Seyed Oveis
Mirabootalebi
and
Yang
Liu
*
College of Science and Engineering, James Cook University, Townsville, Queensland 4811, Australia. E-mail: yang.liu11@jcu.edu.au
First published on 4th June 2024
Solid-contact ion-selective electrodes (SC-ISEs) are advanced potentiometric sensors with great capability to detect a wide range of ions for the monitoring of industrial processes and environmental pollutants, as well as the determination of electrolytes for clinical analysis. Over the past decades, the innovative design of ion-selective electrodes (ISEs), specifically SC-ISEs, to improve potential stability and miniaturization for in situ/real-time analysis, has attracted considerable interest. Recently, the utilisation of nanomaterials was particularly prominent in SC-ISEs due to their excellent physical and chemical properties. In this article, we review the recent applications of various types of nanostructured materials that are composed of carbon, metals and polymers for the development of SC-ISEs. The challenges and opportunities in this field, along with the prospects for future applications of nanomaterials in SC-ISEs are also discussed.
In recent years, the innovative design of ISEs by integrating advanced materials into different components such as membranes, transducers and their interfaces has attracted considerable interest to improve the performance of ISEs.10–13 Nanomaterials are an important group of advanced materials, encompassing a wide range of structures such as nanospheres, nanotubes, nanorods, nanowires, nanofibers, nanoclusters, and quantum dots, in which at least one of their dimensions is less than 100 nm.14 Due to ultra-high surface areas, nanomaterials show unique properties compared with bulk materials,15 making them promising candidates for the development of the next generation of SC-ISEs. For example, nanomaterials with high conductivity, light insensitivity, and low interference by O2 and CO2 have better signal stability than the bulk materials.
The application of nanomaterials as a component of ISEs was initiated in 1994 when Diamond et al.16 predicted the development of new generations of intelligent sensing systems, with the rapid progress in nanotechnology. In 1997, Ghadiri et al.17 introduced a novel diffusion-limited size-selective ion sensor based on monolayer-supported peptide nanotubes on a gold film. Nowadays, many types of nanomaterials including carbon nanotubes (CNTs), graphene, fullerene, carbon black, three-dimensionally ordered microporous (3DOM) carbon, metal and polymer nanomaterials, as well as nanocomposites have been investigated for the development of the next generation of ISEs.18–21
In recent years, although there have been several review papers published in the area of SC-ISEs, which have provided comprehensive information on mechanisms and applications of SC-ISEs,11,22,23 none of them have focused on the evaluation, discussion, and identification of the challenges in nanomaterial-based SC-ISEs, which are of significant importance for the future development of low-cost and high-performance chemical sensors. Additionally, regarding SC-ISEs based on nanomaterials, most of these reviews focus on the use of nanomaterials as transducers of SC-ISEs. Although it is the most common way to improve the sensing performance,11,22,23 the benefit of using nanomaterials in other parts of ISEs should not be ignored. In this review, an overview of recent advances in nanomaterial-based SC-ISEs over the last five years is provided, with an emphasis on the applications of various types of nanomaterials as different components of SC-ISEs. Furthermore, comments on the main progress and challenges, as well as the potential of using nanomaterials for the design of SC-ISEs with improved sensing performance are also provided and discussed.
The principles of ion-selective sensors are rooted in the relationship between ion activity and output voltage, as defined by the Nernst equation.6,27 At electrochemical equilibrium of an ISE, the solution/membrane interface undergoes the formation of an electrical double layer, leading to the generation of an electrical potential difference. The following equation predicts a linear dependence of the potentiometric sensor response (E) based on a function of the ion's activity:28
(1) |
Converting the ionic signal to the detectable electrical signal, which is performed through the transducer layer, plays a significant role in the potential stability, lifetime, sensitivity, response time, and reproducibility of SC-ISEs. The ion-to-electron transition mechanism in traditional liquid-contact ISEs is based on an internal reference electrode and an inner filling solution, which enable the ion transition via a reversible redox reaction. While this approach offers advantages such as the simplicity of the process, it encounters limitations like portability, lifetime, and fragility due to the nature of the liquid contact. In SC-ISEs, the charge transition from an ion to an electron is governed by the transducer functional material. There are two main types of SC-ISEs, requiring distinct ion-to-electron transducer materials. The transducer based on redox capacitance typically involves the use of redox polymers or other electroactive materials, such as redox couple doped gold nanostructures. The ion-to-electron procedure is like the process for the internal reference electrode of traditional ISEs as described by:
M+A− + e ⇄ M + A− |
Although the SC-ISEs exhibit improved performance, there are several problems to address, including potential instability and poor reproducibility. For example, dissolved O2 and CO2, which may reach the solid contact, can generate side reactions in parallel to the ion-to-electron transduction procedure.33 These reactions can interfere with the electrochemical processes involved in the detection of target ions. In a similar way, exposure to light during measurements, especially in harsh environmental situations, e.g., intense sunlight, can induce photochemical reactions that may interfere with the electrochemical processes, resulting in a poor response accuracy of SC-ISEs. Furthermore, a water layer formed between the membrane and the solid contact of SC-ISEs serves as an electrolyte reservoir, which equilibrates after every change in the sample's composition, leading to high limits of detection (LODs) and low potential stability.34,35
To overcome these challenges, different types of nanomaterials have been employed in SC-ISEs. Although nanomaterials have been employed in the components of external reference electrodes and ion-selective membranes,36–38 most studies have concentrated on enhancing the performance of transducers with nanomaterials. Fig. 1 provides a schematic diagram of the components of a typical nanomaterial-based SC-ISE, in which nanomaterials are integrated into the transducer layer of the ISE by drop casting. The emphasis lies on improving the ion-to-electron transduction process by increasing electrical conductivity, capacitance, hydrophobicity, and surface area.
