Yulin
Chen
,
Junwei
Qiu
,
Kai
Xu
,
Huijun
Zhu
,
Shuo
Zhang
,
Xinxin
Lu
and
Xiaoxu
Li
*
School of Mechanical and Electrical Engineering, Soochow University, Suzhou, 215021, China. E-mail: xxli@suda.edu.cn
First published on 3rd January 2025
A portable gas chromatograph–mass spectrometer (GC-MS) is an effective instrument for rapid on-site detection of volatile organic compounds (VOCs). Current instruments typically adsorb samples at ambient temperature, challenging the detection of low-boiling VOCs. In this study, a low-temperature adsorption thermal desorption method is proposed for sample enrichment in a portable GC-MS. The refrigeration module adopts a thermoelectric cooler (TEC), and a heating wire directly heats the adsorption tube to reduce the heat capacity. The miniaturization and low-power design make this module integrable into portable GC-MS devices. This module can reduce the temperature to around 0 °C within ten minutes for sample enrichment, and the heating system can increase the temperature to 260 °C within 20 seconds to ensure rapid desorption and injection of samples. Due to the miniaturization design, the total weight of the portable GC-MS is 21.7 kg, and the volume is 48 cm × 38 cm × 17 cm. Within merely 10 minutes, it completely separated and detected 65 VOCs in the TO-15 standard substance, with a detection limit down to 0.12 μg L−1 for toluene. The detection performance for low-boiling substances could be enhanced by up to 17 times compared to ambient temperature adsorption thermal desorption, such as 1,3-butadiene. Moreover, the results demonstrated long-term stability (RSD < 10% for 98% of the substances, with recovery rates from 91.66% to 109.12%). This study provides a feasible strategy for the rapid and reliable detection of VOCs in the air, holding great potential in the field of environmental monitoring.
The most commonly used methods for detecting VOCs include gas chromatography and gas chromatography–mass spectrometry.20 Gas chromatography can effectively separate complex mixtures based on differences21 in adsorption properties and reduce interference between compounds. The detection of speciated VOCs can be achieved by a range of instruments, such as a mass spectrometer (MS), flame ionization detector (FID)22 and photoionization detector (PID). Compared to other detectors, a mass spectrometer offers high sensitivity and resolution. It provides information regarding the molecular structure of a compound, which is useful for identifying unknown compounds.23 In recent years, breakthroughs have been achieved in miniaturization technology, leading to the development of portable gas chromatography and gas chromatography–mass spectrometry systems for on-site detection of VOCs. While these instruments offer the advantages of rapid analysis, ease of operation, and low power consumption, their performance and sensitivity are challenging to match with laboratory-grade instruments. Therefore, in practical use, sample preconcentration techniques are indispensable.
Sample preconcentration techniques are employed to increase the concentration of target analytes, facilitating the detection and quantification of trace or low-concentration compounds. Prominent methods employed for this purpose include solvent extraction,24 headspace techniques,25,26 adsorption on solid sorbents,27 and the full evaporation technique. Adsorption on solid sorbents is a commonly used method, which includes ambient temperature adsorption sampling and low-temperature adsorption sampling. Currently, the commonly employed adsorption thermal desorption technology in portable gas chromatography–mass spectrometry instruments relies on ambient temperature adsorption sampling, followed by the heating of the adsorbent to release the adsorbed target compounds for analysis. This technology has now been widely applied. For instance, Liu et al. developed a novel method for detecting Huanglongbing (HLB),28 which combines headspace solid-phase microextraction with a portable GC-MS for the in situ field detection of volatile metabolites in citrus leaves. Additionally, Wang et al. presented a method for rapid in situ detection of VOCs in soil gas and groundwater using a portable GC-MS coupled with solid-phase microextraction,29 demonstrated through a case study of a trichloroethene-contaminated site in Australia. Although the application of a portable GC-MS using