Silvia García-Ruiz, Mariella Moldovan and J. Ignacio García Alonso*
Department of Physical and Analytical Chemistry, University of Oviedo, Julián Clavería 8, 33006, Oviedo, Spain. E-mail: jiga@uniovi.es; Fax: +34-985103125; Tel: +34-985103484
First published on 17th August 2007
Large volume injection ion chromatography has been evaluated for the precise and accurate measurement of strontium isotope ratios by multicollector ICP-MS. On-line coupling of the ion chromatograph to the MC-ICP-MS instrument allowed the complete separation of isobaric rubidium and other matrix elements (Ca, Mg, Na and K) and the measurement of strontium isotope ratios on flat-topped chromatographic peaks of ca. 3 minutes duration. No drift in the strontium isotope ratios was observed during strontium elution and mass bias could be corrected using either the 86Sr/88Sr isotope ratio (internal correction) or the observed 87Rb/85Rb isotope ratio (external correction) in the real samples. Typical internal precisions obtained with the selected measurement and correction conditions were about 2 times lower than the precisions obtained with continuous sample introduction for a solution 50 ng g–1 of Sr NIST SRM 987 in HNO3 2% (typical standard deviations of 0.00022–0.00036 for the corrected 87Sr/86Sr ratio in comparison to 0.00014–0.00020 for continuous sample introduction). The method was applied to the measurement of strontium isotope ratios in cider samples and the results agreed with previous data obtained after an off-line Rb–Sr separation procedure using the Sr-Spec chelating material. The precision obtained for the real samples was poorer than previous data due to the low concentration of strontium in the measured samples. The method was also applied to other sample types (soil leachates, apple leaves and apples) to study the transference of strontium isotope ratios in the soil–tree–cider system.
High precision strontium isotope ratio measurements have been routinely achieved with thermal ionisation mass spectrometry (TIMS) which can be considered as the reference technique with reported values of 0.002% RSD for 87Sr/86Sr isotope ratios.8–10 Recently, multicollector (MC) ICP-MS instruments have also provided high precision strontium isotope ratio measurements (0.002% RSD)11–13 with improved sample throughput in comparison to TIMS. For the accurate determination of Sr isotope ratios by TIMS or MC-ICP-MS, an effective separation of Rb from Sr is necessary before isotope ratio measurement in order to eliminate the isobaric interference of 87Rb on 87Sr. Additionally, for MC-ICP-MS instruments, matrix elements may interfere with the measurements by modifying the mass discrimination factor used for correction, particularly when external mass bias correction is used. In this sense, most pre-treatment methods employed traditionally for TIMS strontium isotope ratio measurements are now applied for MC-ICP-MS.14–17 However, all these off-line separations need to be carried out under ultra-clean conditions in order to minimise sample contamination risks.
The on-line coupling of a chromatographic separation with a multicollector ICP-MS instrument could overcome some of the problems of the off-line separations, providing additional advantages such as improved sample throughput and the use of alternative mass bias correction procedures instead of the traditional 86Sr/88Sr internal correction. This coupling has been evaluated by different authors using either gas chromatography18–20 or liquid chromatography.21,22 The results obtained for short transient signals (typically from 3 s for GC up to about 60 s for HPLC) using MC-ICP-MS detection showed a drift in the isotope ratios during peak elution18,20–22 together with a worsening of the isotope ratio precision up to a factor of ten21 in comparison to continuous sample introduction. This isotope ratio drift was reported to be due to a slow response of the Faraday cup amplifiers used in current MC-ICP-MS instruments18,22,23 or to isotopic fractionation effects during chromatographic separations20 or thermal desorption processes.24 Hence, it would be advisable to develop chromatographic procedures which showed no isotopic fractionation effects and eliminate the need for fast response in the Faraday cup amplifiers. Recent work in our laboratory25 on the on-line separation of Rb and Sr using ion chromatography showed that flat-topped chromatographic peaks of several minutes could be obtained using large volume injection (3 mL) and a mobile phase containing 900 mM nitric acid in the presence of 5 mM 18-crown-6 ether. The chromatographic method was optimised using a quadrupole ICP-MS instrument and showed that Rb and other matrix elements (such as Ca, Mg, K and Na) eluted at the dead volume, while Sr was strongly retained and eluted as a flat-topped chromatographic peak. From the developed chromatographic method,25 it was clear that isotope fractionation effects would be absent and no fast response amplifiers would be required at the flat top of the peak.
Therefore, the aim of the present work was the evaluation of this procedure for the measurement of Sr isotope ratios using a MC-ICP-MS instrument. Evaluated samples included ciders with different geographical origin, vegetal materials and associated soil extracts within the framework of an origin authentication project using Sr isotope ratio measurements.
On-line Sr isotope ratio determinations were carried out by coupling the chromatographic system to a Neptune high resolution MC-ICP-MS (Thermo Fisher Scientific, Bremen, Germany), with a forward Nier–Johnson geometry and equipped with 9 Faraday collectors along the focal plane, eight movable cups and one fixed centre channel. HPLC coupling was made through a 0.3 mm id PFA tube connected to the sample introduction system consisting of a PFA-100 microconcentric nebuliser and a cyclonic spray chamber. Sampler and skimmer Ni cones were employed. Lens settings were optimised daily for maximum analyte sensitivity. Collector positions were adjusted for peak overlap of signals in L2 (84Sr, 84Kr), L1 (85Rb), C (86Sr, 86Kr), H1 (87Sr, 87Rb) and H2 (88Sr) cups. Instrumental operating conditions and data acquisition parameters are shown in Table 1. Electronic baseline was measured before each block of measurements.
DX-120 Chromatographic parameters | |||
---|---|---|---|
Column | Dionex Ion Pac CS2 | ||
Eluent | 900 mM HNO3, 5 mM 18-crown-6 ether | ||
Mode | isocratic | ||
Flow rate | 1 mL min–1 | ||
Sample injection volume | 3 mL |
Neptune MC-ICP-MS parameters | |||
---|---|---|---|
Instrument settings | |||
Rf power | 1200 W | ||
Cool gas flow | 14 L min–1 Ar | ||
Auxiliary gas flow | 0.9 L min–1 Ar | ||
Sample gas flow | 0.935 mL min–1 Ar | ||
Sampler and skimmer cones | Ni | ||
Data acquisition parameters | |||
Collection mode | Static | ||
Cup configuration | L4 | 82 | Kr |
L3 | 83 | Kr | |
L2 | 84 | Sr/Kr | |
L1 | 85 | Rb | |
C | 86 | Sr/Kr | |
H1 | 87 | Sr/Rb | |
H2 | 88 | Sr | |
Resolution mode | low | ||
Continuous sample introduction | |||
Acquisition method | 5 blocks, 10 cycles, 4.194 s integration, 3 s idle | ||
HPLC coupling | |||
Acquisition method | 1 block, 280 cycles, 4.194 s integration, 0.5 s idle |
Off-line Sr isotope ratio measurements on cider samples were carried out using a VG Axiom MC-ICP-MS (Thermo-Elemental, Winsford, England) according to the procedure described previously.7
The chromatographic eluent was freshly prepared using subboiled nitric acid and puriss. p.a. grade 18-crown-6 ether (Fluka, Buchs, Switzerland) and degassed with helium prior its use. PVDF (polyvinylidene fluoride) 0.45 µm syringe filters (Teknokroma, Barcelona, Spain) were used for sample filtration.
Subboiled nitric acid, suprapur hydrogen peroxide 30% (v/v) and suprapur hydrofluoric acid 48% (v/v) were employed for the apple and apple leaves digestion using a microwave digestion apparatus Milestone mls 1200 mega (Milestone, Leutkirch, Germany). Extraction of soils was performed using subboiled nitric acid or analytical grade ammonium nitrate. All these reagents for sample digestion or extraction were provided by Merck.
Soil samples were air-dried and sieved (<300 µm). Extraction was performed by shaking, for 24 hours, 1 g of soil sample and 10 g of 1 M NH4NO3 or 0.2 M HNO3. The ammonium nitrate extract represented the mobile fraction, which includes water soluble and non-specifically adsorbed metals and easily soluble metallo-organic complexes. The extracts were further diluted (1 : 10) and the concentration of nitric acid and 18-crown-6 was adjusted to match that in the eluent.
Procedural blanks were evaluated following all the sample measurement process but using ultra-pure water instead of the sample.
In spite of improved precision of multicollector ICP-MS compared to scanning mass spectrometers for short transient signals, current procedures published for the on-line coupling of chromatographic separations to MC-ICP-MS instruments showed that internal precision worsened up to a factor of ten21 in comparison to steady-state signals and a drift on isotope ratio measurements could be observed during peak elution. In order to overcome these problems, we have developed an IC-MC-ICP-MS procedure in which the large volume injection (3 mL) led to efficient Rb and Sr separation with flat-topped peaks of about 3 min of stable plateau (for both Rb and Sr)25. As an example, Fig. 1 shows the chromatogram obtained for the separation of a Rb standard (Merck) and the Sr NIST SRM 987. Under the chromatographic conditions used, Rb is practically unretained and elutes at the dead volume whereas Sr elutes at a retention time between 8–14 min. As can be observed in the chromatogram, no dilution of Rb or Sr in the mobile phase occurs, in contrast to traditional LC separations. Other matrix elements, such as Ca, Mg, Na and K, also elute at the dead volume showing an efficient separation from Sr. Flat-topped peaks provided a steady-state signal on the plateau that allowed isotope ratio measurements (point to point) during ca. 3 minutes. Under these conditions, only data points corresponding to signal intensities higher than 90% of the maximum intensity were considered, so that between 20 and 122 strontium isotope ratios can be measured on the signal plateau depending on the integration and idle time used. Therefore, isotope ratios for a single injection were determined as the mean obtained from all data points on the plateau of the flat-topped peak. The internal precision was calculated, similarly for continuous sample introduction, as the standard deviation of the isotope ratio values for each point on the plateau of the flat-topped peak within the same chromatographic run. Additionally, external precision can be evaluated by repeated injections of the same sample.
![]() | ||
Fig. 1 Separation of a standard solution containing 50 ng g–1 of Rb (Merck) and 50 ng g–1 of Sr NIST SRM 987 by IC-MC-ICP-MS under the chromatographic conditions shown in Table 1. |
(a) Rb baseline correction.
![]() | (1) |
(b) Rb tail correction.
87Sr = S87 – S85R87/85Rb | (2) |
![]() | ||
Fig. 2 87Sr/86Sr isotope ratios across the 50 ng g–1 Sr NIST SRM 987 transient signal separated from Rb by IC-MC-ICP-MS: experimental data (filled diamond); corrected for baseline (open triangles); corrected for baseline and Rb interference (open diamond); and, corrected for baseline, Rb interference and mass bias (open circles). The Sr elution is represented by the 88Sr intensity signal. Only intensity data above 90% of the maximum at mass 88 were used for isotope ratio measurements. |
(a) Determination of the mass bias factor.
Mass bias per mass unit, K, was calculated for internal mass bias correction according to the exponential model equation:
![]() | (3) |
(b) Correction of mass bias on the strontium isotope ratios.
The mass bias corrected 87Sr/86Sr isotope ratio, R87/86Sr,cor, was then calculated as:
R87/86Sr,cor = R87/86Sr,meas × exp(K × ΔmSr87) | (4) |
Additionally, the on-line coupling of the chromatographic separation to the MC-ICP-MS instrument allowed the measurement of all three Sr isotope ratios and the 87Rb/85Rb isotope ratio in the sample within the same chromatographic run. The measured 87Rb/85Rb isotope ratio could be used to calculate the mass discrimination factor, acting as a reference value for external mass bias correction of Sr isotope ratios assuming that the variability of the 85Rb/87Rb value in nature (87Rb/85Rb = 0.38562 ± 0.00017, obtained from the IUPAC Table 2) is negligible. Recently, Nebel et al.32 showed that no natural variations in the Rb isotope composition of terrestrial samples could be observed within ±0.5 δ87Rb. Therefore, this alternative procedure will be also evaluated for the real samples.
Integration time/s | 0.1 s idle time | 3 s idle time | ||||
---|---|---|---|---|---|---|
87Sr/86Sr | sa | nb | 87Sr/86Sr | sa | nb | |
a Standard deviation of the corrected 87Sr/86Sr isotope ratios across the flat-topped peak.b Number of acquisition points considered for the calculation of the Sr isotope ratios (intensities for 88Sr > 90% of maximum). | ||||||
0.131 | — | — | — | 0.71047 ± 0.00053 | 0.00167 | 39 |
1.049 | 0.71036 ± 0.00013 | 0.00071 | 122 | 0.71034 ± 0.00032 | 0.00094 | 35 |
2.097 | 0.71033 ± 0.00014 | 0.00057 | 69 | 0.71060 ± 0.00018 | 0.00047 | 28 |
4.194 | 0.71044 ± 0.00011 | 0.00030 | 30 | 0.71045 ± 0.00018 | 0.00041 | 20 |
Although precision of Sr isotope ratios was dependent on analyte concentration, concentrations of 50 ng g–1 Rb (Merck) and 50 ng g–1 Sr NIST SRM 987 were selected as representative of typical contents of these elements in the evaluated samples. Strontium isotope ratios and internal precision were evaluated by on-line IC-MC-ICP-MS for integration times of 0.13, 1, 2.1 and 4.2 s, and for idle times of 0.1 and 3 s, the minimum and recommended value for dynamic measurements, respectively, with a standard solution containing Rb (Merck) and Sr NIST SRM 987. The idle time affects the total acquisition time of the method and the number of points per acquisition run, even though it is not critical for static measurements by MC-ICP-MS, as is the case of our measurements. Table 2 shows the obtained results. As can be observed, after spectral interferences and internal mass bias correction using 88Sr/86Sr, all 87Sr/86Sr isotope ratio values and their uncertainties overlapped with the certified interval given in the certificate of the reference material NIST SRM 987 (87Sr/86Sr = 0.71034 ± 0.00026, 95% confidence). However, internal precision for the 87Sr/86Sr isotope ratio was improved for longer integration times, between 0.13 and 4.2 s, and at constant idle time. As reported in Table 2, the standard deviation (s) of data points on the plateau of the flat-topped peak decreased for increasing integration times, as well as the standard deviation on the mean (2s/√n) even though the reduced number of data points. For a fixed integration time, standard deviation was similar for both (0.1 and 3 s) idle times, although standard deviation on the mean was lower for 0.1 s idle since more data points were obtained. Nevertheless, signal fluctuations were observed for the short idle time of 0.1 s which affected the correction procedures applied. This effect was not observed at 0.5 s idle time, so an integration time of 4.2 s and idle time of 0.5 s were selected for precise isotope ratio measurements and stable signal in the plateau of the flat-topped peak.
Under the selected data acquisition parameters, between 25 and 30 data points, which showed intensities above 90% of the maximum, were obtained in all replicate chromatograms measured for the standard solution containing Rb and Sr NIST SRM 987. Typical internal precisions obtained with the selected measurement and correction conditions were about 2 times lower than the precisions obtained with continuous sample introduction for a solution of the same concentration of Sr NIST SRM 987 in HNO3 2% (typical standard deviations of 0.00022–0.00036 for the corrected 87Sr/86Sr ratio in comparison to 0.00014–0.00020 for continuous sample introduction). It should be noted that the integration time used for steady-state signals from continuous sample introduction was 4.2 s integration and 3 s idle times, measuring 50 data points per acquisition run.