Rius et al.58 pioneered the application of CNTs in SC-ISEs by using single-walled carbon nanotubes (SWCNTs) as an ion-to-electron transducer, which exhibited good signal stability (25 days) and a short response time (fewer than 10 s for activities over 10−5.5 M) for the detection of K+. Although it is more challenging to design SC-ISEs for the determination of anions than cations due to the smaller charge-to-radius ratio of anions, their propensity to become protonated at low pH, the influence of solvation effects, and the variability in their geometric sizes,59 CNTs used as the inner transducer layer in SC-ISEs exhibit high capabilities for detecting both anions and cations.60 The capacitance of the transducer layer of ISEs plays a significant role in potential stability and can also have an impact on important parameters like reproducibility and long-term performance. Zdrachek and Bakker61 reported that the double-layer capacitance increased linearly by increasing the number of deposited CNT-based transducer layers, which led to a reduced potential drift. This method could also serve as a new way to estimate the double-layer capacitance of the SWCNT layer in the deposition process.
Serving as a reference electrode is another application of CNTs in SC-ISEs. The most commonly used reference electrode, the Ag/AgCl electrode, is not very appropriate for sweat analysis, as the potential of the electrode is highly affected by the concentration of chloride ions. Toor et al.62 embedded a layer of CNTs between the membrane and Ag/AgCl layer to adsorb and retain chloride ions for the detection of Na+, NH4+, and lactate in sweat. This method resulted in a reproducible and stable reference electrode that exhibited a negligible potential fluctuation up to 0.08 mV in a solution with chloride concentrations varying between 0.1 and 100 mM, while practical applications such as real-time monitoring of sweat on skin are expected to be investigated, considering the common issues of reproducibility and long-term stability associated with interference from the electrolytes in sweat. Hanein et al.63 developed a wearable CNT-based SC-ISE for Na+ detection by using CNT electrode arrays in both the solid contact and reference electrode. CNT electrodes were prepared by the synthesis of CNTs via chemical vapor deposition (CVD) on SiO2/Si substrates. Subsequently, reference electrodes were prepared by coating the CNT electrodes with a colloidal dispersion of Ag/AgCl, agarose, NaCl, and polyvinyl chloride (PVC). The CNT-based reference electrodes showed high repeatability and low sensitivity (−1.7 ± 1.2 mV per decade) in the NaCl solution. Despite the fact that some important characterization studies including potential drift were not provided, this paper was pioneering in the application of CNTs in wearable potentiometric sensors and opened a new avenue for subsequent studies.
Carbon paste is an attractive type of electrode due to its good reusability, simple and versatile preparation procedure, and low cost.64,65 It is generally composed of a wide range of carbon materials, additives and solvents66 and has been used in ISEs for several decades.67 In recent years, nanocarbon-based paste, particularly CNT paste, has been introduced into ISEs, which have exhibited improved performance for ion sensing.68–71 For example, Ramezani and co-workers72 developed multi-walled carbon nanotube (MWCNT)-grafted 2,6-bis[2-(amino methyl)phenol]pyridine (BAPP) synthetic ligands as novel ion carriers and signal boosters in carbon paste for the simultaneous detection of Hg2+ and Cu2+. The modified ISE displayed satisfactory reproducibility, with LODs of 2.0 × 10−9 and 7.9 × 10−10 mol L−1 for Hg2+ and Cu2+, respectively.
Wu and co-workers85 proposed an SC-ISE using electrochemically reduced graphene oxide as the ion-to-electron transducer for Ca2+ detection, which exhibited a Nernstian slope of 29.1 mV per decade and a fast response time of less than 10 s. Moreover, a low detection limit of 10−5.8 M and high hydrophobicity were obtained. Additionally, the effects of defects in graphene on capacitance and hydrophobicity were investigated by Gan et al.86Fig. 2 illustrates the use of defective or high-quality graphene as the solid contact in an SC-ISE. The photograph of the contact angle test for reduced graphene oxide shows a contact angle (θ = 113°) after 3 hours of reduction at 160 °C. It was found that there was a negative correlation between the hydrophobicity of graphene and its capacitance. Notably, reduced graphene oxide with a moderate level of defects demonstrated the highest potential stability by achieving a balance between hydrophobicity and capacitance.
Fig. 2 Graphical summary of defective and high-quality graphene used as the solid contact material in an ISE, along with a photograph of the contact angle of water on reduced graphene oxide, and a schematic figure of the interfacial EDL capacitance of ISE.86 Copyright 2021, Elsevier. |
Recently, a new method based on laser-induced graphene was proposed for the preparation of SC-ISEs for NO2− detection.87 These sensors exhibited a broad linear sensing range with high potential stability without the formation of the water layer due to the hydrophobic characteristics of the electrode. The outstanding properties can be attributed to the double-laser treatment of the graphene surface, resulting in facile electron transfer kinetics, high surface porosity, and superior conductivity.88,89 Additionally, Gomes and co-workers90 employed a double-laser process to produce hydrophobic laser-induced graphene, which was coated with an ion-selective membrane to form an SC-ISE for the detection of NO3−. The proposed ISE exhibited a low LOD (6.01 ± 1.44 μM), and the performance remained almost stable over a period of 5 weeks. The application of different types of graphene and CNTs as a transducer layer in SC-ISEs is well developed for a wide range of target analytes. However, their high costs and tedious preparation procedures become barriers in the further advancement of these SC-ISEs. Therefore, cost-effective carbon-based nanomaterials with similar physical and chemical properties to those of graphene and CNTs were investigated to circumvent these limitations.