ambient temperature adsorption thermal desorption technology is currently extensive, it also has certain limitations. Sampling with adsorbents at ambient temperature requires an extended sampling duration, and it is prone to breakthrough30–34 for less volatile substances, exhibiting limited capture efficiency. Low-temperature adsorption sampling is an effective means of pre-concentrating an ambient air sample collected from continuous flow for online measurements on adsorbent materials at a specific temperature significantly lower than the ambient temperature, a process also known as “cryofocusing”,35 which significantly enhances the adsorption efficiency.36–38
In traditional approaches, the low temperature in adsorption tubes is usually achieved by liquid nitrogen or a coolant.38 Apel et al. developed a fast response GC-MS for onboard aircraft39 to measure carbonyl compounds from C2 to C4 and methanol, using liquid nitrogen for cryofocusing. The coolant used to circulate the refrigeration provides excellent cooling. Wang et al. developed an automated GC-MS/FID system22 that eliminates the need for liquid nitrogen, using a multi-component refrigerant blend to achieve ultra-low temperatures through a cascade of refrigeration cycles for VOC enrichment. All these methods have achieved cryofocusing and are sensitive enough to detect low pptv levels of all target analytes. However, none of them can achieve miniaturization, and it is not suitable to integrate them into a portable GC-MS. Peltier coolers, commonly referred to as thermoelectric coolers,40 exhibit a significant value in cooling technology by harnessing the Peltier effect for heat transfer. This technology generates a cooling effect through the application of an electric current across dissimilar conductive materials, thereby creating a temperature differential. Peltier coolers have the advantages of simple structure and miniaturization and are fully electrically driven.40 They do not require refrigerants and do not involve any potentially dangerous working fluids (only helium), making them suitable for building miniaturized cryofocusing devices that meet the requirements of compact instruments. Although their cooling capacity and minimum temperature cannot compete with refrigerant-based systems, they can still provide effective cooling in applications that do not require extremely low temperatures. In particular, Peltier coolers can respond quickly, rapidly cooling from ambient temperature to the target temperature. This feature is of great significance for instruments that require rapid temperature changes or maintaining specific temperature ranges. Guan et al. have used a Peltier cooler as a cold modulator for 2D gas chromatography,2,41 reducing cost and power consumption without affecting the performance of the device.
Based on the above content, it could be seen that cryofocusing is widely used in desktop GC-MSs, but there is little related research on portable GC-MSs. This study proposes an innovative idea of using the adsorption thermal desorption technology of ‘low-temperature + adsorbent’ in a portable GC-MS, and develops a miniaturized fast gas chromatograph–linear ion trap–mass spectrometer (GC-LIT-MS) for the detection of VOCs in the air. By embedding a low-temperature adsorption thermal desorption module in the existing portable GC-MS, the system completes the separation and detection of 65 VOCs in the air within 10 minutes, with a minimum detection limit of 0.12 μg L−1 for toluene, and a maximum of 17 times increase in comparison with that of using ambient temperature adsorption thermal desorption in the peak area for the detection performance of low-boiling substances. In addition, the measurement results show long-term stability and reliability, providing a feasible strategy for the rapid and reliable detection of VOCs in the air that has great potential in the field of environmental monitoring.
The Summa canisters were cleaned with a canister cleaner at elevated temperatures prior to use, and the canisters were evacuated (<50 mTorr) after cleaning and set aside for use. A high-precision static diluter was used to dilute the TO-15 standard with high-purity nitrogen to a mixture of standard gases at concentrations of 2, 10, 25, and 50 ppb for storage in the Summa canisters, with care taken to humidify the canisters to minimize adsorption of compounds into the canisters.
