Issue 10, 2018

Internal standards in inductively coupled plasma mass spectrometry using kinetic energy discrimination and dynamic reaction cells

Abstract

ICP-MS is a sensitive element analysis technique used for analyzing several different sample types. This can result in difficult matrixes which can affect both physical parameters and create overlaps of analyte elements. Some of the possible overlaps can be reduced by the use of reaction and/or collision cells, while the use of internal standards can help with reducing the physical interferences caused by a matrix. While both internal standardization and the use of cells have been studied separately, their effects on each other have not been investigated earlier. In this study ICP-MS was used to analyze 24Mg, 27Al, 47Ti, 49Ti, 51V, 52Cr, 53Cr, 55Mn, 57Fe, 59Co, 60Ni, 61Ni, 62Ni, 63Cu, 65Cu, 66Zn, 67Zn, 75As, 78Se, 82Se, 111Cd, and 208Pb with 9Be, 89Y, 69Ga, 103Rh, 115In, 193Ir, and 205Tl as internal standards with high concentrations of either HNO3, PBS-buffer, or Triton X-100 as the matrix, in reaction-, collision- and standard-cell modes. This was done to investigate which internal standards would compensate matrix effects in different cell modes. All internal standards, except Be, compensated fairly well (relative sensitivity RSD < 10%) even for severe matrix effects for most elements regardless of similarity in mass in the different cell modes. For Zn, As and Se no proper internal standard could be found, of the ones investigated.

Graphical abstract: Internal standards in inductively coupled plasma mass spectrometry using kinetic energy discrimination and dynamic reaction cells

Supplementary files

Article information

Article type
Paper
Submitted
30 Mey 2018
Accepted
28 Aga 2018
First published
28 Aga 2018
This article is Open Access
Creative Commons BY license

J. Anal. At. Spectrom., 2018,33, 1770-1776

Internal standards in inductively coupled plasma mass spectrometry using kinetic energy discrimination and dynamic reaction cells

M. Korvela, M. Andersson and J. Pettersson, J. Anal. At. Spectrom., 2018, 33, 1770 DOI: 10.1039/C8JA00171E

This article is licensed under a Creative Commons Attribution 3.0 Unported Licence. You can use material from this article in other publications without requesting further permissions from the RSC, provided that the correct acknowledgement is given.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements