Nina Z.
Janković
abc,
Wei Lee
Leong
d,
Andrew I.
Ryan
e,
Omar N.
Tantawi
b,
Brian S.
Smith
d and
Desiree L.
Plata
*bc
aDepartment of Chemical and Environmental Engineering, Yale University, New Haven, CT 06511, USA
bDepartment of Civil and Environmental Engineering, Massachusetts Institute of Technology, Cambridge, MA 02142, USA. E-mail: dplata@mit.edu; Tel: +1 617 258 8596
cCenter for Green Chemistry and Green Engineering, Yale University, 225 Prospect Street, New Haven, CT 06520, USA
dEnvironmental Health and Safety, Massachusetts Institute of Technology, Cambridge, MA 02142, USA
eMIT Central Machine Shop, Massachusetts Institute of Technology, Cambridge, MA 02142, USA
First published on 13th February 2024
Split tube furnaces, which rely on insulation commonly made of refractory ceramic fiber (RCF) material, are routinely used in nanotechnology laboratories to generate carbon-based nanomaterials and other manmade materials through chemical vapor deposition (CVD) processes. RCF aerosols can pose a use-phase inhalation risk to operators. We quantified the inhalation exposure risk and designed, built, and tested the impact of a benchtop ventilated enclosure for a common split tube furnace. Direct real-time measurements revealed that traditional use of the furnace could result in peak RCF total and respirable fraction particle mean concentrations of 25 ± 10 mg m−3 and 11 ± 4 mg m−3, respectively (n = 50). Employment of the ventilated enclosure reduces instantaneous exposure to total RCF dust and the respirable fraction to approximately baseline values: 0.006 mg m−3 ± 0.003 mg m−3, and 0.003 mg m−3 ± 0.002 mg m−3, respectively (n = 30). The peak concentration of suspended particulate matter is highly variable over uniform release triggers, ranging from 5–50 mg m−3 for PMTOTAL and 2–18 mg m−3 for PMRESPIRABLE. Electron microscopic examinations of collected airborne materials were conducted to count the airborne number concentrations of RCFs greater than 5 μm in length, less than 3 μm in width, and that met a 5:1 length:width aspect ratio minimum, which are of toxicological concern. Concentrations of those RCFs were similarly reduced when the enclosure was in place. Technical drawings and specifications of the split tube furnace enclosure design are available for ready recreation and implementation, in light industry or laboratory settings, thereby providing low-cost modification to protect the health of workers and researchers.
Environmental significanceSplit tube furnaces are among the most common laboratory devices used to synthesize carbon-based nanomaterials by chemical vapor deposition. This work illustrates that certain use scenarios can lead to unnecessary exposure to furnace insulation materials that may lead to health concerns. We present a low-cost mitigation technology to protect occupational health. |
RCFs are amorphous fibers that belong to the class of synthetic vitreous fibers (SVFs), also known as manmade mineral fibers (MMMF). RCFs are produced by melting calcined kaolinite (Al2Si2O5[OH]4), or a mixture of alumina (Al2O3) and silica (SiO2) in approximately equal portions.9 Other oxides, such as ZrO2, TiO2, B2O3, and Cr2O3 are occasionally added in smaller amounts to modify fiber properties, such as durability and heat resistance.10,11 Subsequently, bulk fibers are formed by either blowing or spinning the molten material, commonly yielding up to 80% w/w fibers (product), and 20–50% w/w nonfibrous particulate (byproduct; i.e., “shot”).9,11,12 The resultant chemical compositions vary with approximately 50% w/w silica, 35–50% w/w alumina, and less than 1–15% w/w other oxides, most notably zirconia.9 At service temperatures in excess of approximately 980 °C over extended periods of time, RCFs partially convert to the silica polymorph cristobalite through devitrification.11,13,14 The devitrification rate is dependent on temperature and time, and the resultant material is commonly referred to as “after-service fiber”.9,15
Fibrous particles are physical toxicants, whose toxicity uniquely depends on their geometry, where fibers are characterized by large aspect ratios (length-to-diameter ratios).16 The pathogenicity of inhaled fibers is controlled by the coordinated effects of the dose, dimension, and durability of fibrous particles (known as the “3D paradigm” of fiber toxicity).16–18 These 3Ds are interrelated; for example, the length of inhalable fibers deposited in the lung is a primary determinant of fiber residence time.16,18 Biopersistence is a measure of fiber residence time in the lung, and it incorporates breakage, dissolution, and mechanical clearance phenomenon.16,17 Natural mineral fibers, such as asbestos minerals, are crystalline and fracture longitudinally along cleavage planes, under mechanical stress, resulting in acicular (i.e., needle-shaped) fragments.18 Consequently, fragmentation of asbestos fibers produces a larger number of asbestos fibers with decreased diameters, while retaining their lengths. Conversely, RCFs are amorphous fibers that fracture transversely, producing shorter fibers with the same diameters, which may be cleared by macrophages.18,19 This capacity to fracture transversely after deposition in the lower respiratory tract translates to a lower biopersistence for RCF (88 days in rodents) than for asbestos fibers (1000 days for amosite and crocidolite);16,20 in spite of the difference, RCF are not toxicologically benign.
Several animal studies of chronic nose-inhalation showed that rodents exposed to high levels of RCFs developed fibrosis and tumors (lung cancer or mesothelioma), indicating RCFs were persistent long enough to cause lung pathologies.10,11,20–22 The International Agency for Research on Cancer (IARC), classified RCF as possibly carcinogenic to humans (Group 2B) based on sufficient evidence for carcinogenicity in experimental animals, but inadequate evidence in humans.12 Similarly, “[o]n the basis of a weight-of-evidence carcinogenic risk assessment, the EPA23 [also] classified RCFs as a Group B2 carcinogen (probable human carcinogen based on sufficient animal data).”9 As such, prudent guidelines for limiting occupational RCF exposure have been in place for over three decades.10,23,24 National Institute for Occupational Safety and Health (NIOSH) established a recommended exposure limit (REL) of 0.5 f cm−3 as a time-weighted average (TWA) for up to a 10 h work shift in a 40 h workweek, where “reasonable efforts” should be made to reduce exposure below 0.2 f cm−3.9 In spite of this recommendation, no legally enforceable, Occupational Safety and Health Administration (OSHA) permissible exposure limit (PEL) exists for RCFs, apart from the California Division of Occupational Safety and Health (Cal/OSHA) PEL of 0.2 f cm−3 in California.25 However, RCF exposures have successfully been monitored and controlled through a voluntary product stewardship program, an industry-wide program developed by RCF manufacturers (overseen by OSHA) to identify and control risks associated with manufacture and use of RCF-containing products.10 While the stewardship program successfully decreased average fiber exposure concentrations for both manufacturers and customers,10,26 the many RCF-lined furnaces are found in research laboratory settings are not part of the program. Considering just our own university as a representative sample, we found the use of furnaces to be prevalent, with 218 registered furnaces spread over 1021 principal investigators (including faculty without labs), on Massachusetts Institute of Technology's (MIT) campus alone.
In campus settings, Environmental Health and Safety (EHS) departments are not required or compelled to monitor the release of RCFs from these commonly used furnaces, because it is not obvious that there is an imminent occupational hazard associated with use of clamshell furnaces. (Here we note that many campus EHS units do recommend enclosure, ventilation, exhaust for furnaces that use or generate hazardous chemicals or vapors). However, certain furnace configurations have exposed insulation or have operation routine that requires access to insulated compartments, and this motion can disturb the insulation, incidentally releasing fibers into the air. In particular, benchtops below split-tube furnaces often have a thin coating of white dust, which could contain RCFs and alludes to the potential for user inhalation exposure.
To determine the possible dust and RCF exposures during laboratory-scale nanomaterial synthesis, we used a representative model furnace to investigate the possible peak dust and average RCF concentrations in the breathing zone during furnace use. We simulated use with agitation events and deployed both instantaneous and filter-based measurement techniques to gauge episodic and integrated exposures, as well as to characterize released materials. Finding some undesirable exposure scenarios, we present designs for low-cost enclosures with dynamic motion accommodation to mitigate occupational risk to researchers associated with CVD nanomaterial production in split-tube clamshell furnaces. Importantly, this work provides some of the first measures of RCF exposure during the routine utilization of tube furnaces and provides a route for the safe utilization of these critical tools for discovery and advanced manufacturing.
Two Lindberg/Blue M Mini-Mite furnaces (part number TF55030A-1) were used for aerosol and bulk RCF sampling. The furnace's exterior and insulation dimensions were 41L × 28W × 38H centimeters and 36L × 16W × 20H centimeters, respectively. The furnace can accommodate a 2.54 cm process tube diameter and has an internal heated zone of 30.5 cm in length. The age of the furnaces was approximately ten years.
A custom-built box enclosure (Fig. 1) was designed and constructed from clear acrylic and aluminum. The front, back, left, and right sides of the enclosure were made from clear acrylic for transparency. Note that polycarbonate is an alternative material that is less susceptible to melting or burning than acrylic, and that some fire codes place restrictions on the use of acrylic panels. The bottom of the enclosure was made using an aluminum plate connected to a block of plywood for mechanical stability. An aluminum plate was used for the top of the enclosure to ensure resistance to high-heat events associated with an open clamshell furnace. The benchtop enclosure was connected to local exhaust ventilation (i.e., a snorkel).
Fig. 1 Schematic of furnace inside ventilated enclosure. (a) 3-D view; (b) front view; (c) side view. Note the enclosure can accommodate two lengths of the Lindberg/blue M mini-mite furnace. |
Two features provide dynamic motion of the furnace within the enclosure while it is sealed, slide rails and lid lifting handles. The slide rails allow for rolling of the furnace along the quartz tube (e.g., to control the insertion of a substrate into the heated zone). The enclosure length can easily be modified to accommodate a traditional stationary furnace setup. Furnace lid lifting handles were fashioned from an aluminum rod and attached to the top of the furnace. These handles protrude from the left and right sides of the enclosure and are guided by a curved opening which follows the arched trajectory of lifting the furnace lid. Since vertical alignment of the furnace tube is sometimes desired or necessary with both the feed system at the tube inlet and the collection or further processing system at the tube outlet (each of which might be constrained in their vertical alignment due to weight, stability, or other limitations), the underside of the split tube furnace can be equipped with rugged, adjustable laboratory jacks. In such cases, the inlet and outlet openings of the enclosure could also be fabricated as vertical slots to readily allow vertical adjustments of the tube height.
An opening with a detachable cover was added to the front of the enclosure to provide full access to the inside of the box, which is intended for occasional use. This opening was sealed with a foam gasket and screws. Two 1.25 inch openings located at the front of the enclosure serve as inlets for air and are covered with MERV13 and high efficiency particulate air (HEPA) filters (to prevent particle discharge into the room in the case of air flow disturbance). All other openings (i.e., furnace lid lifting handles, furnace rolling rod, quartz tube, furnace cable) were sealed with rubber sheets to control and limit the air flow to the designated air flow openings. The effluent from the quartz tube reactor is also directed to the enclosure for discharge. (Technical drawings and CAD files of the design are available by contacting the author).
In designing the enclosure to mitigate exposure to releases of RCF insulation material and facilitate furnace operation as described above, it is prudent Industrial Hygiene and Occupational EHS (IH/OEHS) practice27 to ensure that the proposed enclosure design would not introduce or increase risks from other hazards. Other hazards associated with split tube furnace operations include (but are not necessarily limited to) pinch or crush injuries during positioning or manipulation of the furnace body and clamshell; electrical shock; thermal burns; cuts, puncture wounds, or eye injuries from contact with potentially broken or damaged ceramic tube materials; inhalation or dermal exposure to releases of RCF insulation material during furnace setup, operation, maintenance, or decommissioning; and inhalation or dermal exposure to fugitive airborne releases of process materials from within the tube furnace during pre-operational, operational, and post-operational activities. Risks of exposure to these other hazards are not increased by the selected design.
The mass concentration was reported in mg m−3 for PM2.5 using a laser light-scattering photometer. The photometric signal combined with single particle optical sizing allows for the following size segregated mass distributions to be calculated: PM1, PM2.5, respirable (PM4), PM10, and total (greater than PM10).28,29 The DustTrak DRX monitor has a mass concentration range of 0.001–150 mg m−3 and a particle size range of approximately 0.1–15 μm.30
The mass of individual counted fibers was estimated from the volume and density. For the calculation of volume, the lengths and widths of the fibers were directly measured, and the fibers were assumed to be cylindrical, as the third dimension could not be directly measured. The mass of individual particles (including aggregates) was estimated for approximately 60 particles or aggregates.
Fig. 2 Bulk material characterization by SEM and TEM. (a) and (b) are SEM images and (c) is a TEM images, centered around representative RCFs. |
In RCF manufacturing, the particle-to-fiber ratios range from 20:80 to 50:50 percent by weight.9 Typically, the product (fibers) yield is 50% by weight, and the rest is byproduct (nonfibrous particulate).9,11 To estimate the particle to fiber ratios of our bulk material, we analyzed 56 particles, by TEM; the fibrous content is approximately 80% by weight. The presence of nonfibrous nanoparticle agglomerates (mostly SiO2) falsely increases the cumulative estimated mass; the thickness of particles is assumed to be equal to the width, since thickness cannot be directly measured. All particles were analyzed by SAED; only aluminosilicate particles diffracted. We note that these composition estimates agree with the disclosed chemical composition from the furnace manufacturer, under the trade name Moldatherm, (i.e., 79–99% aluminosilicate and 1–21% amorphous silica).14
Exposure limits are established based on specific fiber size criteria. For an airborne particle to be counted as a fiber it must meet the following criteria: greater than 5 μm long, less than 3 μm diameter, and with an aspect ratio greater than or equal to 5:1 (NIOSH B rules). To investigate the size distribution of fibers, we measured the lengths and widths of RCFs for portions of the bulk sample and airborne fibers collected on filters (Fig. 3). EDS confirmed fibers in the airborne sample are RCFs with a chemical composition of 47 wt% Al2O3 and 53 wt% SiO2 (approximately 1Al2O3·2SiO2). Nearly all RCF lengths cluster below 50 μm for all measured fibers; 78% of fibers for the bulk sample are between greater than 0.1 and less than or equal to 1 μm in width, while 80% of fiber widths for the airborne sample without enclosure are between greater than 0.5 and less than or equal to 1.5 μm in width (Fig. 3). Overall, the fiber size distribution is similar between the bulk and airborne samples, which supports the assumption that the bulk material is representative of the airborne sample (Fig. 3).
Fig. 3 Aerosol and bulk fiber size distribution. (a) RCFs greater than 5 μm in length, less than 3 μm in width, and with a greater than or equal to 5:1 length:width aspect ratio counted by TEM. Grey dotted line highlights the 5 μm length minimum. Purple curve represents 5:1 length:width aspect ratio for each measured particle width. (b) Fiber size distribution represented as a fraction across different size bins based on length or width of the fiber (1/2 fibers omitted, see ESI† for figure including 1/2 fibers). |
In order to determine exposure to RCF, we analyzed the airborne samples to determine RCF concentration. We sampled across 10 agitation events (approximately 15 min), following NIOSH 7402 to collect and analyze the samples (expanded for RCF; using the “B” counting rules).31–33 The average RCF concentration without the enclosure was 0.83 f cm−3; this equates to an 8 h TWA of 0.0256 f cm−3 which approaches the ATSDR IMRC of 0.03 f cm−3. This result suggests that if the set of agitations described were repeated 10 times throughout an 8 hour workday, the ACGIH TLV 0.2 f cm−3 exposure limit would be breached. No fibers were detected when the furnace enclosure was employed, demonstrating the effectiveness of the engineering control to contain RCFs.
Fig. 4 RCF aerosol concentrations mitigated by ventilated enclosure. Particle concentration was measured with a DustTrak DRX in ISO 12103-1 A1 ultrafine test dust equivalents. Red dotted lines denote the OSHA PEL (8 hour TWA) for total nuisance dust (15 mg m−3) and respirable fraction of particulates not otherwise regulated (PNOR) or “nuisance dust” (5 mg m−3).36 Box plots represent peak suspended particulate matter concentrations, where the crosses represent mean values, the line in the middle of the box represents the median, the hinges (top and bottom) of the box indicate the inner quartiles (25–75th percentile), and the whiskers represent the spread (minimum and maximum values) of all data. PM1, PM2.5, PM10, indicated the particle count for particulate matter (PM) less than 1, 2.5, and 10 micrometers in aerodynamic diameter assuming a spherical particle. RESP indicates the total respirable fraction (defined as less than 4 micrometers). TOTAL is the sum of all measured particulate matter. |
Real-time measurements using DustTrak DRX are useful to estimate RCF exposure, but do not distinguish between fibrous and non-fibrous suspended particulate matter. The geometry of the particle has important implications for the toxicity. Thus, to estimate peak fiber concentrations, we converted the total peak mass concentration (Fig. 4) to estimated fiber concentrations using the median (0.02 ng f−1) or the 97% confidence intervals (lower 0.01 ng f−1 or upper 0.03 ng f−1) or the minimum (0.002 ng f−1) or the maximum (0.4 ng f−1) airborne fiber mass and a range of particle-to-fiber ratios. The particle-to-fiber ratios were chosen based on TEM measurements as discussed earlier (20:80 wt%) and shot content (20–50 wt%).9,19 Mean airborne fiber mass was not considered because the standard deviation was larger than the mean. The mass of individual fibers was estimated from the volume and density. To determine volume, the lengths and widths of the fibers were measured via imaging and assumed to be cylindrical. The airborne fiber mass (ng f−1) statistics were calculated from a subset of fibers, which all meet particular fiber size criteria (greater than 5 μm long, less than 3 μm in diameter with an aspect ratio greater than or equal to 5:1). Converting from mass to fiber concentration involves two key limiting assumptions: (1) the subset of fiber masses considered is a representative sample of fiber mass distribution, and (2) a sample size of 85 fibers is representative of the fiber distribution.
The sample of fibers considered does not have a normal distribution, which is evidenced by a standard deviation greater than the mean fiber mass. Therefore, we estimate the mass per fiber using the median fiber mass. Further, we bound the median by considering the upper and lower confidence intervals at 97%. Additionally, to show the range of values considered, we utilize the minimum and maximum mass per fiber.
We calculated that peak total fiber concentrations were on the order of 103 f/cm3 based on median mass fiber and mean peak mass concertation (DustTrak). The lowest calculated peak fiber concentration was 30 ± 10 f cm−3 and ranged 6–60 f cm−3 (Table 2). Even the lowest values exceed the proposed excursion limit for RCFs in the range of 0.3–5 f cm−3. The proposed excursion limit is discussed in greater detail below.
We only measured fibers that met the criteria defined under NIOSH 7402 and 7400 B rules,31–33 therefore the fiber mass used was not representative of the true fiber mass distribution. Nevertheless, we find that even the lowest fiber concentrations surpass proposed short-term exposure limits proposed below. Further to partially resolve the constraints on the fiber mass, we use a wide range on fiber masses to calculate possible peak fiber concentrations (Table 2).
In the absence of an OSHA PEL for RCFs, we can refer to the following 8 h TWA exposure limits NIOSH REL of 0.5 f cm−3, American Conference of Governmental Industrial Hygienists (ACGIH) threshold limit values (TLV) of 0.2 f cm−3, and agency for toxic substances disease registry (ATSDR) inhalation minimal risk concentration (IMRC) 0.03 f cm−3 exposure limits.9 Note that the NIOSH REL was established as a TWA for up to a 10 h work shift in a 40 h workweek. Estimated peak fiber concentrations (Table 1) notably exceed the recommended TWA exposure limits. While the TWA exposure limits cannot be compared directly to peak values, exceeding time weighted limits for any amount of time, particularly for a highly variable fiber release process, suggests that it is possible to exceed the limit under some use scenarios or circumstances.
RCF concentration (f cm−3) | Sensitivity (structures per cm3) | 95% CI (low–high) | 8 h TWA (f cm−3) | |
---|---|---|---|---|
Without enclosure | 0.83 | 0.0081 | 0.57–1.2 | 0.0256 |
With enclosure | <0.0079 | 0.0079 | 0–<0.035 | <0.0002 |
Inside enclosure | 0.40 | 0.0067 | 0.26–0.60 | 0.0112 |
PMtotal (mg m−3) | Mass per fiber (ng f−1) considered | Fibers total (f cm−3) | Fibers total (f cm−3) | |
---|---|---|---|---|
Particle-to-fiber ratio 20:80 (wt%) | Particle-to-fiber ratio 50:50 (wt%) | |||
25 ± 10 (mean ± st dev) | Median | 0.02 | 1000 ± 400 | 700 ± 200 |
97% CI (lower bound) | 0.01 | 1000 ± 500 | 900 ± 300 | |
97% CI (lower bound) | 0.03 | 700 ± 300 | 400 ± 200 | |
Min | 0.002 | 10000 ± 4000 | 6000 ± 2000 | |
Max | 0.4 | 50 ± 20 | 30 ± 10 | |
5–50 (range: min–max) | Median | 0.02 | 200–2000 | 100–1000 |
97% CI (lower bound) | 0.01 | 300–3000 | 200–2000 | |
97% CI (lower bound) | 0.03 | 100–1000 | 90–900 | |
Min | 0.002 | 2000–20000 | 1000–10000 | |
Max | 0.4 | 10–100 | 6–60 |
Further, in the absence of a ceiling, short-term exposure limits (STEL), or immediately dangerous to life or health (IDLH) limits for RCFs, we may consider the way in which asbestos is regulated. The 8 h TWA exposure limit OSHA PEL for asbestos is 0.1 f cm−3, with an excursion limit of 1.0 f cm−3 averaged over any 30 minute sampling period.37 The excursion limit for asbestos is an order of magnitude greater than the PEL, and we could employ this as a proxy to estimate the STEL for RCFs. Consequently, an estimate for a recommended excursion limit for RCFs may be in the range of 0.3–5 f cm−3 based on the recommended time-weighted exposure limits from various agencies listed above. Peak fiber concentration estimates are well above the upper limit of the approximate excursion limit (raised tenfold above the 8 h TWA).
Conversely, the NIOSH method 7402 for sampling and analysis of fibrous material, provides an integrated sample from which the average fiber concentration is determined, which masks the peak concentrations found in episodic exposures. Concurrently, this method enables exact and explicit determination of fiber concentration for a subset of fibers with particular sizes defined by NIOSH A or B rules for counting fibers.
While sampling by NIOSH 7402 method, DustTrak was run in parallel, which permitted a direct comparison of the resultant average fiber concentration to the average mass concentration. We related peak mass concentrations to peak fiber concentrations through a conversion factor, mg f−1. We found a peak total fiber concentration of 6 ± 2 f cm−3 (equivalent to PMTOTAL 25 ± 10 mg m−3), and a range of 1.2–12 f cm−3 (equivalent to PMTOTAL 5–50 mg m−3).
To protect the furnace operators from potential excursions in (1) fiber concentrations above the approximate excursion limits for short term exposures, (2) the time weighted exposure limits for fibers, and (3) the OSHA PEL for PNOR for both total dust and respirable dust fractions, we built a custom ventilated enclosure (Fig. 1 and S1†).
We found that peak particle and fiber concentrations exceeded time-weighted permissible limits, which indicated that it is possible to surpass these limits depending on use scenarios. We established that uncovered furnace use poses an occupational risk through episodic exposure, which was highly variable from between agitation events. Aerosol particles and fibers were generated through use and their concentrations had a wide range of values. We measured these concentrations under one type of use scenario in one environment. However, other use scenarios and operator practices may result in higher or lower RCF aerosol concentrations. Other types of scenarios include use of multiple furnaces in the same space and use of fans for fast cooling of the insulation, which we reason would affect resultant fiber concentrations. Further, differences in air flow environments, such as ventilation rate, may affect operator exposure in the absence of an engineering control. With this publication we seek to promote environmental health and safety across multiple disciplines by making our economical custom engineering control design accessible to researchers who use furnaces that contain RCFs. The furnace enclosure computer-aided design (CAD) files and technical drawings are available by contacting the corresponding author for ready reproduction of this adaptable engineering control.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3en00041a |