Andrew
Taylor
*a,
Nicola
Barlow
b,
Martin P.
Day
c,
Sarah
Hill
d,
Nicholas
Martin
e and
Marina
Patriarca
f
aSupra-regional Assay Service, Trace Element Laboratory, Royal Surrey County Hospital, 15 Frederick Sanger Road, Guildford, Surrey, GU2 7YD, UK. E-mail: andrewtaylor4@nhs.net
bTrace Elements Laboratory, Department of Clinical Biochemistry, Sandwell General Hospital, West Bromwich, West Midlands, B71 4HJ, UK
cThe Australian Wine Research Institute, PO Box 197, Glen Osmond, SA 5064, Australia
dLGC, Queens Road, Teddington, Middlesex, TW11 0LY, UK
eDepartment of Clinical Biochemistry, Charing Cross Hospital, 8th Floor Lab Block, Fulham Palace Road, London, W6 8RF, UK
fIstituto Superiore di Sanitá, Viale Regina Elena 299, 00161 Roma, Italy
First published on 14th February 2018
This review covers publications from the second half of 2016 to the middle of 2017. Techniques and applications relevant to clinical and biological materials, foods and beverages are discussed in the text, presenting the key aspects of the work referenced, while the tables provide a summary of the publications considered. Further to the observations made in the 2017 Update, we have noted the increase in publications featuring nanomaterials. Topics relate to exposure and absorption of nanoparticles; potential toxicity, use as reagents in analytical methods and the analysis or characterisation of these materials. A growing number of techniques are being applied to bio-imaging. In particular, those involving XRF spectrometry are being applied to a large range of biological sample types. A gradual trend during recent years, that is now quite apparent, has been the increasing proportion of publications that involve foods and beverages compared with clinical samples. Not necessarily related, but another trend is the growing number of reports from laboratories in the Republic of China.
Sample preparation is crucial to all analytical procedures and features extensively in all our Updates. Much of the recent interest has focused on dispersive liquid–liquid microextraction (DLLME). Campillo et al.7 produced a review of the different variations and examples of the technique that have developed during the last ten years. With a more specific objective, de la Calle et al.8 discussed how DLLME may be coupled with ETAAS, and the applications made possible when using this approach.
In our recent reviews we have drawn attention to the technical developments and the varied applications associated with bio-imaging techniques. As a consequence, a number of review articles featuring different aspects of imaging, were published in the last year. Collingwood and Adams9 prepared a tutorial review on X-ray methods applied to imaging analysis of the brain. Methods covered included XRF spectrometry for elemental imaging, XRA spectrometry for speciation imaging, X-ray diffraction for structural imaging, phase contrast for enhanced contrast imaging and scanning transmission X-ray microscopy for spectromicroscopy. Two- and three-dimensional (confocal and tomographic) imaging methods were considered as well as the correlation of X-ray microscopy with other imaging tools. They pointed out that X-rays have the advantage to penetrate through a thick biological sample providing imaging of metals in brain tissues at regional, cellular, and sub-cellular spatial resolution. Two publications described how LA-ICP-MS was employed to investigate clinical disorders. Pozebon et al.10 concentrated on the technical improvements during the last three years, referring to fast wash-out laser ablation cells, novel calibration strategies, nanoparticle uptake, applications and mapping of elemental distribution in biological samples. Meanwhile, Becker and Sussulini explained the formation of qualitative and quantitative maps showing distributions of elements in tissue sections of brain, cartilage, spinal cord, etc. in studies of neurodegenerative diseases.11 Other MS approaches, reviewed by Lee et al.,12 have been applied to metal-based anticancer drug research. Further discussion of developments in imaging is found in Section 6.2 of this Update. Cellular distribution of Zn has long been investigated in relation to its role in processes such as synaptic activity in neuronal cells. Earlier studies exploited fluorescent sensors to identify labile Zn pools. Carpenter et al.13 reviewed work to show that mapping using MS and XRF techniques complement and strengthen data elaborated from sensors, with examples that included fertilisation of eggs and carcinogenesis.
As seen throughout this review and in Section 5, much innovation in the last few years has involved nanomaterials. Karadjova et al.14 considered how materials such as magnetic nanoparticles, carbon nanostructures, metal oxides, noble metal nanoparticles, and ion imprinted polymers have been used for speciation applications such as analysis of food and biological samples. Single particle ICP-MS (spICP-MS) is a powerful technique to detect and characterise aqueous dispersions of metal and metalloid-containing nanomaterials. Recent advances in data acquisition, signal processing, and the use of alternative mass analysers (e.g. TOF-MS) provide a larger toolbox offering progress toward overcoming many of the challenges in the quantitative determination of nanoparticles. A review by Montano et al.15 gave an overview of spICP-MS, discussing key issues for quantitative analysis of increasingly complex environmental and biological samples.
The feasibility of utilising LIBS for food analysis, was published by Markiewicz-Keszycka et al.16 who suggested that the technique provided an alternative approach to time-consuming analytical methods. The review contained an overview of LIBS fundamentals, instrumentation and statistical data analysis and described recent progress and applications to assess the quality and composition of food products. It concluded that while the technology shows many advantages, sample preparation, matrix effects, spectral pre-processing and calibration still present challenges that require resolution.
Other reviews relating to food analysis are of interest. Taylor et al.17 considered combining the distribution of organic As species in seafood with analytical aspects and consumption and toxicity data to identify gaps in current understanding. They highlighted that for organic As species further work is required: more complete As speciation across seafood types, particularly for arsenolipids; more RMs; more advanced toxicity testing is necessary to assess their effects; studies on metabolic pathways and individual variability, and distribution of As compounds in the body, to examine their toxicokinetics; and identification of biomarkers for As in seafood. To inform regulatory practices, a crucial missing link to understanding exposure is a full assessment of As in food that considers organic As compounds in seafood. Moreda-Piñeiro et al.18 discussed the determination of Se and Se-compounds bioavailability, using both in vivo and in vitro tests. It was shown that results depended on the experimental conditions selected and the authors proposed that an internationally agreed standardised protocol for such experiments was necessary. Food traceability is of considerable commercial importance and we report on relevant work in these Updates (Section 8.3). Baffi and Trincherini19 discussed recent results using the 87Sr:86Sr isotopic ratio technique applied to vegetables, beverages, dairy products, meat and fish products. Summaries of work considered for this ASU are shown in Tables 1 and 2.
Element | Matrix | Technique | Sample treatment/comments | Reference |
---|---|---|---|---|
Ag | Pulmonary epithelial cells | SF-ICP-MS | Study of the effect of NP diameter and tissue culture medium composition on AgNP uptake rates in cultured 16HBE14o cells | 55 |
Ag | T84 human colonic epithelial cells | ICP-MS | Total Ag was measured in conjunction with other analyses following exposure of basal cells to AgNPs of different sizes and at different doses | 103 |
Ag | Serum, environmental waters | SEC-ICP-MS | A rapid 8 min method for measurement of elemental mass size distribution of Ag and AuNPs. High ionic strength mobile phase (2% FL-70™ (strong alkaline detergent)/2 mM Na2S2O3) ensured size independent elution of NPs from the column. The NP size distribution obtained agreed closely with that from TEM and this method had a higher mass sensitivity (1 pg) and comparable size discrimination (0.27 nm) | 114 |
Al | Simulated body fluid, artificial saliva | ICP-MS | Evaluation of metal ion (Al, Ti, V) release from plasma electrolytic oxidation coatings used for dental and orthopaedic implants while immersed in normal and inflammatory simulated body fluids | 142 |
Al | Human brain tissue | GF-AAS | Following microwave-assisted digestion, brain tissue from 12 patients with familial Alzheimer’s disease was found to have Al content ranging from 0.01 to 35.65 μg g−1 dw with 58% of tissues having pathologically significant concentrations ≥3.00 μg g−1 | 149 |
As | Erythrocytes, whole blood, plasma, urine and liver tissue | ICP-MS, HPLC-ICP-MS | In a case study of poisoning with iAs and CrVI, three days post exposure, blood As was 206 μg L−1, erythrocyte CrVI was 4540 μg L−1, blood and plasma total Cr were 2180 and 1070 μg L−1, and liver As and Cr content was 0.9 and 11.7 mg kg−1 respectively. Speciation analysis of As showed the predominant species to be iAs, MMA and DMA | 151 |
As | Urine | HPLC-HG-ICP-MS | Urinary concentrations of iAs, MMA and DMA in 9 year old Bangladeshi children (n = 488) showed a higher percentage DMA and lower iAs with respect to their mothers. The association of these concentrations with erythrocyte Se and folate was also investigated | 121 |
As | Urine | HPLC-ICP-MS | In a Spanish population (n = 124), MMA and AsIII were detected in 5.6% and 1.6% of subjects respectively while AsV was not found. Total As concentrations were higher than reported by other European studies but were composed of primarily AB and DMA | 183 |
As | Urine | LC-ICP-MS | Concentrations of five As species were higher in infertile men with normal semen (n = 101) than fertile controls (n = 61), with AsV being 20-fold higher. The primary As methylation index (MMA/iAs) and AsV were most significantly associated with unexplained male infertility | 184 |
As | Urine | HPLC-ICP-MS | Sum concentrations of iAs and MMA were assessed as a biological indicator of total iAs exposure in 330 exposed and 172 non-exposed subjects | 150 |
As | Urine (NIST SRM 2669) | LC-ICP-MS | Fast speciation of As in 5 min using a short-length mixed bed ion exchange column with (NH4)2CO3 mobile phase coupled to ICP-MS. The AsIII was quantitatively converted to AsV with H2O2. The method achieved RSDs of 2.0 to 4.8% for all species and LODs from 0.003 to 0.051 ng g−1. In urine samples, the sum of species measured corresponded to 62 to 125% of total As | 185 |
As | Soft tissue and bone phantoms | SR-XRF, portable XRF | A comparison of results obtained using SR-XRF (monochromatic beam energy 15.8 keV) and portable XRF (Rh anode X-ray tube operated at 40 kV) for As in soft tissue phantoms and Pb in bone phantoms. Lower LODs were achieved for SR-XRF | 93 |
Au | Serum, environmental waters | SEC-ICP-MS | See Ag, ref. 114 | 114 |
Au | Mouse organs | ICP-MS | Following intravenous injection of Tween® 20-conjugated AuNPs for 21 days in mice, quantification of total Au found highest concentrations in the bladder and hematopoietic bone tissue | 108 |
Au | Fibroblasts, mouse organs | ICP-MS | In vitro uptake by fibroblasts of functionalised AuNPs and Au nanoclusters was evaluated by ICP-MS in conjunction with fluorescence confocal microscopy. In vivo studies of 9 mice treated for 24 h with 1.70 μg g−1 AuNPs showed different distributions of accumulation depending on the functionality of the NP | 104 |
Au | Human placental co-culture microtissues | ICP-MS, LA-ICP-MS, TEM | Higher uptake and deeper penetration of AuNPs was demonstrated for smaller particles and those modified with Na carboxylate rather than PEGylate in an in vitro study | 102 |
Au | Human skin hair follicles | ICP-OES, LA-ICP-MS | Evaluation of the distribution of hydrophobic, neutral and charged Au nanorods in hair follicles and skin | 105 |
Au | Macrophage-like RAW 264.7 cells | Cluster SIMS | Chemical imaging of AuNPs using C60+ primary ions in a single cell system. Fluorophore-functionalised AuNPs allowed visual confirmation of the intra-cellular location of NPs | 133 |
B | Serum, whole blood, urine and drinking water | ICP-MS | Serum B concentrations in 194 pregnant women in northern Argentina were 0.73 to 605 μg L−1 (median 133 μg L−1) correlating strongly with blood and urine concentrations and to a lesser extent with drinking water B (377 to 10929 μg L−1). Serum B >80 μg L−1 was inversely associated with birth length | 153 |
B | Urine, rye grass (ERM-CD281) | ICP-OES | A fast and simple method for urine B biomonitoring using ICP-OES, which achieved precision ≤2% and recoveries of 102 to 109%. When compared with results obtained by Q-ICP-MS and SF-ICP-MS at low and medium resolution, there was good agreement for 11B but suspected Ne2+ (an Ar impurity) interference on 10B with all methods except medium resolution SF-ICP-MS | 152 |
Ca | Serum | ICP-MS | A candidate reference method for serum Ca involved first spiking samples gravimetrically with Al internal standard then digestion with concentrated HNO3 before dilution to measure concentration. The 44Ca/27Al ratios were measured in H2 mode. Analytical figures of merit were CVs of 0.27 and 0.16% and recovery, 100.24%. Analysis of CRMs and comparison with a reference method were satisfactory | 27 |
Ca | Anterior pituitary secretory granules | XRF microscopy, ICP-MS | A study of rat pituitary sections by XRF microscopy and subcellular sections by ICP-MS to localise and quantify Ca, Cu and Zn. The Ca concentration was approximately 10-fold higher than Zn and 50 times higher than Cu | 135 |
Cd | Blood | GF-AAS | Reference ranges for blood Cd and Pb were determined for residents of Athens. The 95th percentile was 1.8 μg L−1 for blood Cd with a median of 1.2 μg L−1 in smokers versus 0.46 μg L−1 in non-smokers | 147 |
Ce | Rat blood, urine faeces and organs | ICP-MS | Investigation of the bio-distribution and pharmacokinetic modelling of inhaled fresh and environmentally aged CeO2 NPs. Most Ce was recovered in the lungs and faeces with no differences between fresh and aged NPs | 100 |
Co | Biological tissues; lobster hepatopancreas (NRC TORT-2 and TORT-3) | VG-ICP-OES | Photochemical VG of volatile Co species dissolved in 50% CH2O2 through exposure to UV radiation (254 and 185 nm) for 10 s before detection by ICP-OES at 238.892 nm afforded a 27-fold increase in sensitivity compared with pneumatic nebulisation. Optimum generation efficiency was 42 ± 2%, the LOD was 0.4 μg L−1 and precision was 3% at 100 μg L−1 | 73 |
Co | Hip fluid simulant | ICP-MS | A modified sample preparation procedure, utilising nano-porous ultra-filters to separate particles from free ions and complexes in solution was employed to investigate the release of ions (Co, Cr and Mo) from Co, Cr, Mo alloys during wear simulation in different media | 137 |
Co | Whole blood | ICP-MS | A method for routine analysis of blood Co and Cr achieved an LOD for Co of 0.06 μg L−1. Spiked whole blood and historical PT material were used to demonstrate method accuracy, precision and recovery. Short term room temperature stability (64 h) of blood samples was shown to be acceptable | 138 |
Co | Urine, blood and plasma | ICP-MS | Wide fluctuations in urinary Co concentrations were observed through the working week in 34 occupationally exposed individuals. Mean urine Co concentrations were significantly correlated with levels of Co-haemoglobin adducts and whole blood Co. Half-lives of Co were determined to be 12.3, 9.1 and 5.3 days in blood, plasma and urine respectively | 155 |
Co | Human heart tissue | ICP-MS | Heart tissue for Co, Cr and V measurement was prepared by closed-vessel microwave digestion. Accuracy was verified using RMs and the within run CV was <20%. Reference ranges were established from 80 non-metal hip implant heart tissues. The reference range for Co (mean ± SD) was 0.1039 ± 0.1305 μg g−1 | 143 |
Cr | Hip fluid simulant | ICP-MS | See Co, ref. 137 | 137 |
Cr | Whole blood | ICP-MS | See Co, ref. 138. The LOD for Cr was 0.41 μg L−1 | 138 |
Cr | Human heart tissue | ICP-MS | See Co, ref. 143. The reference range (mean ± SD) for heart Cr was 0.1523 ± 0.2157 μg g−1 | 143 |
Cr | Erythrocytes, whole blood, plasma, urine and liver tissue | ICP-MS, HPLC-ICP-MS | See As, ref. 151 | 151 |
Cr | Exhaled breath condensate (EBC), urine | ICP-MS, LC-ICP-MS | Measurement of CrIII and CrVI in EBC with separation of the Cr species via a micro-sized column in a cohort of 58 occupationally exposed workers demonstrated significantly higher concentrations compared with 22 controls | 154 |
Cu | Anterior pituitary secretory granules | XRF microscopy, ICP-MS | See Ca, ref. 135 | 135 |
Cu | Erythrocytes | HPLC-ID-ICP-MS | Development of a candidate reference method for quantification of Cu, Zn-SOD in erythrocytes using species specific double and triple ID-MS. Sample preparation involved precipitation of Hb by addition of 3:1 EtOH:CHCl3 followed by concentration of the supernatant using micro-centrifugation filters. The protein was then separated and quantified by strong anion exchange HPLC-ICP-MS. An expanded measurement uncertainty of ≤1.50% was obtained | 65 |
Cu | Serum | MC-ICP-MS | Application of high precision Cu isotopic analysis to monitoring patients post liver transplant. Lighter isotope ratios were associated with persistence of poor liver function as well as seeming to predict disease recurrence | 157 |
Cu | Serum | MC-ICP-MS | No significant differences were found between Cu isotope ratios measured following a new Cu isolation procedure, involving a Chelex-100 chelating exchange resin followed by an AG® MP-1 anion exchange resin, and the conventional procedure, which consisted of two passes through the AG® MP-1 resin. The method was applied to study diurnal variation of Cu isotopic composition (n = 5) | 38 |
Cu | Human bone and tooth enamel | LC-MC-ICP-MS | A report of Cu and Zn isotope ratios in historical French bone and teeth remains found no sex-related differences in Cu isotope ratios but observed a decrease in Zn isotope ratios related to increased meat and fish consumption during the 20th century | 186 |
Fe | Serum | MC-ICP-MS | High precision Fe isotopic analysis of serum showed heavier Fe isotope ratios in patients with chronic kidney disease and Fe disorders compared with controls but not in erythropoietin-related anaemia | 158 |