Fig. 3 (A) Schematic diagram of SC-ISEs based on NMC, RGO, and CNTs. (B) SEM images of NMC, RGO, and CNTs. (C) Chronopotentiometric curves obtained for the NMC-based, RGO-based, and CNT-based SC-ISEs for the detection of Ca2+.96 Copyright 2022, MDPI. |
Fig. 4 (A) The process of preparing a home-made TMA sensor for vaginosis analysis. (B) A SEM cross-section image of the membrane coated TMA sensor (scale bar is 50 μm). (C and D) Photographs of the sensor depicting its portability and size.102 Copyright 2023, MDPI. |
Nanostructured carbon black is a cost-effective nanomaterial that can be obtained through straightforward manufacturing approaches and further modified to increase its surface area.97,98 It has been successfully employed as a solid contact in ISEs due to its high porosity, good conductivity, hydrophobicity, and inert nature. Paczosa-Bator first developed an SC-ISE based on carbon black for determining K+.99 In a recent study, Michalska et al.100 introduced a simple method to prepare a 3D-drawn substrate using a 3D pen containing carbon black/polymer for the preparation of disposable ISEs. This process enables electrodes to be made with desirable shapes and high device-to-device reproducibility. Moreover, carbon black has been explored to produce paste electrodes for ISE applications. Paczosa-Bator et al.101 recently proposed a paste electrode consisting of carbon black, ruthenium, iridium, and poly(3-octylthiophene-2,5-diyl) to monitor NO3− in soil. The obtained results were close to the Nernst equation and showed a negligible response to interfering ions such as Cl−, CH3COO−, SO42−, and HPO42−, making this method suitable for analysing real samples. Furthermore, Mousavi and co-workers102 developed a home-made SC-ISE for the rapid detection of trimethylamine (TMA) using a carbon black cotton string and a TMA selective membrane, which could be applied for the clinical diagnosis of bacterial vaginosis. Fig. 4 shows the main steps for preparing the TMA sensor, its morphological image, as well as photos of the portable sensing device. Carbon black has also been used as a solid contact in SC-ISEs for the determination of Na+, K+, NO3−, and acetylcholine,103–108 exhibiting a Nernstian response with low LODs.
Additionally, hollow carbon nanospheres were used as the solid contact to significantly enhance the ion-to-electron transduction between the conductive substrate and the ion-selective membrane for the monitoring of Ca2+.109 The hydrophobicity, high surface area, and unique hollow structure of carbon nanospheres as a solid contact are the main reasons for the enhancement of ion-to-electron transduction. Carbon nanohorns,110 carbon nanofibers,111,112 and graphdiyne oxide113 were also employed as the ion-to-electron transducer in ISEs for the detection of Ca2+, Pb2+, moxifloxacin, and K+, which achieved good potential stability and sensitivity (close to a Nernstian response). Pretsch et al.114 applied redox-active and lipophilic fullerene, which is a spherical form of carbon composed of carbon atoms arranged in a network of hexagonal and pentagonal rings,115–118 for the preparation of an SC-ISE via self-assembly. They obtained Nernstian responses to K+ with a linearity down to 10−5 M, without the O2 interface alongside high stability.114 However, in recent years, the application of fullerene in ISEs has significantly decreased due to the rapid development of state-of-the-art nanomaterials that exhibit higher performance and lower cost for potentiometric ion sensing.
Fig. 5 (A) A graphical depiction of the preparation process of (a) a sensing array of multicalibrated urea potential, and (b) Au@urease nanoparticles.129 Copyright 2022, ACS publications. (B) Schematic illustration of the modification process for a GCE with metal oxide nanoparticles and membrane, alongside a SEM image of zinc oxide nanoparticles and calibration curves of unmodified and different metal oxide nanoparticle-based-ISEs (ZnO, CuO, and Fe2O3) in KNO3 solution over the concentration range of 1 × 10−8–1 × 10−1 mol L−1 after two months of measurements.120 Copyright 2022, Elsevier. |
Nanoporous metal materials have attracted considerable interest for the development of SC-ISEs due to their numerous interconnected nano-sized cavities that can provide a large area for ion interaction and adsorption.137–139 For example, nanoporous gold film and platinum oxide were used widely as the solid contact in ISEs, which demonstrated high potential stability with a short response time.140–142 The impact of different types of metal oxide nanoparticles as the solid contact, including zinc oxides, iron oxides, and copper oxides, on the sensing performance of ISEs for K+ detection was studied by Wardak et al.120 It was found that the integration of metal oxide nanoparticles into the electrode enhanced its performance, particularly in terms of electrical parameters and potential stability; this could be attributed to excellent electrochemical activity and a high surface-to-volume ratio. In addition, the best performance was achieved by using zinc oxide nanoparticles, with a low LOD (3.66 × 10−6 mol L−1), a fast response time (4–6 s), and sustained functionality over 5 months. The superior characteristics of electrodes utilizing zinc oxide nanoparticles might be attributed to greater homogeneity. A graphical depiction of the modification of a glassy carbon electrode (GCE) with metal oxide nanoparticles and a membrane, alongside a SEM image of the zinc oxide nanoparticles and calibration curve for the zinc oxide electrode, is shown in Fig. 5B.