The adsorption tube, filled with a solid phase adsorbent, is used to pre-concentrate the sample. Features to be considered include the size and internal volume, the mass and type of adsorbent used, and the temperature of operation. These properties can be tailored for optimum performance, depending on the analyte of interest and the amount of preconcentration required to bring that compound into the detection range of the detector. To prepare the adsorption tube, stainless steel tubing was cut into a 90 mm length, cleaned by repeated flushing with ethanol and acetone, and oven dried. Quartz wool was inserted into the tube and was secured by packing the end of the tube with compacted quartz wool. The tube is integrated into the gas circuit by securing polyetheretherketone (PEEK) end caps with internal threads. The sealing between the two is achieved through an O-shaped fluororubber ring. The heating wire and thermocouple are evenly wound on the surface of the stainless steel tube, achieving a heating rate up to 28 °C s−1 with precise and rapid temperature control.
Different types of adsorbents were used when trapping VOCs from the atmosphere,43 including graphitized carbon materials such as Carbotrap and carbon-based molecular sieves (CMS). Tenax is a polytetrafluoroethylene (PTFE)-based adsorbent and one of the most widely used for the pre-desorption of VOCs. It is characterized by high thermal stability, oxidation resistance, and low adsorption for water, and good adsorption for most of the substances with medium and high boiling points, making it suitable for the enrichment of VOCs in the range of C6 to C26.
The thermoelectric cooling module comprises a Peltier cooler, a dissipator, and a fan, and is constructed as shown in Fig. 1. An alumina plate is utilized to secure and encase the tube. The cold junction of the Peltier cooler is coated with a thermally conductive resin and firmly attached to an aluminum oxide block. When current passes through, the cold side absorbs heat, while the hot side releases heat, in accordance with the Peltier effect. By integrating a thermoelectric cooling system with an adsorption tube, the tube is housed within a cavity structure for non-contact air cooling. This configuration ensures uniform cooling and reduced energy consumption. The temperature of the tube can be lowered from ambient temperature to zero degrees in six minutes.
![]() | ||
| Fig. 1 The structure of the thermoelectric cooling module with the TEC system: (a) schematic view of the device; (b) picture of the device embedded in a portable GC-MS. | ||
The thermal behavior of the adsorption tube by experimental observation was studied. Fig. 2 shows the heating and cooling performance of the adsorption tube with a combination of natural convection cooling and forced convection cooling.
It could be observed that the adsorption tube reaches a temperature of 240 °C within 15 seconds, which is sufficient for complete desorption. Once the fan is activated, the temperature of the adsorption tube cools down to the set temperature of 2 °C within approximately 10 minutes. It is worth noting that in the absence of the cooling device, the temperature of the adsorption tube only slowly decreases to ambient temperature over a period of 10 minutes.
The EI method adopted by the instrument is suitable for thermally stable and volatile compounds and is compatible with chromatographic techniques. The outlet of the chromatographic column is inserted into the ion source through a heated sleeve for correct alignment and connection. The gas flow rate is adjusted to 0.2 ml min−1 to adapt to the vacuum conditions of the MS section. For the EI source and linear ion trap, the operating vacuum degree is required to be between 10−4 Torr and 10−5 Torr. Different from the original instrument, this instrument selects the turbomolecular pump (HiPace 80Neo, Pfeiffer Vacuum Technology Inc., Germany) as the vacuum pump. This pump has the advantages of small size, light weight, and anti-aging, and is suitable for integration into miniaturized and portable devices. Moreover, its pumping speed for nitrogen can reach up to 67 L s−1, achieving a vacuum degree that is one order of magnitude higher than that of the original instrument.
Considering the wide boiling point range and complex composition of VOCs, this study used a DB-5 chromatographic column (size 30 m × 0.25 mm × 0.25 μm) produced by Agilent Technologies (USA) to separate the compounds in the TO-15 standard substance. Compared with the column used in the original instrument, the column capacity of the new column is increased, thus better fulfilling the requirements of the increased pumping speed of the vacuum pump on the system carrier gas flow rate. Meanwhile, the novel chromatographic column exhibits an enhanced column performance, which enhances the separation effect of the sample, thereby improving the sensitivity of the instrument detection.
