Patrick J.
Parsons
*ab,
Ying
Zhou
a,
Christopher D.
Palmer
a,
Kenneth M.
Aldous
ab and
Paul
Brockman
c
aWadsworth Center, New York State Department of Health, P.O. Box 509, Albany, New York 12201-0509, USA
bDepartment of Environmental Health and Toxicology, School of Public Health, The University at Albany, PO Box 509, Albany, New York 12201-0509, USA
cExeter Analytical Inc., 7 Doris Drive, North Chelmsford, Massachusetts 01863, USA
First published on 29th November 2002
In an effort to understand the pyrolysis and atomization behavior of Pb in a dilute blood matrix when determined using a tungsten filament electrothermal atomic absorption spectrometer (W-filament ETAAS), we directly coupled a W-filament ETAAS instrument (J. Anal. At. Spectrom., 2001, 16, 82) to an inductively coupled plasma mass spectrometer (ICP-MS). The W-filament instrument was operated both as an electrothermal atomizer (ETA) and as an electrothermal vaporizer (ETV). The experimental arrangement was used to monitor simultaneously the analyte (208Pb), the permanent modifier (103Rh) coating the W surface, a surrogate for the organic components of the blood matrix (40Ar12C), and the metal atomizer (180W, 183W and 184W16O), while also measuring Pb by AAS pseudo-simultaneously. Electrothermal parameters used in the ETV/ETA experiments were precisely the same as those described previously (Spectrochim. Acta, Part B, 2002, 57, 727) for ETAAS work. This was necessary in order that the ICP-MS data could be used to observe the vaporization of analyte, matrix, modifier, and metallic atomizer throughout the electrothermal cycle. The purge gas used to transport vaporized material from the atomization cell to the ICP was Ar containing 6% H2. Pseudo-simultaneous ETAAS and ETV-ICP-MS measurements of Pb in blood were achieved using this experimental arrangement, with the MS signals appearing roughly 2 s after those by AAS, consistent with the delay caused by the 15 cm long Teflon transfer line connecting the two instruments. During pyrolysis, a strong signal observed at mass 52 is shown to be due to 40Ar12C; this signal increases as filament power for pyrolysis increases. When plotted, the data show the effective removal of much of the carbon-based matrix during pyrolysis, but without significant loss of analyte. In contrast, the pre-coated permanent modifier, Rh, is only lost during the final cleaning stage of the program in which maximum power is necessary to remove the carbonaceous matrix. However, when Rh is added to the diluent, it is lost even during pyrolysis, and at a much lower temperature. This explains why addition of Rh directly to the diluent yields no modification benefits, and why periodic re-coating of the filament surface for the analysis is necessary.
The graphite furnace is currently the most widely used atomizer for electrothermal atomic absorption spectrometry (ETAAS) and vaporizer for ETV-ICP-MS. Most commercially available ETV systems for ICP-MS instruments are actually modified graphite furnace AAS atomizers. Instead of being used as the atomizer (as it is in ETAAS), the graphite furnace serves as a vaporizer in ETV-ICP-MS, whereby the sample is dried, pyrolyzed, vaporized and subsequently transported into the plasma. Atomization is not required, since the analyte need only be vaporized and introduced by the carrier gas stream into the plasma where atomization and ionization processes occur. Thus, vaporization temperatures required in ETV-ICP-MS are typically lower than the atomization temperatures used in graphite furnace ETAAS.
ETV-ICP-MS arrangements have also been used for some fundamental studies of vaporization and atomization processes that occur within the graphite furnace. Such studies are important not only for determining the optimum experimental conditions required for graphite furnace ETV-ICP-MS, but they may also be useful for probing atomization mechanisms in graphite furnace ETAAS.
