Hakan
Ceylan
ab,
Joshua
Giltinan
abc,
Kristen
Kozielski
a and
Metin
Sitti
*abc
aPhysical Intelligence Department, Max Planck Institute for Intelligent Systems, 70569 Stuttgart, Germany. E-mail: sitti@is.mpg.de
bMax Planck ETH Center for Learning Systems, 70569 Stuttgart, Germany
cDepartment of Mechanical Engineering, Carnegie Mellon University, Pittsburgh, USA
First published on 28th April 2017
Untethered micron-scale mobile robots can navigate and non-invasively perform specific tasks inside unprecedented and hard-to-reach inner human body sites and inside enclosed organ-on-a-chip microfluidic devices with live cells. They are aimed to operate robustly and safely in complex physiological environments where they will have a transforming impact in bioengineering and healthcare. Research along this line has already demonstrated significant progress, increasing attention, and high promise over the past several years. The first-generation microrobots, which could deliver therapeutics and other cargo to targeted specific body sites, have just been started to be tested inside small animals toward clinical use. Here, we review frontline advances in design, fabrication, and testing of untethered mobile microrobots for bioengineering applications. We convey the most impactful and recent strategies in actuation, mobility, sensing, and other functional capabilities of mobile microrobots, and discuss their potential advantages and drawbacks to operate inside complex, enclosed and physiologically relevant environments. We lastly draw an outlook to provide directions in the veins of more sophisticated designs and applications, considering biodegradability, immunogenicity, mobility, sensing, and possible medical interventions in complex microenvironments.
Use of microrobots for lab-on-a-chip devices has already proved to be a powerful tool. Handling small objects in very small fluid volumes for manipulating, moving, and reconfiguring components in 3D by means of microrobots make this route highly attractive. Assembly of 3D heterogeneous microobjects, which require orientation and positional control, would be best addressed using microrobotic assembly.11,12 Organ-on-a-chip applications could benefit from microrobotic operations, in which complex cellular materials with 3D microscale features may need to be positioned to better recapitulate the native physiological status.13 Additionally, preclinical characterizations of microrobots for drug release profiles and their interactions with living tissues could be tested in organ-on-a-chip platforms.
Active and targeted delivery of therapeutic cargos, such as drugs, imaging agents, and genetic materials, are the major objectives of the first-generation microrobotic systems. Active navigation inside the body to a specific target site with a controllable cargo carrier is superior to relatively limited cargo delivery and distribution efficiencies provided by current passive routes of administrations, such as intravenous delivery and local diffusion.14 Using active, drivable carriers, it is possible to minimize systemic side effects by achieving targeted local treatment options. For example, intravenously administered interleukin-12 caused lethal systemic toxicities in a clinical trial.15 Active delivery and controlled on-site release schemes increase the overall bioavailability of single dose administration. Sensitive cargo types, such as proteins, peptides, or nucleic acids, are better protected from degradation inside a carrier, as they otherwise have very short half-lives in serum.16 Autonomous, real-time control over cargo release dynamics would perhaps represent the state-of-the-art of the in situ therapeutic and diagnostic strategy. To this end, microrobots that are able to navigate inside the human body, act intelligently in response to changing conditions, carry, deliver, and release therapeutics, and perform complicated tasks in semi- or fully autonomous manners could revolutionize many clinical practices.
The first concept of miniaturized machines for bioengineering was artistically visualized in the popular science fiction movie Fantastic Voyage (1966). In the movie, the brain clot of a nearly dying scientist had to be removed in one hour by a submarine shrunken to microscopic size and injected into his blood stream with a small crew. In view of the scaling laws, which we discuss in the following section, a macroscale submarine design in the microscopic dimensions is highly inefficient to operate inside human body. However, the venture of the crew has significant overlaps with premises and vision of microrobotics field, which constitute the main theme of the present review.
Here, we review frontline advances in the design and testing of microrobots for and toward bioengineering applications. In the following section, we provide a brief overview of micron-scale robotics and founding principles of this emerging field. We then make a higher-level classification of microrobots using on-board and off-board actuation, powering, and control approaches. We further subdivide each approach based on the mechanism of actuation/propulsion. We critically discuss design and operational capabilities of each type of microrobot in physiological environments, and provide potential pitfalls and parts to be optimized toward bioengineering applications (Table 1). In the final section, we provide an outlook toward more complex designs and applications. We layout future challenges and critical directions to consider for rendering intelligence to microrobotic systems.
Design approach | Principal source of powering | Actuation method | Key design features for effective powering | Advantages for bioengineering applications | Major limitations |
---|---|---|---|---|---|
Off board (externally actuated and guided) | Magnetic fields | Rotating magnetic fields and gradient pulling | Flagellum-mimetic rigid and flexible38 helices; cilia and sperm-mimetic undulating synthetic tails;42 gradient pulling of magnetically active designs | Wireless powering, actuation and maneuverability; biocompatible energy source; reliable for in vitro, in vivo, and lab-on-a-chip applications | Difficulty in selective agent addressability, high cost requirements for medical instrumentation |
Acoustic fields | Acoustic radiation force and acoustic streaming | Bubble-integrated bodies; flexible tail;84 microcannon;87 passive particles89 | Biocompatible energy source; both 2D and 3D assemblies can be realized; reliable for lab-on-a-chip and in vitro applications | In vivo use requires development of proper instrumentation; microrobot material composition and shape needs further investigation | |
Light | Light-induced formation of thermal gradient around the microrobot | Generation of an air–liquid;91 stable bubble encapsulation into the microrobot body100 | Sub-micron resolution, multiple pathways for energy transfer (e.g., thermal, nematic alignment); can produce traveling waves; simple selective agent addressability; reliable for lab-on-a-chip and in vitro applications | Limited to 2D; not applicable to in vivo conditions; limited workspace size; requires line of sight | |
On board (self-propelled) | Chemical energy | Bubble propulsion | Catalytic formation of gas bubbles by breaking down fuel in a confined reactor is ejected with high speed102 | High swimming speed (up to 103 body lengths per second); robust power output | Toxic hydrogen peroxide is the main fuel source. High power output is possible with toxic catalyst like Pt. Action of Mg and Ni-driven systems elevate local pH |
Local chemical gradients formed around the microrobot | Asymmetric distribution of the catalyst, e.g., Janus colloids113 | Biocompatible fuels and catalysts, such as glucose and glucose oxidase, respectively, are possible101 | Movement is extremely sensitive to the ionic strength; lack of long-range directional motion due to the Brownian effect | ||
Activation of cellular receptors | Physical attachment of live bacteria, sperm or muscle cells with synthetically engineered bodies27 | Integrated sensing and mobility; inherently compatible with physiological fluids; comparatively high efficiency in power output | Live cells function only in delicate conditions (37 °C, 5% CO2, nutrients etc.) to survive |
For microrobotic swimmers in fluids, the ratio of inertial forces to the viscous drag forces, a dimensionless quantity called Reynolds number, determines the fundamental characteristics of their swimming fluid dynamics. When Reynolds number is much less than 1, viscous forces predominate inertial forces; therefore, in order for a microswimmer to propel, it has to do time-irreversible, i.e., non-reciprocal, shape changes with its body.19 As inertia plays an insignificant role in low Reynolds number, reciprocal motion does not lead to a displacement. In other words, the movement due to the forward component of the motion will be cancelled out by the backward component of the motion. In order to comply with such physical requirement, microorganisms have evolved elaborate swimming strategies, such as continuous rotation of helical bacterial flagella and non-reciprocal beating of a sperm tail and paramecium cilia. Such biological swimmers have already inspired a number of microswimmer designs, some of which are covered in the present review. Nevertheless, the size of the biological microswimmers tend to be limited by around 1 μm, because molecular diffusion predominates the advection of material to the microorganisms over the active search by swimming at the sub-micron scale.20 As the size of the swimmer goes down to below 0.3 μm, the stochastic Brownian effect contributes to the most of the motion dynamics, thereby hindering long-range directional propulsion inside bulk fluid.21 Below 0.1 μm, the continuum hydrodynamics does not safely apply any longer, and the effects of quantum mechanics take over. This is the domain of nanorobotics, for which the interested readers are directed to the reviews toward this direction.22,23 Altogether, understanding the scaling forces by taking into account how physical forces are experienced by entities in the lowered dimensions is a vital aspect for the design and successful operation of microrobots.
Conventionally, a robot is made to perceive and learn by means of on-board sensing and computational capabilities, so that it can decide an appropriate response in given environmental conditions. Consequently, a prime question is how this is going to be achieved at the smaller dimensions, where such computational capabilities do not exist. Programmable physical and chemical properties of microrobots, dynamically interacting with its surrounding world, sensing, and adapting to the changes in the environment can enable robust design routes for making sophisticated systems at the microscale. In nature, organisms without brains, such as slime molds, bacteria, and plants, already use physical intelligence as the main route of making decisions and adaptations to complex and evolving conditions.24–26 As a result, natural systems could provide a plethora of inspiration to create similarly performing artificial miniature systems based on their physical design, processing, adaptation, reconfiguration, self-organization, and control. Early attempts toward this direction have been realized as biohybrid systems where live cells are physically integrated to artificial materials to exploit their on-board integrated powering, actuation, sensing, and control capabilities.27 This concept has recently drawn a special attention in the contexts of biohybrid microrobotics and biological soft robotics,28 which will be discussed in detail later.
Using magnetic actuation, much design inspiration has so far arisen from the observation of bacteria, which use helical rotation of flexible flagella for propulsion, providing a method to both propel forward and to change direction (Fig. 1a). While early research often yielded rigid artificial flagella, flexible DNA strands have also been shown to act as flagella for magnetic particles.34,35 Other flexible structures have been investigated such as swimming sheets that can be controlled independently of each other36,37 and rotating swimming microrobots with multiple flexible flagella.38 Cellular motility structures, such as cilia and sperm-like undulations, have also served as source of inspiration for human-engineered microswimmer mobility, as well as functional microchannel coatings that provide controlled fluid motion or mixing (Fig. 1b).39–43 Propeller-shaped microswimmers can either be human-engineered21 or dynamically self-assembled.44 In addition, they can have multiple modalities, alternating between swimming, rolling, and propelling, depending on the substrate and control signal.45
Fig. 1 Off-board magnetic actuation and powering methods for mobile microrobots. (a) Illustration of a functionalized helical swimmer, where the magnetic core is covered by a coating that can have biomedical functions, such as sensing, tissue drilling, or therapeutic drug release. (b) Programmable matter: elastomer beams, impregnated with magnetic microparticles, can be programmed to have different magnetization direction and strength locally to produce desired programmed motions, such as the undulatory motion of an artificial cilium (reprinted with permission from ref. 42. Copyright 2016 from the Proceedings of the National Academy of Sciences). (c) (i) A custom eight electromagnet coil system capable of 5- and 6-DOF control; (ii) permanent magnets mounted on stepper motors (reprinted with permission from ref. 47. Copyright 2016 from the IEEE); (iii) a four electromagnet coil system, which is capable of 3D motion control and accommodating a patient in the central bore (reprinted with permission from ref. 48. Copyright 2015 from the IEEE); (iv) a small-animal research MRI system (Bruker). (d) A body with three magnets, which create a 6-DOF controllable microrobot.46 (e) A side-by-side illustration of an H. pylori bacterium and a synthetic swimming microrobot, which uses a similar mechanism to penetrate the mucin gel (reprinted with permission from ref. 53. Copyright 2015 from the American Association for the Advancement of Science). (f) The basic configurations for determining separation distance using two-agent control (reprinted with permission from ref. 66. Copyright 2017 from the Springer). |
