Miroslav Kovačeviča, Regina Leberb, Sepp D. Kohlweinb and Walter Goessler*c
aNational Institute of Chemistry, Hajdrihova 19, SI-1000 Ljubljana, Slovenia
bInstitute of Molecular Biology, Biochemistry and Microbiology, Karl Franzens University Graz, Schubertstrasse 1, A-8010 Graz, Austria
cInstitute of Chemistry-Analytical Chemistry, Karl Franzens University Graz, Universitaetsplatz 1, A-8010 Graz, Austria. E-mail: walter.goessler@uni-graz.at
First published on 25th November 2003
Phospholipids are the main constituents of membranes in all types of prokaryotic and eukariotic cells. Due to their complexity and heterogeneity in biological samples, qualitative and quantitative analyses of membrane phospholipids in cellular extracts represent major analytical challenges, mainly due to suitable and sensitive detection methods. The inductively coupled plasma mass spectrometer (ICP-MS) is a suitable detector for selective determination of phospholipids as they all contain phosphorus. Phospholipids are extractable with organic solvents, therefore liquid chromatography with an organic mobile phase was used for separation of different lipid species. Solvent load to the plasma was reduced by splitting the mobile phase prior to reaching the nebulizer, by chilling the spray chamber to −5 °C and by optimisation of carrier gas flow for maximum condensation of organic vapours. Despite desolvation, oxygen was added to prevent carbon deposition on interface cones. To reduce polyatomic interferences at m/z ratio 31 (e.g.31CH3O+) and to improve detection limits, helium was used as a collision gas. The achieved absolute detection limits were between 0.21 and 1.2 ng of phosphorus and were superior to those obtained by an evaporative light scattering detector, which provides an alternative detection system for lipid analysis. The usefulness of the developed method was demonstrated by analysis of lipid extracts from the yeast Saccharomyces cerevisiae.
![]() | ||
Fig. 1 Chemical structures of most common phospholipids and their abbreviations* (R1 and R2 are different fatty acids groups).1 |
Phospholipid classes are characterized by the alcohol residue esterified to the phosphate group. The basic PL structure is phosphatidic acid, or 1,2 diacylglycerol-3-phosphate. For example, phosphatidylcholine (PtdCho) describes a phospholipid harboring choline (N,N,N-trimethylaminoethanol-2) esterified to the phosphate residue. The diversity of phospholipid molecular species is brought about by the diversity of fatty acids esterified to the glycerol backbone. Yeast, for instance, produces some 20 different molecular species of phosphatidylcholines.
Pure synthetic phospholipids are chemically defined and they can be designated by their systematic name. For example, abbreviation “DOPC” stands for 1,2-dioleoyl-phosphatidylcholine.2
A major goal of research on phospholipids is to understand the significance of these compounds in the functioning of cellular membranes. The majority of studies are based on HPLC analysis of lipid extracts. Since phospholipids are soluble only in organic solvents, normal phase liquid chromatography is the most frequently used separation method.3 The columns are packed either with silica-, amino- or with diol-type stationary phases and all have in common, that they separate only different chemical classes of phospholipids, rather than molecular species differing in the fatty acid composition.4 Mobile phases commonly employed are usually mixtures of different solvents, such as methanol, acetone, hexane, chloroform, tetrahydrofurane, 2-propanol or acetonitrile. For better separation, gradient elution should be applied, starting from a weak (nonpolar) mobile phase to a strong (more polar) mobile phase. However, reverse phase liquid chromatography can also be used for their separation.5,6
Due to the absence of intrinsic molecular properties and the heterogeneity of the substance classes, detection of phospholipids is a major analytical problem. Different detection systems have been described for phospholipid analysis.7 The refractive index detection suffers from poor detection limits and is only useful for simple mixtures, since gradient elution cannot be applied due to a baseline drift.8 Reported detection limits are approximately 20 ng of phosphorus.9 Because many solvents used for liquid chromatography are non-transparent in the UV range (200–210 nm, useful for detection of phospholipids) these detectors have serious constraints with respect to the selection of mobile phases.10 To overcome lack of chromophores in the phospholipid molecules, post-column derivatisation was applied by Rastegar et al.6 They derivatised the sample with Naproxen and derivatives were subjected to HPLC with UV absorption measurements at 230 nm. The reported detection limit was 0.3 ng (expressed as phosphorus). With the introduction of the evaporative light scattering detector, analysis of all lipid species became easier.8,11,12 However, this detector suffers from a limited linear range, with 400 ng of phosphorus as a lowest sample amount giving linear response.13,14 Lower sample quantities can be detected (10 ng of phosphorus), however the calibration response is not linear.15 Another possibility for detection of phospholipids is hyphenation of HPLC to mass spectrometry, giving better selectivity towards the compounds of interest.16 With electrospray ionisation mass spectrometry a quantification limit for phosphatidylserine was reported to be 0.05 ng of phosphorus.17 Electrospray ionisation used together with tandem mass spectrometry can be used to obtain molecular information about the phospholipids without prior chromatographic separation.18–20
It is a common property of all described detectors that their responses are depending on the molecular structure of species being analysed and, therefore, standard compounds are required for their quantification. To overcome this problem, an inductively coupled plasma mass spectrometer (ICP-MS) can be used as a detection technique, since its response is theoretically dependant only on the element in question. The hyphenation of ICP-MS to LC for phospholipid analysis was first described by Axelsson et al.21 The authors have reported successful normal phase HPLC separation and ICP-MS detection of three standard compounds belonging to PC, PE and PG phospholipid classes. However, authors did not provide any details on analysis procedure and they did not test this method on real complex biological samples. Additionally they used a low flow membrane desolvation unit for removal of interfering organic solvents, which is still not widely accessible in laboratories. It should be noted, that the determination of phosphorus and its compounds by an ICP-MS is not an easy task because phosphorus has a high ionisation potential and consequently is poorly ionised in the plasma. Additionally, it suffers from polyatomic interferences at m/z ratio 31 when low mass resolution instruments are used. Therefore not only 31P+ ions are measured, but also polyatomic ions such as 12C1H316O+.22,23 As a result, poorer detection limits are achieved when carbon compounds are in the sample matrix. Jiang and Houk24 reported a decrease of the mass spectrometer response with increasing concentration of organic modifiers in mobile phase of the LC-ICP-MS system when polyphosphates and adenosine phosphates were analysed.
In this work we present a detailed study of processes occurring during the determination of phospholipids by a quadrupole ICP-MS instrument equipped with a conventional double pass spray chamber for desolvation. Additionally, we are demonstrating its use for the characterization and detection of phospholipids from yeast lipid extracts by applying a modified chromatographic separation published by Sas et al.3
The standard mixture of six phospholipids was prepared by diluting each phospholipid standard in a chloroform/methanol mixture (2/1, v/v). The concentrations of each expressed as phosphorus were as follows: 3.3 mg l−1 DOPA, 2.9 mg l−1 DOPG, 2.7 mg l−1 PI, 3.1 mg l−1 DOPE, 3.0 mg l−1 DOPS and 2.9 mg l−1 DOPC. Approximately 1 g of wild-type yeast cells (Saccharomyces cerevisiae W303, MATα, leu2, ura3, his3, ade2, trp1) were resuspended in 10 ml of deionised water and cells were disrupted in a glass bead homogeniser (B. Braun Melsungen, Germany). Then 80 ml of chloroform/methanol mixture (2/1, v/v) was added and the suspension was stirred for 30 min at room temperature, additionally 20 ml of a MgCl2 solution (0.034%) was added for phase separation and it was again stirred for 30 min. After centrifugation for 5 min, the organic phase was transferred to a round bottom flask and evaporated to dryness. Lipids were dissolved in 2 ml of chloroform/methanol mixture (2/1, v/v) and the solution was transferred to a HPLC vial. Solutions were stored at −20 °C.
The system was optimised by pumping mobile phase A containing 2 mg l−1 of phosphorus as a DOPE. During the tuning procedure, such conditions were chosen that the signal at m/z 31 was as high as possible and that the mass peak at m/z 31 and neighbouring peaks m/z 30 and m/z 32 were clearly separated. Further optimisation of detector response was performed in flow injection analysis mode with mobile phase A as a carrier solution and by injection of 2 µl of 90 mg l−1 of phosphorus as a DOPE diluted in mobile phase A. To evaluate results, signal to background ratios (SB) were calculated as a quotient between height of phosphorus signal and height of background signal. After having optimal conditions for detection of phosphorus, helium was added as a collision gas to reduce the background signal.
For the detection of the phospholipids the following optimised conditions were used: Plasma gas 15 l min−1, auxiliary gas 1.0 l min−1, carrier gas 0.50 l min−1, optional gas flow rate 24% (of carrier gas flow rate), rf power 1600 W (reflected power ≤10 W), ORC gas (helium) 4.0 ml min−1, spray chamber temperature −5 °C and sample depth (torch-interface distance) 10 mm. All chromatograms were smoothed before integration.
The first parameter we optimised was the rf power. We tested powers from 1200 to 1600 W by recording flow injection signal. The experiment was performed at three different sample depths (6, 9 and 12 mm) at constant carrier gas flow rates of 0.68 l min−1 and without any make-up gas flow. As expected, when highest rf power was used, SB ratio was highest at all three tested sample depths, since high temperature is required to decompose organic matrix and to improve ionisation of phosphorus. According to this results rf power of 1600 W was chosen for further optimisation and routine measurements.
In order to find optimum carrier and make-up gas flows, several combinations of these gas flows were tested by flow injection technique at 1600 W of rf power. The whole experiment was conducted at three different sample depths (6, 9 and 12 mm). Carrier gas flows were tested in the range from 0.2 to 0.70 l min−1 at four different make-up gas flows (0, 0.10, 0.15 and 0.20 l min−1). The signal to background ratios are not improved by the addition of a make-up gas. The usage of a make-up gas only shifts the optimal carrier gas flow rate to lower values. That means the signal to background ratio is only depending on total flow of gases through the spray chamber and the most important process in spray chamber is condensation of organic vapours. When too high total gas flows are applied, the residence time of the aerosol in the spray chamber is shorter, the amount of condensed organic matter is lower and therefore the background derived from organic vapours in the form of carbon based polyatomic ions is increased. However, when too low flow rates are applied, condensation is efficient, but transport of aerosol particles is compromised and SB ratios are decreased. This fact was also confirmed by observation of the colour of the plasma and the required amount of oxygen to prevent carbon deposition. At higher carrier or make-up gas flow rates the plasma was greener and more oxygen was needed. On the basis of these results it was decided not to use any make-up gas.
The optimum sampling point of ions in the plasma is of course influenced by the gas flows and sample depth. Therefore both parameters were optimised, while no make-up gas flow was used. The tested carrier gas flow rates were in the range from 0.35 to 0.55 l min−1 and sampling depths were in the range from 5 to 11 mm. The results are presented in Fig. 2 as a two-dimensional plot, where colour intensity is representing SB ratio. The highest SB ratios were observed when carrier gas flow rate of 0.50 l min−1 was applied and sample depths were from 5 to 9 mm. These conditions are therefore giving highest sensitivity for detection of phospholipids. However, there is an empirical rule that the intensive green coloured zone on the front end of the plasma should end before reaching the tip of the sampler cone.25 This requires higher amounts of oxygen and means shorter life time of the cones. As a compromise between good sensitivity and system robustness, a carrier gas flow rate of 0.50 l min−1 and sample depth of 10 mm were chosen for routine work.
![]() | ||
Fig. 2 Signal to background ratios (SB) at different carrier gas flow rates (x-axis) and different sample depths (y-axis) (1600 W, no make-up gas). |
Despite optimal conditions for the desolvation process in the spray chamber, the background signal was still high and was negatively influencing the limits of detection. Another negative consequence was noticeable drift of baseline in the chromatogram due to change of mobile phase composition during chromatography. This problem was solved with helium as a collision gas. We explored helium flow rates through the ORC in the range from 0 to 6 ml min−1. At each tested helium flow rate, the flow injection signal was recorded and SB ratios were calculated (Fig. 3). As can be seen, helium improved SB ratio at flow rates between 4 and 6 ml min−1. For the practical applications a flow rate of 4 ml min−1 was applied, giving five-times higher SB ratio than without using any collision gas. At higher helium flow rates, suppression of both phosphorus and background signal was too strong and the noise of the signal was hindering improved SB ratios. At the same time, background was efficiently reduced for about two orders of magnitude, minimising the problem of baseline drift during the chromatography.