Gd | Bone phantoms | XRF | An in vivo XRF system, based on a 109Cd source (5 GBq) was investigated for detection of Gd in doped bone phantoms. The LOD was 0.8 ppm. For phantoms coated with lucite as a tissue equivalent, LODs were 1.81 to 3.47 ppm | 160 |
Gd | Brain and bone tissue | ICP-MS | In 9 patients with normal renal function given a single dose of non-group 1 Gd-based contrast agents, higher concentrations of Gd were measured in bone tissue and these correlated with Gd in brain | 161 |
Gd | Rat brain and blood | ICP-MS | Concentrations of Gd were measured in 12 rats after high and low dosing regimens with gadodiamide contrast agents. Dose-dependent low levels of Gd were detected in brain one week after dosing (mean 2.49 nmol g−1), which diminished to approximately 50% by 20 weeks | 162 |
Gd | Rat blood, CSF and brain tissue | ICP-MS | Measurement of Gd in rat blood, CSF and brain tissue, 24 h post administration of different high dose Gd-based contrast agents (1.8 mmol kg−1), yielded no agent dependent differences. The pattern of Gd concentrations in CSF and brain tissue suggested CSF as an entry point into the brain | 163 |
Hg | Hair | LA-ICP-MS, GC-ICP-MS | In 5 individuals exposed to Hg vapour 3 months previously, single point LA of hair samples (50 μm diameter at 1 mm intervals), using a 34S internal standard followed by speciation analysis showed up to a 1000-fold increase in iHg concentrations at corresponding distances from the root. This was explained by direct adsorption of Hg vapour onto hair at the time of exposure | 187 |
Hg | Human whole blood, hair and urine; caprine blood (NIST SRM 955c level 3); human hair (IAEA 085/086) | GC-ICP-MS | A method for simultaneous determination of MeHg, EtHg and iHg in various biological matrices with identical sample preparation and GC-ICP-MS measurement conditions was validated using the CRMs listed opposite. Species interconversion, particularly of EtHg into HgII, was observed and found to be dependent on the amount of Na tetrapropylborate added for derivatisation | 122 |
Hg | Blood | ICP-MS | A UK study in 4044 pregnant women showed no significant association of maternal blood Hg concentration and anthropometric variables or gestational age at delivery. A negative association with birthweight was observed in non-fish eaters | 188 |
Hg | Umbilical cord blood | ICP-MS | Mean total Hg, MeHg and iHg concentrations were 1.40, 0.95 and 0.13 μg L−1 respectively in umbilical cord blood from 263 pregnant women. Increased concentrations of MeHg were associated with higher pulse blood pressure while a similar increase of iHg was associated with lower systolic and pulse blood pressure | 165 |
Hg | Urine (NIST SRM 3668, levels 1 and 2) | FI-CV-AAS | Online use of a Fe3+ catalyst in an FI system increased the vaporisation efficiency of organic Hg and eliminated sample pre-treatment. Optimisation enabled Hg2+ to be used as the calibration standard for both organic and iHg. An LOD of 0.14 μg L−1 was reported. Recoveries from CRMs and Hg2+/MeHg-spiked urine (n = 101) were between 93.4 and 107.7% | 86 |
Hg | Urine | AFS | The Hg in urine was extracted by SPE via active nano-Au coated Si material prior to thermal desorption and detection by AFS. From spiked urine, recoveries were 96.13 ± 5.34%. Results from occupationally exposed individuals were in good agreement with those obtained by conventional CV-AAS and CV-AFS. The LOD was 0.004 μg L−1 and RSDs were ≤4.2% (n = 6) | 36 |
Hg | Bone (human proximal femur) | CF-AFS | In 96 patients undergoing total hip replacement, mean Hg content of the femoral neck and head were 37.1 ± 35.0 and 24.2 ± 35.0 ng g−1 respectively. An association was described with body mass index, differences in body anatomy and gender | 189 |
Hg | Human teeth | fs-LIBS | Diffusion of Hg and other elements in the tissue surrounding amalgam tooth restorations was found to be more extensive in permanent teeth (n = 8) than in deciduous teeth (n = 7). Mapping of element distribution was achieved with a spatial resolution <100 μm | 84 |
In | Urine, blood, air | ICP-MS | Samples from 494 Chinese smelting plant workers were analysed for In. Results were indicative of In exposure and there was a positive correlation between personal air and both blood and urine In concentrations | 190 |
Mo | Hip fluid simulant | ICP-MS | See Co, ref. 137 | 137 |
Os | Human ovarian cancer cells | SR-XRF | Mapping of an Os-containing chemotherapy drug candidate in ovarian cancer cells using an SR-XRF nanoprobe | 97 |
Os | Mouse organs, cell culture | ICP-MS | An Os-containing anticancer drug was quantified in organs of drug-treated mice by ICP-MS. Sample preparation involved addition of a stabilising solution (ascorbic acid, thiourea and EDTA) and avoidance of oxidising conditions. The LOD was 0.02 μg kg−1. In spiked liver tissue, the RSD was 4% and recovery was 92%. When applied to cellular accumulation studies in vitro, cells were lysed with a non-oxidising buffer before addition of the stabilising solution | 191 |
Pb | Blood | GF-AAS | See Cd, ref. 147. The 95th percentile for blood Pb in residents of Athens was 77 μg L−1 | 147 |
Pb | Soft tissue and bone phantoms | SR-XRF, portable XRF | See As, ref. 93 | 93 |
Pb | Bone | XRF | A study demonstrating the feasibility of high-energy X-ray tube based XRF with a Cd Zn telluride detector for metal exposure assessment, achieving an LOD for Pb of 6.91 μg g−1 bone | 94 |
Pb | Human teeth | SR-μXRF | Determination of the micro-distribution of Pb within teeth demonstrated that in incisors, Pb was primarily deposited in the secondary dentine region close to the pulp, co-localised with Zn, while in molars, Pb accumulated in the pulp and was co-localised with Ca | 164 |
Pt | Plasma ultrafiltrate | ICP-MS | A method for measuring the chemotherapy drug, carboplatin, in plasma ultrafiltrate, prepared using Amicon Ultra 30000 Da cut-off filters, followed by dilution in NH3 EDTA. Bias was <6% and 12% at the LOQ. Precision CV’s were <3%. The assay was applied to 10 clinical samples | 166 |
Pt | Human lung carcinoma cells | HPLC-SF-ICP-MS | Quantification of carboplatin–DNA adducts was achieved using a species-specific double ID strategy with micro-flow HPLC RP separation coupled to SF-ICP-MS. The LOD was 0.2 ng Pt mg−1 DNA. Following validation using reference carboplatin–DNA spiked cell samples, the methodology was applied to cultured lung carcinoma cells exposed to carboplatin | 167 |
Pu | Urine | LC-ICP-MS | Following addition of a 239Pu internal standard to acidified urine and microwave digestion, 242Pu was separated by extraction chromatography and detected by ICP-MS. A 30-fold improvement in sensitivity was achieved compared with a direct “dilute and shoot” method | 39 |
Ru | Artificial cytosolic environment | CE-ICP-MS | Study of intravenous anticancer drugs, indazolium Ru–transport protein adducts, in an intracellular model containing glutathione (1–10 mM), ascorbic acid (10 mM), and citric acid (50 mM), demonstrated no transformation of the Ru compounds | 192 |
Ru | Mouse blood cells, plasma and solid organs | ICP-MS, CE-ICP-MS, SEC-IC-ICP-MS | Distribution of a soluble Ru complex following intravenous administration in mice at different time points was assessed using separation techniques, CE and SEC-IC, followed by ICP-MS | 116 |
S | Snake venom | ICP-QQQ-MS, ESI-QTOF-MS | Use of a hybrid elemental and molecular MS configuration to identify and quantify intact S-containing snake venom proteins via32S/34S ID analysis | 64 |
Se | Blood | CSDF-ME-GF-AAS | Continuous sample drop flow microextraction (CSDF-ME) coupled to GF-AAS was reported for speciation of Se in blood. The method involved extraction of SeIV-ammonium 1-pyrrolidine dithiocarbamate complex by passing aqueous sample (5 mL) through carbon tetrachloride (32 μL) before injection into the graphite tube. Following optimisation, the LOD was 0.02 μg L−1, between batch RSD (n = 7) was 4.2% and good enrichment and enhancement factors of 106 and 91 respectively were achieved. Total iSe was estimated after reduction of SeVI with gentle boiling in 5 M HCl and subtraction of SeIV from total Se concentration | 123 |
Ti | Simulated body fluid, artificial saliva | ICP-MS | See Al, ref. 142 | 142 |
Ti | Blood, serum and hip fluid | ICP-OES | Validation of a method in clinical samples for Ti by ICP-OES using emission lines 336.1 and 371.0 nm and Y internal standard. The LOD and LOQ were 0.6 and 1.9 μg L−1 respectively. Bias was 3.3% and −1.0% for whole blood and serum respectively while intermediate precision was 10.9% and 8.6%. Higher Ti concentrations were measured in hip fluid samples (mean 5.1 μg L−1) compared with whole blood and serum (median 2.4 μg L−1) | 75 |
Ti | Serum | SF-ICP-MS | A prospective longitudinal study of serum Ti concentrations in 30 patients implanted with Ti ceramic composite cervical disc replacements. The LOD for Ti was 0.2 ng mL−1 | 141 |
Ti | Human jawbones | ICP-OES, LA-ICP-MS, SEM-EDX | The Ti content of jawbones with dental implants (n = 7) assessed by ICP-OES was found to be significantly higher (mean 1940 versus 634 μg kg−1) than controls. Particles (0.5 to 40 μm) were identified in bone marrow by SEM-EDX | 128 |
Ti | Rat shinbones | SR-XRF | Diffusion of Ti in Ti-implanted shinbones of 6 adult rats was demonstrated by SR-XRF | 139 |
Ti | Rat organs | ICP-MS | Report of bio-distribution of Ti in healthy young adult and elderly rats exposed via inhalation to 10 mg m−3 of a TiO2 NP aerosol daily over 4 weeks. Immediately after exposure, high concentrations of Ti were found in the lungs but this was followed by translocation to extra-pulmonary organs (spleen and liver), particularly in elderly rats | 98 |
Ti | Rat blood, tissues and excreta | ICP-MS | Measurement of Ti concentrations in rats following inhalation exposure to 15 mg m−3 of TiO2 NPs over 6 h demonstrated initial accumulation in the lungs before translocation to the systemic circulation. Higher Ti concentrations were found in faeces versus urine | 99 |
V | Urine | ICP-MS | Median concentrations of V measured in 816 Chinese maternal urine samples were higher in those with low birthweight babies compared with controls (3.04 vs. 1.93 μg g−1 creatinine). The odds of low birthweight were found to be increased 2.23 fold in mothers with urinary V ≥2.91 μg g−1 creatinine | 193 |
V | Simulated body fluid, artificial saliva | ICP-MS | See Al, ref. 142 | 142 |
V | Human heart tissue | ICP-MS | See Co, ref. 143. The reference range (mean ± SD) for heart V was 0.0094 ± 0.0211 μg g−1 | 143 |
Y | Cancer cells | ICP-MS | Development of a multifunctional probe to allow specific binding to cancer cells with subsequent dual imaging (down-conversion fluorescence and up-conversion luminescence) and quantification by ICP-MS via89Y detection. An LOD of 100 HepG2 cells and an RSD of 7.1% at 1000 HepG2 cells (n = 7) was reported | 57 |
Zn | Serum and plasma | GF-AAS | Chloride interference and intolerance to high pyrolysis temperatures in a GF-AAS method for Zn was overcome by employing two matrix modifiers (Mg/Pa mixture combined with 1% (w/v) aqueous EDTA) and a high pyrolysis temperature of 1000 °C. Accuracy of the method was verified by direct comparison with ICP-MS and FAAS for analysis of an RM and analytical recovery studies | 69 |
Zn | Anterior pituitary secretory granules | XRF microscopy, ICP-MS | See Ca, ref. 135 | 135 |
Zn | Human bone and enamel | LC-MC-ICP-MS | See Cu, ref. 186 | 186 |
Zn | CRMs; fish muscle, bovine muscle, pig kidney and human hair, serum and urine | MC-ICP-MS | Analysis of six biological CRMs to determine their stable Zn isotope ratios using the double-spike technique and high precision MC-ICP-MS (RSDs 0.04 to 0.13%) | 25 |
Various (19) | Urine | ICP-MS | A study of the seasonal variation in urine metal concentrations involving 7 participants. Twelve elements were found to have significantly higher concentrations in the summer months | 148 |
Various (9) | Urine | ICP-MS, LC-AFS | A study to determine reference values for first morning spot urines for Al, As, Ba, Ce, Mn, Sc, Ti, V and Y in 210 Chinese children, aged 2 to 12 years. The creatinine adjusted metal concentrations were significantly correlated with those in 24 h collections | 31 |
Various (14) | Blood and urine | ICP-MS | A survey of metal concentrations in a cross section of 2000 individuals from a former heavily industrialised area of Northern France. Concentrations of most metals in urine were higher than those reported in the French National Survey for the same period although blood Pb levels were lower | 30 |
Various (14) | Serum | SF-ICP-MS | A reference range study of 1400 pregnant women selected at random from the China Nutrition and Health Survey 2010 to 2012 | 29 |
Various (13) | Serum | TXRF | A study to determine serum concentrations of elements in 105 healthy volunteers from the Swietokrzyskie region in Poland by TXRF | 194 |
Various (10) | Blood, serum and plasma | PIXE-PIGE | Techniques PIXE and PIGE were applied in external (in-air) microbeam configuration to determine concentrations of elements Br, Ca, Cl, Cu, Fe, K, Na, P, S and Zn in whole blood, serum and plasma. Very small volumes (10 μL) could be analysed directly with no sample preparation | 96 |
Various (6) | Serum | ID-ICP-MS, ICP-OES, AAS, IC, ISE | Human serum CRMs for Ca, Cl, K, Li, Mg and Na at three concentration levels were developed and certified by a number of analytical methods. The expanded uncertainties ranged from 2.2 to 3.9% | 26 |
Various (4) | Serum | ICP-MS | A sandwich-type immunoassay for serum neuron specific enolase, alpha-fetoprotein and C-reactive protein consisted of a Cs-doped magnetic multi-core NP for magnetic extraction and ratiometric ICP-MS measurement and three metal (Au, Cd, Y)/dye-doped SiNP probes for ICP-MS multiplex detection. The dye moieties also allowed confocal laser scanning microscopy imaging of the probes. No sample pre-treatment was required. The LODs varied from 0.35 to 77 ng mL−1 and recoveries from 83 to 125% | 56 |
Various (10) | CSF | ICP-MS, ICP-AES | Concentrations of Ca, K, Mg, Na, P, Sr and Ti in CSF of patients with chronic neuropathic pain were found to be significantly higher than controls but the use of spinal cord stimulation had no effect | 145 |
Various (4) | Saliva | TXRF | Measurement of salivary Cr, Cu, Fe and Ni demonstrated significantly higher Cr and Ni concentrations in patients with metal (n = 30) compared with aesthetic (n = 30) fixed orthodontic appliances but this did not reach significance versus controls (n = 30) | 144 |
Various (4) | Liver | μXRF, SR-μXRF, XANES | Elemental distribution maps of Cu, Fe, Mn and Zn with resolution at the cellular level were obtained from a Wilson’s disease liver biopsy specimen by SR-μXRF (beam size 4 μm). Oxidation states of Cu were identified by XANES | 195 |
Various (4) | Bone and mucosal tissues | SR-XRF | A prospective pilot study investigating the incidence of elements (Ca, Fe, P, Ti) in bone and mucosal tissues around dental Ti implants with signs of peri-implantitis (n = 12). Analysis by SR-XRF revealed presence of Ti and an associated occurrence of Fe in 75% of samples | 140 |
Various (4) | Breast cancer tissues | LA-ICP-MS | Application of LA-ICP-MS to assessment of Ca, Cu, Fe and Zn in breast cancer tissues found significantly higher concentrations in tumour areas compared with normal tissue and the distribution of elements was observed to be heterogeneous | 127 |
Various (8) | Porcine nerve and fat tissues | LIBS | Study of elemental composition (C, Ca, H, K, Mg, N, Na and O) of fat and nerve tissue using LIBS with an Nd: YAG laser in open air and under normal stray light conditions | 196 |
Various (9) | Xylem and embryo sac liquid of green pea (Pisum sativum) | HILIC-ICP-MS/ESI-FTMS, SEC-ICP-MS/ESI-FTMS | A general analytical methodology for large scale identification and quantification of metals, Ca, Co, Cu, Fe, Mg, Mn, Mo, Ni and Zn, and their complexes, in pea fluids involving two chromatographic techniques in conjunction with elemental and molecular MS detection | 59 |
Various (5) | Plants; aquatic plant (BCR-670) | ICP-MS | Various digestion procedures for plant tissues were compared (dry-ashing, dry-ashing including a HF step and wet digestion using H2O2 and HNO3) to enable measurement of rare earth elements Eu, Gd, Ho, La, Lu. Accuracy of the method was verified using the CRM opposite, RSDs varied from 1 to 4% and LODs were between 0.02 (Lu) and 13 ng L−1 (La) | 34 |
Various (6) | Unicellular green algae (Chlamydomonas reinhardtii) | TEM, TEM-EDX, nanoSIMS | Chemical imaging by nanoSIMS of Ca, Cu, Fe, Mg, P, and Zn, employing a recently developed rf plasma oxygen primary ion source, achieved lateral resolution <50 nm. Parallel TEM of the same cell allowed relation of the ultrastructure to the spatial distribution of metals | 136 |
Element | Matrix | Technique(s) | Sample preparation/comments | Ref. |
---|---|---|---|---|
Ag | Chicken (muscle, organs) | AAS; ICP-MS; SEM-EDX | An in vivo investigation of the distribution of AgNPs in hens following oral administration. After identification of target organs by AAS, single particle ICP-MS and SEM-EDX were applied to livers and yolks to confirm the presence of AgNPs | 51 |
As | Cereals | XRF; ETAAS | Matrix-matched calibration standards were prepared by incubating 10 g of white rice grains with 100 mL MeOH containing DMAV in the range 2–100 μg As. After drying at 150 °C, 5.0 g of each calibration standard were sealed into polyethylene cups. The calibration curve was linear over the concentration range 0.21–5.00 mg kg−1 As (r = 0.999). The LOD was 0.080 mg kg−1. The results obtained on samples of grain cereals (6) and flour (2) agreed with those obtained by ETAAS after acid decomposition | 23 |
As | CRM | INAA; LC-ICP-MS | An As-specific mass balance method was established and applied to the certification of an AsB standard solution CRM, metrologically traceable to the IS. This can be used for calibration in quantitative analysis of AsB, e.g. in fish products | 20 |
As | CRM (apple juice) | ID-ICP-MS | An apple juice CRM was developed and certified for total As, As species (iAs and DMA), Cd and Pb | 21 |
As | Food RMs | FI-HG-AAS | 0.5–1.0 g samples were digested with conc. HCl, followed by 2 mL HBr and 1 mL of hydrazine sulphate. iAs species were extracted in CHCl3 and back-extracted into 10 mL HCl 1 mol L−1, prior to analysis by FI-HG-AAS. The LOQ was 0.010 mg kg−1. In a collaborative method study, in the concentration range 0.074–7.55 mg kg−1, repeatability (RSDr%) and reproducibility (RSDR%) values ranged from 4.1% to 10.3% and from 6.1 to 22.8%, respectively | 22 |