Uncertainty (s) | 0.000029 | 0.000049 | 0.000032 | 0.000035 | 0.000058 | 0.000070 | 0.000228 | 0.00185 | 0.00162 | |
---|---|---|---|---|---|---|---|---|---|---|
Mean value | 84B | 85B | 86B | 87B | 88B | 87Rb/85Rb | 87Sr/86Sr | 88Sr/86Sr | 88Sr/86Sr ref | |
84B (V) | 0.000105 | 0.000134 | 0.000105 | 0.000105 | 0.000105 | 0.000105 | 0.000105 | 0.000105 | 0.000105 | 0.000105 |
85B (V) | 0.004717 | 0.004717 | 0.004766 | 0.004717 | 0.004717 | 0.004717 | 0.004717 | 0.004717 | 0.004717 | 0.004717 |
86B (V) | 0.000591 | 0.000591 | 0.000591 | 0.000623 | 0.000591 | 0.000591 | 0.000591 | 0.000591 | 0.000591 | 0.000591 |
87B (V) | 0.002238 | 0.002238 | 0.002238 | 0.002238 | 0.002273 | 0.002238 | 0.002238 | 0.002238 | 0.002238 | 0.002238 |
88B (V) | 0.004906 | 0.004906 | 0.004906 | 0.004906 | 0.004906 | 0.004963 | 0.004906 | 0.004906 | 0.004906 | 0.004906 |
87Rb/85Rb | 0.403374 | 0.403374 | 0.403374 | 0.403374 | 0.403374 | 0.403374 | 0.403444 | 0.403374 | 0.403374 | 0.403374 |
87Sr/86Sr | 0.726030 | 0.726030 | 0.726030 | 0.726030 | 0.726030 | 0.726030 | 0.726030 | 0.726258 | 0.726030 | 0.726030 |
88Sr/86Sr | 8.74701 | 8.74701 | 8.74701 | 8.74701 | 8.74701 | 8.74701 | 8.74701 | 8.74701 | 8.74886 | 8.74701 |
88Sr/86Sr ref | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.37861 | 8.38023 |
87Sr/86Sr cor | 0.71058 | 0.710576 | 0.710706 | 0.710652 | 0.710352 | 0.710592 | 0.710575 | 0.710805 | 0.710501 | 0.710645 |
UR | 0.00037 | 0.00 | 1.67 × 10–8 | 5.66 × 10–9 | 5.04 × 10–8 | 2.44 × 10–10 | 3.00 × 10–12 | 5.19 × 10–8 | 5.67 × 10–9 | 4.75 × 10–9 |
Contribution (%) | 0.0 | 12.4 | 4.2 | 37.2 | 0.2 | 0.0 | 38.4 | 4.2 | 3.5 |
![]() | (5) |
In the first experiments, using the Merck Rb standard and the NIST SRM 987, Sr isotope ratio values obtained by external mass bias correction were significantly lower than those obtained by internal mass bias correction and well below the certified interval given for the reference material NIST SRM 987. However, when cider samples were injected, the values obtained for 87Sr/86Sr using the internal correction were in agreement with those values obtained using external Rb correction. This disagreement was initially attributed to isotopic effects in the Merck standard which would be absent for the real samples. As no Rb isotopically certified standard was available, it was decided to use the NIST SRM 987 as reference to calculate the isotopic composition of different Rb standards present in the laboratory. In this way, two different lots of Merck certified Rb standard, a solid RbNO3 standard, and a CPI certified Rb standard were measured by MC-ICP-MS under continuous nebulisation using the NIST SRM 987 standard as reference for mass bias correction. The results obtained, compared with the published IUPAC values, are given in Fig. 3. It was observed that the mass bias corrected 87Rb/85Rb isotope ratio value for all Merck Rb standards was, on average, 0.38649 ± 0.00008 which was higher than the IUPAC value of 0.38562 ± 0.00017. Additionally, Rb fractionation effects on the chromatographic procedure used here were rejected since the obtained value for the 87Rb/85Rb isotope ratio after the chromatographic separation of the Merck standard and the NIST SRM 987 Sr reference material, calculated using the NIST as reference, was 0.38639 ± 0.00008, in agreement with the value obtained by continuous nebulisation. Finally, cider samples injected under the same conditions as the Merck standard, but using the found 88Sr/86Sr as reference for mass bias correction, gave rubidium isotope ratios of 0.38557 ± 0.00018, in agreement with the IUPAC values. Such deviation of 87Rb/85Rb isotope ratio from IUPAC value can be attributed to possible fractionation effects during the process of purification of the Rb standard. Although other publications on measurement of Rb isotope ratios by MC-ICP-MS32,35 did not provide isotopic data on non-certified Rb standard solutions, fractionation of Rb isotopes in cation exchange chromatography was already described on the literature.36