Lamoureux et al.4,5 used a modified commercial graphite furnace atomizer coupled to an ICP-MS to record pseudo-simultaneous AAS and ICP-MS measurements. This instrument arrangement was employed to investigate a MgCl2 interference reported for Mn determinations by graphite ETAAS.4 Although this approach is a potentially powerful tool for elucidating fundamental processes within the graphite furnace, conditions within the ETV device are not quite comparable to those in ETAAS. For example, during atomization in graphite furnace ETAAS, the Ar gas flow is normally stopped to prolong the residence time of ground state atoms in the optical path, thus improving sensitivity. This is considered a fundamental aspect of the stabilized temperature platform furnace (STPF) concept.6 However, during the vaporization step in graphite furnace ETV, Ar gas is continually purged to transport the vaporized sample into the plasma. Lamoureux et al.4,5 used only a very small gas flow rate (50 mL min−1) in their ETV experimental arrangement to transfer vaporized material combined with a make-up gas to achieve a total Ar flow rate of 1 L min−1 into the plasma. In a series of reports, Grégoire and his co-workers used a commercial graphite furnace ETV coupled with an ICP-MS to investigate the vaporization (atomization) of several analytes including W,7 B,8 Y and rare earth elements,9 U,10 the Pt group elements,11 and Ra and alkaline earth elements.12 Some studies focused on the vaporization process itself7,9,12 while others used ETV data to elucidate the atomization mechanisms in graphite furnace AAS.8,10,11
Metal surfaces, such as W-filaments and the W-tube furnace, have also been used as ETV devices for ICP-MS. Park et al.13 used a W or a Re metal ribbon filament as an ETV device with a 5 µL sample capacity. Hauptkorn et al.14 reported coupling a W-filament ETV device to an ICP-MS to determine trace impurities in high-purity quartz samples using a slurry sampling technique. Their W-filament ETV-ICP-MS arrangement was also utilized to identify combinatorial chemistry compounds from elemental tagging procedures used in drug discovery research.15 Other groups report using a W boat furnace as an ETV device for ICP-MS.16,17 The W boat furnace was used for the determination of rare-earth elements in aqueous solution,18 metal impurities in sulfamic acid,19 Pb in biological materials,17 Pb in rock samples,20 and F in aqueous samples.21 It has also been reported that strips of W, Ta, Mo or Re have been inserted in a graphite tube and used as vaporization platforms in ETV-ICP-MS.22,23
We have previously described the feasibility of using W-filaments as electrothermal atomizers for the determination of Pb in blood by AAS.24 A compact prototype AAS instrument equipped with a novel W-filament and a self-reversal background correction system was assembled specifically for blood Pb measurements and was shown to meet current US regulatory requirements for this analysis.25 Further improvements were reported more recently using a Rh-coated W-filament.26 However, the Rh coating is not permanent, and periodic re-coating is required, thus diminishing its utility in non-technical hands. The principal objective of this study was to explore the pyrolysis and atomization behavior of Pb from Rh-coated W filaments by directly coupling a prototype W-filament AAS instrument to an ICP-MS. Using this experimental arrangement made it possible to monitor simultaneously (a) the analyte (208Pb), (b) the permanent modifier (103Rh) that coats the W-filament, and (c) a possible surrogate for the organic components of the blood matrix (40Ar12C) that appears at mass 52, while also measuring Pb pseudo-simultaneously by AAS.
For the purposes of this study, the PBA-10 prototype AAS instrument was interfaced to an ELAN 6100 DRC® inductively coupled plasma mass spectrometer (ICP-MS) (PerkinElmer Sciex, Shelton, CT, USA) running version 2.3.2 of the ELAN software. Although the ELAN 6100 DRC is equipped with dynamic reaction cell (DRC) technology for reducing polyatomic interferences in ICP-MS, this feature was not necessary for this work; therefore, the ELAN was operated in standard, i.e., non-DRC, mode only. The ICP-MS instrumentation in our laboratory is also equipped with an HGA 600MS graphite furnace ETV and an AS 60 autosampler (PerkinElmer Instruments, Inc., Shelton, CT, USA). Initial optimization of the ICP-MS was carried out using a cross-flow nebulizer (PE Part # N8120516), a Scott-type spray chamber (PE Part # N8120124), and aqueous tuning solutions. No further optimization was performed after switching to ETV mode. The carrier gas flow rate was fixed at 1 L min−1 (LPM), consistent with the flow rate used in the W-filament AAS cell.
On the ICP-MS, peak-hopping mode was used for data acquisition, and the optimum dwell time was found to be 10 ms. The length of the actual read time was determined by the number of readings per replicate. Integrated ion intensities, rather than the maximum ion counts, were used for ICP-MS data evaluation. No more than six isotopes were measured in a single ETV event to ensure reliable integration of the transient signals. On the PBA-10 instrument, AAS data were collected pseudo-simultaneously as integrated absorbance (Ai) measurements. The operating parameters of the ICP-MS and the PBA-10 instrumentation are listed in Table 1 unless otherwise stated.
PE SCIEX ELAN 6100 DRC® ICP-MS | |
ICP rf power: | 1100 W |
Lens voltage: | 7.65 V |
ICP supporting gas (Ar): | 15 LPM |
Auxiliary gas (Ar): | 1.1 LPM |
Carrier gas (6% H2 in Ar): | 1.0 LPM |
Dwell time: | 10 ms |
Scan mode: | Peak hopping |
Signal measurement: | Integrated counts |
Ions monitored: | 208Pb, 40Ar12C, 13C, 103Rh, 183W |
EAI PBA 10 W-filament ETV/ETA | |
Sample volume: | 15 µL |
Matrix modifier for W-filament ETAAS: | Rh permanent modifier |
Heating program: | Dry—160 s at 0.9 W |
Pyrolysis—30 s at 2.6 W | |
Cool—10 s at 0 W | |
Atomization—2 s at 30 W | |
Clean—20 s at 46 W | |
Purge gas (6% H2 in Ar): | 1.0 LPM |
Sample transfer from ETV to ICP-MS: | 15 cm Teflon tube (6 mm id) |
For investigations of the W-filament ETAAS coupled to ETV-ICP-MS, we used a New York State Blood Reference Material (NYS-RM Lot 049) with a certified Pb concentration of 42.2 ± 1.8 µg dL−1.‡ These RMs consist of lyophilized whole blood that is obtained from Pb-dosed animals and are certified through interlaboratory studies.27 The lyophilized materials are reconstituted using double-deionized water, and an aliquot is diluted 1 + 4 with a diluent containing 0.5% (v/v) Triton X-100 and 0.2% (v/v) HNO3. Diluted blood samples (15 µL) are manually deposited on the W-filament using a double-action micropipette (Finnpipette, Helsinki, Finland) or injected into the graphite furnace ETV using the AS-60 autosampler.
The vertical position of the W-filament atomizer relative to the plasma-torch injector tube was also optimized with respect to 208Pb intensity. Although the integrated counts for 208Pb did not appear to be very sensitive to this parameter, we did observe a small improvement in sensitivity (∼10%) when the atomization cell outlet and the injector tube inlet were aligned, i.e., when the Teflon tube connecting the two instruments was level. However, this improvement may not be statistically significant.
We also varied the id of the Teflon tube connecting the atomization cell to the ICP-MS. Replacement of a 1.5 mm id Teflon tube with a 6 mm id diameter tube markedly improved the sensitivity for 208Pb by ICP-MS. This is consistent with observations reported in the literature for graphite furnace ETV instrumentation.28
The HGA 600MS furnace ETV system was coupled to the ICP-MS using a 6 mm id Teflon connecting tube of a similar tube length. We compared pure Ar with a mixture of 6% H2 in Ar as the carrier gas using 15 µL of 1 + 4 diluted NYS RM Lot 049 as the test solution. Sensitivity was monitored at mass 208 and evaluated using integrated ion counts. Addition of 6% H2 to the Ar carrier gas suppressed the sensitivity for 208Pb by an order of magnitude, from 7.5 × 106 counts·s to 5.1 × 105 counts·s.
We repeated this experiment, comparing pure Ar with 6% H2 in Ar as the carrier gas, with the W-filament ETV system. Again, the integrated ion intensity for 208Pb, using the same diluted blood sample, was reduced by around an order of magnitude, from 1.5 × 107 counts·s with pure Ar to 1.5 × 106 counts·s with 6% H2 in Ar as the carrier gas. In a third experiment, we used direct nebulization sample introduction into the ICP-MS with 6% H2 in Ar as the intermediate carrier gas. The sensitivity for 208Pb in the same 1 + 4 diluted blood sample was reduced by four orders of magnitude, from 2 × 107 counts with pure Ar to around 8 × 103 counts with 6% H2 in Ar. Thus, introduction of as little as 6% H2 in Ar into an Ar plasma ICP-MS greatly reduces the sensitivity for 208Pb, regardless of the sample introduction arrangement.
The effects of molecular gases such as H2 on Ar plasmas were previously reported by Sesi et al.29 Different impacts were described, depending on whether the molecular gases were added to the plasma outer gas, the auxiliary gas, or the intermediate carrier gas. Sesi et al.29 argue that addition of up to 16.7% H2 in the Ar carrier gas increases both the electron temperature and the gas-kinetic temperature, thus providing a hotter plasma due both to the much higher thermal conductivity of H2 than Ar and to the constricted plasma volume after H2 is added. In contrast, these authors also contend that the concentration of electrons in the plasma is decreased. Sesi et al.29 explain these seemingly contradictory observations as the thermal pinch effect of the plasma. That is, when H2 is added to the carrier gas, the plasma begins to move away from the torch injector tube, i.e., the distance between the tip of the injector tube and the bottom of the plasma increases. They suggest that the increase in gas temperature induced by addition of H2 is not sufficient to compensate for the increased residence time of the sample aerosol in the “cooler” torch region, and thus ionization, and consequently sensitivity, is suppressed. Our observations here of reduced sensitivity for 208Pb upon introduction of 6% H2 to the Ar carrier gas are consistent with those reported by Sesi et al.29
In W-filament AAS, addition of H2 to the purge gas is considered desirable because it enhances sensitivity for many elements by promoting a reducing environment that improves the signal-to-noise ratio,30 as well as protecting the W-filament from oxidation, thus extending its lifetime. In this study, Pb absorbance signals obtained using a W-filament AAS instrument show a modest improvement (∼10%) in sensitivity, from 264 to 238 pg, upon addition of 6% H2 to the purge gas. Since a key objective in interfacing a W-filament instrument to the ICP-MS was to investigate the atomization/vaporization processes of Pb in blood, it was necessary to maintain identical conditions to those used in previous ETAAS studies, including the use of 6% H2 in Ar as the purge gas even though this resulted in less than maximal sensitivity for 208Pb by ICP-MS.