The adsorption tube was cooled down to approximately 0 °C for sample enrichment. In the injection process, the sampling time was set to 3 min, the spiking time was 0.15 min, the pre-desorption time was 0.2 min, and the desorption time was 0.7 min. In the separation process, the initial temperature (50 °C) was held for 2 min, then increased to 120 °C at a rate of 25 °C min−1, and finally raised to a final temperature of 260 °C at a rate of 30 °C min−1, with a hold time of 1 min. The whole process lasted for 10.47 min.
This instrument utilizes resistance heating for the chromatographic column. The chromatographic column, heating wire, and temperature sensor are arranged in a circular configuration to maximize the heat exchange area while minimizing component exposure. A sealing tape is applied around the ring surface to ensure airtightness and facilitate uniform heat conduction, thereby maintaining a stable column temperature.
Using K-type thermocouples to measure temperature, the thermoelectric potential signal generated by the temperature gradient is amplified and converted into a digital signal. According to the PID control principle, the collected signal is compared with the set value. By integrating the parameters of proportional, integral, and derivative, automatic temperature control is achieved.
Temperature also significantly affects the desorption process, with higher temperatures promoting desorption of adsorbed VOCs. Therefore, the desorption temperature should be optimized to ensure complete desorption without thermal decomposition or degradation. Temperature control is also critical in preventing breakthrough phenomena, where high temperatures may cause the permeation of compounds through the adsorbent, reducing trapping efficiency.
Understanding the influence of temperature on adsorption and desorption is crucial for the successful application of low-temperature preconcentration techniques in a portable GC-MS to achieve accurate and sensitive VOC analysis. Experimental optimization of temperature parameters enables efficient adsorption and desorption, thereby improving the performance and capabilities of the instrument for environmental monitoring and analytical applications.
To evaluate the adsorption efficiency of the low-temperature preconcentration device, TO-15 standard samples with a concentration of 50 ppb were analyzed at adsorption temperatures of 0 °C and 35 °C, as shown in the total ion chromatograms in Fig. 4. It can be observed from the figure that, under the influence of the low-temperature preconcentration device, the overall peak intensity of VOCs is significantly enhanced, particularly for low-boiling VOCs with retention times of less than 2 minutes and 30 seconds, which exhibited a five-fold or greater increase in peak intensity. As the temperature increases, the adsorption efficiency of the adsorbent for VOCs decreases, and breakthrough phenomena become more pronounced. Therefore, the low-temperature device effectively prevents the permeation of low-boiling substances during the adsorption process, leading to increased peak intensities. For high-boiling substances, the concentration effect of the low-temperature preconcentration device noticeably diminishes, resulting in only a slight increase in peak intensity.
To further investigate the device's detection performance for low-concentration samples, TO-15 standard samples with a concentration of 10 ppb were analyzed at 0 °C and 35 °C, and the peak areas of some of the low-boiling compounds were compiled to observe the trapping of five specific low-boiling compounds (chloroethylene, 1,3-butadiene, monobromomethane, dichloromethane, and n-hexane), as shown in Fig. 5. The results indicate that under low-temperature adsorption conditions, the five substances exhibited increases in the peak area of 15.88-fold, 17.45-fold, 14.82-fold, 11.57-fold, and 13.91-fold, respectively (see Table 1). This suggests that for lower-concentration low-boiling samples, the effect of the low-temperature device becomes more pronounced, with enhancements of up to 11–17-fold. When the sample concentration is below 10 ppb, certain substances cannot be detected at ambient temperature, but only at low temperatures.