Recently, MOFs consisting of metal ions or clusters coordinated with organic ligands to form an ultra-high porous structure143,144 have been highly favored for the development of SC-ISEs. Mirica et al.145 proposed an SC-ISE using a conductive MOF as the ion-to-electron transducer. This electrode was prepared by modifying a GCE with MOFs through drop casting, followed by modification of a K+ or NO3− selective membrane. The ISE showed a very good performance close to Nernstian behavior (54.1–58.2 mV s−1) with a satisfactory signal stability (15 ± 1 μV s−1) and a wide dynamic range spanning from millimolar to nanomolar concentrations for the detection of K+ and NO3−. Recently, Asadnia and co-workers146 investigated the impact of using Ni-HAB MOFs as the ion-to-electron transducer in an SC-ISE for the detection of Li+. It was found that the capacitance of the sensor was raised by increasing the thickness of MOFs, which led to long-term stability with a 1.15 × 10−6 mV h−1 drift over 12 h. In principle, the high capacitance of the ion-to-electron transducer layer helps to prevent polarization due to a minimal electrical current, resulting in a quicker and more stable response.146 Although the low electrical conductivity of MOFs may limit their application in SC-ISEs, their high proton conductivity and compatibility with highly conductive materials enable the development of effective nanocomposite-based solid contacts. Moreover, their interconnected and tunable pore sizes make them promising candidates as solid contacts or ionophores for the preparation of SC-ISEs.147–151
Recently, Fouad et al.152 synthesized Mullite nanoceramics using the sol–gel method; these were used as ionophores in carbon paste of ISEs for the detection of Cd2+. The mechanism of action was based on ions binding to the surface of the electrode containing Mullite nanoparticles. Since the mesoporous structure of Mullite significantly speeds up electron transfer processes, this work can be considered a good example of the high potential of porous nanomaterials for ionophore preparation. Chen, Ge et al.128 developed a portable Ca2+ ISE by using phenylboronic acid-functionalized nanometer-sized CaCO3 particles as the recognition and signal elements for monitoring carcinoembryonic antigen glycoprotein. In order to maintain the bioactivity, the carcinoembryonic antigen aptamers were immobilized on the Au substrate. In addition, phenylboronic acid-functionalized nanometer-sized CaCO3 enhanced the storage stability by facilitating the retention of organoboron-conjugated nanomaterials and aptamers on the Au substrate.
Michalska et al.159 reported modified polyvinylidene fluoride nanofibers with bis(2-ethylhexyl)sebacate as a receptor for K+ detection. The thickness of the ion-selective layer was below 10 nm, but the electrolyte diffusion coefficient for the nanofiber phase was estimated to be 10−10 cm2 s−1, which was considerably lower than the values typically associated with ion transport through traditional poly vinyl chloride (PVC)-based membranes. This could be related to the effect of air trapped in the pores of the nanofiber mats, which hinders the penetration of this phase by water, indicating that the rate limiting step in this process is filling the nanofiber mat pores with electrolyte.
Among various types of polymer nanomaterials, polyaniline (PANI) is more favourable for the design of high-performance ISEs owing to its high conductivity, environmental stability, low cost, versatile production methods along with simple setup.160–163 In addition, it was found that the morphology of PANI had a significant effect on the sensing performance of SC-ISEs.164,165 For instance, 3D PANI nanowire arrays (NWA) used as the ion-to-electron transducer layer offered a better performance, including faster response (approximately 1 s) and lower detection limit (2.5 × 10−8 M) for the selective detection of Pb2+, than PANI micro-/nanowire networks (MNWN) and micro-/nanosheets (MNS).166 This performance can be attributed to the high double-layer capacitance and enhanced diffusion of Pb2+ through the transducer layer due to the open structure of nanowire arrays, as shown in Fig. 6A and B.166Fig. 6C compares the stability of the potential response of the various Pb2+ ISEs, indicating that the potential drift significantly decreased by introducing PANI micro-/nanomaterials. Furthermore, the PANI-NWA can inhibit the formation of the water layer most efficiently due to its hydrophobic character, as observed in the water contact angle measurement. Additionally, PANI nanoparticles and nanofibers were utilized as the solid contact of ISEs, which achieved fast and stable Nernstian responses for the sensing of flucarbazone herbicide, Cr3+, Cr6+, NH4+, NO3−, alcaftadine, and cephalosporin.160,167–170 In general, nanostructured polymers have good flexibility and elasticity, besides the typical properties of nanomaterials, making them promising candidates for the development of wearable and stretchable SC-ISEs in real-time analysis.
Fig. 6 (A) SEM images of PANI-NWA (a), PANI-MNWN (b), and PANI-MNS (c) deposited on a glassy carbon electrode by direct electrodeposition. (B) Sensing mechanism for a NWA-based ISE. (C) Potential curves of different Pb2+-ISEs in 10−3 M Pb(NO3)2 for 3 h, 0.1 M NaCl for 3 h, and then 10−3 M Pb(NO3)2 again for 12.5 h.166 Copyright 2021, Elsevier. |
Ni2O3/RGO nanocomposites exhibit high electrochemical performances as a transducer element for the sensing of NO3− due to the synergistic effect between RGO and Ni2O3 nanostructures.175 Generally, the combination and interaction of the components play a major role in nanochannels, leading to an attractive sensing performance, such as good selectivity and long-term stability.30,176,177 For example, graphene/PVC and gold–CNT–gold nanocomposites were developed as a transducer material for the detection of tenofovir disoproxil fumarate178 and Na+,179 respectively. Many studies reported that CNTs were promising components in nanocomposites for enhancing the analytical performance of the ISEs due to their high aspect ratios and high functionalization ability.174,180–182 Recently, Hernández-Varela et al.181 developed an ISE based on cellulose/MWCNTs as the transducer for detecting K+. The homogeneous dispersion of cellulose aerogel (CA) and CNT ink increased the electrical and conductive behaviour of the ISE, and the resulting solid network exhibited good mechanical properties. The ISE shows a near Nernstian response (52.04 ± 0.26 mV) with a linear range from 10−4 to 10−1 M, while the LOD is 8.49 × 10−3 M, which is higher than that of many other K+ sensors. Additionally, a detailed evaluation of important parameters such as long-term potential stability and capacitance was not provided. Fig. 7A shows a scheme of the potentiometric cell and photos of the prepared K+ ISE. The potentiometric response, the calibration plot, and preparation steps for the cellulose–CNT nanocomposite aerogel are shown in Fig. 7B, C and D, respectively. While using specific types of nanomaterials like an aerogel could offer potential improvements in ISEs, the LOD of the prepared sensor was 8.49 × 10−3, which was higher than those reported in similar studies. Also, the authors did not provide information on important parameters such as long-term potential stability and capacitance.