As | Cucumber (Cucumis sativus L. cv Joker) | XANES | AsIII and AsV were detected in cucumber samples cooled using a liquid N2 flow to maintain species integrity | 92 |
As | Diet | HPLC-ICP-MS | The dietary exposure of the Italian population to iAs was investigated in the national Total Diet Study (2012–2014) | 197 |
As | Drinking water | ETAAS | Using DLLME, the AsIII complex formed with 1,2,6-hexanetriol trithioglycolate at neutral pH was separated from AsV. Total As (following the reduction of AsV to AsIII by L-cysteine) and AsIII were determined by ETAAS and AsV by difference. With an enhancement factor of 87, the LOD (AsIII) was 0.03 μg L−1, linearity was up to 3.0 μg L−1 and RSD% was 5.9%. Recovery from fortified samples ranged from 92.0% to 113.3% | 119 |
As | Food and water, CRMs | HG-AAS | Using pressure-induced ionic liquid-based DLLME, the DDC–As complex formed at pH 4 was extracted into 1-hexyl-3-methylimidazolium bis(trifluoromethylsul-fonyl)-imide as the ionic liquid. With an enhancement factor of 87.5, an LOD of 0.016 μg L−1, LOQ of 0.055 μg L−1, linearity range up to 1.25 μg L−1 and RSD% of 4.89% were reported | 43 |
As | Fruit juices (apple and orange) | ICP-QQQ-MS | As was determined in fruit juices, after microwave-assisted acid digestion in closed vessels with 2 mol L−1 HNO3–H2O2, by means of ICP-QQQ-MS. To overcome the spectral interference of 40Ar35Cl+, determinations were carried out at mass 91 (75As16O+) after cell reaction with O2. The LOD was 0.013 μg L−1 and recovery ranged from 88% to 109% | 61 |
As | Marine fish oil | HPLC-ICP-MS; ESI-MS | Five new arsenolipids were identified in raw marine fish oil, mainly from Peruvian anchoveta (Engraulis ringens), using RP-HPLC-ICP-MS and ESI-MS | 63 |
As | Rice | HPLC-ICP-MS | 4 As species (AsIII, AsV, MMAIII and DMAV) and 3 Hg (HgII, MeHg and EtHg) were determined simultaneously, within a 20 min run, by ion-pairing RP-HPLC-ICP-MS | 118 |
As | Rice | ICP-MS; HPLC-ICP-MS | The concentration of total As was determined by ICP-MS and 5 As species (AsIII, AsV, MMAIII and DMAV) by HPLC-ICP-MS, using extraction with 5 mmol L−1 malonic acid and an isocratic mobile phase | 198 |
As | Rice, canned tuna, water, CRMs | ETAAS | A novel sorbent, based on a magnetic nanoparticle composite, was synthesised and applied to the speciation of AsIII and AsV | 110 |
As | Rice, RMs | HPLC-ICP-MS; ICP-MS; ETAAS | Extraction of iAs from rice samples (0.2 g dry mass) was achieved in a microwave oven with 9 + 1 mL of H2O—30% v/v H2O2 at 200 °C and 25 Bar. It gave comparable results to extraction with 2% v/v HNO3 (y = 1.046 + 0.010, r2 = 0.997, p < 0.0001), but with simplified chromatograms | 174 |
As | Rice, water, SRM | HG-AFS | Sulfydryl-containing substances (L-cysteine and glutathione) were applied to modify the performance of the graphite electrode used for electrolytic HG-AFS. The modifier and applied current allowed the selective and quantitative conversion of AsIII or AsIII + AsV to AsH3, in the presence of MMA and DMA, as well as the determination of total As and DMA. In aqueous solutions, LODs (3 s) for AsIII, iAs and total As were 0.25 μg L−1, 0.22 μg L−1 and 0.10 μg L−1, respectively | 88 |
As | Rice, wine | ICP-QQQ-MS | AsIII was oxidised to AsV with H2O2 during sample preparation, followed by separation of the sum of iAs (as AsV) from other As species using a short, narrow bore, 5 μm chromatographic column. Determinations were carried out by ICP-QQQ-MS using O2 as the reaction gas | 62 |
As | Seafood | AFS | 5 As species (AB, AsIII, AsV, DMA and MMA, extracted from seafood samples using 2% HCOOH and ultrasonication, were separated in <8 min on a Hamilton PRP-X100 anion-exchange column using gradient elution with 33 mmol L−1 CH3COONH4 and 15 mmol L−1 Na2CO3—10 mL EtOH at pH 8.4. Recoveries from spiked samples ranged from 72.6% and 109% and RSD% was <3.6% | 199 |
As | Seafood (fish, mussels, shrimps, edible algae), rice-based products (Jasmine and Arborio rice, spaghetti, flour, crackers), CRMs | HPLC-ICP-MS | After extraction with a 1 + 4 MeOH–H2O solution, the separation of AsIII, AsV, MMAIII and DMAV was achieved in 7 min using an anion-exchange narrow-bore Nucleosil 100 SB column and 12 mmol L−1 NH4H2PO3 (pH 5.2) as the mobile phase, at a flow rate of 0.3 mL min−1. Instrumental LODs were in the range 0.3–0.4 ng As mL−1 and RSD% was <3% | 200 |
As | Seaweed (red laver, green laver, sea tangle, sea mustard, and hizikia), seafood (snow crab, red snow crab, octopus, octopus minor, and squid) | IC-ICP-AES | Seaweed and seafood samples were dried and ground to fine powder prior to extraction and determination of total As and As species (AsIII, AsV, AsB, DMA, MMA) by IC coupled with end-on-ICP-AES. Recoveries ranged from 98.2 to 99.1% | 201 |
As | Seaweeds (Dulse, Nori, Kombu, Wakame, Sea Spaghetti, canned seaweed), CRM | HPLC-ICP-MS | Water soluble As species were extracted from edible seaweeds within 7 min using pressurised hot water. The LOQs were in the range of 47 to 101 ng g−1 and repeatability RSDs% were <18% | 202 |
Ca | Powdered milk, solid dietary supplements | LIBS; ICP-AES | The results of the direct determination of Ca, K, and Mg by LIBS, using two calibration strategies, were compared with ICP-AES after acid digestion | 80 |
Cd | CRM (apple juice) | ID-ICP-MS | See As, ref. 21 | 21 |
Cd | Food (bread, rice and fruit juice) SRM (rice flour) | FAAS | Rice (1 g), fruit juice (200 ml) and bread (5 g) samples were digested by wet oxidation with conc. HNO3—conc. HClO4—30% H2O2. The residues were re-dissolved in 2 mL of 0.5 M HNO3, then made up to 50 mL with triple distilled water. After pre-concentration on a newly synthesised chelating resin, Cd and Zn were determined by FAAS with LODs and LOQs of 1.5 and 5.1 μg L−1 (Cd) and 1.2 and 4.1 μg L−1 (Zn), respectively. The resin’s capacity was determined as 53.96 mg g−1 (Cd) and 24.19 mg g−1 (Zn) at pH 8.0. The analysis of a SRM gave results in agreement with the certified values and recovery from spiked samples ranged from 96.1% to 100.3% (Cd) and from 97.67% to 105.7% (Zn) | 203 |
Cd | Food and water CRMs | GD-AES | A method based on atmospheric pressure GD-AES with a miniaturised liquid drop anode was developed and applied to the direct determination of Cd in 50 μL samples. The LOD ranged between 0.20 and 0.40 μg L−1, linearity was up to 1000 μg L−1 and RSD% was between 2 and 5%. Recovery for CRM values and spiked water samples ranged from 95.2% and 99.6% | 71 |
Cd | Herbs | ETAAS | Pre-concentration of Cd from herb samples was achieved using solidified floating organic drop micro-extraction with 1-(2-pyridylazo)-2-naphthol (PAN) as the chelating agent, with an RSD% (n = 6) of 2.67% at 1.0 μg L−1. Recovery ranged from 94.5% to 110.2% | 46 |
Cd | Soft drinks | ETAAS | The extraction and pre-concentration of sub-ppt levels of Cd and Pb from soft drinks was achieved by DLLME, with 500 μL of (CH3)2CO (Cd) or 700 μL of CH3CN (Pb) as dispersive solvents, 60 μL of CCl4 as extraction solvent for both analytes and 500 μL of complexing agent (1.5% DDTC). Recovery ranged from 91% to 113% for Cd and from 95% to 108% for Pb. RSD% were <10% (Cd) and <7% (Pb). LODs were 0.006 and 0.072 ng L−1 for Cd and Pb, respectively | 204 |
Ce | Drinking water, surface water | ICP-MS | ZnO and CeO2NPs were detected in surface and drinking water using spICP-MS, with a dwell time of 0.1 ms or lower. Conventional drinking water treatments were found to be effective in removing most (from 60% to >95%) of these NPs | 53 |
Ce | Edible plants (cucumber, tomato, soybean, and pumpkin) | ICP-MS | The size and size distribution of CeO2NPs, as well as the concentration of Ce and CeO2NPs, were determined simultaneously, using spICP-MS, after enzymatic digestion of plant shoots with macerozyme R-10 | 205 |
Cl | Milk, CRM (waste water) | HR-CS-ETAAS | Cl was determined indirectly by measuring the molecular absorption of SrCl at 635.862 nm, with a Zr-coated GF and 25 μg Sr, with an LOD of 1.76 μg mL−1. Pyrolysis and molecule-forming temperatures were 600 °C and 2300 °C, respectively | 206 |
Co | Vitamin B12 | ICP-AES | The metrological traceable quantitation of cyanocobalamin (vitamin B12) was achieved via measurements of Co and P, carried out by ICP-AES and IC. The expanded uncertainty (95% CL) ranged from 0.3 to 1% | 156 |
Cr | Cereal products, dairy products | IC-ICP-MS | 3 mL milk or 0.3 g of solid samples were extracted with NH4OH in an ultrasonic bath. After centrifugation and filtration of the supernatant, samples were analysed by IC-HPLC-ICP-MS, using IonPac CG5A and IonPac CS5A columns. Gradient elution was achieved using 2 mmol L−1 HNO3 and ultrapure water as the mobile phase. The LOQs ranged from 0.6 μg kg−1 (dairy products) to 0.8 μg kg−1 (cereal products). The relative bias observed on spiked samples ranged from −3% to 14% over the concentration range 1–90 μg kg−1. CrVI was not detected, even in the presence of measurable total Cr values | 175 |
Cr | Drinking water, SRMs | ETAAS; IC-ICP-MS | By adding in situ 2-thenoyltrifluoracetone, the CrIII complex vaporised at 400 °C, whereas CrVI was released at 1200 °C, during ETAAS. CrIII concentration was obtained as the difference between total Cr and CrVI concentrations in the sample. LODs were 0.046 μg L−1 (CrVI) and 0.039 μg L−1 (total Cr), with RSDs% (n = 5) of 3.1% (1.0 μg L−1 CrVI) and 2.8% (1.0 μg L−1 Cr). A comparison was carried out with IC-ICP-MS | 120 |
Cr | Milk and baby foods, SRM | ETAAS | Slurries were prepared by 10 min sonication with ultrapure water, whereas liquid samples were only diluted, prior to analysis by ETAAS, with 500 μg Rh as the permanent modifier. An LOD of 1.43 μg L−1 was achieved, linearity was up to 50.0 μg L−1 and RSD% (n = 8) ranged from 1.5% to 11.8% | 207 |
Cu | Beef | LIBS; ETAAS | Cu measurements by LIBS, coupled with modelling techniques, were assessed, in combination with ETAAS determinations as a reference, as a tool to detect offal adulteration in beef | 78 |
Cu | Distilled alcoholic beverages (palinka) | HR-CS-ETAAS; XRF | After direct injection, followed by a fast heating program, Cu was determined with an LOD of 0.03 mg L−1. The linearity was up to 10–15 mg L−1 and average recovery was 100.9 ± 8.5%. A comparison with TXRF was carried out | 66 |
Cu | Food samples (wheat flour, corn flour, rice flour and tahini), SRMs | FAAS | For the determination of Cu, Fe and Zn in food samples by FAAS, microwave-aided digestion with a HNO3–H2O2 mixture provided the best results over other sample preparation methods, including wet digestion and other reagents (HCl, HNO3). Analysis of CRMs gave results with percent relative errors from −2.5% to +3.0%. LODs and LOQs were 14.5 and 48.3 μg L−1for Cu, 16.4 and 54.6 μg L−1 for Fe, and 10.8 and 36.0 μg L−1 for Zn | 208 |
Cu | Infant formula (rice, oats, bovine milk and soybean based) | HR-CS-ETAAS | After direct solid sampling, pyrolysis, 1300 °C, was applied for both analytes, whereas atomisation was carried out at 2300 °C and 2100 °C for Cu and Mn, respectively. For Cu (sample mass 3.0 mg), LOD and LOQ were 0.003 μg g−1 and 0.01 μg g−1, respectively, whereas, for Mn (sample mass 1 mg) LOD and LOQ were 0.01 μg g−1 and 0.04 μg g−1. Analysis of a CRM (rice flour) showed no significant differences between measured and certified values | 67 |
Eu (as label) | Seafood | CE-ICP-MS | Diethylenetriamine-N,N,N′,N′′, N′′-pentaacetic acid was coupled with saxitonin, a paralysing shellfish toxin, while simultaneously forming a complex with Eu3+, to provide a label for the indirect detection of saxitonin in seafood. An absolute LOD of only 0.38 fmol (using 100 nL sample injection) and an RSD% (n = 5) <7% were achieved | 58 |
F | Food CRMs (14), branded tea samples (13) | ICP-QQQ-MS | A mass shift strategy was applied to the determination of F in food and tea samples by ICP-QQQ-MS. F was converted to BaF+ and selectively extracted into the reaction cell with other species at m/z 157, then reacted with 10% NH3—90% He (8.0 mL min−1), and determined at m/z 208 | 60 |
Fe | Chicken | LIBS; LIF | LIBS and LIF measurements were applied to detect spoilage in chicken meat over 3 days after slaughter, using Fe and Mg as markers | 79 |
Fe | Food samples (wheat flour, corn flour, rice flour and tahini), SRMs | FAAS | See Cu, ref. 208 | 208 |
Fe (histamine) | Fish, dairy products, alcoholic beverages | FAAS | Histamine was determined indirectly by FAAS, with an LOD of 0.25 μg L−1. Sample preparation consisted of formation of a complex of histamine with FeIII (25 mmol L−1) and 2′,4′,5′,7′-tetrabromofluorescein (eosin, 5 mmol L−1) at pH 4.5, followed by extraction, using ultrasonic assisted CPE, into the micellar phase of polyethyleneglycoldodecylether (Brij 35, 10 mmol L−1) | 70 |
Hg | Environmental water, wastewater, tap water, fish, CRM (DORM-2) | HPLC-ICP-MS | A novel procedure for the speciation of Hg2+, MeHg+ and PhHg+ was developed using SPE with Fe3O4@SiO2@gamma-mercaptopropyltrimethoxysilane magnetic NPs. Species separation was achieved in 8 min. LODs for Hg2+, MeHg+ and PhHg+ were in the range of 0.49–0.74 ng L−1. Intra- and inter-day RSDs% (n = 5) were <9.0% and 12%, respectively | 109 |
Hg | Fish (common carp: muscle, gill, intestine, liver, gallbladder, brain, eye), CRMs | CV-AFS | Hg species in 1 mg sample were extracted into 100 μL eluent by MWCNT assisted matrix solid-phase dispersion and transported to AFS after single-drop solution electrode GD-induced CV generation | 35 |
Hg | Rice | HPLC-ICP-MS | See As, ref. 118 | 118 |
Hg | Seafood (muscle, liver, gonads and eggs), CRMs (hair, cod fish tissue, swordfish tissue) | CV-AAS | Samples (0.6–1.5 g) were digested with 1 mL 0.1% L-cysteine—1 mL 5 M NaOH at 80 °C for 1–2 h, then made up to 5 mL with distilled water. After extraction of fat with 6 mL IBMK and washing of the remaining solution with C6H14, determination of total Hg was carried out, whereas, for the measurement of MeHg concentrations, further extraction into 6 mL C6H5CH3 was necessary using 2 mL of the aqueous extract, 2 mL of 5 M HBr and 0.5 mL of 2 M CuCl2. The organic layer was then back-extracted into a 0.2% L-cysteine—2% NaCH3COO− solution and analysed. Analysis of CRMs, recovery tests and a comparison with GC-ECD (MeHg) confirmed the reliability of the procedure | 87 |
Hg | Seafood, CRMs (12) | VG-CCP-AES | Samples were subjected to a treatment with 47% HBr, followed by extraction of MeHg+ in C6H5CH3 and back-extraction in 1% L-cysteine. Hg determination was achieved, after FI UV photo-reduction of MeHg+ in 0.6 mol L−1 HCOOH, by VG-CCP-AES. The LOD and LOQ were 2 μg kg−1 and 6 μg kg−1, respectively, and recovery was 99 ± 8% | 74 |
I | Drinking water | ICP-MS | The content of total organic I in water was determined by ICP-MS after adsorption of the sample (pH < 1) on granular activated carbon, combustion, and trapping of the combustion products using a 2% (v/v) TMAH solution. In between samples, washing of the ICP-MS with 0.1% (v/v) TMAH was necessary to remove any carry-over | 176 |
I | Fish, CRM | ICP-MS | A novel sample preparation, consisting of ashing fish samples with KOH and extracting I in water from the ash, was evaluated. The LOQ was 0.20 mg kg−1, repeatability and intermediate precision RSDs% were 7.7% and recovery ranged from 85 ± 9% to 96 ± 7% | 177 |
I | Food and dietary supplements | ICP-MS Colorimetry | As part of the process to build a national database of I content in food and food supplements, two US bodies established the comparability of results obtained by either ICP-MS or a standard colorimetric method, thus allowing the pooling of data obtained independently by the two organisations | 171 |
K | Powdered milk, solid dietary supplements | LIBS; ICP-AES | See Ca, ref. 80 | 80 |
Lanthanides | Milk | ICP-MS | The distribution of lanthanides was identified as a fingerprint for traceability purposes in the milk production chain | 181 |
Mg | Chicken | LIBS; LIF | See Fe, ref. 79 | 79 |
Mg | Powdered milk, solid dietary supplements | LIBS; ICP-AES | See Ca, ref. 80 | 80 |
Mn | Fish muscle, oyster, bovine liver, CRMs | HR-CS ETAAS | Three elements (Mn, Ni and Rb) were determined in food samples by solid sampling and HR-CS-ETAAS. LODs were 0.005 μg g−1 (Mn), 0.002 μg g−1 (Ni) and 0.1 μg g−1 f(Rb) | 68 |
Mn | Infant formula (rice, oats, bovine milk and soybean based) | HR-CS-ETAAS | See Cu, ref. 67 | 67 |
Ni | Fish muscle, oyster, bovine liver, CRMs | HR-CS ETAAS | See Mn, ref. 68 | 68 |
Pb | CRM (apple juice) | ID-ICP-MS | See As, ref. 21 | 21 |
Pb | Food (grains, vegetables, seafood), CRMs | ETV-AFS; ICP-MS | The concentration of Pb in mg food samples was determined in less than 6 min, by solid sampling and ETV-AFS. The LOD was 3.0 pg and LOQ was 10.0 pg. The RSD% observed on CRMs was <5.0% and recovery ranged from 90.0% to 110.0%. A comparison with microwave-aided digestion and ICP-MS showed no significant difference | 90 |
Pb | Game meat | ICP-MS | spICP-MS was applied to detect PbNPs (40 to 750 nm) in game meat, from animals shot with Pb-containing bullets | 54 |
Pb | Soft drinks | ETAAS | See Cd, ref. 204 | 204 |
Rb | Fish muscle, oyster, bovine liver, CRMs | HR-CS ETAAS | See Mn, ref. 68 | 68 |
REEs | Plant samples, CRM | ICP-MS | Dry-ashing was considered adequate for plant sample digestion, in comparison with dry-ashing with a HF step or wet digestion with H2O2–HNO3. LODs between 0.02 ng L−1 (Lu) and 13 ng L−1 (La) were achieved. RSD% ranged from 1% to 4% | 34 |
REEs (10) | Indian basmati rice, soil, water | ICP-QMS; MC-ICP-MS | The concentrations of ten REEs (Ce, Dy, Er, Eu, Gd, La, Nd, Pr, Sm, Yb) were determined by Q-ICP-MS in Indian basmati rice and local soil samples, as potential markers of geographical authenticity. In addition, the 87Sr:86Sr isotopic ratio was studied by MC-ICP-MS, in the same samples and in samples of the water used for irrigation, after extraction on a Sr-specific resin. The 87Sr:86Sr isotopic ratio could discriminate between Indian Basmati rice and other rice samples, except Australian and Thailand rice | 180 |
S | Edible salts | LIBS; LA-ICP-AES; LA-ICP-MS; ICP-AES | Different analytical techniques were compared for the determination of S in edible salts from different geographical origins. LIBS and LA-ICP-AES showed the best accuracy and LIBS could identify S chemical forms (S2− or SO42−) associated with origin and quality of edible salt | 209 |
Se | Dietary supplements | XRF | XRF was applied for the rapid screening of total Se for label verification, whereas the selenoamino acid fraction was quantified using HR-MS/MS. This approach required <40 min per sample | 91 |
Se | Fish tissue (muscle, liver), CRMs | HPLC-ICP-MS | Several sample preparation methods were evaluated, using CRMs, for the determination of SeCys and/or SeMet in marine fish tissues. After enzymatic hydrolysis, SeMet was measured by anion or cation exchange chromatography coupled with ICP-MS. After a complexation step for SeCys, followed by enzymatic hydrolysis and SEC, SeCys and SeMet were determined by means of C8 ion-pair RP-HPLC-ICP-MS. Recoveries for spiked samples were 84% ± 9% (SeCys) and 97% ± 9% (SeMet), respectively | 40 |
Se | Food samples (including cereals, edible seaweeds, mushrooms, soya and tea leaf), CRMs | HG-AFS | 0.5–1 g of finely ground food samples were acid digested with 15 mL of 3 + 1 HNO3–HClO4, on a hot plate, followed by addition of 5 mL of 6 mol L−1 HCl to reduce SeVI to SeIV, and dilution to 50 mL. A double CPE was performed, first, using 5% (m/v) Triton X-114—1 g L−1 APDC, to extract the SeIV-APDC complex into the surfactant rich phase, then adding a mixture of 16% (v/v) HCl—20% (v/v) H2O2 to convert SeIV-APDC into free SeIV and back extract it into the aqueous phase. Recovery ranged from 95.6% to 105.2% | 210 |
Se | Vegetables (garlic, onion, leek, broccoli and cauliflower) | ETAAS | After extraction of SeIV and SeVI from freeze-dried and powdered samples with 0.1 mol L−1 HCl, one portion of the extract was treated with 37% HCl to reduce SeVI to SeIV prior to analysis. Another portion of the extract was further pre-concentrated by LLME of the SeIV-APDC complex with 50% (w/v) trihexyl(tetradecyl)phosphonium decanoate in CHCl3 as the ionic liquid. Recovery from samples spiked with either SeIV or SeVI ranged from 94% to 102% | 44 |
Sn | Edible vegetable oil | HPLC-ICP-MS | 4 organotin compounds were determined simultaneously in edible vegetable oil samples, after freezing with MeOH and purification by matrix solid-phase dispersion with graphitised carbon black. LODs and LOQs ranged from 0.28 to 0.59 μg L−1 and from 0.93 to 1.8 μg L−1, respectively | 49 |
Sr | Indian Basmati rice, soil, water | ICP-QMS; MC-ICP-MS | See REEs, ref. 180 | 180 |
Tl | Water, spinach, CRMs | ETAAS | Tl pre-concentration was achieved with DLLME, by forming a complex of Tl3+ with Br− and cetylpyridinium chloride, followed by extraction into a mixture of 1.5 mL MeOH—150 μL CHCl3. The LOD was 0.03 μg L−1, linearity was up to 2 μg L−1 (r = 0.9983) and RSD% at 1 μg L−1 TlIII was 4.1% (n = 6) | 45 |
Zn | CRMs | MC-ICP-MS | 6 biological CRMs (fish muscle, bovine muscle, pig kidney, human hair, human blood serum and human urine) with certified Zn concentrations were characterised for their 66Zn:64Zn isotope ratio | 25 |
Zn | Drinking and surface water | ICP-MS | See Ce, ref. 53 | 53 |
Zn | Food (bread, rice and fruit juice) SRM (rice flour) | FAAS | See Cd, ref. 203 | 203 |
Zn | Food samples (wheat flour, corn flour, rice flour and tahini), CRMs | FAAS | See Cu, ref. 208 | 208 |
Zn | Infant formula | SEC-ICP-MS | SEC-ICP-MS was applied to investigate the availability of Zn from infant formula and the results compared with those obtained by in vitro methods (solubility and dialysability) | 117 |
Various | Red wine samples | LIBS | The combined application of LIBS and neural networks to 38 red wine samples allowed classification of them according to protected designation of origin. A new sample pre-treatment with a dry collagen gel avoided working with liquid samples | 81 |
Various | Vegetables (Daucus carota, Spinacia oleracea, Cynara scolymus, Raphanus staivus, Coriandrum sativum) | EDXRF; XRF | EDXRF and μXRF were applied to investigate the presence and distribution of major and trace elements in edible vegetables from soils irrigated with municipal treated wastewater | 211 |
Various (4) | Single rice seed | ICP-MS | The simultaneous determination of As, Cd, Cr and Pb in single rice grains was achieved using matrix solid phase dispersion, with MWCNT as the dispersing agent, and ICP-MS detection. Comparison with microwave-aided digestion (n = 30) did not show significant differences. LODs were 5.0 ng g−1 (As), 0.6 ng g−1 (Cd), 10 ng g−1 (Cr) and 2.1 ng g−1 (Pb), respectively. RSDs% (n = 6) were <7.7% | 47 |
Various (5) | Canned vegetables | Various | The outcome of a proficiency test (IMEP-118) for the determination of total As, iAs, Cd, Hg, Pb and Sn mass fractions in peas in brine was reported, addressing critical issues related to harmonised compliance assessment for such samples | 28 |
Various (5) | Tap water, mineral water, Yellow River water and waste water | GD-AES | A liquid cathode GD-AE spectrometer was constructed and applied to the simultaneous determination of Ca, K, Mg, Na and Zn. The LOD for Ca was 0.19 mg L−1 whereas lower values (from 0.02 to 0.05 mg L−1) were obtained for K, Na, Mg and Zn | 72 |
Various (6) | Insects (grasshopper, cricket, buffalo worms, mealworms), Sirloin beef | ICP-AES | The availability of Fe and other elements (Ca, Cu, Mg, Mn and Zn) from 4 commonly eaten insect species were compared to that from Sirloin beef using in vitro methods and ICP-AES | 179 |
Various (6) | Milk, SRM | FAAS; SEM | Cd, Cu, CrIII, Mn, Ni and Pb were quantitatively sorbed on Aspergillus niger loaded on TiO2NPs at pH 6.0 and flow rate of 3.0 mL min−1. Elution was obtained with 2.0 mL of 0.5 mol L−1 HNO3, with a pre-concentration factor of 600. LODs (3 s) ranged from 0.08 μg L−1 (Mn) to 0.61 μg L−1 (Pb) and RSDs were <3% | 111 |
Various (7) | Drinking water | XRF | The on-site determination of Cd, Cu, Fe, Hg, Mn, Pb and Zn in drinking water was achieved by filtering a 50 mL sample, adjusted to pH 5 with acetate buffer, through an iminodiacetate chelating disk, which was then coated with adhesive cellophane tape, dried at 100 °C for 60 s and examined by XRF. The LODs were 7.6 μg L−1 (Cd), 4.0 μg L−1 (Cu), 9.4 μg L−1 (Fe), 4.4 μg L−1 (Hg), 6.6 μg L−1 (Mn), 3.2 μg L−1 (Pb) and 3.6 μg L−1 (Zn). Recoveries from spiked samples ranged from 89.0% to 110% | 212 |
Various (10) | Ethiopian green coffee (Coffea arabica L.) | ICP-AES | The content of Ba, Ca, Cu, Fe, K, Mg, Mn, P, S and Si was determined in 49 coffee bean samples, representing varieties from different Ethiopian regions. Statistical analysis of the results (PCA and linear discriminant analysis) provided models for origin traceability | 182 |
Various (12) | Edible vegetables, CRMs | XRF; ICP-MS | A sample preparation procedure for TXRF, consisting of suspension of 20 mg of powdered sample in 1 mL deionised water containing 10 μg Ga as internal standard, followed by 5 min sonication in an ultrasonic bath, was proposed and compared with microwave-aided acid digestion and acid digestion in a tube block heater. No statistically significant differences were observed | 213 |
Various (13) | Bee pollen | ICP-AES | 13 elements (Ba, Ca, Cd, Cr, Cu, Fe, Li, Mn, Mo, P, Pb, V and Zn) were determined in 5 bee pollen samples, dried at 42 °C for 20 h. Prior to analysis, samples were treated with 20% HNO3 for 24 h, to exclude any contaminated material, then wet acid digested with a 65% HNO3–H2O2 mixture at 150 °C for 5 days and the digests made up to 10 mL with deionised H2O | 214 |
Various (13) | Juice and milk | ICP-AES | The performance of a high-pressure microwave-assisted flow digestion system was evaluated. Juice samples were digested with 3.7 mol L−1 HNO3—0.3 mol L−1 HCl, whereas for partially skimmed and whole milk only conc. HNO3 was used. No significant differences were observed with the results from other conventional, batch digestion systems | 41 |
Various (20) | Chitosan (food grade) | ICP-MS | Chitosan was dissolved in 3% citric acid—HCl for the determination of 20 trace elements by DRC-ICP-MS using Bi, In, Sc and Y as internal standards and matrix matching for calibration. Instrumental LODs ranged from 0.6 to 19.0 ng L−1. Recovery from spiked samples varied from 89.50% to 109.00% and RSD% from 1.17% to 4.05% | 215 |
Various (23) | Dietary supplements | ICP-MS | The content of 23 elements in 67 food supplements based on algae, garlic, yeast, fish and krill oils was determined and the maximum daily intake calculated. No risk to healthy adults was revealed | 216 |
Chronic exposure to As, through the diet, is a potential health risk for the general population, however As toxicity depends on the chemical structure of the As species, with iAs being the most toxic. In food of marine origin, relatively high levels of As can be observed, but mostly as AB, a nearly non-toxic As species. The speciation of the five most common As species observed in food (AsIII, AsV, AB, DMA, MMA) generally requires complex analytical procedures, for which calibration standards traceable to the SI are only available as iAs solutions. Owing to the non-equivalent response of As compounds during speciation procedures, specific standard solutions for individual As species are necessary for calibration, to assure the metrological traceability, and hence comparability, of the measurement results. To this aim, Lee and co-workers20 at the Korean Metrology Institute, prepared and characterised a CRM for AB, starting from AB bromide (98% purity). The total As content in the solution was determined by INAA, recognised as a primary method for the measurement of elemental As. Trace impurities of other As compounds were quantified by HPLC-ICP-MS. To verify the complete separation of AB from other As species, the content of AB was further determined by the same chromatographic technique using a different column (Zorbax 300 SCX) and 4 mmol L−1 pyridine buffer at pH 2.3, as the mobile phase. The content of As as AB was determined by an As-specific mass balance method. This involved calculating the difference between the total mass fraction of As, determined by INAA, and the sum of the impurities, as mass fractions of As, determined by HPLC-ICP-MS. The uncertainty budget, including the contributions of all analytical steps, was calculated according to the “Guide to the expression of uncertainty in measurement” (JCGM 100). The content of AB, as As mass fraction, was certified as 51.28 mg kg−1 with an expanded uncertainty of ±1.17 mg kg−1 (k = 2, approximately at 95% CL). This value was verified successfully by submitting the CRM to an international comparison among National Metrology Institutes (CCQM-K97 “Measurement of arsenobetaine standard solution and arsenobetaine content in marine fish.”). The report is available at http://iopscience.iop.org/article/10.1088/0026-1394/54/1A/08003.
Besides being calibrators, matrix specific CRMs are also important to assess the overall method performance during validation and to monitor the quality of results over time. Shin and co-workers21 developed an apple juice CRM, including As and As species (AsIII, AsV and DMA), Cd and Pb as the reference analytes, starting from a commercially available apple juice, containing negligible amounts of the target elements. The item preparation and assessment were carried out according to the requirements of ISO Guides 34 and 35. After filtration and addition of 1.2 g L−1 Na benzoate as a preservative, the material was spiked with known amounts of the analytes. After homogenisation, 100 mL aliquots were placed in 125 mL HDPE bottles, flushed with N2 for 15 s, sealed and stored frozen at −20 °C. Homogeneity and stability studies (as well as the assignment of the reference values) were carried out in a single laboratory by SF–ICP-MS (As), HPLC-ICP-MS (As species) and ID-ICP-MS (Cd, Pb). The method for the determination of As species was validated in-house using rice flour CRMs, owing to the lack of similar matrix CRMs. Combined relative standard uncertainties (uc%) were evaluated as 2.45% (Cd), 2.84% (Pb), 3.95% (As), 4.74% (iAs) and 4.30% (DMA), respectively. The uncertainty budgets indicated long term stability (4.5 months) as the major uncertainty contributor for As (3.88%), Cd (2.40%) and Pb (2.76%). For iAs and DMA, the uncertainty associated with the characterisation (4.07% and 3.66%) had the largest impact. The certified mass fractions (±expanded uncertainty, k = 2, 95% CL) were: 0.185 ± 0.015 mg kg−1 (As), 0.124 ± 0.012 mg kg−1 (iAs), 0.220 ± 0.011 mg kg−1 (Cd), 0.0601 ± 0.0052 mg kg−1 (DMA) and 0.245 ± 0.014 mg kg−1 (Pb). Unfortunately, only 54 units were prepared, but the information gained may be useful for future batch production.
Since iAs was identified as the most toxic As species in food and drinking water, the need has emerged for establishing harmonised analytical methods for iAs in various food commodities to be applied in assessment and pre-regulatory stages. de la Calle et al.22 reported the inter-laboratory validation, by means of a collaborative trial (IMEP-41), of an analytical method for the quantification of iAs in food by FI-HG-AAS. Sample pre-treatment consisted of releasing As species from the matrix by incubation with concentrated HCl, followed by extraction of iAs with CHCl3 and back-extraction into an acidic medium. Analytical performances were assessed according to the statistical protocol described in ISO 5725. The test items were seven RMs, covering a broad range of food matrices (cabbage, fish protein, mushrooms, mussels, rice, seaweed (hijiki) and wheat) and concentrations ranging from 0.074 to 7.55 mg kg−1. Assigned values for iAs mass fractions were either provided by the respective CRM producers (rice and seaweed), derived as consensus value from a proficiency test (mushrooms) or determined by expert laboratories with demonstrated measurement capabilities in this field of analysis. The repeatability RSD% ranged from 4.1% to 10.3%, while the reproducibility RSD% ranged from 6.1 to 22.8%. No significant bias could be identified for the matrices investigated.
Acceptable limits for the content of iAs in rice, recently established by CODEX and other international legislation, together with the large consumption of this food commodity, impose the need for rapid methods to determine compliance of rice batches with the stated limit values before they are distributed to markets. As XRF spectrometry is able to directly analyse solid samples the technique has been used to assess levels of Cd in rice grain samples. Although limited to the determination of total As content, XRF spectrometry also carries the potential for rapid screening for the As content in rice and several rice flour CRMs exist. However, in XRF spectrometry, responses are affected by differences between the sample matrices and/or physical form of the sample. With these caveats in mind and, following from their previous work on the preparation of rice grain calibration standards for the determination of Cd, Nakamura et al.23 applied the same procedure to the calibration of As in cereal grains by XRF spectrometry. A preliminary assessment of suitable As species for spiking, led to the choice of DMA as the best option, providing approximately 100% recovery, whereas AsIII was lost during the treatment. Aliquots of rice grains (10 g) were added to 100 mL MeOH containing DMA, to achieve final mass fractions of As between 0 and 5 mg kg−1. The solvent was then removed by heating on a hot plate at 150 °C for 1 h. After cooling at room temperature, 5.0 g aliquots of each calibration standard were packed into polyethylene cups and covered with a 6 μm thick polypropylene film. Calibration curves produced with these materials showed good linearity (r = 0.999). The results obtained by XRF in commercially available samples of brown and white rice grains and As spiked barley grains compared well with those obtained by ETAAS after acid digestion.
Uribe and co-workers24 reported the preparation and certification of a CRM for the content (ng g−1) of Al, As, Cd, Cr, Cu, Mn, Ni, Pb and Zn in drinking water, produced by the Peru National Institute of Quality (INACAL), with expanded relative uncertainties between 0.8% and 1.9%. The target mass fractions were set according to national maximum permissible limits. The assigned values were derived from both the gravimetric preparation and the subsequent analytical measurements by means of ETAAS or FAAS, using the bracketing calibration technique, in a single laboratory. The measurement capabilities of INACAL for these analytes in drinking water were confirmed by successful participation in a comparison with primary methods organised by the Inter-American Metrology System (SIM-QM-S2). The assigned values were monitored for 24 months using ETAAS or ICP-AES. The CRM production was carried out according to the requirements of the ISO Guides 34 and 35 and uncertainties were evaluated according to established international guidance (JCGM 100 “Guide to the expression of uncertainty in measurement” and Eurachem guides).
Isotopic ratios are increasingly investigated in biomedical research, together with elemental concentrations, to highlight small variations in the metabolic patterns of essential and toxic elements, although the cost, time and skill required for such measurements may hamper their wider application. To aid such research, Moore and co-workers25 undertook high precision measurements of the 66Zn:64Zn ratio in several biological CRMs and RMs, of human and animal origin, such as human hair, serum and urine, fish muscle, bovine muscle and pig kidney, using MC-ICP-MS with an Aridus II sample introduction system and the double spike technique. A complex multi-step sample preparation procedure was necessary, consisting of microwave-assisted digestion with a 3:2 mixture of 15.4 mol L−1 HNO3–H2O2. This was followed by evaporation of the digests to dryness and re-solubilisation with 2 mol L−1 HCl to achieve approximate concentrations of 1 μg mL−1 Zn. Aliquots of these diluted solutions, containing either ∼250 or ∼800 ng Zn, were mixed with known amounts of the tracer, highly enriched with 64Zn and 67Zn, to yield a molar ratio of tracer to natural Zn of 1 ± 0.05. The resulting solutions were dried, re-dissolved in 6 mol L−1 HCl, left at 130 °C overnight to equilibrate, then dried again and re-dissolved in 1 mol L−1 HCL prior to undergoing anion exchange chromatography to separate Zn from other matrix components. The eluate (0.01 mol L−1 HCl) was dried, then treated twice with 50 μL of 15.4 mol L−1 HNO3, followed by evaporation to dryness, to substitute Cl− with NO3− ions. The solutions presented for instrumental analysis were prepared in 0.1 mol L−1 HNO3. The relative differences (‰) between the 66Zn:64Zn ratios, in the RM samples relative to the certified isotopic standard IRMM-3702 indicated limited variability for five of the tested RMs (−0.30 ± 0.10‰ for ERM-BB184 bovine liver; −0.34 ± 0.04‰ for ERM-BB422 fish muscle; −0.65 ± 0.13‰ for ERM-BB186 pig kidney; −0.31 ± 0.04‰ for ERM-DB001 human hair and −0.05 ± 0.04‰ for Seronorm Trace Elements Urine L−1). For the blood serum RM (BCR-639) a larger difference was observed (−3.29 ± 0.12‰) confirming previous literature data.
The measurement of electrolytes in human serum is a routine task in clinical laboratories, necessary for the diagnosis and monitoring of pathological conditions and therapeutic treatments. A number of CRMs exist to support these measurements, however, only few seem to simultaneously fulfil the requirements for range of analytes, levels of concentrations and physical status similar to those in patient samples. A new set of three serum CRMs was prepared by adding high purity chloride salts to pooled human serum and certified for the mass fractions of Ca, Cl, K, Li, Mg and Na, by the group of Feng et al.26 at the Chinese National Metrology Institute. The target concentration levels were within ±20% of the values observed in serum of healthy subjects. The certified values for Ca, K, Li and Mg were obtained by in-house primary methods, based on double ID-ICP-MS. To overcome polyatomic interferences affecting the accurate measurements of one or more isotopes of Ca, K and Mg, measurements were carried out using “cold” plasma conditions (RF power set at 750 W) and H2 as the collision gas, at a flow rate of 6.0 mL min−1. The assigned values for each of the two other elements were determined from the results of two independent reference methods (AAS and ICP-AES, for Na; combustion-hydrolysis IC and ICP-MS for Cl). An informative value was provided for the mass fraction of ionised Ca2+, obtained by ISE. A 14 month stability study carried out under storage conditions (−20 °C) did not show significant differences in the measured values for any of the elements or levels. The uncertainties of the values assigned to the CRMs were estimated by combining the contributions of the measurement step, between-bottle homogeneity and long-term stability, to give expanded uncertainties for the certified values ranging from 2.2% to 3.9%. Another Chinese group, led by Zhang,27 developed and validated, through analysis of CRMs and successful participation in international inter-laboratory comparisons, an ICP-MS method for the determination of Ca in serum. The procedure consists of gravimetric spiking of the samples with Al, as internal standard, followed by digestion with 69% HNO3, dilution of the digest and measurement of the 44Ca:27Al ratio using H2 as the collision gas at a flow rate of 4 mL min−1. The bracketing approach was applied for quantification. The RSD% was <0.27% and analysis of two NIST SRMs (4 concentrations levels) yielded bias values between −0.54% and 0.01%. The advantages of the proposed method were claimed as simplicity and low cost, in comparison with ID methods, which require highly enriched isotopes, expensive instrumentation, such as TIMS or SF-ICP-MS, and very skilled staff. In addition, the proposed method overcomes the potential bias problems associated with flame temperature, sample introduction and matrix effects in FAAS, achieved a better LOD than ICP-AES and the analytical step takes only 20 s in comparison with almost 10 min for analysis by IC.
Besides the accuracy of analytical measurements, an appropriate and harmonised choice of the “sample” to be analysed can be a critical issue for the comparability of measurement results and related judgements, e.g. compliance statements in regulated areas of testing. Most of the available PT items and CRMs represent homogenised matrices, either as lyophilised powders or pastes. The organisers of the proficiency test IMEP-118,28 for the determination of the As, iAs, Cd, Hg, Pb and Sn mass fractions in canned vegetables, addressed this issue by choosing to prepare a test item (peas in brine in a glass jar) identical to those available on the market and potentially subjected to official food controls. After preliminary studies of elemental uptake, acid cleaned glass jars were manually filled with ∼99 g of frozen green peas and 75 mL of a brine solution, spiked with the elements of interest (0.01 mol L−1 HCl, 0.3 mg L−1 AsV, 0.3 mg L−1 Cd, 0.2 mg L−1 Pb, 470 mg L−1 Sn and 6.9 g L−1 NaCl). Assigned values, for both the drained and whole material, were determined by five expert laboratories, selected on the basis of their experience and having accreditation for elemental analysis in food matrices. Both the total As and the iAs mass fractions were determined, using methods based on mild extraction procedures and HPLC-ICP-MS. Depending on the analyte and the choice of “sample” to be analysed between 74% and 98% of the participants achieved satisfactory analytical performances. Criticalities were identified: the potential binding of AsV to proteins, thus reducing its availability under mild extraction conditions, the limited number of participants (41 out of 123) reporting results for iAs and the lack of a harmonised approach on the choice of the “sample” (drained peas or whole sample) to be analysed.
Reference ranges for essential nutrients and for potentially toxic agents in biological fluids and tissues have been recommended for many years. However, several factors, such as ethnicity, residence, lifestyle habits, physiological conditions, age and sex, are known to influence such values and, therefore, additional, well designed studies, taking account of these variables are important. In this review year, three papers reported reference ranges addressing specific population groups or areas. The China Nutrition and Health Survey is a national cross-sectional study involving all areas of the country. As part of the 2010–2012 survey, Xiaoguang Yang et al.29 determined the concentrations of 14 trace elements (As, Cd, Co, Cr, Cu, Fe, Mn, Mo, Pb, Rb, Se, Sr, V and Zn) in serum samples collected from 1400, randomly chosen, pregnant women, aged between 18 and 44 years (mean ± SD: 27.0 ± 4.5 years). Measurements were carried out by SF-ICP-MS, after 1 + 19 dilution with 0.5% (v/v) HNO3. Data were analysed according to age, residence in urban or rural regions, anthropometric status (by the 4 categories: underweight, normal weight, overweight and obese) and duration of pregnancy (<12 weeks; 12 to <28 weeks; >28 weeks), using the Kruskal–Wallis test and linear model factorial analysis. The authors concluded that, for some elements, remarkable changes were observed, according to the factors considered, but did not address their relevance in comparison with established findings or potential exposure. For example, women resident in rural regions showed significantly (p < 0.05) lower serum concentrations of As, Cd, Se, Pb and V, in comparison with those living in urban areas, and higher levels of Mn, Mo and Sr. To our knowledge, the observed ranges (including variations associated with age or pregnancy) for Cu, Fe, Se and Zn were similar to those generally applied in hospitals in other countries. On the other hand, the concentrations (2.5th–97.5th percentiles) reported for Co (57.0–1130.0 ng L−1), Cr (219.8–4287.7 ng L−1) and Mn (1.2–8.4 μg L−1) were significantly higher than the currently accepted reference ranges. Given the notably higher levels of these elements in erythrocytes, sample contamination or significant exposure must be considered. Since Cd and Pb are usually determined in whole blood for the same reason, the observed values for these two elements may also need further investigation.
Another cross-sectional study (Nisse et al.30) examined the levels of 14 elements (Al, As, Be, Cd, Co, Cr, Hg, Mn, Ni, Pb, Sb, Tl, V and Zn) determined by ICP-MS in blood and urine samples from 2000 adult subjects representing the population of Northern France (IMEPOGE, 2008–2010). Age, sex and smoking habit were the determinants included in the study, whereas no exclusion was made for subjects who may have been occupationally exposed. Potential contamination from the test tubes used for the collection of blood samples was highlighted for Sb, leading to the exclusion of most of the data from further analysis. The survey results were compared with those obtained in several similar studies in other countries as well as in the French National Survey on Nutrition and Health, all carried out over approximately the same period. For most elements, the mean blood and urine levels were generally higher than those observed in other countries except for South Korea and China, owing to persisting environmental contamination from previous industrial activities.
Xiong et al.32 investigated possible trace element sample contamination from 13 different blood collection tube types available in China. The vials were tested by addition of water and 10% HNO3 and 20 elements determined by ICP-MS. With this protocol, all tube types would potentially cause the concentration of at least one element in the blood sample to be significantly increased above the reference interval. Most tubes would influence the results for several elements. It should be noted that the solutions used are not equivalent to human blood and that 10% HNO3 is an aggressive medium likely to extract more than might happen with blood. Nevertheless, experienced analysts will be aware that collection tubes can contaminate samples. Certain elements are always liable to be affected by impurities are sometimes unpredictable with variations from batch to batch and even from tube to tube within a batch.
As part of a project to develop a procedure to measure Pt in plasma, urine and tissues, Zhang et al.33 found that Pt concentrations in calibration solutions, stock solutions and processed samples were stable after three freeze–thaw cycles and for up to three months.
Carbon materials for solid phase extraction have long proved popular. A procedure to prepare limited quantities of material was developed by Chen et al.35 for determination of Hg. Milligram amounts of sample solution in a volume of 100 μL were extracted with multiwall-carbon tubes (MWCNT) and this single drop taken for solution electrode GD-CV-AFS. With a 1 mg sample the LOD was determined as 0.01 mg L−1. Advantages of the method included minimal sample dilution, low blank, high sample introduction efficiency, high sensitivity and minimal toxic chemicals.
Two publications refer to the use of gold nanoparticles for the extraction of Hg from biological samples. In the method developed by Schlathauer et al.36 Hg in urine was separated with a nanogold-coated silica material and determined by AFS. Recovery from Hg spiked real samples was 96.13 ± 6.34% and samples analysed using this procedure and by CV-AAS and CV-AFS gave results that were in very good agreement. The LOD for the method was reported as 0.004 μg L−1 and the RSD% was <4.2%. Treated samples showed no loss of Hg when stored at 30 °C for up to a week. Hsu et al.37 developed a similarly innovative approach to extract Hg2+ from CSF, dialysates and water. Nanoparticles with a gold core and iron oxide shell, functionalised with L-cysteine were used to adsorb the analyte ions prior to determination by ICP-MS. Validation of the method was demonstrated by analysis of a CRM and by recovery studies with results of 101.5–109.3%. Compared with conventional SPE this procedure was rapid and did not involve a desorption step.
In our recent Updates we have referred to the research work of Vanhaecke and colleagues directed at determination of Cu isotopes in serum. To achieve accurate results in a matrix in which elements such as Ca and Na are at much higher concentrations than the Cu, Lauwens et al.38 extracted the alkali and alkaline earth metals using Chelex-100 ion exchange resin, followed by further purification of the Cu-fraction with AG MP1 anion exchange resin. The Cu isotope ratios were then determined by MC-ICP-MS. This procedure afforded low blank values, associated with lower amounts of acid required, and improved precision compared to other methods. Analysis of samples collected at time intervals throughout the day showed no diurnal variation in the serum Cu isotopic ratios.
Occupational exposure to Pu is monitored by analysis of urine samples. Current methods for separations and analysis by alpha spectroscopy are lengthy and require long count times. To avoid this, Gallardo et al.39 proposed using ICP-MS after microwave heating digestion and separation of the Pu using an EichromTRU resin column. Urine samples were acidified and 239Pu added as an internal standard. The pre-concentration step provided a 30-fold improvement in sensitivity compared to a simple dilution approach.
Various modifications to liquid–liquid microextraction for heavy metals determination in food continue to be developed for use with less-sensitive spectrometric techniques. Magnetic ionic liquids (IL) were used by Wang et al.42 to facilitate the separation of a Se-containing complex formed with 2,3-diaminonaphthalene. A vertical shaker allowed mixing of the aqueous rice sample with the extractants and a magnet was employed to retain the IL (1-butyl-3-methylimidazolium tetrachloroferrate) at the bottom of the microtube for determination by ETAAS. An LOD of 0.016 μg L−1 was reported with a RSD of <3% for SeIV. The methodology was used for speciation of SeIV and SeVI in different rice samples. A pressure-induced IL, namely {1-hexyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide} extraction approach was used to extract the hydrophobic complex As-DDTC at pH 4.43 This increased the phase transfer ratio which significantly increased the recovery of the As–DDTC complex to give an enhancement factor of 87.5. Determination of As by HG-AAS gave an LOD of 0.016 μg L−1 and RSD of 4.89%. Another approach used the IL trihexyl(tetradecyl)phosphonium decanoate to extract a SeIV–APDC complex with a 90% efficiency and 100-fold pre-concentration factor.44 The LOD of the Se determination by ETAAS was 5 ng L−1 and the method was applied to the determination of Se in allium and brassica species. Javedani-Asleh et al.45 described a ligand-free extraction process where Br− was used to complex TlIII in the presence of cetylpyridinium chloride which was then extracted into MeOH–CHCl3. Analysis by ETAAS, achieved an LOD of 0.03 μg L−1 and an RSD of 4.1%. Good liquid mixing of a chelating agent and easier separation of the chelated analyte when cold solidified, was used to determine Cd in herb samples.46 After digestion of dried herbs with HNO3 and H2O2, 1-(2-pyridylazo)-2-naphthol (PAN) in the liquid phase was used to extract Cd. The solvent was undecan-1-ol whose melting point was quoted as 19 °C. Optimised extraction conditions were a PAN concentration of 5 mmol L−1, and an extraction time of 20 min at 65 °C with stirring at 1000 rpm. Buffer was then added to achieve pH 9. The LOD and LOQ using ETAAS were 0.0052 and 0.017 μg L−1, respectively; the RSD at 1 μg L−1 Cd was 2.67% with a pre-concentration factor of 22.
Multiwall carbon nanotubes (MWCNT) continue to be popular for SPE although they themselves may contain transition metal catalyst contaminants such as Co, Fe and Ni. However, He et al.47 combined MWCNT with matrix solid-phase dispersion (MSPD) techniques for the determination of As, Cd, Cr and Pb in single rice grain samples with ICP-MS. Equal masses (ca. 0.02 g) of rice and MWCNTs, previously washed with 1:1 HNO3:H2SO4, were ground to a paste then blended with 1% HNO3:HCl (3:1, v/v, 2 mL) and centrifuged. The supernatant was analysed directly by ICP-MS yielding LODs of 5.0, 0.6, 10 and 2.1 ng g−1, for As, Cd, Cr and Pb respectively, with RSDs <7.7%. Four CRMs were analysed with no significant difference (at 95% CL) between the certified and measured concentrations. The method compared well (R2 > 0.94) with standard microwave digestion determined across 30 samples while the new MWCNT-MSPD method allowed a 100-fold reduction in sample size. Other workers in China48 also used the same MWCNT-MSPD preparation techniques to facilitate the determination of Hg in limited size fish organ samples. The Hg species contained in 1 mg tissue were extracted into 100 μL of solvent (0.5% L-cysteine–4% HCOOH), and total Hg quantified by AFS following conversion to Hg0 using a single-drop discharge-induced CV-GD system. An LOD of 0.01 μg L−1 was achieved with RSDs of 5.2% and 4.6% at 2 and 20 μg L−1, respectively. Another application of the MSPD technique combined with graphitised carbon black, was employed by Liu et al.49 to clean up cold-precipitated extracts of vegetable oil–MeOH for the determination of organotin compounds. Filtered extracts were separated by C18 RP-HPLC coupled to ICP-MS allowing quantification of four Sn species (DBT, TBT, DPhT, TPhT) in various edible vegetable oils, with LODs ranging from 0.28 to 0.59 μg L−1 and LOQs of 0.93 to 1.8 μg L−1.
Two papers in this review period used NP probes in sandwich-type immunoassays with ICP-MS detection for biological analysis. Multiplex measurement of serum biomarkers, alpha-fetoprotein (AFP), neuron specific enolase (NSE) and C-reactive protein (CRP), was achieved via Cs-doped multicore magnetic NPs immobilised with polyclonal antibodies to extract the biomarkers from the sample and three different metal (Au, Cd, Y)/dye-doped SiNPs immobilised with monoclonal antibodies for detection by ICP-MS and visualisation by fluorescence imaging.56 The design had several advantages over previously reported work, including Cs within the magnetic NP core to allow ratiometric measurement in ICP-MS and incorporation of different metal tags whilst maintaining an NP probe size <90 nm, which is required for good extraction efficiency and reproducibility. Slightly superior LODs to those typically seen with automated immunoassay methodologies were achieved (0.37, 0.36 and 77 ng mL−1 for AFP, NSE and CRP respectively). Linearity was satisfactory up to 100 ng mL−1 for AFP/NSE and up to 10000 ng mL−1 for CRP. Recoveries from spiked samples were 83% to 125%. Although no prior sample preparation was required, assay run times were long (>4 h) compared with automated immunoassays. Furthermore, expansion to include more metal tags, while maintaining a small probe size, may be a challenge. A similar approach involved a novel multifunctional NP probe for both cell counting and multimodal imaging of cancer cells.57 The probe consisted of an antibody to bind anti-EPCAM antibody-labelled cells, a fluorescent dye moiety for down-conversion fluorescence and an up-conversion NP tag, incorporating 89Y, for luminescence and ICP-MS quantification. Under optimised conditions, excellent linearity (R2 of 0.9955 for 400 to 30000 HepG2 cells) and sensitivity was demonstrated with an LOD of 100 HepG2 cells and RSD of 7.1% at 1000 HepG2 cells. Simple metal tags may also be utilised to label and detect biomolecules and He et al.58 reported use of Eu3+ chelate labelling to determine trace concentrations of the paralytic toxin, saxitoxin, in mussels using CE-ICP-MS with a run time of 13 minutes. Mussel sample preparation involved vacuum freeze drying, pulverisation and extraction of saxitoxin into DMSO before Eu3+ labelling via the coupler, pentetic acid. The merits of the described method were strong anti-interference ability, good stability and very good sensitivity with an LOD of 3.8 nmol L−1 and RSDs <7%. Recoveries from spiked mussels were 93–100%.
Dual analysis, combining accurate quantification by ICP-MS with the identification capabilities of molecular MS, offers versatility for both biological and food analysis. The combined application of HILIC-ICP-MS/ESI-FT-MS and SEC-ICP-MS/ESI-FT-MS to large scale detection of >50 metal complexes in green pea plant fluids was reported by Flis et al.59 Data were collected using two chromatographic systems with different mobile phase conditions and in positive and negative ionisation modes to enable confirmation of the findings and elimination of minor artefacts. The column recovery of metal complexes was 60% to 90%. Interestingly for multi-elemental ICP-MS analysis, H2 reaction gas rather than He collision gas was employed. Several isotopes were monitored by ICP-MS to improve quantification accuracy and use of HR-mode ESI-FT-MS enabled isobaric interferences to be overcome. Inclusion of ICP-MS afforded increased sensitivity and obviated the need for individual species standards for quantification, facilitating identification of previously unreported species.
Last year saw the application of ICP-QQQMS to halide analysis in blood plasma and this year the technique has been used to determine F in tea.60 Due to its low IP, F was converted to BaF+ in the plasma before reaction with gas mixture 10:90 NH3:He at 8.0 mL min−1 to form BaF(NH3)3+, which was monitored at m/z 157 → 208. The mass shift strategy avoided Gd interference although the relevance of this in tea is questionable. The RF power, which affects the competition between BaO and BaF in the plasma, was optimised to 1500 W. An LOD of 0.022 μg mL−1 was achieved and good agreement with certified values for food SRMs was observed. In 13 branded tea samples, F concentrations (39.2 to 93.2 μg mL−1) agreed well with those obtained by ISE. Utilisation of ICP-QQQ-MS with O2 reaction gas for high sensitivity As analysis, thus avoiding interference on m/z 75, was this year extended to food and beverage matrices. Pinheiro et al.61 measured total As in fruit juices, preparing the samples by closed vessel microwave assisted digestion and achieving an LOD of 0.03 μg L−1 with recoveries ranging from 88% to 109%. Another group developed a fast As speciation method (<2 min) for DMA, MMA and iAs in wine and rice,62 which featured a short narrow bore 5 μm column and decreased mobile phase flow rate to reduce sample dilution and increase transport efficiency. Consideration was given to mitigating the well-known C-enhancement effect in the EtOH matrix which was achieved through a low sample injection volume (5 μL) and addition of 3% MeOH to the mobile phase. Prior oxidation of AsIII to AsV by H2O2 facilitated faster separation and improved S/N for AsV while pre-column injection of AB internal standard increased accuracy through drift correction. The LODs achieved were 0.022, 0.018 and 0.026 μg kg−1 for iAs, DMA and MMA respectively. Results were within 13% of the reference method target in wine and spiked recoveries were 93 to 123% in rice.
The combination of ICP-QQQ-MS and dual analysis was applied to the determination of arsenolipids adsorbed onto bentonite,63 a mineral used to reduce contaminants during production of commercial fish oil. A clean-up procedure to remove phospholipids, which cause ion suppression in ESI-MS, consisted of sequential solvent extraction before each fraction was run on a normal phase Si-gel column with an increasingly polar eluent. Quantification and identification of arsenolipids was accomplished by RP-HPLC followed by ICP-QQQ-MS against a DMAV standard and Ge internal standard, and HR-ESI-MS. Clean-up column recovery was 89 to 95% and total mass recovery was 88%, indicating minimal loss of As during the extraction, clean up and speciation steps. Five new arsenolipids were identified including three As-containing medium chain fatty acids, thought to be degradation species resulting from the clean-up procedure. In the more exotic matrix of snake venom, Calderón-Celis et al.64 described the μHPLC separation of intact S-containing proteins with quantification using ICP-QQQ-MS and 32S:34S ID and identification by ESI-QTOF-MS. The approach required only one 34S-containing standard (added continuously post-column) to quantify all cysteine- and methionine-containing proteins with known S:protein stoichiometry. Samples were diluted and centrifuged prior to separation by RP-μHPLC with gradient elution consisting of 0.2% aqueous formic acid and ACN. Addition of BOC-methionine internal standard, significantly reduced the complexity of the ID equations rendering correction of post-column flow and a mass discrimination factor unnecessary. Total protein recovery was 92.6% to 97.4%. The LODs achieved for ICP-QQQ-MS were 15 to 25 fmol. When applied to proteomic analysis of venom from three snakes, results were in fairly good agreement with previously reported work with some discrepancies for minor proteins.
Another report making use of IDMS techniques presented the development of a high precision candidate reference method for speciation and quantification of metalloprotein, Cu–Zn–superoxide dismutase (SOD1), via Cu detection in red cells, using species-specific double and triple IDMS with strong anion exchange HPLC-ICP-MS.65 The isotope enriched spike, prepared from a commercially available SOD1 calibration standard and 65Cu and 67Zn enriched standard solutions, was added to the sample prior to preparation and separation to compensate for any losses or changes during these processes. While double IDMS used a reference spike blend with a 1:1 65Cu:63Cu ratio and obviated the need to know the purity of the species specific spike used, the triple IDMS involved two reference spike blends with 65Cu:63Cu ratios of 0.9 and 1.1 and did not require the isotopic composition of the spike to be determined, although at the cost of longer measurement times. The calculated mass fractions of SOD1 were in good agreement for double and triple IDMS; 63.94 ± 0.93 μg g−1 and 64.02 ± 0.96 μg g−1 respectively. Excellent precision was demonstrated with an expanded uncertainty of ≤1.50% k = 2, 95% CL).
Last year, an improved procedure for the direct isotopic ratio analysis of Fe in whole blood was discussed. A report this year38 described the development of a Cu isolation procedure in serum, prior to MC-ICP-MS analysis, to reduce concentrations of matrix elements that have an effect on instrumental mass discrimination. Following acid digestion of serum, the new procedure consisted of removal of alkali and alkaline earth metals via 1 mL Chelex-100 chelating ion exchange resin before further purification of the Cu fraction using 250 μL AGMP1 strong anion exchange resin. This was compared to the conventional sample preparation. Advantages of the modified procedure were decreased volumes of high purity acid, cost and time required and lower blanks. Using standards of known isotopic composition and serum RMs, no statistically significant differences were found between δ65Cu values obtained by the proposed and conventional procedures. Use of the propriety jet interface system, rather than standard MC-ICP-MS interface cones led to superior precision for δ65Cu (0.01‰ to 0.03‰) and reduced effects of instrumental mass discrimination for aqueous standard solutions but a stronger influence of matrix elements on Cu isotope ratios was noted.
Those with personal experience of measuring Zn concentrations by ETAAS will be well aware of problems associated with contamination of equipment and reagents and interfering species. Concentrations in plasma are usually high enough to be determined by FAAS or other techniques but where the available sample volume is very small, such as from young babies, ETAAS may be required. The work by Stevens et al.69 utilised two chemical modifiers, a Mg/Pd mixture combined with 1% (w/v) EDTA, which allowed a pyrolysis temperature of 1000 °C for effective removal of the Cl− interference. Samples were and standards were simply diluted 20-fold using 0.05% (v/v) Antifoam B Emulsion and Triton-X-100 prior to measurement. Recovery studies, comparison with ICP-MS and FAAS methods and analysis of a commercial RM all confirmed the accuracy of results.
An original application of FAAS was reported by Altunay et al.70 who developed an indirect method to measure concentrations of histamine in fish, dairy products and alcoholic beverages. Histamine forms a complex with Fe(III) and 2′,4′,5′,7′-tetrabromofluorescein (eosin) at pH 4.5 which can be extracted into the micellar phase of polyethylene glycol dodecyl ether (Brij 35). Ultrasonic assisted CPE was used to separate and pre-concentrate the histamine from the sample matrices, followed by FAAS measurement of the associated Fe. The optimum reagent concentrations and other conditions were evaluated and the procedure achieved an LOD of 0.25 μg L−1.
Photochemical vapour generation for ICP-AES featured in a novel application in which a volatile Co species was generated from acid digested biological tissues. In this work de Jesus et al.73 took Co2+ solutions in a pH 3.3 medium of 50% formic acid for exposure to deep UV (254 and 185 nm) radiation generated within a 19 W flow-through low pressure mercury discharge lamp. A phase separation device allowed the analyte vapour to be transported to the ICP-AES system for detection at the 238.892 nm emission line. Investigations of the performance indicated that in a continuous mode, with an irradiation time of 10 s, the generation efficiency was 42% and the LOD was 0.4 μg L−1. Matrix interferences from some digested samples required longer irradiation times, additional dilution and standard additions calibration to give accurate results. A second application with photochemical vapour generation was reported by Covaci et al.74 Seafood samples were prepared using the procedure recommended in the JRC Technical Report of the European Commission for the determination of MeHg. The MeHg extracted from samples was directed to a flow injection UV photo-reduction system to generate Hg vapour for AE at the 253.652 nm line with a low power/Ar consumption plasma microtorch. The LOD and LOQ, 2 μg kg−1 and 6 μg kg−1, respectively and RSD of 2.8% meet the requirement of the European legislation for this application. Analysis of CRMs confirmed the trueness of results.
The concern around metal-on-metal hip replacements has been discussed in recent Updates with a focus on Co and Cr but other elements such as Ti are present in the prostheses and may be relevant to the problems experienced by some patients. Titanium cannot be easily determined by Q-ICP-MS due to interferences from Ca isotopes. Therefore, Harrington et al.75 developed and validated a simple method to measure concentrations of Ti in whole blood, serum and hip fluid by ICP-OES using the Ti 336.1 nm emission line and an Y internal standard. Increased levels were measured in samples from patients where Ti was a component in the femoral stem or total hip replacement prostheses.
Santos et al.76 reported an ETV-ICP-OES method said to give results for eight elements in powdered medicinal plants in less than 5 minutes, including sample preparation time. The optimised procedure used the rather unusual choice of Freon R12 (CCl2F2) as the modifier and temperatures of 300 and 1800 °C for pyrolysis and vaporisation, respectively. Standard additions calibration was necessary to compensate for residual matrix effects. The LODs for As, Cd, Cr, Cu, Mn, Ni, Pb and Zn were from 0.01 to 0.48 μg g−1. Results obtained from the analysis of two NIST CRMs were in good agreement with the certified values.
Developments with elemental imaging using MS and X-ray spectrometry are discussed in Section 6.2 but Thyssen et al.77 propose that LA-ICP-OES is an alternative technique for bio-imaging of the easily excited alkali and alkali earth elements.
A number of publications have highlighted LIBS as a useful system for rapid analysis in food and beverage products for a variety of purposes. Casado-Gavalda et al.78 described the application of LIBS to detect offal adulteration in beef. The concentration of Cu was used as an indicator because offal can contain over 100-times higher levels than in meat. Samples of minced beef and liver were dried, homogenised into powder before mixing together and pressed into pellets at different ratios to generate the in-house calibration standards. These were characterised by ETAAS and the values incorporated into the PLS regression calibration model using the LIBS spectral data. Two separate sample sets were employed to train the system, with a third independent set applied as the test, which achieved a high Rcv2 of 0.85 which was considered reasonable for LIBS. Furthermore, the spatial homogeneity of the pellets was assessed by taking 100 spots across the surface. Two additional sample pellets were produced containing sections with two different ratios of beef:liver. A visual inspection of the maps generated showed a fair level of homogeneity in the calibrants and some regional distinction in the multi-surface pellets, though no formal statistical analysis was presented. Overall, the method did demonstrate the potential to identify offal adulteration in meat. Chicken breast fillets were the subject of a study by Abdel-Salam and co-workers79 who investigated the feasibility of using LIBS and LIF as a rapid and cost-effective system to identify meat spoilage. Samples were analysed by LIBS for the cyanide and carbon molecular bands to monitor protein content and the ratio of ‘ionic to atomic’ spectral lines for Mg and Fe as an indicator of meat tenderness. The LIF data showed an inverse relationship between the fluorescence band intensity and the storage time. After comparing the results against a conventional meat analyser, the authors concluded that both LIBS and LIF were promising techniques to detect degradation in the meat processing industry. Augusto et al.80 utilised LIBS for the direct determination of Ca, K and Mg in milk powders and dietary supplements. The calibration strategies relied on additional measurements by ICP-OES after acid digestion for the PLS regression calibration. The method accuracy ranged between 60% to 168% for Ca, 77% to 152% for K and 76% to 131% for Mg. Considering the LIBS approach could achieve a throughput of one sample per minute with no preparation, the results were encouraging.
Typically, the direct analysis of samples with high moisture content or liquids/solutions by LIBS can suffer significant problems but a number of publications have attempted to solve this issue. Moncayo et al.81 developed an approach to classify red wines based on the ‘protected designation of origin’ region. To minimise the difficulties associated with aqueous samples, the group presented a novel and simple preparation procedure which converted the liquid wine into a solid by generating a dry collagen gel. With the use of neural networks, a model was designed and tested using genuine and ‘fake’ wines to successfully assign the correct categories. The method was simple, quick and could be used for authenticity purposes and quality control screening. Pasquini and Farias Filho82 described an instrumental development of LIBS which incorporated a ring-oven as another technique for liquid-to-solid conversion and as an online pre-concentration approach. In this work, further improvements were made to the ring-oven set-up and inclusion of an internal standard (Li) enabled correction for problems associated with repeatability and laser pulses missing the target. Once optimised, it was used to determine Cu concentrations in samples of cachaça (a Brazilian spirit derived from sugarcane juice). The LIBS method achieved an LOD and LOQ of 0.3 and 1.0 mg L−1 respectively. It was compared with AAS as the reference method and in general, the results obtained were in good agreement, given the reported RSDs, with the exception of one sample which was approximately 44% higher by LIBS. The authors performed further experiments which demonstrated changes in the slope of the calibration curve for the ring-oven/LIBS method in the presence of sugar. Overall, the study showed that LIBS with integrated sample preparation and pre-concentration had potential as a fast screening technique (6 minutes per sample). Hedwig et al.83 described the combination of ps-LIBS with a vacuum sample chamber pressurised at 2 kPa with He which was shown to significantly improve emission intensities in plant materials when compared against ambient air. With the optimised setup, slices of carrot and meatballs were spiked with Pb and B to assess the emission characteristics in samples with a water content of approximately 15–20%. The LOD was estimated at 100 ppb for Pb and 200 ppb B which was respectable given the direct analytical approach.
A biological application of fs-LIBS to detect Hg migration from amalgam in teeth was presented by Bello et al.84 who optimised the instrumental parameters to detect Hg plus Ag, Ca, Cu, Mg and Sn as markers of the amalgam, tooth and interface regions. Deciduous and permanent teeth were also compared and an assessment of elemental penetration depth was calculated. The fs-LIBS spectra for each tooth section showed expected behaviours where Ag, Cu, Hg and Sn were at the highest intensity in the amalgam which decreased towards the tooth whereas Ca and Mg on the other hand were only observed in the tooth and interface areas. A difference in Hg penetration depth was found between deciduous and permanent teeth (157 ± 7 μm versus 173 ± 6 μm respectively) which could be a factor of the residence time of the amalgam, however the age of the restorations was not known. Considering the regions analysed, the authors hypothesised that the diffusion mechanism was linked to liquid transport through the dentinal tubules.
Two other procedures for Hg vapour generation followed by AAS were reported. Sabouri and Nouroozi86 described an online method to directly determine iHg and MeHg in urine determined Hg in urine. An FI system was used which included an Fe3+ catalyst to ensure equal vaporisation efficiencies for MeHg and iHg, enabling calibration with iHg only. Concentrations of Hg in two CRMs were determined at 107.7% and 97.3% of the certified values and spikes of Hg2+ and MeHg were recovered in the range of 93.4% to 104.5%. A sample throughput of 108 h−1 was reported. Total Hg and MeHg in seafood were measured by Yoshimoto et al.87 using IBMK to remove fat, leaving the Hg in the aqueous phase. The MeHg was then extracted from an aliquot of the aqueous phase with HBr, CuCl2 and toluene, and back-extracted into an L-cysteine–sodium acetate solution. Heating-vaporisation AAS was used to detect and quantify the concentration of Hg. Analysis of CRMs demonstrated accuracy of the procedure. The approach was applied to determine the level and ratios of total Hg and MeHg in glands of squid and sea urchins. Most of the Hg in squid was MeHg whereas in sea urchin it represented less than 4% of the total concentration.
Atomic fluorescence spectrometry was used in a number of different applications. Continuing with the theme of Hg, the procedure to prepare small amounts of biological samples for determination by glow-discharge CV-AFS35 was described in Section 3.2.1. A non-chromatographic method to determine four As species by modified graphite electrode-based electrolytic HG-AFS was developed by Yang et al.88 The electrode was modified with either L-cysteine or glutathione and, depending on the applied current, AsIII, AsV, MMA and total As were selectively measured. Zeng et al.89 were able to generate volatile species of Mn through reduction of acidified analyte solutions in the presence of 0.1% APDC and 0.1% 1-butyl-3-methylimidazolium hexafluorophosphate with 2.0% KBH4 and 0.1% phenanthroline. Efficiency of vaporisation was calculated to be 44%. With optimised parameters the LOD was 1.5 ng mL−1. The method was successfully applied to the determination of Mn in water sample and environmental CRMs. Electrothermal vaporisation from a quartz furnace was introduced by Shang et al.90 to determine Pb in foods with detection by AFS. Temperatures for ashing and vaporisation were optimised to remove matrix interferences and the accuracy of the method was assessed by analysis of CRMs.
The past year has shown an overall reduction in the number of noteworthy X-ray spectrometry papers in comparison to the previous Update.1 However, several publications have highlighted the use of X-ray based techniques for food analysis. Hanley et al.91 reported a rapid screening method with XRF for total Se determination in dietary supplements. Both solid and liquid formulations were tested which required different calibrations to account for the matrix effect: serial dilutions of a standard Se solution for the liquid samples and the CRM SELM-1 was mixed gravimetrically with cellulose to produce the solid calibrants. Additionally, the authors developed an approach for Se speciation using direct analysis in real time high resolution accurate mass–mass spectrometry (DART-HRAM-MS). Both total and species data were compared against conventional ICP-MS and HPLC-ICP-MS. The XRF method had an LOD and LOQ of 4 μg g−1 and 30 μg g−1 respectively, which is more than adequate for supplement testing, with good agreement was observed against the ICP-MS data when a quadratic fit calibration equation was used for XRF. Similarly, the DART-HRAM-MS results produced comparable values to HPLC-ICP-MS for non-protein bound SeMet and MeSeCys, achieving an LOD of 100 μg Se species per capsule. An attempt was made to detect iSe but practical issues such as sensitivity and carry-over prevented this. However, given that no sample preparation was required and the fast measurement times (40 minutes for both methods), the combination of XRF and DART-HRAM-MS offered significantly increased throughput with suitable accuracy levels. Gerenyi and co-workers92 published an interesting application of XANES to investigate the uptake of As in cucumbers. Liquid nitrogen ‘steam’ was used to cryogenically preserve the vegetable which enabled the researchers to determine the As oxidation state in vivo to understand the metabolic pathways of As.
The use of X-ray techniques for clinical applications, particularly for in vivo bone measurements, was the focus of several papers. Groskopf et al.93 prepared bone and skin phantoms doped with As and Pb to evaluate SR-XRF against portable XRF. The thickness of the skin layer was also considered. Although the LODs using SR were better, the study demonstrated similar results for both techniques, adding evidence to support the use of portable XRF systems as in vivo measurement devices. Specht and co-workers94 described a theoretical model using Monte Carlo simulations to assess the feasibility of a portable XRF device with a high-energy X-ray tube and room temperature detector, for the measurement of Pb in bone. The results were compared with experimental data and good agreement was observed. Byrnes and Bush95 evaluated the ability of portable XRF for forensic bone analysis. Factors, such as bone depth, were investigated using equine bone of known thickness. Historic specimens were then tested before and after sanding the periosteal surface to determine any changes in the XRF response due to potential diagenesis. It was concluded that portable XRF is a convenient tool for onsite analysis but still has some limitations. Huszank et al.96 published a study utilising PIXE and PIGE for the direct determination of various elements (Br, Ca, Cl, Cu, Fe, K, Na, P, S and Zn) in liquid blood samples. The combination of techniques using the ‘in air’ microbeam configuration enabled the measurement of small sample volumes (∼10 μL) in serum, plasma and whole blood with no preparation. The authors suggested that the approach was suitable for routine use in the ppm range but added that levels closer to 1 ppm required longer measurement times with poorer accuracy.
A paper by Sanchez-Cano et al.97 described the use of a new nanoprobe beamline for SR-XRF to investigate the cellular distribution of a new osmium-based candidate cancer drug. Human ovarian carcinoma cells A2870 were exposed to 1 μM of the Os complex, purported to be 49 times more potent than cisplatin, for 24 h before fixation with paraformaldehyde. The SR-XRF beamline was 27 × 37 nm2, scanned at step sizes of either 50 × 50 nm2 or 20 × 20 nm2 and a dwell time of 50 ms. Detailed image maps were generated for Ca, Os, P and Zn which allowed the researchers to gain insight into the mechanism of action. Localisation of Os in the mitochondria effected reactive oxygen species which appeared to induce Ca release from the endoplasmic reticulum, a signalling event for cell death. The study highlighted important and crucial findings in the search for new cancer therapies.
