Abstract

Optimum conditions for the rapid qualitative and quantitative determination of trace levels of halides as copper halides by molecular emission cavity analysis with a plain copper cavity are presented. A light green emission was observed for CuCl at 532 nm, an intense blue emission for CuBr at 494 nm and a deep green emission for Cul at 510 nm. Resolution of mixtures of the halides is feasible using an optimised nitrogen-air-hydrogen flame. Linear calibration graphs were obtained for trace concentrations of halides over the ranges 3 × 10–3–1 × 10–2M Cl, 1 × 10–3-6 × 10–3M Br and 1 × 10–3–3 × 10–2M l, respectively.

Article information

Article type
Paper

Analyst, 1989,114, 1483-1487

Determination of trace levels of halides as copper halides by molecular emission cavity analysis

O. Osibanjo, A. O. Bankole and S. O. Ajayi, Analyst, 1989, 114, 1483 DOI: 10.1039/AN9891401483

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.