Regarding the three-dimensional (3D) maneuverability of magnetic microrobots, 5-degrees-of-freedom (DOF) control of microrobots was introduced by Kummer et al.8 For an arbitrary arrangement of electromagnetic coils, they showed how to solve for the forces and rigid body torques on a single magnetic microrobot (Fig. 1ci). To avoid the need for orientation feedback, it was assumed that the microrobot had a uniform magnetization, which due to the magnetic torque, would always align with the magnetic field. Thus, only the magnetic field and magnetic gradients, which create the forces, were required to achieve the 5-DOF control. However, rotation about the magnetization axis, as the vector that aligns with the field, could not be specified. By adding known, perpendicular magnetization to a magnetic microrobot, motion about all 6-DOF was shown (Fig. 1d).46 However, as this method requires careful design and feedback, 5-DOF control is still the common method in the literature. Seeking to avoid the heat and power requirements of electromagnets, a permanent magnet-based 5-DOF control system could also provide similar path-following accuracy as an electromagnet system (Fig. 1cii).47 3-DOF control is also popular as many microrobots are spherical, and thus do not require any orientation control, allowing for more design freedom, such as an electromagnet configuration which incorporates a central bore to accommodate a patient (Fig. 1ciii).48
Gradient pulling-based magnetic microswimmers carrying plasmid DNA-loaded liposomes have been demonstrated to undertake transfection of cells for selective gene editing.49 A microrobot coated with a 10 nm silver layer can directly disintegrate the membrane of an E. coli bacterium by physical contact.50 Hyperthermia induced by microrobots is a promising method to target cancer cells, without causing damage to surrounding healthy tissue. A small rod-like microrobot, rotated at sufficiently high frequencies, can feasibly generate heat from fluid drag, but has not been shown experimentally.51 Alternatively, magnetotactic bacteria were shown to be able to sufficiently heat to kill Staphylococcus bacteria in a span of 60 minutes.52 Microrobots can also be made to penetrate through some tissues by chemical assistance, such as the Helicobacter pylori-inspired method of using urease activity to locally degrade gastric mucin gel (Fig. 1e).53
Recent works have tackled the more difficult challenge of assembling structures and transporting cargo in 3D, which cannot be based on contact-based pushing mechanisms used in 2D. Swimming or rolling microrobots induce fluid flow by their propulsion mechanism, and can use these vortices to trap objects and transport them.54–57 Flexure-based gripping mechanisms have been shown for magnetic microrobots, where the magnitude of the magnetic field is used to control the opening and closing of the gripping arms.58 This was shown to be highly repeatable, assembling up to 10 layers of hydrogels around a post.12 While these untethered mechanical grippers prefer parts that match the gripper geometry, microrobots with a bubble on their surfaces could pick and place a wide range of 3D and various material parts using capillary forces.11
Magnetic fields have also been used for maneuverability while alternative methods serve as the source of the mobility. For example, 3D-printed microfish exhibit chemically powered propulsion while being magnetically guided.59 Microrobots guided by magnetic fields and propelled by ultrasound were demonstrated to be simultaneously loaded with quantum dots, an anti-cancer drug, and magnetic nanoparticles.60 An alternative propulsion method is the creation of fluid flows by the resonant oscillations of air bubbles within a fluid.61,62 These fluid flows can additionally trap objects, and then the system can be carried by a magnetic transporter.63,64
Microrobot teams and swarms remain a critical research topic, because such an organization paves the way of parallel and distributed complex operations.65 Swarm control is of particular importance as commands might be given to the whole population, subsets, individuals, or a combination of these. Control of multiple magnetic microrobots remains an ongoing topic of research, as all microrobots will receive the same global control signal. This control signal causes all microrobots to move in a coupled manner. In addition, the microrobots will either attract or repel each other due to the magnetic field gradients generated by each. By controlling the orientation of the magnets with an external field, the magnetic attraction and repulsion can be controlled to dictate the motion of the individuals (Fig. 1f).66 Techniques to use non-uniformities in the field67 or selective electrostatic clamps to address select individuals68,69 have been shown, but require patterned and structured surfaces and environments. Even with these problems solved, interference of individual microrobots in a swarm yields a lower net velocity than a solitary swimmer.70 This behavior was observed when a swarm of magnetic microrobots was driven in vivo, in the intra-peritoneal cavity of a mouse. However, 80000 swimmers were required to generate a sufficient fluorescence signal to track and when coupled with the complex biological fluid medium, the mean speed decreased to 6.8 μm s−1, roughly half of the length of the microrobot body per second.70
While swarm microrobotics is promising for minimally invasive surgery, research in the functionality of individual micororobots for in vivo use has progressed. An artificial arterial thromboembolism was penetrated inside a live porcine model by a centimeter long microrobot using fluoroscopic imaging for position control.71 Recent in vitro efforts have focused on improving functionality of microrobots that will eventually operate inside the body. Intra-ocular surgery is still on track to be the first use of a medical microrobot in humans, as a camera above the patient will allow for fast, non-invasive feedback of the microrobot position. Recent advancements have enhanced the mobility of microrobots in vitreous fluid and tested online measuring of the viscoelastic properties of the vitreous, vital for control optimization.9,72
The electromagnet coil systems discussed are often customized for in vitro research. To transition to clinical in vivo applications, commercial systems must either be developed or adapted from existing technologies. A magnetic resonance imaging (MRI) device is a diagnostic tool that is utilized by most hospitals and adapting this technology could provide an affordable therapeutic modality for many clinical settings. An MRI device can both localize and actuate a microrobot toward their clinical use (Fig. 1civ). To this end, the existing MRI devices are promising, but advances in the state-of-the-art are limited due to both the high static magnetic field and the high cost-of-entry for research. Static, multi-DOF MR-compatible devices have been developed, but mobile device control remains in its infancy.73 Recent work has been able to operate such a microrobot inside a phantom with over 100 Hz feedback by limiting imaging to the direction of the applied gradients, so that all imaging sequences will yield propulsive force.74 A mobile injection system has been developed in use of an MRI by using the transfer of magnetic energy in a Gauss gun configuration.75 Magnetic particles can be tracked in an MRI to visualize blood vessels that are smaller than the current resolution limits of an MRI.76 If the microrobot is non-magnetic, ongoing research is increasingly miniaturizing electromagnetic coils to be placed on-board a device for visualization by the MRI, though on-board miniaturized power remains an ongoing challenge.77 If the object is magnetic, a new technique, dipole field navigation, was developed for tracking and actuation in an MRI. This technique allows for actuation that requires large spatial gradients, provided by a placing a magnetic sphere outside of the patient. By placing the magnet near the targeted microrobot, actuation can be achieved, and by moving the sphere with the MRI gradient coils, the MRI can automatically image and actuate the microrobot; however, optimization and in vivo validation and testing remain as future works.78,79
Fig. 2 Off-board acoustic and optical actuation, powering, and control strategies for microrobots. (a) Low-power acoustic field driven oscillatory motion of a bubble enables directional motion in water. Two microswimmers with bubbles of different size enable independent control of each body (reprinted with permission from ref. 83. Copyright 2015 from Nature Publishing Group). (b) A flexible tail attached to a rigid body is actuated by travelling acoustic waves that result in directional motion (reprinted with permission from ref. 84. Copyright 2016 from the American Chemical Society). (c) Ultrasound triggered high-speed ejection of microparticle cargo from cannon-like microholes (reprinted with permission from ref. 87. Copyright 2016 from the American Chemical Society). (d) Hologram-based patterning of acoustic waves drives the assembly of microparticles into predictable reconstructions in 3D (reprinted with permission from ref. 89. Copyright 2016 from Nature Publishing Group). (e) The automated simultaneous manipulation of four heterogeneous bubbles by thermocapillary flow enables the bubbles to surround and transport a microobject (reprinted with permission from ref. 92. Copyright 2016 from the IEEE). (f) A light controlled microrobot with a particle ejecting function. By illuminating the top hole, which has a thin layer of gold for heat transduction, the resulting thermocapillary flow ejects a bead (reprinted with permission from ref. 100. Copyright 2016 from Nature Publishing Group). |
In the context of in vivo applications, the standing wave method is limited, because standing waves cannot be established in the human body in a predictable manner. For in vivo efficient propulsion, traveling wave-mediated mobility is the most advantageous strategy. In this context, an acoustically activated flagellum was recently shown to propel an artificial microswimmer through an aqueous solution by small amplitude oscillation of the tail in the presence of travelling acoustic waves (Fig. 2b).84 Acoustically responsive, sperm-shaped microswimmers can also be fabricated in situ within a microchannel and therein actuated via flagella oscillation.85 On the other hand, this method falls short for size scalability and maneuverability, which remain as the major improvements in the future.
Acoustic radiation force can also be generated by focusing an ultrasonic beam to physically trap an individual or a group of microobjects.86 The trapped microobject can then be transferred by moving the ultrasonic transducer that generates the ultrasound. However, strong ultrasound at the focal point may cause a high temperature rise resulting in tissue damage. A noteworthy study turned this local energy buildup into advantage and proposed a potential method of drug delivery. Focused ultrasound pulse vaporized perfluorocarbon emulsion within a microcannon that resulted in rapid ejection of microparticles (Fig. 2c).87 During the ejection, the speed of microparticles reached exceptionally high speeds of a few meters per second. Acoustic streaming can also generate directional propulsion force in fluids. Ultrasound actuation of a simple structure consisting of an array of comb-like cavities with trapped air bubbles oscillating at the resonant frequency of microbubbles propels the structure in the direction opposite to the acoustic streaming.88
Holographic acoustic elements have recently been introduced as a strong tool for complex manipulation of microobjects wirelessly. A remarkably simple method introduced by Fischer et al. demonstrated a 3D-printed surface profile used to encode the acoustic phases of the desired wavefront for enabling intricate particle patterning and trajectory control (Fig. 2d).89
Light can also directly propel untethered microdevices. A soft device comprised of a liquid crystal elastomer exhibits peristaltic motion under a modulated travelling light wave.95 Similarly, a liquid crystalline microrobot can crawl on surfaces by contracting when heating above 100 °C and expanding with directional friction due to asymmetric legs.96 This can be scaled up to a few millimeters, and by patterning the liquid crystal elastomer, the contraction can produce travelling waves in the structures, allowing the microrobot to move by propelling the wave down the length of the microrobot.97 Optical trapping, using a laser source to hold the position of a microparticle in 3D, was developed over 30 years ago. However it cannot directly manipulate objects greater than a few microns, and has been extensively covered in the literature, though novel bioengineering uses are being developed.98 If the particle is engineered asymmetrically, the beam does not have to be directional or focused, instead the thermophoretic effect propels the particle in a direction depending on the particle design and light wavelength.99 Recently a mobile microrobot with a syringe function was developed by exploiting four optical traps manipulating four handles on the microrobot. In the center, they fabricated a metal layer that could generate bubbles and thus thermocapillary convection flow, which is used to capture and “inject” microobjects (Fig. 2f).100 Optical trapping can also bend an air–liquid interface, and the resulting stored elastic energy is able to propel a bubble like a catapult when released.101
Bubble-propelled microswimmers are mainly fabricated in tubular shapes, which contain the catalytic material in the inner layer. This enables accumulation and accelerated ejection of the bubbles in well-confined and controlled reactor (Fig. 3a).114 As the gas accumulates and the internal gas pressure grows up, the resulting bubble accelerates towards one end to be ejected from an opening to the outside environment. It was suggested that the momentum change during the bubble ejection plays a critical role to drive the motion in the opposite direction.109,114 Tuning the magnitude of the momentum change by external stimuli such as temperature, light, and ultrasound modulated the microswimmer speed in a dynamic way. For example, ultrasound induced disruption of normal bubble evolution and ejection was an interesting method for microswimmer speed control. By this means, it was possible to achieve very fast changes in the micromotor speed in less than 0.1 s as well as reproducible on/off control (Fig. 3b).115 Another strategy toward this direction was folding and unfolding of the rolled-up polymer-platinum tubes by means of temperature changes, which resulted in speed control between 100–250 μm s−1 (Fig. 3c).116 In a recent study, temperature-sensitive polymer brushes were chemically grown onto the bubble-propelled stomatocytes, where the change in the temperature narrowed or enlarged the opening into the catalytic site. Control over the catalytic turnover rate by tuning the access to hydrogen peroxide fuel regulated thrust force and the resulting swimming speed.117 UV irradiation was used in another study to activate tubular micromotors made of TiO2 that propel themselves in the presence of hydrogen peroxide by generating oxygen bubbles.118 The intensity of UV light allowed a facile speed control between 50–250 μm s−1 with 0.2 s response time. However, these methods are external, off-board approaches, and no on-board mechanism has been reported for self-controlled catalytic activity of the micromotors to control the swimming speed so far.