![]() | ||
| Fig. 5 Comparison of the peak area of the samples at 0 °C and 35 °C: (a) chloroethylene; (b) 1,3-butadiene; (c) monobromomethane; (d) dichloromethane; and (e) n-hexane. | ||
| Compound | Retention time | Peak area (0 °C) | Peak area (35 °C) |
|---|---|---|---|
| Chloroethylene | 1.539 | 2208 | 139 |
| 1,3-Butadiene | 1.545 | 8638 | 495 |
| Monobromomethane | 1.582 | 2564 | 173 |
| Dichloromethane | 1.768 | 2303 | 199 |
| n-Hexane | 1.999 | 39 071 |
2808 |
In a series of experiments with varying sample concentrations, after each analysis, a blank was performed immediately after desorption to evaluate any carryover in the adsorbent within the adsorption tube.
| No. | Compounds | Retention time | RSD | LOD | Recovery |
|---|---|---|---|---|---|
| 1 | Propylene | 1.317 | 3.44% | 0.26 | 101.28% |
| 2 | 1,1,2,2-Tetrafluoro-1,2-dichloroethane | 1.352 | 9.16% | 0.55 | 91.88% |
| 3 | Dichlorodifluoromethane | 1.364 | 9.52% | 0.58 | 97.14% |
| 4 | Chloroethylene | 1.384 | 6.96% | 0.46 | 100.96% |
| 5 | 1,3-Butadiene | 1.404 | 9.97% | 0.58 | 92.27% |
| 6 | Monochloromethane | 1.404 | 7.14% | 0.46 | 102.33% |
| 7 | Ethyl chloride | 1.421 | 9.73% | 0.13 | 98.67% |
| 8 | Monobromomethane | 1.436 | 8.33% | 0.58 | 108.53% |
| 9 | Ethanol | 1.488 | 9.72% | 0.59 | 94.91% |
| 10 | Trichloromethane monofluoride (Freon 11) | 1.500 | 5.91% | 0.38 | 101.87% |
| 11 | Acrolein | 1.508 | 2.12% | 0.16 | 100.16% |
| 12 | Acetone | 1.524 | 6.44% | 0.40 | 98.82% |
| 13 | Isopropyl alcohol | 1.548 | 3.10% | 0.19 | 100.50% |
| 14 | 1,1-Dichloroethylene | 1.564 | 9.99% | 0.65 | 103.24% |
| 15 | 1,2,2-Trifluoro-1,1,2-trichloroethane (Freon 113) | 1.572 | 5.22% | 0.24 | 100.41% |
| 16 | Dichloromethane | 1.608 | 9.34% | 0.60 | 102.99% |
| 17 | Carbon disulfide | 1.628 | 8.04% | 0.56 | 97.07% |
| 18 | trans-1,2-Dichloroethylene | 1.697 | 3.57% | 0.24 | 107.77% |
| 19 | Methyl tert-butyl methyl ether | 1.701 | 9.83% | 0.55 | 91.66% |
| 20 | Vinyl acetate | 1.741 | 8.75% | 0.56 | 102.20% |
| 21 | 1,1-Dichloroethane | 1.745 | 3.89% | 0.29 | 100.02% |
| 22 | n-Hexane | 1.805 | 3.21% | 0.20 | 100.02% |
| 23 | 2-Butanone | 1.817 | 8.51% | 0.51 | 99.08% |
| 24 | cis-1,2-Dichloroethylene | 1.857 | 6.70% | 0.43 | 101.01% |
| 25 | Ethyl acetate | 1.886 | 4.22% | 0.26 | 98.37% |
| 26 | Chloroform | 1.946 | 7.82% | 0.57 | 97.32% |
| 27 | Tetrahydrofuran | 2.002 | 6.52% | 0.40 | 98.61% |
| 28 | 1,1,1-Trichloroethane | 2.086 | 3.80% | 0.25 | 104.20% |
| 29 | 1,2-Dichloroethane | 2.135 | 2.07% | 0.13 | 99.05% |
| 30 | Benzene | 2.195 | 7.75% | 0.56 | 99.25% |
| 31 | Carbon tetrachloride | 2.203 | 8.17% | 0.56 | 109.12% |