Fig. 7 (A) Graphical representation of an ISE based on cellulose/MWCNT nanocomposites as a component of a potentiometric cell with accompanying photographs, (B) the potentiometric response time trace, (C) calibration curve, and (D) graphical depiction of the preparation steps for the nanocomposite, including the preparation of CA, addition of CNTs (CA1), and incorporation of CNTs and NaDDBS ink to prepare the functionalized aerogel.181 Copyright 2023, Wiley. |
The application of MOF-based nanocomposites as a solid contact has been on the rise in recent years, owing to their high electrical conductivity, charge mobility, and surface area.32,183,184 For the same reasons, MOFs are also widely used as a component of carbon paste in ISEs to improve their sensing performance. For example, 2D Cu-MOF:tricresyl phosphate (TCP):graphite mixture in a ratio of 2.67:30.54:66.79 (wt/wt%) was prepared in the form of carbon paste for the detection of Cu2, where graphite, 2D Cu-MOF, and TCP acted as the matrix, the electroactive material, and the plasticizer, respectively.185 The results displayed Nernstian behavior with a fast response time (3–5 s) and pH independency in real samples, including red and mate tea, sesame seeds, human hair, and tap water (non-drinking water).
In recent years, the development of flexible and miniature SC-ISEs has been highly desirable to meet the increasing demand for wearable devices. Zhang et al.32 synthesized a nickel triphenylene-fused metal catecholate (NiCAT) nanowire array on carbon nanotube fibers (CNTF) for the development of a wearable SC-ISE for sweat sensing. Fig. 8A shows photographs of the attachment of the wearable sensor to the forehead, construction details and a schematic diagram of the NiCAT@CNTF based ISE and RE in an electrochemical cell, as well as SEM images of the NiCAT@CNTF material, indicating that MOF nanowires aligned the carbon nanotube fibers as the transducer. This sensor exhibited a stable potential response alongside a good calibration curve and real-time sweat monitoring, as shown in Fig. 8A. The extremely high double-layer capacitance (34.39 mF cm−2) of the solid contact transducer can be attributed to the small contact impedance and ordered porous structure of the nanomaterials. In addition, biomimetic zirconium MOF/polyvinylidene fluoride (PVDF) was used as the solid contact for the detection of Cl−.186 The different components of the miniaturized MOF-integrated potentiometric sensor, alongside the sensor setup and portable potentiostat are depicted in Fig. 8B. The miniaturized wireless electrochemical sensing system demonstrates the high portability of the sensor, making it suitable for real-time, on-site and remote detection of Cl−.
Fig. 8 (A) SEM images of the NiCAT nanowire array, along with a photograph of a flexible and wearable ISE based on MOF nanowires aligned on carbon nanotube fibers (CNTF), including their position in the sensor, potential response, calibration curves, and typical real-time sweat monitoring.32 Copyright 2022, Elsevier. (B) Graphical representation of the miniaturized MOF-integrated solid contact, the sensor components, and its sulfur mustard detection mechanism using a miniaturized portable potentiostat for simulant analysis on a Bluetooth-interfaced smartphone.186 Copyright 2023, Elsevier. |
Core–shell nanomaterials exhibit unique characteristics by combining the properties of the inner and outer parts when applied to ISEs.187,188 Kaushal and Singh et al.189 developed a highly selective core–shell nanocomposite electrode based on boron doped graphene oxide–aluminium fumarate metal–organic framework (BGO/AlFu MOF core–shell) for the detection of Br−, showing a close to Nernstian response at low concentrations (1 × 10−7–1 × 10−1 M). The LOD and average slope of the calibration curve were 7.1 × 10−8 M and 54.53 ± 0.15 mV per decade change of concentration, respectively. Fullerene-based nanocomposites, owing to the high compatibility of fullerene with many materials, also serve as the solid contacts or membranes in SC-ISEs.117,190,191 Over the past few years, a substantial number of studies on SC-ISEs focused on the exploration of novel nanocomposites with tailored structures and features as solid contact transducer layers or ion-selective membranes for the sensing of various ions.135,167,192–195
Wearable SC-ISEs have attracted considerable interest for the non-invasive and real-time monitoring of ions. Nanocomposites have been proposed as fascinating candidates for wearable sensors due to good biocompatibility and mechanical stability with high flexibility on the skin's surface, and good ion-to-electron transduction.196–198 Recently, Doan and co-workers199 developed NH4+-ISEs based on gold-nanoparticles–reduced graphene oxide as an ion-to-electron transducer. The prepared sensor exhibited high potential stability and a fast response time (<10 s) due to good hydrophobicity, conductivity, and an increased double-layer capacitance at the membrane/solid-contact interface (from 1.22 to 8.41 μF) due to the AuNP–RGO transducer layer. The linear range and detection limit were 10−5–10−2 M and 3.80 × 10−6 M, respectively, without interference from light, carbon dioxide, oxygen, and redox species. The potentiometric response of the sensor in NH4Cl solutions spanned the range from 10−7 to 10−2 M, with high potential stability and reversibility at low concentrations. However, since the composition of the nanocomposites in the solid contact was not characterized in detail, the mechanisms associated with the synergistic effect of the materials on the analytical performance were not well understood. A wearable all-in-one SC-ISE was developed for Na+ detection based on nanocomposites of Au and CNTs (Fig. 9).179 As shown in Fig. 9A, the Au/CNT/Au sensors and thin-layer circuits are fully encapsulated within a soft and biocompatible silicone elastomer membrane, which can be attached to different substrates. Fig. 9B indicates the preparation process for the nanocomposite transducer composed of CNTs and Au nanoparticles, along with a schematic and SEM images. The prepared nanocomposites exhibited an enhanced surface contact area and improved signal stability, leading to a sensitivity of 55.5 ± 0.3 mV per decade for Na+. Although it was observed that the capacitance of the Au/CNT/Au electrode increased to 620.5 μF as compared to 75 μF for the Au/CNT electrode and 2.4 μF for the Au electrode, studies on the transduction mechanisms were not provided. The prepared thin electronic device with a thickness of 2 mm is promising for manufacturing miniaturized and wearable sensors. In addition, as reported by Bakker and Bobacka,200,201 thin layers, as a class of nanomaterials, could also be used as membranes due to fast diffusion processes, despite some possible drawbacks such as low robustness, limited lifetime, and potential leakage of membrane components. Table 1 summarizes the sensing performance of various nanomaterial-based SC-ISEs developed over the past 5 years.