Pulmonary exposure to NPs is of particular interest. Gate et al.98 measured Ti concentrations in young and elderly rats by ICP-MS, following exposure to 10 mg m−3 aerosols for 6 h day−1, 5 days week−1 for 4 weeks. Titanium in the lung slowly cleared post-exposure and was found in the spleen, and liver with higher concentrations in the elderly animals. No increase in Ti was detected in blood, kidney or brain. Pujalte et al.99 studied rats exposed to 15 mg m−3 aerosols, for just 6 h. Peak concentrations in the lung were found at 48 h and decreased over 14 days. Titanium was present in blood, lymph nodes, liver, kidney and spleen. Measurable amounts were also determined in olfactory organs and brain. Large amounts of Ti were present in faeces suggesting that inhaled NPs were eliminated mainly by mucociliary clearance. These two reports indicate that absorption and further metabolism is complex and further work is required to understand the implication of exposure to TiNPs. Cerium oxide NPs in diesel fuel additives can ultimately lead to human inhalation. The emitted NPs undergo an aging process by interaction with other airborne pollutants and this may influence the biodistribution after inhalation. Therefore, Li and colleagues100 investigated inhalation exposure of Sprague Dawley rats to combustion-generated CeO2NPs. Test samples were aged in a chamber representing ambient urban air conditions, with UV lights. Exposure was to freshly emitted or aged CeO2NPs for 15 min, 24 h, and 7 days followed by measurement of Ce in the blood, lungs, gastrointestinal tract, liver, spleen, kidneys, heart, brain, olfactory bulb, urine, and faeces. Cerium was found in the lungs and faeces, with extrapulmonary organs contributing less than 4% to the recovery rate at 24 h post exposure. No significant differences were found between fresh and aged CeO2NPs. The findings from all of these studies of Ce and Ti in olfactory organs and brain are similar to observations with Hg vapour and may indicate translocation via neural pathways.
Two reports refer to metabolism of NPs during pregnancy. Austin et al.101 examined the distribution of Ag in pregnant mice and embryos/foetuses following intravenous injections of 10 nm AgNPs or 66 mg Ag/mouse on gestation days 7, 8 and 9. Highest concentrations of Ag were found in maternal liver, spleen and visceral yolk sac. Levels in the yolk sac of the AgNO3 treated mice were twice those of the NP exposed animals but there were no significant differences in other tissues examined. One or more embryos from eight of the ten AgNP-treated mice were small for their age, an effect that was seen in embryos from only one mouse in the control group. Silver accumulation in embryos/foetuses was negligible. Muoth et al.102 determined the uptake of AuNPs into human placental co-culture microtissues in the context of NPs drug carriers. Uptake was found to be greater with smaller (3–4 nm) or sodium carboxylate-modified AuNPs than with larger (13–14 nm) or PEGylate AuNPs, for which there was minimal passage through the trophoblast barrier layer. The authors suggest that the effectiveness of AuNP-drug carriers can be optimised by tailoring their properties.
A number of investigations having relevance to intestinal absorption were reported in the last year. Gallocchio et al.51 administered AgNPs to hens and showed that Ag accumulated in liver and egg yolk. Single particle ICP-MS (spICP-MS) and SEM-EDX spectrometry results that 5–10% of the Ag in liver was as the NP form. In an in vitro experiment, Ramos et al.50 measured Ag after samples of chicken meat containing 40 nm AgNPs were treated with either saliva, gastric or intestinal fluids. Single particle ICP-MS indicated that 13% of the AgNPs would be available to the intestinal cells while up to 44% of the initial material would be in bioaccessible dissolved forms. Previous Updates have noted work that has shown transfer of metals from food contact materials, containers and cooking utensils into food products. Some of these studies refer specifically to AgNPs incorporated into containers etc. because of their antimicrobial properties. Williams et al.103 pointed out that AgNPs may, therefore become available for intestinal exposure with potential implications on health of consumers. These workers assessed the effects of AgNPs of different sizes and doses T84 human colonic epithelial cells. Uptake into cells was demonstrated by ICP-MS analysis and TEM, together with perturbation of parameters indicative of cell function. The smallest AgNPs (10 nm) had the greatest effects.
There is considerable interest in the pharmacological applications of nanomaterials. Escudero-Francos et al.104 used a variety of techniques, including ICP-MS to assess the in vitro uptake into fibroblasts and the in vivo biodistribution of AuNPs functionalised into nanoclusters with mercaptoundecanoic acid (MUA), GSH or BSA. The lack of toxicity and the preferential tissue uptake of MAU, GSH and BSA Au-nanoclusters into brain, kidney, liver and spleen, respectively point to potential organ-specific carriers of drugs. In a similar study, Mahmoud and colleagues105 assessed Au nanorods (AuNR) functionalised with neutral, anionic, cationic, and hydrophobic ligands. Using ICP-OES and LA-ICP-MS, accumulation into hair follicles and other skin compartments of human skin sheets was measured. The lipophilic properties of hair follicles enhanced the accumulation of hydrophobic polystyrene (PS)–AuNR into hair follicles while neutral polyethylene glycol (PEG)–AuNR were distributed into all skin compartments, especially the dermis. Charged AuNR showed a negligible penetration into any of the skin layers. These properties could be exploited for targeted skin disease treatment and transdermal administration of drugs. Auoxo3 is a cytotoxic AuIII compound which exhibits promising anticancer activity. When encapsulated in a ferritin nanocage there is significantly less damage to non-tumour cells.106
Despite the many positive features associated with nanomaterials, acute exposure studies with TiO2have shown that there can be harmful consequences. In the work reported by Armand et al.107 possible toxicity from long-term exposure to low concentrations of TiO2NPs was assessed using A549 alveolar epithelial cells. Genotoxicity was determined by alkaline and Fpg-modified comet assay, immunostaining of 53BP1 foci and the cytokinesis-blocked micronucleus assay and the impact of a subsequent exposure of these cells to the alkylating agent methyl methanesulfonate. Cytotoxicity and intracellular accumulation were determined using conventional tests and ICP-MS. The results showed that while long-term exposure to TiO2NPs does not affect cell viability, it does cause DNA damage, particularly oxidative disruption to DNA and increased 53BP1 foci counts. These results correlated with increased intracellular accumulation of NPs. Following intra-venous injections of Tween® 20-gold AuNP conjugates to mice, Berce et al.108 determined haematological parameters, Au concentrations in tissues and the appearance of bone marrow cells. The results indicated accumulation of Au in the bladder and bone marrow, proliferation of white blood cells and platelets and liver toxicity. From this experiment it was uncertain whether there was simply a reaction to the presence of Au or if a severe haematological condition could later develop. The authors warned that the potential use of AuNPs in cancer therapy must be evaluated against possible harm.
Nanomaterials feature prominently in new developments for analytical procedures. A review by Karadjova et al.14 summarised the different materials that have been used for elemental speciation. The authors considered size, surface to volume ratio and chemical reactivity and how the properties were exploited for the analysis of environmental, food and biological samples. The work of Zhu et al.109 using Fe3O4@SiO2@gamma-mercaptopropyltrimethoxysilane magnetic NPs to successfully speciate Hg in samples of water and fish is described in Section 6.1 Metallomics, below. In the same vein, Abbaszadeh and Tadjarodi110 developed what they termed a magnetic metal–organic framework for As speciatiation in water, rice and tuna samples.
To extract and concentrate iHg from samples of water and CSF, Hsu et al.37 also exploited magnetic NPs. The SPE procedure employed NPs with a Fe3O4 shell, Au core and an L-cysteine functional group. Unlike conventional SPE methods, desorption of Hg for measurement by ICP-MS, was not necessary. A novel sorbent was developed by Alpdogan et al.111 to measure Cd, Cr, Cu, Mn, Ni and Pb from milk, with an impressive preconcentration factor of 600. These workers used a column containing TiO2NPs, functionalised with Aspergillus niger. The selected elements were retained at pH 6.0, eluted with 0.5 mol L−1 HNO3, and determined by FAAS.
In recent years we have seen a small number of publications describing immunoassays in which the final analytical step is the measurement of a metal label by ICP-MS. Two publications report original analytical approaches in which NPs have a role. Ko and Lim56 synthesised three metal/dye-doped silica NPs as probes for simultaneous detection of the disease markers C-reactive protein, alpha-fetoprotein and neuron-specific enolase. This sandwich immunoassay also used Cs-doped multicore magnetic NPs for magnetic extraction of the analytes and quantification using ICP-MS. No sample treatment was necessary prior to the extraction and the sensitive method provided short analysis time and convenience for the determination of these biomarkers. Detection limits ranged from 0.35 ng mL−1 to 77 ng mL−1. An immunoassay for cancer cells, counting and imaging with the aid of a multifunctional probe, was described by Yang et al.57 The probe consisted of a recognition unit of goat anti-mouse IgG to label the anti-EpCAM antibody attached cells, a fluorescent dye (Cy3) moiety for fluorescence imaging and upconversion NPs tag for both ICP-MS quantification and imaging of cancer cells. Excellent linearity and sensitivity were achieved due to the signal amplification effect of the NPs.
Signal amplification due to inclusion of NPs in assays was also discussed in a review of laser desorption/ionisation coupled with MS, by Unnikrishnan et al.112 Nanomaterials, such as AuNPs, with high photoabsorption, photothermal conversion, and heat-transfer efficiency, reduce the interference of matrix molecules, and improve the reproducibility and sensitivity. Monitoring the ligand ions and Au cluster ions amplifies the mass spectral signals and allows analysis of macromolecules with minimum errors.
With the increasing application of NPs in a wide range of manufacturing and other sectors, analysis of NPs in clinical, foods and environmental samples is required. Two articles reviewed approaches to the analysis of nanomaterials. De la Calle et al.113 discussed the techniques for extraction and preconcentration from the sample matrix, separation of NPs from ionic species and larger particulates and size fractionation. Applications to biological tissues, food, environmental samples, cosmetics and other consumer products were described with discussion of the advantages and inconveniences of the procedures mentioned. In their review of analysis of nanomaterials, Montano et al.15 referred specifically to the technique of spICP-MS. The extensive discussion included advances in data acquisition, signal processing and different mass analysers. The key features for quantitative analysis are explained with examples of analysis of increasingly complex environmental and biological samples. A method with SEC and ICP-MS was described by Zhou and colleague114 to characterise particle size, mass and composition of trace NPs in serum and water. Complete elution from the column of different-sized NPs was achieved using a mobile phase with a relatively high ionic strength, 2% FL-70 and 2 mM Na2S2O3. The size distribution of NPs agreed closely with that obtained using TEM. Donovan et al.53 reported the measurement of Ce- and Zn-containing NPs in surface and drinking water by spICP-MS. Laboratory experiments with river water were conducted in which sequential treatment by lime softening, alum coagulation, powdered activated carbon sorption and disinfection by free chlorine, were applied. The results showed that most of the CeNPs were captured during lime softening and that ZnNPs were progressively removed by the additional treatments. Analysis of real source and drinking water samples showed at least 60%, and in most cases >95%, of NPs were removed by the treatment processes.
The metabolism of metal-containing anticancer agents is investigated in studies to determine modes of action, drug resistance and toxicity. Most notable is the work with Pt-agents but also other compounds, of As and Ru for example. Sodium trans[tetrachloridobis(1H-indazole)ruthenate(III)] is a new drug being trialled. Distribution and accumulation of Ru after administration of this drug to a normal mouse model was investigated by Byztek et al.116 After dosing, Ru concentrations in the blood rose, with plasma levels three times greater than those of the cells. Within the plasma there was initial binding to albumin and another high molecular weight species with the albumin:Ru ratio declining as the metal entered tissues and/or was excreted. Ruthenium was detected in a wide range of tissues, especially the thymus, colon, lung, liver and kidney.
Snake venom is a topic not previously discussed in these Updates. As part of a project to identify and quantify the toxic proteins, with a view towards producing therapeutic anti-venom agents Calderón-Celis and colleagues64 developed a protocol to separate venom from three species of snake by HPLC, with parallel ESI-QTOF-MS and ICP-QQQ-MS. It will be of interest to learn what this approach will reveal and how the results can be applied to the protection of individual who may be bitten by these reptiles.
Gomez et al.117 determined the distribution of Zn in infant milk formulas using a combination of solubility properties, SEC and ICP-MS. The work showed that Zn is present entirely within the protein fraction. When the protein fraction was further examined, the aqueous soluble protein fraction of milk formula included 90% of the zinc but only 7% and 24% were in the same fraction from soy and lactose-free formulas, respectively. Separation of samples by SEC indicated that, in all formulas tested, Zn is bound to compounds of around 10 kDa.
Most of the speciation work mentioned in these Updates refers to applications involving just a single element. Fang et al.,118 developed a procedure to determine four As and three Hg species in the same assay. Reverse-phase chromatography, with ICP-MS for detection, separated AsIII, AsV, MMA, DMA, iHg, EtHg and MeHg from rice. Samples (0.5 g) were extracted into 1% HNO3 with microwave heating and diluted to a volume of 25 mL. The LODs using these solutions were 0.84–2.41 and 0.01–0.04 μg L−1 for As and Hg species, respectively. The concentrations in a rice flour CRM and locally purchased rice were reported.
Much of the elemental speciation work we report refers to chromatographic separation of species. Two other approaches may also be mentioned. Zhang et al.119 used 1,2,6-hexanetriol trithioglycolate, at neutral pH, to form an AsIII hydrophobic complex, leaving the AsV in the aqueous phase. Orthoganol array design was adopted to optimise the liquid–liquid extraction and the analyte enhancement factor from drinking water was 87, giving an LOD of 0.03 μg L−1. Drinking water samples were also analysed by Cui et al.120 The work of these authors involved speciation of CrIII and CrVI with the measurements being made by ETAAS. A selective reaction between CrIII and 2-thenoyltrifluoracetone forms a complex that vaporises at 400 °C, while CrVI is stable up to a temperature of 1200 °C. Measurements are made of the total Cr and the CrVI concentrations with the difference reported as the CrIII. Two drinking water CRMs evidenced the accuracy of the method and concentrations in tap, bottled and well water samples were reported.
Arsenic metabolism is associated with methylation efficiency, which is known to be influenced by gender, nutrition and genetics. Löveborn et al.121 elected to investigate whether there are differences in methylation efficiency between children and adults. For this study, iAs, MMA and DMA were determined by ICP-MS in urine samples from 488 nine year old children. The percentage of DMA was higher and that of iAs was lower in samples from adults with the same exposure. Erythrocyte Se concentrations were also found to be positively associated with As methylation, indicating that a number of additional factors may be relevant to the metabolism of As.
Two interesting publications report on work involving speciation of Hg. In recent Updates magnetic solid phase extraction procedures to pre-concentrate analytes during sample preparation have been covered. Zhu et al.109 added to this experience by describing the preparation of Fe3O4@SiO2@gamma-mercaptopropyltrimethoxysilane magnetic nanoparticles which were then used to extract and concentrate Hg species from samples of water and fish. Rapid separation of iHg, MeHg and PhHg by HPLC-ICP-MS was effected using a mobile phase with a greater than usual concentration of MeOH. The LODs were calculated to be 0.49 to 0.74 ng L−1. Analysis of CRMs provided results to confirm the accuracy of the method. Many examples of methods to analyse blood, urine and hair samples have been published in abundance but Abad et al.122 have developed a common procedure which is applicable to the simultaneous determination of iHg, MeHg and EtHg, and inorganic mercury in all three sample types. The method requires 0.15 g blood, 0.5 g urine or 0.1 g hair and the same sample preparation and GC-ICP-MS conditions were employed. Results obtained from the analysis of NIST and IAEA CRMs were in agreement with the certified values. In addition, the authors reported the Hg species concentrations for SRM 955c (caprine blood) levels 2 and 4 for the first time. However, it was shown that conditions required to obtain acceptable values for caprine and human blood are not the same.
The complex metabolism of Se has been intensively investigated and aspects are still unclear as new intermediates continue to be identified. Akramipour et al.123 claim to have developed a new methodology to determine trace elements in biological fluids and apply this innovation to the speciation of iSe in blood samples from children. The method was described as continuous sample drop flow microextraction followed by ETAAS using an Ir-modified graphite furnace. A sample, 5 mL, was slowly passed through 32 μL CCl4 during which hydrophobic SeIV-APDC is formed. The organic phase was then taken for measurement of the Se concentration by ETAAS. The work reported by Kwak and colleagues124 addressed urinary Se metabolites. These authors proposed that volatile Se species may account for some of the unknown compounds that can be lost during analytical procedures. In the course of this study to discover potentially volatile pheromones in golden hamster urine, dimethyl selenenylsulfide, dimethyl diselenide, dimethyl bis(thio)selenide, and dimethyl selenodisulfide were detected. Whether these species are relevant to Se metabolism in other mammals or in man, or whether they are specific to the golden hamster remains to be determined. An enrichment factor of 106 was reported with an LOD of 0.02 μg L−1. The total Se concentrations were determined after reduction of SeVI to SeIV.
The application of LA-ICP-MS for the analysis of breast cancer tissues was presented by de Vega et al.127 to demonstrate the distribution of Ca, Cu, Fe and Zn in the growth of tumours. Quantitation was made possible with in-house prepared matrix-matched calibration standards using bovine breast tissue spiked with the elements and internal standardisation with Au. Histological samples of breast cancer tumours were obtained and compared with non-cancerous regions. Clear differences in the elemental profiles were found as the tumour contained enhanced concentrations of Ca, Cu, Fe and Zn with a heterogeneous distribution compared with non-tumour areas. The authors concluded that this technique has the potential to assist with the understanding of the role of elements in disease progression and could potentially provide an additional clinical diagnostic tool. He et al.128 applied LA-ICP-MS in the investigation of Ti mobility from dental implants. Human jaw bone (n = 7) from 4 patients with Ti-containing implants were compared against a control group (n = 6), with additional tests including total elemental content using ICP-OES, histological analysis and particle identification via SEM-EDX. The imaging data was only qualitative but did show that the Ti intensity decreased as the distance from implant increased. The ICP-OES data added evidence that Ti levels were elevated in the test group (1940 μg kg−1) compared to the controls (634 μg kg−1). The roundworm Caenorhabditis elegans (C. elegans) was the focus of a review by Aschner et al.129 as the transparent nematode is considered a model organism for studying metal metabolism due to its ease of handling, quick life cycle (3 days) and comparable mammalian gene expression. The authors evaluated the state of the art analytical tools available to visualise the distribution of metals within C. elegans along with the ability to understand metal homoeostasis and toxicity. The advantages and drawbacks of SR-μXRF, LA-ICP-MS and in vivo fluorescent probes were compared, with the authors summarising that the choice of technique should be driven by the study aim as it is unlikely all three can be used in the same experiment. Additionally, it was noted that challenges still arise from different sample preparation methods and with a future view that should consider metallomics approaches to gain species information.
It is not so often that publications featuring the use of ICP-TOF-MS are included in ASU Reviews, however recent instrumental developments have highlighted ICP-TOF-MS as an emerging technology for imaging applications. Bussweiler and co-workers130 described how new low-volume, fast-washout cells for LA, such as the HelEx II from Teledyne/CETAC and the Bloodhound-TwoVol2 cell from ESI, have significant advantages when coupled with an ICP-TOF-MS. The article included examples of a geological sample and a kidney tissue slice. This union demonstrated promising results with faster analysis speed, enhanced resolution and, perhaps most importantly, simultaneous and multi-element capability. Whilst mass cytometry could be considered more of an applied atomic spectroscopic technique, the use of stable metal-tagged antibodies with ICP-TOF-MS has started to gain traction and the latest development in this field has expanded to LA imaging by Chang et al.131 for thin tissue sections. The advantages and applications of the new system, as well as the drawbacks, were discussed. To add to this, Straus et al.132 described in more detail the analytical figures of merit for this new system, showing the potential of the new technology.