Fig. 3 Chemically powered microrobots. (a) A schematic diagram of a bubble-propelled microswimmer motion step. The bubble makes a cyclic and asymmetric change from “bubble inside” into “detached bubble”, which causes the motion of the microswimmer (reprinted with permission from ref. 114. Copyright 2011 from the Royal Society of Chemistry). (b) Speed modulation of hydrogen peroxide-based, bubble-propelled microswimmers by ultrasound. Application of the ultrasound disrupts the bubble evolution and hence reducing the swimming speed (reprinted with permission from ref. 115. Copyright 2014 from the American Society of Chemistry). (c) Maturation of effective bubbles for motion depends on the geometry of the film. The visible oxygen bubbles are formed only in the cavity of a tube. (I) Folding of stimuli-responsive films by cooling below 28 °C. (II) Unfolding by warming up with stopping of bubbles (reprinted with permission from ref. 116. Copyright 2014 from the John Wiley & Sons, Inc.). (d) Time-lapse images of zinc-based microrobots in gastric acid at 37 °C (1 s intervals, I–IV) (reprinted with permission from ref. 125. Copyright 2015 from the American Society of Chemistry). (e) Development of a rationally designed metabolic network for temporally sustained autonomous movement at constant speed (reprinted with permission from ref. 128. Copyright 2016 from the American Society of Chemistry). (f) Velocity control of a Janus micromotor by manipulating the enzymatic activity of urease (reprinted with permission from ref. 131. Copyright 2016 from the American Society of Chemistry). (g) Long-range directional motion of colloidal Janus particles, guided along prescribed topographic pathways (reprinted with permission from ref. 136. Copyright 2016 from Nature Publishing Group). |
The movement of bubble-propelled microswimmers typically follows random trajectories in 3D. Their directional control to date was accomplished by wireless external control, such as magnetic, electric, and acoustic fields, in a way similar to that of off-board approach.119 An autonomous strategy was only recently demonstrated for the first time in the form of chemotaxis, where chemotherapeutic drug doxorubicin-loaded 0.3 μm swimmers move along a hydrogen peroxide concentration gradient toward higher fuel concentration.120
While bubble-propelled microswimmers can robustly operate in high ionic media, the termination of motion may be desired, e.g., when the microswimmers reach a target goal. This can be accomplished by the irreversible inhibition of the catalyst. For example, the presence of sulfhydryl groups in the environment is detrimental to the function of platinum mounted micromotors. Otherwise, a typical bubble-propelled microswimmer has sufficiently long lifetime, allowing continuous operations for weeks without loss of performance.121
Load and transport of cargo mounted on the body of bubble-propelled micromotors have already been demonstrated toward their use in bioengineering applications.102,119 Concerning the fuel type, hydrogen peroxide has been the major fuel type for both bubble-propelled and self-phoretic swimmers, mostly because it provides simple and reproducible conditions to study the basic mechanisms of swimming. For the bioengineering applications, however, it is highly corrosive and thus a major problem. Similarly, hydrazine, strong acids or bases, and bromine/iodine fuels require conditions that are unlikely to be biocompatible. An important direction for self-propelled chemical microrobots is therefore to discover new catalysts that utilize biologically available or biocompatible chemicals. Besides, most of the inorganic catalysts, such as platinum, pose a safety risk in the biological environment. In this regard, enzymes have been recently regarded as biocompatible alternative of inorganic catalysts. To this end, platinum was substituted with catalase, which performs similarly rapid peroxide decomposition for bubble-driven motion.103,122 Alternatively, water-powered magnesium and zinc-based microrobots represent recent examples toward biocompatible application of bubble-propelled microswimmers.123 In these cases, magnesium and zinc are consumed by reacting with water to produce magnesium hydroxide and zinc hydroxide, respectively, along with the evolved hydrogen gas, which forms bubbles to drive motion. In this regard, coating magnesium particles with a red blood cell membrane was demonstrated to detoxify serum-like solution from protein toxins.124 A potential safety risk of this system, however, is rapid elevation of the local pH due to magnesium hydroxide buildup. Nevertheless, depending on the specific body site, this may be an accepted, or even advantageous, route of powering. For example, microrobots based on the magnesium- and zinc-powered micromotors could be well suited for operations in the acidic gastrointestinal tract. To this end, autonomous motion of zinc-based micromotor was demonstrated for the first time in vivo in the stomach of a live mouse (Fig. 3d).125 In the stomach acid, self-propulsion of the micromotors led to increased binding to the stomach wall, and improved retention of the payload in the stomach compared to the passive diffusion. One key advantage of using zinc or magnesium is that the micromotor is self-destructive meaning that it does not leave a toxic residue in the body.126 In the case of intestinal delivery, oral administration of the microrobots is the main route. In order not to prematurely consume the propulsion source in the gastric acid, the micromotors could be shielded by enteric coating, which is insensitive to the pH of gastric acid (pH 1–3) whereas it dissolves in the intestinal fluid (pH 6–7) to expose the micromotors to start the movement.127 It is also noteworthy to add that in contrast to the catalytic approach, consumption of magnesium results in short-lived microrobots, endangering the completion of the operation before running out of its magnesium.
