Mario F.
Mirabelli†
a,
Emanuela
Gionfriddo‡†
b,
Janusz
Pawliszyn
b and
Renato
Zenobi
*a
aETH Zurich, Department of Chemistry and Applied Biosciences, 8093 Zurich, Switzerland. E-mail: zenobi@org.chem.ethz.ch
bDepartment of Chemistry, University of Waterloo, Waterloo, Ontario, Canada
First published on 8th March 2019
A direct and fast method for screening and quantification of illicit drugs in beverages and biological fluids was developed by using dielectric barrier discharge (DBD) as ionization technique, in combination with high-resolution mass spectrometry. Extraction of the targeted analytes was carried out with thin film microextraction (TFME) using ultrasound as agitation method. The targeted analytes were then thermally desorbed and introduced into the source without the need for any cryofocusing apparatus. The performance of this new analytical set up were compared to conventional TFME-TDU-GC/MS, showing enhanced linear dynamic range and lower limits of quantitation (low pg ml−1) achievable at the same extraction conditions (5 min extraction time). The performance of the method was tested in different beverages and body fluids, confirming its applicability for quantitative analysis of the targeted drugs in complex samples.
Recently, the rapid development of various ambient mass spectrometry technologies opened the door to quick and easy direct introduction of samples or extracts into mass spectrometers.6,7 In particular, the dielectric barrier discharge ionization (DBDI) source was already successfully used for the analysis of analytes collected by SPME devices in the fiber8,9 and arrow10 geometries. More recently, capillary atmospheric pressure photoionization (cAPPI) was successfully employed for the direct SPME analysis of nonpolar compounds like PAHs.11
A DBD ionization source consists of a low-temperature plasma, which is ignited between two electrodes separated by a dielectric material.12–15 In recent years, DBDI has been increasingly used for analysis of various sample matrices for direct mass spectrometric16–19 or chromatographic analyses.20–23
The source employed in this work uses an “active capillary” sampling geometry, meaning that the source is directly connected to the mass spectrometer, and the flow of gas and analytes through the source is determined by the vacuum inside the mass spectrometer. This enhances the sensitivity, due to a loss-less transmission of the ions formed into the mass spectrometer. Interfacing of this DBDI source to liquid and gas chromatography was also recently reported.23,24 In addition to its simple design, our DBDI source can also operate with different gases for the plasma ignition, depending on the specific analytical needs. Finally, in contrast to hard ionization methods such as EI, DBDI is a soft ionization method that generates little to no fragmentation, mostly yielding protonated molecules.
In this work, we propose the direct coupling of a self-supported TFME device to a DBD ion source and high-resolution mass spectrometry, as a convenient and rapid analytical setup for quantification of drugs in beverages and biofluids. No cryofocusing system was required since the gas flow rate around the TFME device was sufficiently high to allow for rapid thermal desorption of the analytes. Compared to other approaches,5 the desorption chamber used was directly connected to the ionization source, minimizing desorption time and consequently potential thermal degradation of compounds on their way into the ionization source. This SPME extraction/desorption strategy ensures that most of the matrix components and molecules with very low volatility are either not extracted from the sample or not desorbed from the SPME thin film, thus giving minimal sample carry-over. The method proposed herein also provides the advantage of fast sampling and data acquisition without sacrificing the sensitivity of the analysis.
For the preparation of the TFME devices, a SYLGARD184 silicone elastomer kit was purchased from Dow Corning (Midland, MI, USA). DVB particles with a diameter of 3–5 μm were kindly provided by Millipore Sigma. The TFME devices were prepared according to the procedure proposed by Jiang et al.4 The dimension and shape of the TFME device were customized as a 20 × 3 mm rectangle with a film thickness of 250 ± 10 μm.
The PDMS/DVB TFME devices were supported by a stainless-steel wire (0.125 mm diameter, 60 mm length) and were fully immersed in the sample matrix during extraction from 1.5 mL glass vials. Samples were extracted for 5 minutes at 40 °C. Instead of a classic stirring/shaking approach, an ultrasound-assisted extraction was used since the thin film devices were found to be stable to ultrasound. For this purpose, a lab sonicator was used, and the sonication bath was kept at a constant temperature of 40 °C (±1 °C).
N2 was used as desorption gas, at a temperature of 260 °C and a relative humidity of 90% (at 25 °C). The carrier gas flow was maintained constant at 3 L min−1 by means of a mass-flow controller (Bronkhorst High Tech B.V. Ruurlo, Netherlands). Of the total 3.0 L min−1, only 1.0 L min−1 entered the ionization source. This overflow of nitrogen was used to prevent room air from entering the setup. This prevented the reproducibility of the results from being negatively impacted, considering that the desorption chamber has an open design (no seals/septa used). In addition, the presence of oxygen from the air during desorption should be avoided since it could oxidize the SPME coating at high temperatures.
As previously shown for SPME fibers, a fast desorption is obtained with the use of small I.D. desorption chambers, in order to ensure a quick displacement of the desorbed compounds from the proximity of the fiber.25 We have recently shown the direct coupling of SPME to MS with the use of an open system,8 where the best results were obtained using a desorption chamber with a narrow internal diameter (I.D.) even when employing a high desorption gas flow rate. Since TFME devices can be manufactured in shapes significantly bigger than fibers, a desorption chamber with a proper I.D., able to accommodate the devices, had to be used. In our case, we tried to minimize the dimensions of the thin films, without compromising their extraction capacity deriving from a geometry with higher surface area and coating volume. TFME devices having a width of 3 mm were found to be easy to manufacture and handle for this application. A length of 20 mm was chosen to be compatible with the 1.5 mL glass vials. Shorter films did not offer advantages during desorption. The desorption chamber was equipped with 4 mm I.D. glass liners, passivated by silanization, and capable of accommodating the thin films without bending the devices.
Considering the larger volume of the 4 mm I.D. desorption chamber, and the different geometry of the thin films compared to SPME fibers, it was necessary to verify that the linear gas velocity applied would guarantee efficient desorption of the analytes. Therefore, the desorption kinetics for the target drugs was investigated using different temperatures, ranging from 200 °C to 260 °C (Fig. 3). Considering previous findings on the thermal stability of SPME matrices, higher temperatures were not tested.2,4 Most of the target analytes showed a fast desorption kinetics, with a complete desorption in less than 2 min at 260 °C. Diazepam was found to be strongly retained by the film, resulting in a slower desorption, which was complete only after 6 min at 260 °C.
Considering the flexibility of the self-supported TFME device used, two different sample agitation modes were tested to optimize the extraction conditions: agitation by stir bar and sonication. Generally, the use of sonication is not widespread for SPME, due to concerns regarding the SPME devices’ stability toward ultrasound energy and high variability of the sonication bath temperature during continuous and extensive sonication procedures. However, in this work, we found that the self-supported TFME devices provided enough stability to endure sonication for up to 120 min. We also carefully controlled the temperature of the sonication bath to ensure constant extraction temperature (within ±1 °C). The extraction efficiency and kinetics of the TFME devices were evaluated using both agitation modes, setting the extraction temperature to 40 °C and determining extraction time profiles from 1 to 120 min in ultrapure water. As shown in Fig. 4, for the representative analyte fentanyl, sonication is able to provide a more efficient agitation method with faster extraction kinetics. In light of the results obtained, sonication was selected as the agitation method for this study and an extraction time of 5 min was selected as a good compromise between sensitivity and throughput for the method developed.
Fig. 4 Extraction time profile of fentanyl extracted from water obtained by sonication and stir bar agitation methods. |
In addition to extraction of drugs from water samples, real sample matrices were used to simulate cases of drug analysis in forensic and medical applications, i.e., vodka, wine, cola, human urine, and blood plasma. The beverages chosen for this study present different challenges due to their composition: in particular, the ethanol content of vodka and wine, which constitutes a competing phase for the partitioning of the analytes into the TFME device, and/or the presence of interfering molecules that can affect the quantification of drugs at ultra-trace level. In the case of cola, sugars are present at very high concentrations (ca. 10%). In wine, several polyphenolic compounds with low volatility are present, and the EtOH content ranges usually from 12% to 14% (v/v). In the vodka sample, the concentration of EtOH was even higher (37.5%).
The influence of the amount of alcohol in the sample matrix on the extraction efficiency was investigated by extracting the target analytes from water/ethanol mixtures of various compositions (EtOH concentrations of 0, 10, 20, 40 and 60% (v/v)). The ultrasound-assisted extractions were performed in pre-equilibrium conditions (5 min) at 40 °C. The resulting extracted ion chromatograms were integrated over a 0.5 min time period. The results are reported in Fig. 5.
Most analytes were efficiently extracted up to EtOH concentration of 20%. The extraction efficiency from alcoholic solutions is compound-dependent, and is a function of the partitioning of the analytes between the sample solution and the TFME device, which is influenced both by the analytes’ hydrophobicity and π–π interactions with the extraction phase. In most cases, more than 50% of the signal was still present at 40% EtOH content, when compared to extraction from pure water, with the exception of cocaine and diazepam, which showed a lower extraction efficiency at high alcohol concentration. Since we employed a sensitive DBDI source coupled to a high capacity SPME device, the reduction in the extraction efficiency did not jeopardize the quantification of target analytes at the pg mL−1 level in vodka.
In forensic applications, e.g., a drink spiked with illicit substances, a variable concentration of drugs is expected, depending on the potency of the drug. Potent drugs like fentanyl are effective even at μg levels. In these cases, the drug concentration would be high enough to allow shorter extraction times to be used. To demonstrate the potential of a faster approach, extractions of drinks spiked at 1 μg mL−1 were performed at room temperature without stirring, by inserting the thin film in the liquid samples for only 5 seconds. After extraction, a quick (3 seconds) manual rinsing step in LC-MS grade water was performed to remove potential residues of matrix components, and afterward the thin films were directly desorbed. This approach allows for a quasi-real-time analysis (Fig. 6).
An extraction time of 5 min, together with the accelerated extraction kinetics obtained with the ultrasound-assisted extraction, was sufficient for the quantification of the considered drugs at ultra-trace levels. Calibration curves (1/x weighted) for the drugs in vodka, urine and blood plasma are reported in Table 1.
Compound | I S | Equation | LDR, pg mL−1 | R 2 | LOD, pg mL−1 | |
---|---|---|---|---|---|---|
Vodka | Cocaine | Cocaine-d3 | y = 1.11 × 10−3x − 6.93 × 10−2 | 30–30k | 0.9961 | 10 |
Diazepam | Diazepam-d5 | y = 1.12 × 10−3x − 1.64 × 10−1 | 100–30k | 0.9923 | 30 | |
Desipramine | Fentanyl-d5 | y = 3.49 × 10−4x − 1.07 × 10−2 | 100–30k | 0.9978 | 30 | |
Imipramine | Diazepam-d5 | y = 1.26 × 10−2x + 3.72 × 10−1 | 100–30k | 0.9950 | 10 | |
MDMA | MDMA-d5 | y = 1.07 × 10−3x − 2.06 × 10−2 | 100–30k | 0.9929 | 10 | |
MDEA | MDEA-d5 | y = 9.05 × 10−4x + 4.00 × 10−3 | 100–30k | 0.9998 | 10 | |
Methadone | Methadone-d3 | y = 9.27 × 10−4x − 1.26 × 10−2 | 30–30k | 0.9992 | 10 | |
Fentanyl | Fentanyl-d5 | y = 8.59 × 10−4x − 1.39 × 10−2 | 100–30k | 0.9994 | 30 | |
Urine | Cocaine | Cocaine-d3 | y = 1.18 × 10−3x − 3.58 × 10−2 | 30–10k | 0.9978 | 30 |
Diazepam | Diazepam-d5 | y = 9.79 × 10−4x − 1.45 × 10−3 | 10–10k | 0.9982 | 3 | |
Desipramine | Cocaine-d3 | y = 1.04 × 10−4x − 1.02 × 10−2 | 100–10k | 0.9931 | 100 | |
Imipramine | Cocaine-d3 | y = 2.75 × 10−3x − 6.67 × 10−3 | 10–10k | 0.9998 | 3 | |
MDMA | Diazepam-d5 | y = 1.23 × 10−6x + 1.99 × 10−5 | 100–10k | 0.9882 | 30 | |
MDEA | Diazepam-d5 | y = 5.21 × 10−6x − 3.01 × 10−4 | 300–10k | 0.9982 | 30 | |
Methadone | Methadone-d3 | y = 9.83 × 10−4x − 7.44 × 10−3 | 10–10k | 0.9982 | 3 | |
Fentanyl | Fentanyl-d5 | y = 1.05 × 10−3x − 8.66 × 10−3 | 10–10k | 0.9985 | 10 | |
Blood plasma | Cocaine | Cocaine-d3 | y = 1.18 × 10−3x − 7.45 × 10−2 | 100–30k | 0.9996 | 100 |
Diazepam | Diazepam-d5 | y = 1.17 × 10−3x − 1.69 × 10−1 | 300–30k | 0.9961 | 300 | |
Desipramine | Fentanyl-d5 | y = 1.89 × 10−4x − 5.43 × 10−2 | 300–30k | 0.9986 | 300 | |
Imipramine | Methadone-d3 | y = 2.07 × 10−4x − 2.17 × 10−2 | 300–30k | 0.9993 | 100 | |
MDMA | MDMA-d5 | y = 1.36 × 10−3x + 5.78 × 10−1 | 300–30k | 0.9993 | 300 | |
MDEA | MDEA-d5 | y = 1.27 × 10−3x − 9.79 × 10−3 | 100–30k | 0.9996 | 100 | |
Methadone | Methadone-d3 | y = 1.07 × 10−3x − 2.98 × 10−2 | 30–30k | 0.9998 | 30 | |
Fentanyl | Fentanyl-d5 | y = 1.18 × 10−3x − 1.39 × 10−1 | 300–30k | 0.9997 | 300 |
The results obtained by the TFME-DBDI setup are even more rewarding when compared to what can be obtained by a conventional TDU-GC/MS system with the same extraction conditions. In general, when calibration was attempted in urine, cola, wine, and vodka, using the same extraction conditions, only in a few cases (Table 2) were consistent results obtained. In all cases, LOQs were above 10 ng mL−1. Moreover, compounds like MDMA and MDEA cannot be analyzed by gas-chromatography without a derivatization step, which would add extra effort to the analytical workflow.
Compound | Equation | LDR, ng mL−1 | R 2 | LOQ, ng mL−1 | |
---|---|---|---|---|---|
Cola | Cocaine | — | — | — | — |
Diazepam | y = 0.0106x − 0.0368 | 10–1000 | 0.9993 | 10 | |
Desipramine | — | — | — | — | |
Imipramine | y = 0.00156x − 0.0974 | 100–1000 | 0.9923 | 100 | |
MDMA | — | — | — | — | |
MDEA | — | — | — | — | |
Methadone | — | — | — | — | |
Fentanyl | — | — | — | — | |
Urine | Cocaine | y = 0.0103x + 0.0127 | 10–500 | 0.9998 | 10 |
Diazepam | y = 0.0003x − 0.0324 | 10–1000 | 0.9993 | 10 | |
Desipramine | — | — | — | — | |
Imipramine | y = 0.0198x − 0.444 | 10–1000 | 0.9974 | 10 | |
MDMA | — | — | — | — | |
MDEA | — | — | — | — | |
Methadone | y = 0.0113x − 0.444 | 10–1000 | 0.9912 | 10 | |
Fentanyl | y = 0.0106x − 0.0368 | 10–1000 | 0.9993 | 10 |
The reported direct coupling of TFME with DBDI showed good compatibility with the sample matrices investigated. In particular, the reproducibility data reported in Table 3 show an average intra-day RSD of 6% both for 1 ng mL−1 and 10 ng mL−1 samples, with all values below 21%.
1 ng mL−1 | 10 ng mL−1 | ||
---|---|---|---|
Vodka | Cocaine | 2.0 | 7.1 |
Diazepam | 1.9 | 6.6 | |
Desipramine | 9.7 | 13 | |
Imipramine | 3.5 | 4.8 | |
MDMA | 7.4 | 7.2 | |
MDEA | 3.8 | 8.6 | |
Methadone | 0.1 | 8.1 | |
Fentanyl | 9.7 | 10 | |
Urine | Cocaine | 7.6 | 2.1 |
Diazepam | 2.9 | 1.8 | |
Desipramine | 5.7 | 8.3 | |
Imipramine | 14 | 13 | |
MDMA | 21 | 14 | |
MDEA | 17 | 9.8 | |
Methadone | 3.6 | 1.9 | |
Fentanyl | 7.1 | 2.4 | |
Blood plasma | Cocaine | 1.0 | 1.2 |
Diazepam | 5.7 | 2.3 | |
Desipramine | 11 | 9.8 | |
Imipramine | 4.9 | 8.5 | |
MDMA | 3.6 | 1.8 | |
MDEA | 1.8 | 1.9 | |
Methadone | 1.4 | 1.1 | |
Fentanyl | 1.4 | 2.8 |
Compared to other direct approaches where no sample pretreatment is performed,18,19,26 the use of TFME as a fast and simple technique for liquid sample cleanup/up-concentration is very practical, because it minimizes the amount of interfering compounds ionized together with the analytes of interest (with possible suppression effects) and results in lower LODs due to the pre-concentration provided by the thin film. Although waiving the sample preparation completely would reduce the analysis time to just a few seconds per sample, it appears necessary to perform at least a simple sample pretreatment in order to be able to analyze complex matrices.
The method reported in this work can be fairly directly compared with other SPME-MS direct coupling strategies.27 For example, a method based on the thermal desorption of the analyte prior or during ionization using a DART source was used by Vasiljevic et al. for the analysis of drugs of abuse in oral fluids and urine.28 A sensitivity in the ng mL−1 range was achieved by extracting samples with a coated plastic mesh SPME device, analyzed in transmission mode. In this study, the LODs in urine for cocaine, diazepam, MDMA, methadone and fentany were 0.1, 0.5, 2.5, 1 and 0.5 ng mL−1, respectively. The values reported in this study for the same analytes are, 0.03, 0.003, 0.03, 0.003 and 0.01 ng mL−1, respectively, i.e., more than one order of magnitude lower. Evans-Nguyen et al. also reported the analysis of benzodiazepines in human plasma by gold-wire SPME fibers,29 obtaining a sensitivity in the sub-ppb range.
Solvent-based SPME desorption methods were also successfully used. These methods are generally less comparable to the one described here, because of the different ionization mechanisms and sample desorption. Gómez-Ríos et al. reported the analysis of drugs of abuse in human urine and plasma at the sub-ppb concentration level by using an open port probe (OPP) ionization source.30 Tascon et al. quantified immunosuppressants in whole blood at the sub-ppb level with a microfluidic open interface (MOI) ionization source.31 Coated blade spray32 (CBS) was also used for the quantification od drugs of abuse in human urine and plasma. These techniques, namely OPP, MOI, and CBS, represent a viable alternative to DBDI when an ESI-like ionization is preferred. The use of the abovementioned approaches can be advantageous when only a minor modification of an ESI source is desired (this is the case for OPP or MOI). CBS, although being a more flexible technique, still poses some challenges in terms of automation. The same applies to nano electrospray-based direct SPME-MS couplings,33 where the amount of manual operation still strongly limits the analytical throughput.
The strongest matrix effects were observed in blood plasma, and to a lesser extent, in urine (Fig. 7). In fact, both of these matrices contain proteins, hormones, metabolites and salts. Some of their constituents can bind some of the analytes, reducing the effective amount that the thin film can extract (via free concentration).
Fig. 7 TFME-DBDI-MS ion signals for different drugs at 1 ng mL−1 extracted from vodka, urine, and plasma (1.5 mL, 5 min extraction, n = 4). |
Despite the presence of almost 40% ethanol, vodka showed the lowest matrix effect when compared to urine and plasma. For plasma and urine, which are more complex matrices and contain a greater variety of interfering compounds, ionization suppression effects were observed in addition to competition effects during extraction, as indicated by the higher level of MS signal background from blank extractions. The different response obtained for urine and blood samples also depends on the highest degree of binding of the analytes to matrix components.
The approach was found to be very robust, with an average intra-day RSD of 6% both for 1 ng mL−1 and 10 ng mL−1 samples, with all values below 21%. The figures of merit were obtained with an extraction time of only 5 minutes, although currently with manual handling of the TFME during extraction and desorption. Therefore, room for further improvements in terms of throughput and reproducibility is possible with automation of the analytical workflow. This would also enable the proposed approach to be used in many analytical laboratories, for research and routine measurements.
As future perspectives, the bio-compatible nature of the thin films could also allow for direct sampling of saliva or sweat in a non-invasive way, simplifying sample collection procedures and allowing for more rapid results to be obtained. This might, for example, be helpful for anti-doping investigations, where speed is extremely important to rapidly assess the compliance of the athlete to regulations. In such cases, the extraction time can even be reduced to below 1 minute, allowing for on-site/real time sampling.