Unlike in graphite furnace ETAAS, in which the Ar purge gas flow is stopped during atomization, the purge gas in W-filament ETAAS flows continuously during the atomization event. Thus, conditions for successful W-filament ETAAS measurements are identical to those typically applied in ETV-ICP-MS, in which vaporized species are swept into the ICP-MS at a purge-gas flow rate of 1 L min−1.
![]() | ||
Fig. 1 Pseudo-simultaneous electrothermal vaporization and atomization of Pb from 15 µL of 1 + 4 diluted blood (NYS RM Lot 049, 42.2 µg dL−1) measured by (a) ICP-MS and (b) AAS using a W-filament ETV/ETA instrument coupled to an ICP-MS. Integrated ion intensity for 208Pb by ETV-ICP-MS is 5.5 × 105 counts·s; integrated Pb absorbance by AAS is 0.0222 s, which yields a characteristic mass of (m0) of 251 pg. |
![]() | ||
Fig. 2 A typical W-filament ETV-ICP-MS intensity–time profile for 15 µL of 1 + 4 diluted blood containing Pb (NYS RM Lot 049: 42.2 µg dL−1 Pb). Note that counts for the ions 183W, 40Ar12C and 103Rh are plotted on the right-hand y axis, while those for the analyte 208Pb are plotted on the left-hand y axis. Each ETV step is matched to the corresponding ETA step in the W-filament atomization temperature program. |
Spectral interferences in ICP-MS can occur at precisely the mass selected for the analyte. For example at mass 52, not only is the polyatomic species 40Ar12C expected, but 52Cr also. To ensure that the measurements are not masked by such interferences, we also monitored 13C and calculated the isotope ratio 40Ar12C/13C. Fig. 3 shows the change in the 40Ar12C ion intensity during the pyrolysis event, along with the corresponding 40Ar12C/13C ratio. It is evident that the 40Ar12C/13C ratio remains relatively constant throughout the pyrolysis event independent of ion intensity, suggesting that the ion intensity monitored at mass 52 is primarily contributed by the carbon-containing polyatomic species 40Ar12C that is a marker of the blood matrix. Equally significant, however, is the fact that Cr is a much more refractory element compared with Pb. Given that Pb is not vaporized at the pyrolysis power setting used, it is inconceivable that Cr would be. Thus, 40Ar12C appears to be the major species detected at mass 52, and is an appropriate surrogate for the blood matrix.
![]() | ||
Fig. 3 Plot showing the 40Ar12C ion signal and the ratio of 40Ar12C to13C+ as a function of time during the entire 30 s pyrolysis event for a 15 µL aliquot of 1 + 4 diluted blood (NYS RM Lot 049: 42.2 µg dL−1 Pb). See the text for an explanation of how these data are plotted. |
The amount of W vaporized from the filament and transported into the plasma was monitored at mass 183 (183W, natural abundance 14.3%) throughout the entire electrothermal event. As is shown in Fig. 2, a detectable amount of W is clearly vaporized during pyrolysis, during which the power setting, i.e., temperature, is relatively low in relation to the high melting point (3410 °C) and boiling point (5660 °C) of W metal. No significant vaporization is evident during the ETA/ETV step, or even during the clean step, indicating that a more volatile W species must be formed during the pyrolysis step. By contrast, no significant W loss is observed during pyrolysis with blank firings, i.e., in the absence of a diluted blood matrix, suggesting that the matrix is critical to formation of the volatile W species.