![]() | ||
Fig. 3 87Rb/85Rb isotope ratios measured by continuous nebulisation of different rubidium standards at 200 ppb levels (n = 3) using the NIST SRM 987 for mass bias correction (ratio 88Sr/86Sr). Uncertainties indicate standard deviation of the triplicate measurement. |
Strontium isotope ratios and their propagated uncertainty was evaluated for six consecutive injections of a standard solution containing 50 ng g–1 Sr NIST SRM 987 and 50 ng g–1 Rb (Merck). Both internal and external mass bias correction are given in Table 4. The equation used to calculate the standard uncertainty of the average was:
![]() | (6) |
Injection number | Internally corrected 87Sr/86Sr | UR | Externally corrected 87Sr/86Sr | UR | Externally corrected 88Sr/86Sr | UR |
---|---|---|---|---|---|---|
1 | 0.71058 | 0.00037 | 0.71066 | 0.00037 | 8.3807 | 0.0034 |
2 | 0.71053 | 0.00041 | 0.71059 | 0.00041 | 8.3799 | 0.0035 |
3 | 0.71052 | 0.00033 | 0.71066 | 0.00033 | 8.3818 | 0.0028 |
4 | 0.71057 | 0.00043 | 0.71067 | 0.00043 | 8.3811 | 0.0043 |
5 | 0.71050 | 0.00030 | 0.71060 | 0.00031 | 8.3808 | 0.0032 |
6 | 0.71053 | 0.00042 | 0.71073 | 0.00042 | 8.3833 | 0.0037 |
Average | 0.71054 | 0.00008 | 0.71065 | 0.00009 | 8.3812 | 0.0014 |
Certified | 0.71034 | 0.00026 | 0.71034 | 0.00026 | 8.3786 | 0.0033 |
Error (%) | +0.028 | +0.044 | +0.031 |
Five cider samples originating from England and Spain, and containing different Rb and Sr concentrations, were selected to evaluate the analytical procedure. The obtained 87Sr/86Sr isotope ratios after both off-line7 and on-line separation are shown in Table 5. For on-line separation both internal correction, using the 88Sr/86Sr ratio, and external correction, using the 87Rb/85Rb ratio, are given. The isotope ratio uncertainties for on-line separation were calculated using the Kragten’s method as described before. However, for the off-line separation only the contribution of the 88Sr/86Sr ratio for internal correction was taken into account for the propagation of the uncertainties.7
Sample | Sr (ppb)a | Rb (ppb)a | IC-MC-ICP-MS (Neptune) | Continuous sample introduction MC-ICP-MS (Axiom) | ||||
---|---|---|---|---|---|---|---|---|
Internally corrected 87Sr/86Sr | UR | Externally corrected 87Sr/86Sr | UR | Internally corrected 87Sr/86Sr | UR | |||
a Concentrations in the measured solution after 1:10 dilution of the samples. | ||||||||
England | ||||||||
KX | 36 | 47 | 0.7098 | 0.0006 | 0.7096 | 0.0006 | 0.70951 | 0.00007 |
Gaymer’s Olde | 33 | 13 | 0.7099 | 0.0005 | 0.7097 | 0.0005 | 0.70951 | 0.00007 |
Spain (Asturias) | ||||||||
M. Busto | 9.8 | 74 | 0.7088 | 0.0016 | 0.7086 | 0.0018 | 0.70899 | 0.00007 |
DOP7 | 7.8 | 116 | 0.7131 | 0.0024 | 0.7130 | 0.0024 | 0.71253 | 0.00008 |
Fugitivo | 6.6 | 34 | 0.7097 | 0.0011 | 0.7097 | 0.0011 | 0.70941 | 0.00007 |
