | ||||
RD3F | 8.12 | 938 | < 2 | 59.1 | 1.7 | 2.9 | 100 | 158 | 14 | 8.9 | 98.9 |
59.1 | 1.7 | 2.9 | 200 | 267 | 12 | 4.4 | 104.0 |
For spiking the samples, volumes of 100 μL and 200 μL of nitrate standard solution (10 mM) were added to 10 mL of sample. The calculation of the recovery percentage was made according to IUPAC.16
The SIA methodology provided recovery ratios with an average of 103% (standard deviation 6.2) and a statistical test (t-test) was used to evaluate if that mean recovery value did not significantly differ from 100%.15 Results showed that for a 95% significance level the recovery values did not differ from 100% as the calculated t-value was 1.673 with a correspondent critical value 2.593, thus indicating the absence of multiplicative matrix interference.
After the validation of the developed methodology, the values of nitrite and nitrate of the tested estuaries were assessed. Samples were collected on two different dates for each estuary. The values of nitrite and nitrate as well as the sample characteristics according to the different rivers and locations are summarised in Table 6.
River Estuary | Date | Location | pH (20 °C) | G/μS cm−1 | Salinity | [NO2−]/μM ± SD | [NO3−]/μM ± SD |
---|---|---|---|---|---|---|---|
Douro | 07-02-2008 | 1 | 7.88 | 17870 | 11.8 | 0.403 ± 0.020 | 14.4 ± 2.1 |
2 | 7.56 | 9230 | 5.8 | 0.488 ± 0.084 | 58.7 ± 2.1 | ||
3 | 8.49 | 2010 | <2 | 0.709 ± 0.030 | 40.8 ± 2.0 | ||
20-04-2008 | 1 | 8.04 | 14160 | 9.2 | 0.828 ± 0.011 | 70.9 ± 0.6 | |
2 | 7.72 | 5480 | 3.3 | 0.476 ± 0.011 | 71.9 ± 0.7 | ||
3 | 8.67 | 6540 | 4.0 | 0.391 ± 0.020 | 71.4 ± 1.1 | ||
Cávado | 10-02-2008 | 1 | 8.14 | 20500 | 13.7 | 0.663 ± 0.020 | 54.4 ± 0.5 |
2 | 7.11 | 2250 | <2 | 1.19 ± 0.01 | 65.4 ± 3.1 | ||
3 | 6.92 | 765 | <2 | 1.85 ± 0.02 | 111 ± 2 | ||
06-05-2008 | 1 | 7.63 | 17860 | 11.8 | 0.333 ± 0.020 | 24.3 ± 0.5 | |
2 | 6.97 | 2330 | <2 | 1.03 ± 0.01 | 87.9 ± 0.6 | ||
3 | 7.11 | 695 | <2 | 1.15 ± 0.02 | 76.9 ± 1.5 | ||
Ave | 31-01-2008 | 3 | 8.83 | 334 | <2 | 5.55 ± 0.04 | 211 ± 1 |
05-05-2008 | 3 | 8.38 | 607 | <2 | 3.35 ± 0.04 | 211 ± 1 |
As expected, it was observed that the highest values of both anions were in the estuary of river Ave as it crosses a rural area. Overall, the lowest values were obtained for river Douro, being a dense urban area but with the higher discharge rate.
The methodology was applied to natural samples, from three estuaries in NW Portugal, and the results proved its efficiency. The different composition of the samples did not interfere in the determination as the developed method proved to be quite robust for a wide salinity range, from <2 to 16.7, leaving the possibility for real time measurement.
When compared to previously reported colorimetric flow systems for these determinations, it presents the advantage of direct introduction of estuarine waters, with no need for prior treatments. Other advantages are a low reagent consumption and effluent production. Additionally, as this methodology provides analysis on real time, it could be used for early detection of pollution by fertilizers in freshwater streams and estuaries.