Mobility of the bubble-propelled microswimmers is strongly dependent on the accessible fuel concentration in the immediate microenvironment. Given the fuel amount, the mobility of the microswimmer gradually diminishes as the fuel is consumed, and propulsion ceases once the fuel consumption is complete. To maintain the speed at a constant speed over time, a far-from-equilibrium enzymatic network that actively regulates the fuel supply was demonstrated (Fig. 3e).128
Chemically powered self-phoretic microswimmer designs have been established around making colloidal micro- and nanoparticles with asymmetric distribution of the catalyst.104–106 Asymmetric catalyst creates an asymmetric decomposition of the reactants and formation of products, facilitating self-generated, local chemical gradients that create thrust forces. The resulting microswimmer speed is also proportional to the catalytic turnover rate, and hence the effective surface area of the catalyst.129 On the other hand, a long-range directional motion in bulk fluid using this method is not possible due to strong interference of stochastic Brownian effects. For this reason, such systems are also called as active Brownian particles that exhibit an interplay between random fluctuations and short-range directional swimming, thereby resulting in enhanced effective diffusion coefficient.113 This is an interesting behavior dominated by directed motion on short time scales and by an enhanced random diffusion in long time scales.130 As a result, there is a strong need to understand the behavior of such micromotors in real-life environments, where such randomness could play a particularly useful role to execute certain bioengineering tasks, such as real-time health monitoring and emergency intervention if need be. Nevertheless, in order for self-phoretic particles to move, the ionic strength and viscosity of the fluid are very critical, which puts a major challenge for the current version of the self-phoretic microswimmers in the ionically rich living environment. To control motion, an on/off control mechanism was shown in the enzyme-driven microswimmers, which was achieved by reversibly inhibiting the enzymatic activity (Fig. 3f).131 For maneuverability, external magnetic guidance131,132 and gravitaxis133 were demonstrated.
When self-phoretic chemical microswimmers are around a solid boundary, interesting motion behavior can be observed. The microswimmers can sense the presence of the boundary through chemical and hydrodynamic interactions, and respond to it by exhibiting reflection, steady sliding, and hovering effects.134,135 An attraction potential between the particle and the wall exists so that the particle cannot escape due to rotational diffusion. Using this, self-phoretic micromotors could be guided along prescribed paths on microstructures (Fig. 3g).136
Enzymatic catalysis has broader applicability in self-phoretic swimmers, where more biocompatible fuel sources, such as glucose and urea, were decomposed by the catalytic activity of glucose oxidase and catalase enzymes, respectively.105,131,132 Nevertheless, an important drawback of enzyme catalysis is the short lifetime of the proteins that may fall short to realize the desired outcome with the microswimmer.
Flagellated bacteria species and spermatozoa are motile microorganisms that can generate thrust forces on the orders of 10−3 and 10−1 nN, respectively.139,140 This power is sufficient to actuate the movement of micron-scale particles (Fig. 4a).137 However, as the movement of each cell is stochastic in nature, directionally controlled actuation presents a challenge. In addition, as most bacteria-driven biohybrid particles tend to be functionalized with more than one cell, opposing forces can further limit the speed or directional control of the robot.141 In this regard, one strategy to direct biohybrid robot swimming is to use the cells' innate environmental responses to signals such as pH, temperature, chemicals, or light.142–144 Remote magnetic fields can also be used to steer cell-driven biohybrid microrobots by integrating to them magnetic micro/nanoscale particles or thin films.145–147 Anisotropic cell patterning strategies have also arisen that bias the movement of the biohybrid robots in the direction opposite to the cell functionalization.141
Fig. 4 Biohybrid microrobot design and application strategies for bioengineering applications. (a) Conceptual designs of bacteriobots by attaching motile bacteria to synthetic drug cargos (reprinted with permission from ref. 137. Copyright 2016 from Elsevier). (b) Fluorescent micrographs of bacteriobots demonstrate pH-triggered accumulation over time, indicating a uniquely coupled motion and sensing capability of biohybrid designs (reprinted with permission from ref. 148. Copyright 2015 from Nature Publishing Group). (c) Left: Magneto- and aero-tactic bacteria Magnetococcus marinus strain used to make a nanoliposome-bacteria biohybrid construct to target mouse tumor. Right: Transverse tumor sections actively populated by the bacteria (reprinted with permission from ref. 152. Copyright 2016 from Nature Publishing Group). (d) An immotile sperm is captured by a remotely controlled magnetic helix and delivered to the oocyte for fertilization (reprinted with permission from ref. 1. Copyright 2016 from the American Chemical Society). (e) An elastic 1D filament with a rigid head and a compliant tail, and a small, single cluster of muscle cells generating power to create asymmetric motion for swimming (reprinted with permission from ref. 160. Copyright 2014 from Nature Publishing Group). (f) Musculoskeletal meso- and micro-architecture of a skate, L. erinacea is replicated in a 2D tissue-engineered ray. Upon optical stimulation, the tissue-engineered ray induces sequential muscle activation via serpentine-patterned muscle tissues, generates undulatory locomotion, and sustains steady forward swimming. It changes direction by generating asymmetric undulating motion between left and right fins, modulated by light pulse frequency (reprinted with permission from ref. 158. Copyright 2016 from the American Association for the Advancement of Science). (g) Optogenetic 3D muscle ring-powered biobots (reprinted with permission from ref. 157. Copyright 2016 from the National Academy of Sciences of the U.S.A.). |
Although single bacterium movement in response to environmental cues is well characterized, the coordinated movement of several bacteria attached to a particle has only recently been described. To this end, multiple S. marcescens-functionalized microparticles were demonstrated to conduct uni- and bi-directional pH-taxis towards neutral pH environments (Fig. 4b).148 Moreover, a microrobot model was developed in this work, which was able to isolate swimming direction and speed bias as the major factors that contributed to the pH-taxis behavior. Because a common cellular signaling pathway is believed to be responsible for different bacterial tactic responses,149 it is possible that such models could be used to optimize the directional control of chemotactic or thermotactic bacteriobots as well.
Cancer therapies by bacteria-driven microrobots have recently emerged with promising results. Tumor-targeting bacteria Salmonella typhimurium (S. typhimurium) was attached via biotin-streptavidin linkers to nanoliposomes loaded with chemotherapeutic paclitaxel.150In vitro co-culture experiments demonstrated that the bacteria-based microrobots selectively migrated towards a murine mammary carcinoma line (4T1), rather than toward healthy human fibroblasts (NIH/3T3), and they led to paclitaxel-mediated 4T1 cell killing. Although this work was not in vivo, earlier works with free S. typhimurium showed promising cancer-selective tissue localization,151 suggesting the promise of employing this strategy in in vivo studies in the future. Magneto- and aero-tactic bacteria Magnetococcus marinus strain MC-1 (MC-1) was recently used as an in vivo active drug carrier via cell surface functionalization with drug-loaded nanoliposomes (Fig. 4c).152 In the direction of the constant magnetic field of an MRI system, the active carriers were steered toward a subcutaneous human colorectal tumor (HCT116) region implanted inside a rat. Magnetic guidance enabled millimeter-scale tumor penetration, while the bacteria's aerotactic behavior enabled micron-scale localization into the hypoxic regions of the tumor.
Because microscale fabrication and control are challenging, particularly with biological components, further optimization of biohybrid swimmers is necessary for their realization in applications, such as drug delivery or diagnostic devices. Stochastic bacterial motion on the surface of bacteria-driven microrobots can be directionally controlled using steering at the particle level. Serratia marcescens (S. marcescens) bacteria-driven magnetic microparticles could be steered using remote magnetic fields.145 Since a magnetic field was employed to apply magnetic torque on the robot, rather than a magnetic gradient to apply magnetic force (i.e. pulling), the driving field strength could be kept low (on the order of a few or tens of mT). Using such method, a bovine sperm-driven microrobot was steered using closed-loop control for its precise navigation control.153 The sperm-driven microrobot consisted of a conical tube made of magnetic thin film, which was used to capture a swimming sperm cell inside, such that its tail remained free and was able to provide propulsion.154 By applying low magnetic fields (1.4 mT), the robot was steered at speeds of 25 ± 10 μm s−1 to a target position with a 90 ± 40 μm precision. Also, spinning micron scale helices coated with magnetic thin films were used to trap and transport immotile sperm cells in an in vitro fertilization experiment (Fig. 4d).1 Single sperm cells could also be released from swimming microtubular structures using remote temperature control.155 Finally, electric fields can also be used to steer bacteria-driven microswimmers.156
Unlike bacteria and spermatozoa, muscle cells do not have a specialized swimming component. Skeletal and cardiac muscle cells contract to generate a directional tension on the order of 103–105 nN in magnitude, in response to an artificial excitation stimulus, such as, chemicals or electric field.27 However, these stimuli are invasive and not practically useful toward bioengineering applications. To this end, recently, both skeletal and cardiac muscle cells were genetically engineered for optogenetic light stimulation, and were successfully used in biohybrid soft robotic constructs (Fig. 4f and g).157–159 As the light-mediated excitation is also spatially well controlled, this method further enabled a steering control by local contractions and maneuverability (Fig. 4g).158 Coupled with the actuation frequency, the magnitude and frequency (1–5 Hz) of muscle cell contraction was repeatable and controllable. This property opened up a design window for muscle-based microrobots, or myobots, in which the spatial organization of the cells on the substrate enables transferring of the contraction force, and hence the bending of construct in a predictable manner. In this regard, 1D biohybrid constructs were shown to be actuated by only a few muscle cells (Fig. 4e).160,161 This actuation could generate sufficient force to make it swim in low Reynolds numbers.160 In 2D, careful spatial decoration of muscle cells on a flexible surface enables programmable control of the cell contraction force in both time and space domains, thus allowing for complex designs.158 In 3D, organization of cells is a lot less accurate in terms of force distribution. Nonetheless, a 3D-printed asymmetric shape hydrogel was shown to be powered by skeletal muscle cells that resulted in a stick slip locomotion.162 In this fabrication scheme, cells were packed in high densities embedded in a supportive hydrogel matrix to maintain the overall integrity. Cell-embedded hydrogel could be molded in particular shapes, which was then incorporated to a synthetic polymer for predetermined actuation. In this way, it was possible to generate 300 μN active tension in the final construct.157 In addition to the inefficiency of the cell placement in 3D, another major limitation for a 3D construct is the lack of transport of nutrients through high cell densities in a sustainable manner.
On-board actuated and powered microrobots operate in far-from-equilibrium states, which requires a continuous supply of energy to maintain their activities for extended periods. To tackle a similar problem, biological organisms have evolved dynamic sensing and regulatory systems that impart precision, robustness, and versatility in the execution of intra- and inter-cellular tasks.163 Similarly functioning artificial control pathways, developed for microrobots, would enable dynamic regulation and decision-making processes, such as fuel consumption rate, adjustment of the swimming speed, 3D localization, and functional engagements at the target body sites. Such a closed-loop control system at the microscale has not been available yet. Systems characterized as far from equilibrium can further allow a series of novel behaviors that are not attainable at equilibrium, such as swarming and the emergence of other collective properties.164 Such behavior was covered briefly in the earlier sections, as only a limited number of studies has so far addressed the operations of microrobots in team or swarm organizations. Nevertheless, it is an important direction to take toward bioengineering applications in the form of parallel and distributed operations, depending on the scale and the type of the intervention.
Due to the abovementioned challenges of on-board microrobot designs, the first-generation microrobots developed by the off-board approach are closer to clinical, in vivo and lab-on-a-chip applications. In the off-board systems, wireless powering and guidance of microrobots eliminate the need for local fuel concerns, autonomous sensing and navigation toward the target site. To this end, however, more sophisticated and high-performance control systems need to be developed. This includes powering microrobots in deep regions inside the body without local heating, simultaneous imaging, actuation, and tracking of the physiological environment and the microrobot(s) during a given biomedical operation, and selectively addressing individual microrobots among a group.
