Highly parallel, wash-free, and ultrasensitive centrifugal droplet digital protein detection in sub-microliter blood†
Abstract
The ability to efficiently detect low-abundance protein biomarkers in tiny blood samples is a significant challenge in clinical and laboratory settings. Currently, high-sensitivity approaches require specialized instrumentation, involve multiple washing steps, and lack the ability to parallelize, preventing their widespread implementation. Herein, we developed a parallelized, wash-free, and ultrasensitive centrifugal droplet digital protein detection (CDPro) technology that achieves a femtomolar limit of detection (LoD) of target proteins with sub-microliters of plasma. The CDPro combines two techniques, namely a centrifugal microdroplet generation device and a digital immuno-PCR assay. Miniaturized centrifugal devices enable emulsification of hundreds of samples within 3 minutes using a common centrifuge. The bead-free digital immuno-PCR assay not only eliminates the need for multistep washing, but also possesses ultra-high detection sensitivity and accuracy. We characterized the performance of CDPro using recombinant interleukins (IL-3 and IL-6) as example targets and reported a LoD of 0.0128 pg mL−1. We also quantified IL-6 from 7 human clinical blood samples using the CDPro with only 0.5 μL plasma, which showed excellent agreement with an existing clinical protein diagnostic system with 25 μL plasma from those samples (R2 = 0.98).
- This article is part of the themed collection: Lab on a Chip HOT Articles 2023