In a previous study involving W-filament ETV-ICP-MS, Venable et al.32 detected W at temperatures well below the melting point of the metal and argued that it was due to formation of the more volatile W oxides species. Byrne et al.7 investigated the mechanism of volatilization of W from a graphite substrate using ETV-ICP-MS. They found two distinct vaporization processes occur in the temperature range 800–2700 °C, resulting in two distinct peaks for W. One peak that appeared at temperatures as low as 850 °C was attributed to the volatilization of WO3. Based on our studies, we find it quite conceivable that the blood matrix, and perhaps the nitric acid used in the diluent, could provide a source or sources of oxygen that, during pyrolysis, react(s) with W at the surface of the filament, yielding a mixture of W oxides, such as WO2 and W2O5. These oxide species can undergo sublimation at temperatures between 800 and 900 °C consistent with typical pyrolysis temperatures expected here, thus vaporizing W.33 According to Venable et al.,32 these W species appear as W and WO in the plasma, despite use of 10% H2 in Ar as the purge gas to reduce oxidation and to prolong filament lifetimes. Although we used only 6% H2 in Ar as the purge gas here, we did not detect W in the plasma as W oxide. However, we did detect it as 183W, though only during the pyrolysis step.
![]() | ||
Fig. 4 Pyrolysis studies showing (i) integrated Pb absorbance by AAS and (ii and iii) complementary data obtained pseudo-simultaneously by ICP-MS for a 15 µL aliquot of 1 + 4 diluted blood containing Pb (NYS RM Lot 049: 42.2 µg dL−1 Pb). The ICP-MS data are expressed as the pyrolysis fraction of the total integrated ion signal for (ii) the analyte (208Pb) and (iii) the matrix (40Ar12C) as a function of increasing pyrolysis power. |
As noted above, Rh is only lost during the cleaning step when maximum power (46 W) is applied. The Rh signal, as depicted in Fig. 2, appears to be almost steady-state during the cleaning phase, suggesting that the coating is continuously vaporized throughout the step. Thus, shortening this step will likely improve the durability of the Rh coating and decrease the frequency of recoating. We compared different amounts of blood matrix deposited on the filament. Use of a 15 µL aliquot of 1 + 9 and of 1 + 4 diluted blood yields 1.5 µL and 3 µL of whole blood, respectively. The total integrated ion intensity calculated for 208Pb, based on 1.5 µL whole blood, is 2.8 × 105 counts·s, and is approximately half the integrated intensity found for 3 µL of whole blood (5.5 × 105 counts·s). Similarly, the total integrated ion intensity for 40Ar12C, based on 1.5 µL of whole blood (1.6 × 106 counts·s) is approximately half that found for 3 µL of whole blood (3.3 × 106 counts·s). This observation further supports our inference that the signal at mass 52 is a valid marker of the blood matrix. However, the intensity of the 103Rh signals is roughly the same for both amounts of blood matrix, suggesting that Rh loss is independent of sample loading. Even with blank firings, i.e., in the absence of a diluted blood matrix, significant loss of Rh is observed. These observations are consistent with our previous work that indicated a need for periodic recoating of the W-filament with Rh to sustain the desired performance.26
In our previous study,26 it was observed that the gradual loss of Rh could not be prevented by direct addition of the modifier to the diluent, as is common practice for blood Pb measurements using phosphate modifier. In the present study we were able to probe this characteristic directly. Fig. 5 compares ETV-ICP-MS data collected with (a) Rh pre-coated on the W-filament and (b) Rh added to the blood sample diluent. As noted above, pre-coated Rh is only vaporized during the cleaning step. However, when added to the diluent, Rh is vaporized throughout pyrolysis and atomization, in addition to during the cleaning step. Surprisingly, Rh vaporization appears as numerous, apparently random, spikes throughout the pyrolysis, vaporization/atomization and clean steps. We have no convincing explanation for these observations but they do illustrate why Rh must be thermally pre-coated to, or bound to, the W-surface for successful modification to occur.
![]() | ||
Fig. 5 Intensity–time profiles for 208Pb, 40Ar12C, and 203Rh observed from a 15-µL aliquot of 1 + 4 diluted blood (NYS RM Lot 049: 42.2 µg dL−1 Pb). Compared are (a) use of a permanent Rh modifier thermally precoated on a W-filament and (b) direct addition of Rh modifier to the diluent. Note that ions other than the analyte (208Pb) are plotted on the right hand side y-axis. Note the color legend assigned in panel (a) is the same for (b). |
Footnotes |
† Use of trade names is for informational purposes only and does not imply an endorsement by the New York State Department of Health. |
‡ To convert µg dL−1 into µmol L−1, multiply by 0.04826; i.e., 10 µg dL−1 equals 0.48 µmol L−1. To convert µg L−1 into µmol L−1, multiply by 0.004826. |
This journal is © The Royal Society of Chemistry 2003 |