![]() | ||
Fig. 3 Phosphorus signal (left y-axis, full line), background signal (left y-axis, dotted line) and signal to background ratio (SB) (right y-axis, full bold line) at different helium flow rates through the octopole reaction cell. |
![]() | ||
Fig. 4 Separation of six chemically defined phospholipids in standard mixture on YMC Pack Diol-120 column (250 × 4.6 mm, 5 µm) with ICP-MS detection of phosphorus at m/z ratio 31 (5 µl injected, 0.6 ml min−1, each peak corresponds ∼15 ng of phosphorus). |
![]() | ||
Fig. 5 Separation of phospholipid classes in yeast lipid extract on YMC Pack Diol-120 column (250 × 4.6 mm, 5 µm) with ICP-MS detection of phosphorus at m/z ratio 31 (2 µl injected, 0.6 ml min−1). |
Compound | Retention time/min | Calibration curve | R | Linear range/ng | LOD/ng | Reproducibilitya (%) |
---|---|---|---|---|---|---|
a At lowest point of calibration curve. | ||||||
DOPA | 6.7 | A = 18000 × m − 830 | 0.9998 | 1.6–16 | 0.36 | ±6 |
DOPG | 7.8 | A = 23400 × m − 5650 | 0.9997 | 1.4–14 | 0.21 | ±5 |
PI | 14.2 | A = 25000 × m − 850 | 0.9999 | 1.4–55 | 0.54 | ±7 |
DOPE | 18.3 | A = 21000 × m − 10400 | 0.9999 | 3.0–61 | 1.2 | ±7 |
DOPS | 28.1 | A = 16500 × m − 18600 | 0.9998 | 3.0–59 | 1.2 | ±16 |
DOPC | 35.9 | A = 19500 × m − 230 | 0.9999 | 1.5–59 | 0.50 | ±14 |
The usefulness of the developed method for analysis of phospholipids was demonstrated on a complex lipid extract from yeast as presented in Fig. 5. Each identified compound was quantified by using calibration curves given in Table 1. The results are presented in Table 2 as peak areas and as calculated masses and concentrations of each identified compound in the sample extract. It should be noted, that all masses and concentrations are expressed as phosphorus and that peak coeluting with chemical class of PC was integrated and considered as belonging to PC class. According to the extraction method, calculating the concentration of each class of phospholipids in yeast is impossible, since many steps in the extraction procedure are not quantitative. If such data were required, then the extraction procedure should be modified or an internal standard should be added before the extraction. Since properties of yeast sample are defined by its phospholipid composition, relative amounts of identified compounds were calculated and are given as percentages in Table 2.
Class | Peak area/103 units | Mass/ng | Concentration/mg l−1 | Relative amountsa (%) | Semi-quantiative relative amountsb (%) |
---|---|---|---|---|---|
a Relative amounts of identified compounds.b Relative amounts determined by semi-quantitative procedure. | |||||
PA | 65.5 | 3.7 | 0.74 | 1.6 | 1.2 |
PI | 783 | 31 | 6.3 | 13 | 15 |
PE | 1440 | 69 | 14 | 29 | 27 |
PS | 150 | 10 | 2.0 | 4.3 | 2.8 |
PC | 2380 | 120 | 24 | 51 | 44 |
X1 | 400 | — | — | — | 7.6 |
X2 | 45.6 | — | — | — | 0.9 |
Theoretically, the response of an ICP-MS is element depending, meaning it is the same for all compounds regardless to their structure. The determined response factors, which are part of calibration curves in Table 1, have values from 16500 to 25000 peak area units per ng of phosphorus. Deviations from theory are expected, since we used a gradient elution program, giving a different matrix composition for each compound resulting in different nebulisation efficiencies. Moreover, the changing matrix composition might also change the ionisation efficiency of phosphorus in the plasma. Therefore simplification of quantification procedure by using a calibration curve based only on one compound should be used with critical evaluation of obtained results and it already belongs in the field of semi-quantitative analysis. In cases, when we are interested only in obtaining approximate ratios between classes of phospholipids in the sample, only peak areas without any calibration can be used with an awareness of possible errors, of course. To show the usefulness of such a quick semi-quantitative analysis, a yeast lipid extract was also treated in that way. Peak areas of all peaks found in the chromatogram were summed, their relative amounts were calculated and are also presented in Table 2. Compared to literature data2 one can clearly say that this semi-quantitative approach gives good agreement.
This journal is © The Royal Society of Chemistry 2004 |