Issue 6, 1996

Method for calibrating olfactometer output. Part 1. Acetic and propionic acids

Abstract

An analytical method for measuring acetic acid and propionic acid in air was developed and validated. This method entails collection of acetic acid and propionic acid on XAD-7 sorbent tubes connected to an olfactometer and extraction of the sorbent in a solution of methanol and formic acid. The acids are then determined by gas chromatography with flame ionization detection. This methodology was successfully implemented in calibration of olfactometer output.

Article information

Article type
Paper

Anal. Commun., 1996,33, 199-202

Method for calibrating olfactometer output. Part 1. Acetic and propionic acids

K. C. Maiolo, J. C. Walker and M. W. Ogden, Anal. Commun., 1996, 33, 199 DOI: 10.1039/AC9963300199

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.