Fig. 9 (A) Overview of a wireless all-in-one sodium sensor with a solid-state ISE (SS-ISE) and a reference electrode (RF), attached to different substrates. (B) Graphical depiction of the SS-ISE and RF, and the production steps for the SS-ISE.179 Copyright 2021, Elsevier. |
Nanomaterials & nanocomposites | Component | Transduction mechanism | Analyte | Sensitivity (mV per decade) | E°a (mV) | Response time (s) | Linear range (M) | LOD (M) | Capacitance (μF) | Potential drift (μV h−1) | Ref. |
---|---|---|---|---|---|---|---|---|---|---|---|
a Standard potential, which is the potential recorded when the primary ion activity is set to 1.23,223 | |||||||||||
Carbon nanofibers/nickel–cobalt nanoparticles/CNT layer | Solid contact transducer | Double-layer | K+ | 59.39 ± 0.80 | 413.7 ± 0.9 | — | 10−6–10−1 | 10−6.3 | 330 | 20 | 53 |
Single-walled carbon nanohorns | Solid contact transducer | Double-layer | Ca2+ | 29.69 | — | 4 | 10−6–10−2 | 10−6.1 | 23.26 | 120 | 110 |
Graphene | Solid contact transducer | Double-layer | K+ | 53.53 | ±0.7 | 8 | 10−4–10−1 | 10−4.28 | — | 2844 | 202 |
MOF (Cu3(HHTP)2) | Solid contact transducer | Redox | Cu2+ | 29.5 ± 0.3 | 173.8 ± 3.5 | 10 | 10−2–10−4 | 10−4.5 | 600 | — | 183 |
MWCNTs/Cu(II)-MOF | Solid contact transducer | — | Orphenadrine citrate | 60.05 ± 0.16 | — | 8 ± 1.3 | 10−8–10−2 | 4 × 10−9 | — | — | 203 |
MOF (NH2-MIL-53(Fe))/PANI | Solid contact transducer | Redox | Epinastine hydrochloride | 59.74 ± 1.61 | — | 5 ± 1 | 10−7–10−2 | 7.5 × 10−8 | — | 50 | 150 |
Nano-cordierite (Mg2Al4Si5O18) | Ionophore | — | Sr2+ | 29.1 ± 0.31 | — | 6 | 1.0 × 10−6–1.0 × 10−2 | 5.0 × 10−7 | — | — | 204 |
Gold nanoparticle–reduced graphene oxide | Solid contact transducer | Double-layer | NH4+ | 56.94 (±1.57) | — | <10 | 10−5–10−2 | 3.80 × 10−6 | 123.5 | 8.1 | 199 |
Graphene/PANI | Solid contact transducer | Double-layer | Anti-COVID-19 drug remdesivir | 55.39 ± 0.50 | — | 7 ± 1 | 10−7–10−3 | 1.00 × 10−7 | — | 700 | 205 |
Ruthenium dioxide nanoparticle | Solid contact transducer | Double-layer | H+ | 59 | 571 ± 2 | 5 | 10−12–10−2 | — | 1120 | 3204 | 206 |
Graphene/PVC | Solid contact transducer | — | Metoclopramide | 59.90 ± 0.38 | — | 10 | 10−6–10−2 | 3 × 10−7 | — | 2500 | 207 |
Graphene/PVC | Solid contact transducer | Double-layer | Tedizolid phosphate | −22.0 ± 0.2 | — | 7 | 10−6–10−2 | 7.9 × 10−7 | — | 133 | 208 |
Mullite nano-ceramic in carbon paste | Ionophore | — | Carcinogenic Cd2+ | 29.49 ± 0.24 | — | 6 | 1.0 × 10−8–1.0 × 10−2 | 1.0 × 10−8 | — | — | 152 |
Ag nanoparticles/PANI | Solid contact transducer | Redox | Pb2+ | 29.1 ± 0.3 | — | <5 | 10−3–10−9 | 6.31 × 10−10 | 40 | 90360 | 121 |
Urease enzyme nanoparticles | Membrane | — | Urea | 38 | — | 10–120 | 10−6–80 × 10−5 | 1 × 10−6 | — | — | 209 |
Au nanoparticle/siloxene/graphene/poly(dimethylsiloxane) | Solid contact transducer | Double-layer | Na+ | 60.13 | 421 ± 0.3 | — | 10−5.7 | 5.6 × 10−6.5 | — | 3 | 210 |
Au nanoparticle/siloxene/graphene/poly(dimethylsiloxane) | Solid contact transducer | Double-layer | K+ | 59.4 ± 0.4 | 417 ± 0.5 | — | 10−5.5 | 8.1 × 10−5.7 | — | 5 | 210 |
PANI | Solid contact transducer | Redox | Alcaftadine | 58.8 | — | 5 | 1 × 10−6–1 × 10−2 | 6.3 × 10−7 | 35.5 | 900 | 169 |
Graphene/carbon nanotube | Membrane | Double-layer | NH4+ | 59.6 ± 1.5 | — | — | 10−6–10−1 | 10−6 | — | — | 211 |
Carbon black/soft elastomer | Solid contact transducer | — | Na+ | 57 | — | — | 10−5–1 | 4 × 10−6 | 7.14 | 140 | 212 |
PANI nanofibers/MWCNTs | Solid contact transducer | Double-layer | Cl− | −61.3 | 89.5 ± 1.8 | <10 | 5 × 10−6–1 × 10−1 | 2.56 × 10−6 | 7.16 × 103 | 30 | 180 |
