aInstitute of Biological Information Processing – Bioelectronics, IBI-3, Forschungszentrum Juelich, 52428, Germany. E-mail: francesca.santoro@iit.it
bFaculty of Electrical Engineering and IT, RWTH Aachen, 52074, Germany
cTissue Electronics, Istituto Italiano di Tecnologia, 80125, Naples, Italy
dDepartment of Materials Science and Engineering, Stanford University, Stanford, CA, USA. E-mail: alexander.giovannitti@chalmers.se
First published on 20th December 2023
Organic electrochemical transistors (OECTs) have shown great potential in bioelectronics to transduce small biological signals for applications such as the electrical recording of excitable cells and assessing cell barrier properties. It is imperative that operating the OECT as a biosensor does not affect the biological system. However, bias voltages applied to channel materials such as the conducting polymer (CP) PEDOT:PSS have been shown to induce the formation of hydrogen peroxide (H2O2) which can disrupt the physiology of cells. In this work, we evaluated the impact of H2O2 formation during OECT operation by comparing an oxygen-sensitive CP (PEDOT:PSS) and an oxygen-stable CP (p(gPyDPP-MeOT2)). While both CPs show high biocompatibility in their non-biased, resting state, we observed large differences during the operation of the electrochemical device. OECTs with PEDOT:PSS produce H2O2 where the H2O2 concentration in the electrolyte depends on the channel area and the time of operation. In comparison, OECTs using the oxygen-stable DPP-based polymer showed no sign of H2O2 formation. Further investigation also revealed how the proliferation rate of neuronal cells directly interfaced with such OECTs was affected by the concentration of H2O2. Our work demonstrated the limitations of oxygen-sensitive OECT channel materials for bioelectronic applications and provides guidance for material design strategies to develop safe bioelectronic devices for real-life applications.
Typically, the channel material consists of an organic mixed ionic-electronic conductor (OMIEC)11 which is designed to simultaneously transport electronic and ionic charge carriers for efficient signal transduction. In contact with aqueous electrolytes, solvent molecules, and ions can migrate into the bulk of the channel material (swelling) which opens the structure of the polymer to enable ion penetration into the bulk of the polymer. Furthermore, upon applying an electrochemical potential with respect to the gate (or reference) electrode (Ag/AgCl), electronic charges are injected into the OMIEC and ions migrate from the electrolyte into the bulk of the OMIEC to compensate for the electronic charges injected into the system. Conventionally, the performance of OECTs is benchmarked by the product of the electronic mobility (μ) and the volumetric charge storage capacitance (C*), which are used to calculate the transconductance of the transistor.13 Although the transconductance is an important figure of merit, it only defines the amplification of the device and does not include other aspects such as biocompatibility, electrochemical side reactions, and operational stability. While some work described the operational stability of devices,14 little work has been carried out on investigating the impact of electrochemical side reactions (faradaic reactions) during device operation, forming reactive side products that may be harmful to biological systems.
Currently, PEDOT:PSS is the most used channel material as it achieves high transconductance and is commercially available.15 However, recent reports have demonstrated that OECTs and electrodes based on PEDOT:PSS form hydrogen peroxide (H2O2) during operation in oxygenated aqueous electrolytes16–20 (Fig. 1). These electrochemical side reactions are expected to cause degradation of the transistor material and importantly, also affect the health of cells that are in contact with the device. H2O2 is a reactive oxygen species that plays an essential role in the physiology of cells and tissue.21–23 It is generated naturally through metabolic activity and can diffuse across cellular membranes to act as a molecule messenger and establish redox signaling that initiates direct cellular effects such as changes in shape, proliferation, and differentiation. Under normal conditions, living organisms exhibit antioxidant systems that balance the generation and elimination of hydrogen peroxide. However, external factors can for instance facilitate oxidative stress that disrupts the redox signaling and damages biomolecules such as DNA, proteins, and lipids ultimately leading to cell death.24–26 Furthermore, the oxygen reduction reaction (ORR) can disrupt the cell physiology and deplete oxygen in the electrolyte which may lead to hypoxia for cells near the device. Thus, to engineer inert bioelectronic devices for real-life applications, oxygen-stable OECT materials are needed that function without faradaic side reactions with molecular oxygen to limit the accumulation of H2O2 during device operation.
![]() | ||
| Fig. 1 A schematics of PEDOT:PSS OECT in its resting state and upon Vg operation, showing electron transfer from the polymer channel to the electrolyte and consequent generation of H2O2. | ||
In this work, we evaluated the impact of H2O2 formation during OECT operation in oxygen-containing aqueous electrolytes by comparing an oxygen-sensitive OMIEC (PEDOT:PSS) to an oxygen-stable OMIEC (p(gPyDPP-MeOT2)). For each channel material, the generated H2O2 concentration was determined and the viability of neurons that were exposed to peroxide containing electrolytes was examined. Furthermore, neurons cultured on a PEDOT:PSS channel were visually inspected before and after OECT operation.
The amount of hydrogen peroxide formation during the operation of the devices was measured with a colorimetric assay. In the presence of hydrogen peroxide, the horseradish peroxidase (HRP) catalyzed the conversion of the fluorescent probe into a colorimetric product that changed the absorbance spectra of the solution. A calibration curve was used to determine the concentration of hydrogen peroxide generated during the operation of the OECTs (Fig. S2, ESI†). Operating the PEDOT:PSS OECT at peak gm (Vg = 700 mV vs. Ag/AgCl, with Vd = −500 mV) for 60 seconds caused generation of H2O2 in the electrolyte ranging from 20 to 50 μM (Fig. 2a). On the contrary, no hydrogen peroxide was detected (considering the experimental concentration limit of 5 μM) in the electrolyte of the OECT using polymer p(gPyDPP-MeOT2) (Fig. 2b). For the PEDOT:PSS OECTs, the concentration of generated H2O2 was determined for different durations of the applied gate potential as well as for different channel dimensions. OECTs with millimeter and micrometer scale dimensions were operated for 1 and 5 minutes and the colorimetric assay was used to determine the H2O2 concentration (color map in Fig. 2c). We observed the highest concentration of H2O2 for the large OECT (7 × 4 mm2) when a long gate pulse (5 minutes) was applied indicating that the H2O2 formation depended on the active area of the device and the operation time. The number of charges injected from the gate electrode into the polymer channel correlated to the generated concentration of H2O2, confirming that the by-products from the electrochemical reactions were directly related to the biasing of the OECT (Fig. S3, ESI†).
![]() | ||
| Fig. 2 Current response of the OECT with (a) PEDOT:PSS and (b) p(gPyDPP-MeOT2) upon applying a square voltage pulse at the gate electrode (Vg = 700 mV, and Vd = −500 mV for PEDOT:PSS, Vg = −700 mV, and Vd = −100 mV for p(gPyDPP-MeOT2)). The circular insets show the concentration of peroxide recalling the same color code reported in Fig. 3c. Degradation of p(gPyDPP-MeOT2) is observed at potentials >−650 mV.14 (c) Color map representing the concentration of H2O2 (expressed in μM) produced during the operation of a PEDOT:PSS OECT for different channel areas and gate pulse duration. | ||
The biocompatibility of both polymers was first assessed in their unbiased state (i.e., no gate voltage was applied), by cultivating neuronal cells (HT22) on polymer films and using a live/dead fluorescence assay. After 24 hours in culture, high percentages of green fluorescent cells (>95%) were found indicating excellent viability on both PEDOT:PSS and p(gPyDPP-MeOT2) (Fig. 3a(i)–(iii)). No significant difference in the percentage of live cells was found compared to the glass substrate control, suggesting negligible cytotoxic effects (Fig. 3b). To determine whether the amount of peroxide produced during PEDOT:PSS-based OECT operation affected the cell viability, two concentrations of H2O2 (20 μM and 100 μM) were tested, corresponding to the amount of peroxide produced with large channel area-long pulse duration and small channel area-short pulse duration, respectively. After 1 day from plating, HT22 cells were exposed to hydrogen peroxide by exchanging the cell medium with H2O2-containing medium at the desired concentration after which the metabolic activity was evaluated at different time points through an MTT assay. At low H2O2 concentration (20 μM), a slight decrease in the metabolic activity was observed that was proportionally related to the exposure time to H2O2 (Fig. S4, ESI†). Increasing the H2O2 concentration to 100 μM showed a large change in the absorbance values after 15 minutes, indicating a significant decrease in the metabolic activity of cells (Fig. S4 (ESI†) and Fig. 3c). This was further confirmed by a change in the cell shape as well as by the presence of cell clusters (Fig. 3c(iv)–(vi)). These results demonstrated the sensitivity of HT22 cells hydrogen peroxide produced concentrations during the transistor operation.
To evaluate whether the presence of cells might affect the production of hydrogen peroxide protecting the polymer form undergoing the ORR, HT22 cells were cultivated (60
000 cell per cm2) on top of a PEDOT:PSS OECT, covering the channel area (Fig. 4a). The device was operated at Vg = 700 mV and Vd = −500 mV for 5 minutes. Comparing the amount of H2O2 produced before and after cell plating, no significant difference was found (λ570before = 0.111 ± 0.011 λ570after = 0.099 ± 0.020, n = 3) suggesting that the presence of cells is not limiting the generation, diffusion and detection of peroxide (Fig. 4b).
After device operation, cells were incubated for 1 hour in the same media. The absorbance values evaluated through MTT assay indicated a drastic decrease in the metabolic activity of the HT22 (λ570 = 0.122) with respect to the control (λ570 = 0.348) (not operated device) (Fig. 4c). Similar results were found with 3 different devices (Fig. S5, ESI†). Brightfield images showed round-shaped morphology and cluster formation confirming the adverse effects of oxidative stress on the cells physiology.
:
1) media (Merck Life Science S.r.l., Italy) supplemented with 10% fetal bovine serum (Merck Life Science S.r.l., Italy), 1% penicillin/streptomycin (Merck Life Science S.r.l., Italy), 1% L-glutamine (Merck Life Science S.r.l., Italy), in a 5% CO2 incubator at 37 °C. Confluent HT22 cells were trypsinated using Trypsin-EDTA 0.25% (Life Technologies Italia, Italy) and cells were seeded on the polymeric substrates at a density of 79
000 cells per cm2 and left in the incubator at 37 °C for 24 hours. To assess the biocompatibility cells were stained with a solution of Calcein-AM (Merck Life Science, final concentration 1 μg mL−1) and Ethidium Homodimer (Thermo Fisher, final concentration 10 μg mL−1) in PBS and incubated at 37 °C for 10 minutes.
000 cells per cm2) per well were seeded into a 24-well tissue culture plate for 24 hours. Before plating, substrates were functionalized with poly-L-lysine (PLL, Merck Life Science S.r.l., Italy) 0.01% v/v (aqueous solution), incubated at 37 °C overnight and then washed three times with PBS (Merck Life Science S.r.l., Italy). A solution of MTT 3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyl tetrazolium bromide (Abcam, UK) in 1× solution of PBS (Merck Life Science S.r.l., Italy) (5 mg mL−1) was added to the wells (200 μL in 1 mL of medium) and incubated at 37 °C for 4 hours. After the removal of the medium, 1 mL of DMSO (Merck Life Science S.r.l., Italy) was added to all wells and mixed thoroughly to dissolve the dark blue crystals (insoluble purple formazan formed by cleavage of the tetrazolium ring by dehydrogenase enzymes). After a 1 hour at room temperature, to ensure that all crystals were dissolved, the DMSO solution was measured spectrophotometrically using Dynatech MR580 Microelisa reader, using a test wavelength of 570 nm.
Footnotes |
| † Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3tc02849f |
| ‡ Equal contribution. |
| § Current address: Department of Materials Science and Milano-Bicocca Solar Energy Research Center – MIB-Solar, University of Milano-Bicocca, 20125 Milano, Italy. |
| ¶ Current address: Department of Chemistry and Chemical Engineering, Chalmers University of Technology, Göteborg 412 96, Sweden. |
| This journal is © The Royal Society of Chemistry 2024 |