Use of microrobots for lab-on-a-chip applications is another appealing near-term route, because of powerful micromanipulation and precisely controlled mobile tools provided at a size scale that can non-invasively access to small spaces. For example, individual cells or cell packs organized in unit tissue scaffolds could be assembled by means of a off-board powered microrobot for obtaining 3D heterogeneous tissue-mimetic constructs.2 Particularly, the assembly of 3D parts, which require orientation and positional control, would be best addressed using microrobotic assembly. Handling small objects in very small fluid volumes for manipulation, moving and reconfiguring the components by means of 3D micro robots could make this route highly attractive for organ-on-a-chip applications where complex cellular materials with 3D microscale compositional features are positioned.
Fabrication of microrobots, in both on-board and off-board approaches, presents unique challenges concerning design, fabrication process, and encoding operational capabilities. Conventional microfabrication techniques usually provide relatively simple geometric structures, such as tubes, spheres, and surfaces, with limited design flexibility and function. For example, a bubble-propelled micromotor requires well-compartmentalized placement of the catalyst and effective ejection of the jet bubbles in order to produce efficient propulsion from the catalytic reaction. On the other hand, realization of complex designs with programmable distribution of the catalyst and other functional components is a daunting task at the microscale. For this reason, the majority of the bubble-driven microswimmers are in the form of tubes, made with either electrodeposition or rolled up polymer films, and the catalyst is homogenously present everywhere located in the innermost layer.165 Besides, because the propulsion in the low Reynolds regime suffers greatly from the viscous drag, an optimal 3D microswimmer body design is an important parameter to achieve the maximum propulsion efficiencies.
Integration of computer-aided design (CAD) to microfabrication technologies has been a significant advancement to realize sophisticated 3D designs that could not be conceivable with the existing methods. To this end, application of additive manufacturing processes enabled by two-photon crosslinking, also known as direct laser writing technology, has opened up an unprecedented 3D design and manufacturing freedom at the microscale. In regard to its microrobotic applications, Servant et al. used this technology to develop cork-screw-type magnetic microswimmers.70 Very recently, Ceylan et al. has advanced this technology by adding chemical versatility to the 3D-printed bodies, and thereby realizing the first computer-designed and low drag bubble-propelled microswimmers (Fig. 5a).166 Tailorable local 3D chemical properties would allow advanced programmable functionalities, and hence could lead to novel design opportunities for microrobots.
Fig. 5 Some of the critical future considerations in microrobotics concerning fabrication, materials, and design aspects. (a) CAD-designed and 3D-microprinted microrobots with patterned chemical regions can enable encoding complex microrobotic tasks and functions (reprinted with permission from ref. 166. Copyright 2017 from John Wiley & Sons, Inc.). (b) Red blood cells represent a rich source of making massive numbers of microrobots that are immunogenically safe and patient-specific for targeted cargo delivery applications (reprinted with permission from ref. 167. Copyright 2014 from the American Chemical Society). (c) Designing microrobot mobility components that are efficiently propelling in the complex viscoelastic physiological environment is central. The current designs that are typically optimized to operate in Newtonian fluids, e.g., water, may need revisiting for in vitro and in vivo scenarios, especially taking the particular body location and local properties of the physiological fluid into account (reprinted with permission from ref. 169. Copyright 2014 from Nature Publishing Group). |
Material biocompatibility is an important aspect for devices to operate in the living environment. This has attracted little attention so far, while accumulating knowledge and expertise in tissue engineering and related disciplines could provide a wealth of inspirations regarding type and composition of the materials for fabrication of microrobots. When a microrobot is inside biological fluids it is prone to attacks by the cells of immune system, and its circulation time is closely related with the time of recognition by the host immune system. A possible approach could be the use of patient's own biomaterials to fabricate the microrobot bodies. Such a personalized solution could largely circumvent the immune response, as the body would recognize the microrobots as self. For example, turning natural red blood cells into functional micromotors by loading them with magnetic nanoparticles is a promising example toward this purpose (Fig. 5b).167 Wu et al. developed such systems, and could propel them by ultrasound and guide them by magnetic fields. These cells are vastly available in the blood, as such 2.4 million of those cells are being produced each second.168 Use of red blood cells as the base material could be interesting and useful to enable massive amounts of microrobots that could be hardly achieved by any of the existing microfabrication technologies. Moreover, they are mechanically robust and can change shape under applied stress without undergoing plastic deformation. They have an average of 120 days lifetime, during which they travel around 400 km.168 Design systems based on red blood cells could therefore greatly help robust locomotion in blood vessels by dynamically adjusting their diameters.
In the earlier sections, we emphasized the importance of non-reciprocal motion in the low Reynolds numbers for a microswimmer to propel. However, most biological fluids are non-Newtonian, and thereby exhibiting viscoelastic behavior. This environment is vastly different from what the scallop theory was based upon, i.e., Newtonian fluids. Qiu et al. realized that a microswimmer can also move with reciprocal periodic body-shape changes in non-Newtonian fluids (Fig. 5c).169 The net propulsion here is caused by the modulation of the local fluid viscosity by varying the shear rate exerted by the swimmer body itself. This demonstration opens new design considerations for microrobots that are built to operate in non-Newtonian physiological fluids. Moreover, the existing microswimmers may need to be revisited for their optimal design and performance in the living environment, in which their propulsion speed, energy efficiency, and control may significantly vary.
Microrobot | A reprogrammable, microscopic machine with partly or fully self-contained capabilities entitled by on-board motion, perception, and learning. |
Micromotor | A component of microrobot that can convert energy from various sources, such as magnetic fields, light, or chemical bonds, to do mechanical work in the form of directional motion. |
Microswimmer | A specific locomotion mode of a microrobot that is able to propel and do directional motion in bulk fluid. |
Off-board approach | Remotely actuated, powered, and steered microrobots. |
On-board approach | A microrobot is self-contained with all the components necessary to move, sense, and operate in a manner independent from an external intervention. |
Biohybrid design | Single-celled microrganisms are physically incorporated with synthetic materials to exploit cells' integrated powering, motility, and sensing capabilities. |
Fuel | Chemical that is consumed by the micromotor to produce thrust force. |
Functional component | Components of a microrobot apart from its mobility, such as drug cargo, gripper, controlled release system, and sensing. |
This journal is © The Royal Society of Chemistry 2017 |