As can be observed, English samples contained a higher concentration of strontium which resulted in lower isotope ratio uncertainties in comparison with Spanish ciders. On the other hand, no difference could be observed between internal and external correction both in terms of the measured isotope ratios and their uncertainties indicating that matrix effects on the mass bias measured for rubidium were not present even as rubidium eluted at the dead volumne of the column. The levels of rubidium in all samples were high enough to allow external mass bias correction. Finally, the observed values for on-line separation were in agreement with those obtained previously by off-line separation. The most striking difference between both procedures was the lower uncertainties of the off-line procedure. This is due to the large difference in concentration of strontium in the measured solutions (between 20 and 200 times higher for the off-line procedure) after the off-line Sr-Spec preconcentration. It is clear then that the main disadvantage of the propossed procedure in comparison with the off-line separation of Rb and Sr is the lack of any suitable preconcentration procedure for samples containing low Sr concentrations. As can be observed in Table 4, similar uncertainties can be obtained after repeated injections of the sample.
![]() | ||
Fig. 4 Chromatograms obtained under the selected chromatographic conditions for real samples: (a) cider, (b) apples, and (c) soil extract in ammonium nitrate. |
The obtained 87Sr/86Sr isotope ratios for cider, apples, apple leaves and soil extracts from the same production site are shown in Table 6. For all samples, internal correction is compared with external correction both in terms of 87Sr/86Sr isotope ratios and their uncertainties. For cider and apples, the concentration of strontium in the measured solutions was always below 5 ppb, and that explains the high uncertainty obtained. For apple leaves and soil extracts the strontium concentrations were one order of magnitude higher and, therefore, the uncertainties decreased considerably. For external correction the uncertainties were similar to those found by internal correction except for those samples were the concentration of rubidium was very low as for the nitric acid extract as can be observed in Table 6.
Sample | Sr (ppb)a | Rb (ppb)a | Internally corrected 87Sr/86Sr | UR | Externally corrected 87Sr/86Sr | UR |
---|---|---|---|---|---|---|
a Concentrations in the measured solution after 1 : 10 dilution of the samples. | ||||||
Cider | 2.9 | 75 | 0.7085 | 0.0053 | 0.7074 | 0.0054 |
Apples | 3.9 | 64 | 0.7101 | 0.0090 | 0.7089 | 0.0092 |
Apple leaves | 17 | 6.3 | 0.7105 | 0.0012 | 0.7097 | 0.0011 |
Soil (HNO3 extract) | 41 | 1.9 | 0.7109 | 0.0003 | 0.7106 | 0.0024 |
Soil (NH4NO3 extract) | 48 | 4.7 | 0.7107 | 0.0011 | 0.7086 | 0.0014 |
Using this on-line methodology the uncertainties in the 87Sr/86Sr isotope ratios depended strongly on the concentration of strontium in the samples. In comparison with off-line procedures, where preconcentration factors up to 20 are common, cider samples have to be diluted ten times resulting in much poorer uncertainties. Typical internal precisions obtained with the selected measurement and correction conditions were only 2 times lower than the precisions obtained with continuous sample introduction for a solution of the same concentration of Sr NIST SRM 987. It is clear that samples containing much higher strontium concentrations, in comparison to those measured in this work, will provide comparable isotope ratio uncertainties to off-line procedures with simplified sample preparation. The application of this procedure to other sample types, such as mineral digests and ancient bones is under progress in our laboratory.
This journal is © The Royal Society of Chemistry 2008 |