FexOy nanoparticles/LaF3 | Membrane | — | F− | 50.3–62.4 | — | — | 10−1–3.98 × 10−7 | 7.41 × 10−8 | — | — | 119 |
Graphene nanocomposite | Solid contact transducer | — | Chlorhexidine | 28.8 ± 0.2 | — | 15 | 10−6–10−3 | 4.76 × 10−7 | — | — | 213 |
2D Ni-MOF nanosheets | Membrane cocktail | — | Ni2+ | 29.5 ± 0.09 | — | 6 | 1.0 × 10−5–1.0 10−1 | 2.7 × 10−6 | — | — | 214 |
Acylhydrazone-based MOF | Solid contact transducer | Redox | K+ | 56.30 | — | — | 10−5–10−1 | 2.1 × 10−6 | — | 320 | 215 |
Poly(3-octylthiophene-2,5-diyl)/nanosized ruthenium dioxide | Solid contact transducer | Redox | K+ | 58.64 ± 0.07 | 407 ± 1 | 10−1–10−6 | 10−6.9 | 1170 | 28 | 216 | |
Poly(3-octyl-thiophene)/molybdenum disulfide | Solid contact transducer | Redox | NO3− | 64 | — | — | 1–1500 ppm NO3−–N | 1.4 ppm | — | 95 | 217 |
ZnO/RGO | Solid contact transducer | Redox | K+ | 9.95 μA (μM)−1 cm−2 | — | 2 | (100–500) 10−6 | 956 × 10−6 | — | — | 218 |
TiO2–CuO bimetallic/PANI | Solid contact transducer | — | Vildagliptin | 60.04 ± 1.4 | — | 10 ± 1.3 | 1 × 10−2–1 × 10−8 | 4.5 × 10−9 | — | — | 195 |
MWCNTs | Solid contact transducer | Double-layer | Fluoxetine | 64.4 ± 0.2 | — | <5 (for low concentrations) | 6.0 × 10−7–1.0 × 10−2 | 2.0 × 10−7 | 40.6 ± 2.1 | 88560 ± 5040 | 55 |
SWCNTs/fullerene C60/carbon nanoparticles | Solid contact transducer | Double-layer | Protonated procaine cations | 58.9 ± 0.2 | 40.8 mB | 7 | 5 × 10−7–1 × 10−2 | 10−7.05 | — | ±330 | 190 |
Fullerene C60–Nafion/SWCNT–fullerene C60–PVC | Solid contact transducer | Double-layer | Procaine hydrochloride | 58.0 ± 0.1 | 506.0 mB | ≤5 | 1 × 10−7–1 × 10−2 | 10−7.15 | — | ±140 | 190 |
SWCNTs | Solid contact transducer | Double-layer | Desvenlafaxine | 57.2 ± 0.8 | — | <8 | — | 2.0 × 10−6 | 91.7 ± 3.4 | 10.6 ± 2.1 | 219 |
MWCNTs/PANI | Solid contact transducer | Double-layer | Erythromycin | 54.0 ± 0.5 | — | <10 | 4.6 × 10−6–1.0 × 10−3 | 9.6 ± 0.4 × 10−7 | 324.7 | 11160 | 220 |
Porous NiCo2S4 nanosheet | Solid contact transducer | Redox | Ca2+ | 30.7 ± 0.3 | 0.3–0.7 | — | 1.0 × 10−9–2.9 × 10−2 | 1.6 × 10−7 | 1800 | 1.9 ± 0.5 | 135 |
Polydopamine/Ag nanoparticles | Membrane | — | K+ | 52.5 ± 1.8 | — | 4 | 1 × 10−4–1 × 10−3 | 10−5.4 | — | 257.2 | 122 |
Au/CNT/Au | Solid contact transducer | — | Na+ | 55.5 ± 0.3 | — | — | 10−3–100 | — | 620.5 | 490 | 179 |
MOF nanowires/carbon nanotube fibers | Solid contact transducer | Double-layer | Na+ | 58.69 ± 0.77 | 308.0 ± 3.1 | — | 10−5 to 10−1 | — | 34390 mF cm−2 | 8.1 ± 4.1 | 32 |
Cellulose aerogel/CNTs/sodium dodecylbenzene sulfonate | Solid contact transducer | — | K+ | 52.04 ± 0.26 | — | — | 10−4–10−1 | 8.49 × 10−3 | — | — | 181 |
PANI/NWA | Solid contact transducer | Redox | Pb2+ | 28–29 | — | 1 | 10−7–10−3 | 2.5 × 10−8 | 27.2 | 132120 | 166 |
Au@urease nanoparticles | Sensing material | — | Urea | −88.3 ± 1.5 | — | — | 5.0 × 10−3–1.0 | — | — | — | 129 |
Mesoporous carbon | Solid contact transducer | Double-layer | Ag+ | 55.6 ± 0.8 | — | — | 1.0 × 10−6–1.0 × 10−3 | 10−6.8 | — | 115200–86400 | 93 |
RGO | Solid contact transducer | Double-layer | K+ | 53.34 | — | 6 | 25 × 10−2–6 × 10−5 | 10−4.24 | — | 7200 | 221 |
AuCu nanoparticles/MWCNTs | Solid contact transducer | Double-layer | Ca2+ | 29.0 | 2.5 | <10 | 10−6–10−1 | 6.03 × 10−7 | 54 | 15 ± 3 | 30 |
AuCu nanoparticles/MWCNTs | Solid contact transducer | Double-layer | SO42− | 27.0 | 3.1 | <10 | 10−5–10−1 | 8.91 × 10−6 | 105 | 118 ± 16 | 30 |
Ag@AgCl nanomaterial/1-tetradecyl-3-methylimidazolium chloride | Solid contact transducer | Redox | Ca2+ | 28.3 | — | — | 10−6–10−2 | 10−6.5 | 75.2 | 47880 ± 11160 | 222 |
ZnO nanoparticles | Solid contact transducer | Double-layer | K+ | −56.07 | 350.18 | 4–6 | 1 × 10−5–1 × 10−1 | 3.66 × 10−6 | — | 160 | 120 |
Ni–HAB MOFs | Solid contact transducer | Double-layer | Li+ | 57.6 | — | <1 | — | 9.94 × 10−7 | 1652 (μF cm−2) | 1.15 × 10−3 | 146 |
Laser-induced graphene/PVC | Solid contact transducer | — | NO3− | −58.17 ± 4.21 | — | — | 10−3.5–10−1 | 6.01 ± 1.44 × 10−6 | 95.24 ± 8.24 | 37800 ± 3456 | 90 |
Ordered mesoporous carbon | Solid contact transducer | Double-layer | K+ | 63.5 ± 0.6 | 583 ± 9.1 | 8 | 10−4.19–10−0.21 | 5.4 × 10−6 | 52.93 | 11988 ± 2088 | 92 |
Boron doped graphene oxide–aluminium fumarate metal organic framework | Membrane | — | Br− | 54.53 ± 0.15 | — | 13 | 1 × 10−7–1 × 10−1 | 7.1 × 10−8 | — | — | 189 |
Shelled hollow carbon nanospheres | Solid contact transducer | — | Ca2+ | 28 | — | — | 10−5–0.05 | — | 40 | 20 | 109 |
Molecularly imprinted polymers/PANI | Solid contact transducer | Double-layer | Flucarbazone anion | −45.5 ± 1.3 | — | <5 | 10−2–10−5 | 5.8 × 10−6 | 11.7 ± 0.7 | 41760 | 167 |
Aliquat/PANI | Solid contact transducer | Double-layer | Flucarbazone anion | −56.3 ± 1.5 | — | <5 | 10−2–10−4 | 8.5 × 10−6 | 37.7 ± 1.2 | 141480 | 167 |
Thiol-functionalized reduced graphene oxide | Solid contact transducer | — | K+ | 60.0 ± 0.4 | 2.6 (SD) | <5 | 10−7–10−1 | 2.5 × 10−6 | 12.43 × 10−6 | 1.75 | 83 |
Thiol-functionalized reduced graphene oxide | Solid contact transducer | — | NO3− | −60.0 ± 0.5 | 4.8 (SD) | <5 | 10−7–10−1 | 4.0 × 10−6 | 5.99 × 10−6 | 8.79 | 83 |
Dibutylphthalate/MOF (MIL-53(Al))/PVC | Electroactive material | — | Imipramine hydrochloride | 57.7 | — | <5 | 1.0 × 10−7–1.0 × 10−1 | 5.0 × 10−8 | — | — | 147 |
Although there has been a significant increase in the use of nanomaterials for the development of high-performance SC-ISEs over the past few years, the main challenges in this field include: (1) lack of a fundamental understanding of the structure–property–performance relationships for the rational design of nanomaterials in various SC-ISE applications; (2) poor reproducibility in sensor preparation because of the complex routes for nanomaterial preparation; (3) variations in sensing performance due to their high sensitivity to environmental conditions; (4) the potential toxicity of SC-ISE sensors that include heavy metals or engineered nanoparticles. Recent progress has indicated that nanomaterials are promising candidates for SC-ISE applications, particularly in the development of miniaturized, portable and wearable sensing devices for water quality and health monitoring, while the above-mentioned problems need to be addressed for their commercialisation for practical applications. This requires interdisciplinary efforts from various fields such as materials science, nanotechnology, sensor design and theory advancement. Furthermore, calibration-free ISEs, which are pre-calibrated when manufactured, offer fast measurements and user-friendly properties in environmental and clinical analyses. However, many of them have suffered from poor reproducibility and low stability due to potential variation and drift, resulting in low accuracy and reliability. Recently, advancements have been made in terms of improving the preparation methods and introducing innovative control measures. It is worth noting that the integration of novel materials into SC-ISE components is also a promising approach to address this challenge.
This journal is © The Royal Society of Chemistry 2024 |