Micro-flow-injection analysis (μFIA) immunoassay of herbicide residue 2,6-dichlorobenzamide – towards automated at-line monitoring using modular microfluidics

Abstract

As a part of developing new systems for continuously monitoring the presence of pesticides in groundwater, a microfluidic amperometric immunosensor was developed for detecting the herbicide residue 2,6-dichlorobenzamide (BAM) in water. A competitive immunosorbent assay served as the sensing mechanism and amperometry was applied for detection. Both the immunoreaction chip (IRC) and detection (D) unit are integrated on a modular microfluidic platform with in-built micro-flow-injection analysis (μFIA) function. The immunosorbent, immobilized in the channel of the IRC, was found to have high long-term stability and withstand many regeneration cycles, both of which are key requirements for systems utilized in continuous monitoring. The IRC was regenerated during 51 cycles in a heterogeneous competitive assay out of which 27 were without the analyte (the highest possible signal level) in order to assess the regeneration capability of the immunosorbent. Detection of BAM standard solutions was performed in the concentration range from 62.5 μg L−1 to 0.0008 μg L−1. Non-linear regression of the data using the four-parameter logistic equation generated a sigmoidal standard curve showing an IC50 value (concentration that reduces the signal by 50%) of 0.25 μg L−1. The strongest signal variation is observed in the concentration range between 0.02 and 2.5 μg L−1, which includes the 0.1 μg L−1 threshold limit set by the European Commission for BAM in drinking water. The presented results demonstrate the potential of the constructed μFIA immunosensor as an at-line monitoring system for controlling the quality of ground water supply.

Graphical abstract: Micro-flow-injection analysis (μFIA) immunoassay of herbicide residue 2,6-dichlorobenzamide – towards automated at-line monitoring using modular microfluidics

Supplementary files

Article information

Article type
Paper
Submitted
27 Aug 2014
Accepted
17 Jan 2015
First published
19 Jan 2015

Analyst, 2015,140, 1616-1623

Author version available

Micro-flow-injection analysis (μFIA) immunoassay of herbicide residue 2,6-dichlorobenzamide – towards automated at-line monitoring using modular microfluidics

B. Uthuppu, A. Heiskanen, D. Kofoed, J. Aamand, C. Jørgensen, M. Dufva and M. H. Jakobsen, Analyst, 2015, 140, 1616 DOI: 10.1039/C4AN01576B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements