High-performance liquid chromatographic method for the determination of 9-(4-hydroxy-3-hydroxymethylbut-1-yl)guanine (BRL-39123) in human plasma and urine
Abstract
A rapid, sensitive and reliable reversed-phase high-performance liquid chromatographic (HPLC) method with UV detection has been developed for the assay of a novel anti-herpes agent, 9-(4-hydroxy-3-hydroxymethylbut-1-yl)guanine (BRL-39123), in human plasma and urine. The drug and the internal standard, the structural analogue BRL-42377, were extracted from the biological matrix by adsorption on a cation-exchange column and were subsequently eluted under alkaline conditions prior to HPLC. The method is reproducible, with coefficients of variation of ca. 5%, and linear from 0.1 to at least 30 µg ml–1 in plasma and from 50 to at least 2000 µg ml–1 in urine. The method has been used extensively to measure BRL-39123 in plasma and urine samples generated during clinical studies and is adequate for defining pharmacokinetics at projected therapeutic doses.