| 32 | Cyclohexane | 2.207 | 7.71% | 0.43 | 103.54% |
| 33 | n-Heptane | 2.469 | 11.23% | 0.80 | 98.19% |
| 34 | Trichloroethylene | 2.497 | 7.71% | 0.50 | 102.58% |
| 35 | 1,2-Dichloropropane | 2.529 | 8.76% | 0.70 | 97.63% |
| 36 | Methyl methacrylate | 2.594 | 4.70% | 0.29 | 99.76% |
| 37 | Monobromodichloromethane | 2.622 | 9.90% | 0.76 | 97.22% |
| 38 | 1,4-Dioxane | 2.638 | 9.84% | 0.60 | 96.98% |
| 39 | cis-1,3-Dichloropropene | 2.879 | 3.44% | 0.23 | 107.90% |
| 40 | 4-Methyl-2-pentanone (methyl isobutyl ketone) | 2.891 | 4.99% | 0.61 | 97.93% |
| 41 | trans-1,3-Dichloropropene | 3.156 | 5.48% | 0.36 | 104.89% |
| 42 | Toluene | 3.160 | 1.73% | 0.12 | 106.15% |
| 43 | 1,1,2-Trichloroethane | 3.272 | 3.26% | 0.22 | 102.22% |
| 44 | 2-Hexanone | 3.397 | 8.77% | 0.59 | 106.10% |
| 45 | Dibromochloromethane | 3.546 | 5.96% | 0.35 | 93.93% |
| 46 | Tetrachloroethylene | 3.571 | 5.93% | 0.36 | 96.85% |
| 47 | 1,2-Dichloroethane | 3.648 | 8.07% | 0.54 | 107.18% |
| 48 | Chlorobenzene | 4.033 | 5.99% | 0.40 | 107.34% |
| 49 | Ethylbenzene | 4.180 | 7.04% | 0.49 | 100.60% |
| 50 | m-Xylene | 4.288 | 7.63% | 1.03 | 204.43% |
| 51 | Styrene | 4.521 | 8.60% | 0.56 | 102.91% |
| 52 | o-Xylene | 4.525 | 7.37% | 0.50 | 107.00% |
| 53 | Bromoform | 4.545 | 7.60% | 0.38 | 94.54% |
| 54 | 1,1,2,2-Tetrachloroethane | 4.825 | 4.32% | 0.29 | 106.02% |
| 55,56 | m/p-Ethyltoluene | 5.255 | 9.46% | 0.55 | 92.89% |
| 57 | 1,3,5-Trimethylbenzene | 5.303 | 8.85% | 0.55 | 98.25% |
| 58 | 1,2,4-Trimethylbenzene | 5.538 | 9.26% | 0.55 | 94.01% |
| 59 | m-Dichlorobenzene | 5.657 | 8.18% | 0.48 | 92.61% |
| 60 | Benzyl chloride | 5.725 | 9.30% | 0.58 | 99.94% |
| 61 | p-Dichlorobenzene | 5.741 | 8.84% | 0.43 | 96.88% |
| 62 | o-Dichlorobenzene | 5.902 | 7.53% | 0.37 | 98.39% |
| 63 | 1,2,4-Trichlorobenzene | 7.011 | 6.09% | 0.19 | 100.51% |
| 64 | Naphthalene | 7.099 | 6.45% | 1.08 | 95.47% |
| 65 | Hexachloro-1,3-butadiene | 7.215 | 6.46% | 0.88 | 97.68% |
The peak areas of different samples exhibited moderate variations, while the variations in retention times of different samples were negligible. The peak areas of 98% of the compounds had relative standard deviations (% RSD) of less than 10%, indicating that the portable GC-MS analysis using an integrated thermoelectric cooler is repeatable.
Nevertheless, further testing is necessary to assess potential interferences and other performance-related issues. Future research will focus on field testing, instrument validation, and the utilisation of integrated sampling and pre-processing techniques for the in situ detection of actual polluted exhaust gases in diverse environments.