An article by Bloom et al.133 provided an example of alternative MS techniques for imaging applications. In this work, SIMS with C60+ cluster ions was used to map both Au nanoparticles and those functionalised with pharmaceuticals in single cell systems. As verification of the SIMS results, fluorophore-functionalised Au nanoparticles were also compared for visual confirmation. These results provided confirmation of the uptake ability of functionalised Au-based nanoparticles which could be used as drug delivery systems.
A paper of note which does not fall into either the MS or X-ray categories but warrants inclusion in the Imaging section was published by Thyssen and co-workers77 who utilised LA-ICP-OES for the elemental mapping and distribution of Ca, K and Na in tobacco plant stems and leaf petioles. The instrumental set up was similar to conventional LA-ICP-MS, only requiring Ar make up gas to stabilise the plasma from the LA He flow and connection via PTFE-coated Tygon tubing, whilst the ICP-OES was operated in axial mode for maximum sensitivity. Elemental maps were generated for the two plant tissue types with only Ca quantified using in-house prepared gelatine standards. Good correlations were observed between the Ca, K and Na maps and the structural components, as identified by microscopy. However, whilst a comparison with LA-SF-ICP-MS was performed, only Ca was considered and interference from 40Ar4He+ hampered the detection capability of 44Ca+ plus a lack of peak drift correction, due to the cold plasma conditions of the SF-ICP-MS, led to a qualitative assessment only. Unfortunately this left the paper lacking in terms of critical appraisal of both techniques.
The use of X-ray techniques for imaging applications is usually a significant part of this section of the ASU Review. However, over the past year, a more limited number of examples were found. In a study by Wrobel et al.,134 a combined approach utilising μXRF and TXRF to overcome a common issue with elemental quantification in tissue samples was proposed. The authors presented a calculation scheme to negate the effect of variations in sample density and thickness. External standard μXRF imaging, internal standard TXRF analysis and Compton intensities, enabled determination of the elemental mass fractions in kidney, liver and spleen, taking into account density and tissue thickness. Bonnemaison et al.135 published a paper focused on the quantification of Ca, Cu and Zn within anterior pituitary secretory granules to understand their role in the essential process of enzymatic peptide amidation. Samples of rat pituitaries were simply frozen at −80 °C and sliced with a cryostat at 10 μm thickness, followed by XRF analysis across different sections of the gland. With additional data from the ICP-MS analysis of sub-cellular fractions (obtained from sequential centrifugation), the combination of results provided an insight into the transportation and localisation of Ca, Cu and Zn within the cell structures. Penen et al.136 also described imaging on a sub-cellular level by employing high contrast TEM, TEM-EDX and nanoSIMS techniques to investigate the unicellular algae Chlamydomonas reinhardtii. The TEM-EDX analysis enabled detection of Ca, Fe, K, Mg, P and Zn in sub-cellular granules but the probe size (1 μm) was a limiting factor in the resolution. However, recent developments with nanoSIMS, utilising a RF plasma oxygen primary ion source, have improved the sensitivity and stability leading to a lateral resolution below 50 nm. This permitted mapping of cell organelles such as vacuoles, granules and small vesicles (approximately 200 nm in size). The high contrast TEM images achieved a superior resolution of 1 nm which confirmed the cell structures identified from the elemental mapping. The work demonstrated the capability of nanoSIMS to identify the distribution of key elements at the sub-cellular level.
Compared to hip prostheses, greater interest is now seen in dental implants as questions are now being asked about the possible consequences of implants, which are usually prepared from titanium with traces of other metals. He and colleagues128 analysed jawbones from cadavers. Slices through the jawbone (approximately 500 μm) taken 1 mm apart were prepared and analysed for 11 elements by ICP-OES and LA-ICP-MS. Except for Ti, no differences were seen between the concentrations of any elements in bone from subjects with implants, compared to the controls. Levels of Ti were greatest in the slices adjacent to an implant, which reduced as the distance increased. Particulate materials with sizes of 0.5–40 μm were found in jawbone marrow tissues at distances of 60–700 μm from the implant. In the work reported by Grenon et al.139 diffusion of Ti in titanium-implanted shinbones of adult rats was shown using spatially resolved XRF spectrometry. Consistent with the observations of He et al.,128 concentration decreased with distance from the implant. Dental peri-implantitis is analogous to the reactions seen in some patients with metal-on-metal hip prostheses. Similarly, the role of metals in the aetiology is unclear. Fretwurst et al.140 used SR-XRF spectrometry and polarised light microscopy to examine bone and mucosal tissues taken from regions around titanium implants, in patients with signs of peri-implantitis. Nine of the 12 samples were shown to have detectable levels of Ti and Fe. Microscopy showed metal particles present in soft tissue.
In addition to hip replacements and dental implants, other metallic devices in parts of the skeleton are of interest and Gornet et al.141 measured Ti concentrations in serum from patients with titanium alloy/titanium carbide composite (Ti-6Al-4 V/TiC) cervical disc implants. Insertion of these discs provides a treatment for degenerative cervical disc disease but wear can produce insoluble particulates and soluble metal ions. To assess the release of metal, these authors measured serum Ti concentration at intervals up to 84 months following operation, using SF-ICP-MS. Concentrations at all time points were significantly higher than preoperatively and stabilised at around 1.2 to 1.4 μg L−1 after about 12 months. These concentrations are much lower than in patients with other metal alloy implants.
In consideration of the concerns relating to release of metal ions from implants Matykina et al.142 evaluated in vitro release of ions from high corrosion resistant Ti (c.p. Ti) and Ti6Al4V, with 5–10 μm-thick (PEO) coatings with graded Ca/P ratio. The samples were immersed for 1 or 8 weeks in simulated body fluid or artificial saliva, at 37 °C. Corrosive conditions were mimicked by addition of H2O2 and F− was included in the artificial saliva. There was little difference in results using the different immersion solutions. The coating inhibited release of Ti from both test materials but was more effective on the c.p.Ti. The PEO coating increased loss of Al and V compared to uncoated alloy, which was attributed to these metals being in the coating. Bearing in mind the release of metals from joint prostheses and the known cardiotoxicity associated with Co, Day et al.143 developed a procedure to determine concentrations of Co, Cr and V in heart muscle by ICP-MS following closed vessel microwave digestion. Analysis of CRMs verified the accuracy of the procedure. The analytical range for all three elements was 0.5–100 ng g−1. Reference intervals (mean ± sd, ng g−1) were established from the results of 80 heart tissue samples and were 0.1039 ± 0.1305 Co, 0.1523 ± 0.2157 Cr and 0.0094 ± 0.0211 V. In a study to determine concentrations of Cr, Cu, Fe and Ni in saliva of subjects with fixed orthodontic appliances, Lages et al.144 collected samples from patients with metal or aesthetic appliances and subjects without orthodontic treatment (30 in each group). Measurements were made by TXRF spectrometry. No significant differences among the groups were found.
Two publications described concentrations of trace elements in serum of women during pregnancy. Liu et al.29 reported the reference values for 14 trace elements in serum from 1400 pregnant Chinese women. Samples were collected as part of the 2010–2012 China Nutrition and Health Survey and analysed using SF-ICP-MS. Some significant differences in serum concentrations were observed between different groupings of age intervals, residences, anthropometric status, and duration of pregnancy. Serum Fe, Se and Zn concentrations significantly decreased, whereas serum Cu, Co and Sr levels increased progressively through pregnancy. Choi et al.146 analysed samples from 245 Korean pregnant women and 527 control subjects. They also found that Co and Cu concentrations were increased while Se and Zn were low, compared to control values and that changes were seen through the course of pregnancy. Pregnancy and neonatal outcomes, gestational diabetes, preeclampsia, neonatal birth weight and congenital abnormalities were assessed and the prevalence of preeclampsia was significantly lower in subjects with high serum Cu concentrations. Serum Co concentrations in pregnancy, given in both studies, have not been previously reported but other results are consistent with many studies. Small differences between the two sets of results, possibly associated with timings of collection, were seen but in general there was good agreement between the two data sets.
Projects to establish trace element concentrations in different sample types collected from general populations continue to be presented. Samples of blood and urine were collected from almost 2000 adult subjects living in Northern France and analysed for 14 elements. This work, by Nisse et al.30 was presented as the IMEPOGE Study 2008–2010. Unsurprisingly, blood concentration of Cd and Pb were higher in smokers than in non-smokers. The mean urinary concentrations of most elements found in the general population of Northern France were higher than those found in the French national survey for the same period except for urinary V. However, the mean blood Pb concentration was markedly lower than found for the French national population. Pointing out that, within the European Union, Greeks have the highest proportion of smokers, Sakellari et al.147 measured concentrations of Cd and Pb in blood samples from adults living in metropolitan Athens. The results were evaluated in terms of age, sex, smoking status, alcohol drinking, educational status and nutritional habits. The 95th percentile values for all subjects for Cd and Pb were 2.3 and 88 μg L−1, respectively. Multiple linear regression analysis demonstrated that the predictor variables for Cd in blood were the standardised body weight, smoking, age, alcohol consumption and intake of leafy vegetables, whereas for blood Pb concentrations they were sex and age. Paglia et al.148 measured the urinary excretion of 19 toxic and essential elements during one year to assess seasonal variability in concentrations. Seven participants were recruited, and collected first morning urine samples three times in a week of each month for November, January, April and July. The same balanced diet was followed during the week of collection. Twelve of the 19 elements, both essential and non-essential, had a significant seasonal variation with concentrations that increased during the summer. The authors suggest that time of the year should be considered when interpreting results. The relationship between concentrations of nine elements in early morning urine specimens and 24 h collections31 was included in the discussion in Section 3.1.
Reference intervals for concentrations of Co, Cr and V in normal human heart tissue were discussed in the Section on Metallic implants and biomaterials (6.3.1). The values established were (mean ± sd, μg g−1) 0.1039 ± 0.1305 Co, 0.1523 ± 0.2157 Cr, 0.0094 ± 0.0211 V.143
A common approach to occupational monitoring of As is to sum the concentrations of As species in urine including iAs, MMA and DMA; however, DMA excretion is increased after consumption of certain non-toxic arsenic species found in seafoods. To avoid this problem Hata et al.150 reported the measurement of urine iAs + MMA excluding DMA for As occupational monitoring. The two methods were compared using samples from 330 Bangladeshi villagers with oral As exposure but without seafood intake, and 172 Japanese workers with seafood intake but without As exposure. The two comparisons were used to select cut-offs for monitoring As exposure in a Japanese industrial setting.
Heitland et al.151 described the unfortunate case of a 4 year old boy who was poisoned with CrVI and iAs after ingestion of wood preservative. Concentrations of As and Cr were measured by ICP-MS in urine and blood, with Cr also determined in erythrocytes as an indicator of CrVI exposure. Inorganic and organic As species were measured in urine by HPLC-ICP-MS using previously published HPLC parameters. Extremely high concentrations of these analytes were recorded in the days soon after exposure (day 3: erythrocyte Cr 4540 μg L−1, blood Cr 2180 μg L−1, blood As 206 μg L−1; day 4: urine AsIII 109 μg L−1, urine AsV 115 μg L−1) with further measurements taken during treatment and recovery. Thankfully the child survived but required a liver transplant, DMPS chelation therapy and erythrocyte apheresis.
Igra et al.153 assessed the effect of dietary B intake on pregnancy outcomes in a region of Argentina with high levels of B in drinking water. Serum, whole blood and urine were collected from 194 pregnant women and B, along with As, Cs, Li, were measured in the samples using ICP-MS. Serum B concentrations above 80 μg L−1 were inversely associated with birth size in a multivariable regression model, with a particularly strong association seen with samples collected in the third trimester. The association remained significant after adjustment for possible confounding covariates. This suggests B may be a developmental toxicant in humans and further study in other exposed populations in warranted.
An incident in which a 4 year old boy was poisoned with CrVI and iAs after ingestion of wood preservative,151 was described in Section 6.4.1 Arsenic.
The lack of SI traceable standards and reference methods for vitamin B12 has been addressed by Dash et al.156 who published two novel methods for determination of cyanocobalamin. The approaches, one based on ion chromatography and the other on ICP-AES, measured the Co and P content of cyanocobalamin, allowing traceability to Co and P primary reference materials. Both techniques generated low expanded uncertainties (0.3% to 1.0%, 95% CL) which should allow the development of traceable standards for vitamin B12.
Mostafaei and Neie160 reported the development of a K-XRF method for non-invasive in vivo measurements of Gd in bone, extending work reviewed in our previous Updates. The method used X-rays produced with a 5 GBq 109Cd source, which had increased activity compared to the source in the previous study (0.17 GBq). Bone phantoms spiked with Gd were used to simulate human bone, and were able to achieve an LOD of 0.8 μg g−1. A layer of plastic (lucite) was added to simulate the soft tissue covering the bone, yielding LODs ranging from 1.81 μg g−1 (0.61 mm lucite) to 3.47 μg g−1 (6.13 mm lucite). The authors stated that future studies will use human volunteers who have been exposed to Gd-based contrast agents.
Murata et al.161measured Gd in brain, bone and skin autopsy specimens from 9 patients who had received Gd-based contrast agents. The majority of existing evidence for contrast agent toxicity comes from “group 1 linear agents” so this study selected patients who had only received a single non-group 1 (macrocyclic or linear protein interacting) agent. The samples were digested in HNO3 with microwave assistance and Gd was measured by ICP-MS in no-gas mode with Ir as the internal standard. Gd was found in brain tissue from all but one of the exposed patients but was below or at the LOD in control brain samples. The concentrations were significantly correlated (r = 0.81, p-value 0.022) with brain levels but were on average 23 times higher, leading the authors to suggest that bone Gd may be a useful surrogate for brain Gd deposition.
The review period included two pharmacokinetic studies of Gd-based contrast agents in rats. Smith et al.162 assessed the clearance of Gd from the brains of healthy rats after repeated intravenous dosing gadodiamide, a group 1 linear agent. Blood and tissue samples collected at 1 and 20 weeks post exposure, which were digested using HNO3:H2O2 with microwave assistance and Gd was measured using ICP-MS. After 20 weeks, Gd was below the LOD in blood but remained elevated in brain samples, although there was no histological evidence of neurotoxicity. Jost et al.163 used ICP-MS and imaging with MRI to study the distribution of 7 different Gd-based contrast agents in rats. Four and a half hours after a single intravenous dose the Gd concentrations were higher in the CSF than in the blood but within 24 h, this was reversed. It suggested that Gd was rapidly cleared from the CSF although there was evidence of higher concentrations persisting in brain tissue samples. Interestingly, there were virtually no differences in the distribution of the different agents, indicating that their entry into the CSF was unaffected by their chemical structure.
In similar work Mahmoud et al.105 investigated the targeting of AuNPs to different layers and structures in human skin. The skin was donated by a patient undergoing abdominal plastic surgery and was incubated with Au nanorods that differed in their surface functional groups (neutral, anionic, cationic and hydrophobic). The stratum corneum, hair follicles and deep skin layers were separated and digested in aqua regia before Au was quantitated using ICP-OES. Hydrophobic polystyrene-AuNPs preferentially accumulated in the hair follicles while hydrophilic PEG-AuNPs accumulated in all skin compartments. The charged NPs did not accumulate in any of the skin regions. The distribution of the hydrophilic PEG-AuNPs was also imaged with LA-ICP-MS, confirming the distribution seen with quantitative measurements. Both studies demonstrate an interesting application of quantitative ICP-MS and imaging with LA-ICP-MS, highlighting the growing use of atomic spectroscopy in nanoparticle research.
In the second paper regarding carboplatin, Cuello-Nuñez et al.167 described the development of an SI-traceable assay for Pt-DNA adducts intended to provide reference values for clinical research and trials. The approach measured Pt-GG (bisadduct of cis-diammineplatinum(II) with 2′-deoxyguanylyl-2′-deoxyguanosine) with a complex double IDA strategy using enriched 196Pt and 194Pt-GG with analysis by HPLC-SF-ICP-MS. The procedure was applied to carboplatin treated human lung carcinoma cells and could detect 5.54 ng Pt per mg DNA, with a measurement uncertainty (k = 2) of approximately 20%.
Harrington et al.75 described the development of an ICP-OES method for Ti measurement in clinical specimens. Whole blood, serum and hip aspirated fluid samples were diluted 1:5 with 0.5% (v/v) HNO3 before measurement of Ti using the emission line at 336.1 nm. The intermediate imprecision of the assay was 8.6% in whole blood and 10.9% in serum but a low bias (3.3% serum, −1.0% whole blood) was seen when compared to SF-ICP-MS. The LOD of 0.6 μg L−1 and LOQ of 1.9 μg L−1 were sufficiently low for measurement of Ti in patients with a total hip replacement, but patients with hip resurfacing generally gave results below the LOQ. The measurement of Ti in a cohort of hip replacement patients provided useful information about the distribution of values, although the clinical significance of the results was not discussed.
Gornet et al.141measured Ti in serum from 30 patients with cervical disk spinal implants made from a titanium alloy/titanium carbide composite (Ti-6Al-4 V/TiC). The measurements were made with SF-ICP-MS (LOD of 0.2 ng mL−1), with samples taken before surgery and up to 7 years postoperatively. The median Ti concentration was significantly higher than baseline at all time points, and did not change significantly between the first postoperative measurement (3 months) and the end of follow-up. The authors noted that the concentrations (range, 1.15–1.46 ng mL−1) were lower than those described in studies of spine, knee and hip Ti-implants.
Two studies investigated the toxicokinetics of TiO2NPs after inhalation using rats in conditions designed to simulate occupational exposure. Gate et al.98 exposed young and old rats to TiO2NP aerosols (10 mg m−3) for 6 hours per day, 5 days per week and for 4 weeks. Measurement of Ti in tissues by ICP-MS at 0, 3, 28, 90 and 180 days post exposure was performed. High concentrations were seen in the lungs that cleared slowly over time. Small amounts of Ti were found in the liver and spleen but none in the blood, brain or kidneys. Pujalte et al.99 exposed rats to TiO2NP aerosols (15 mg m−3) for 6 hours and sacrificed the animals at various timepoints up to 14 days. Tissues, urine and faeces were analysed with Ti measured by ICP-MS. Large quantities were found in the lungs and faeces. There was some evidence of translocation of Ti to the blood and liver although the lack of statistical hypothesis testing made the results hard to interpret. The authors also noted that “a slight increase in TiO2 levels in olfactory bulbs and brain was visually apparent”, but whether these findings were statistically significant remains an open question.
Two laboratories in the USA agreed to work together to develop a database to record the concentrations of I in foods and dietary supplements. Prior to this agreement the FDA had, for many years, measured I in foods as part of an annual Total Diet Study while the Nutrient Data Laboratory (NDL) of the USDA Agricultural Research Service analysed foods but did not give results for I concentrations. For the new harmonised initiative, the NDL arranged for a commercial laboratory to establish an ICP-MS method and analyse a series of foods and supplements which were also analysed by the FDA laboratory using a colorimetric method. Results from both procedures were comparable. Henceforth, values from the two organisations will be published in a database for I content of foods and in the Dietary Supplement Ingredient Database of the National Institute for Health Office of Dietary Supplements.171
Another publication from the US FDA reported work to assess Se in dietary supplements.172 In view of the health effects of different Se species, analysis by LC-ICP-MS after heated water extraction was used to measure concentrations of seleno-amino acids, SeIV and SeVI, in non-selenised yeast supplements, and iSe in selenised yeast. Authors from the same laboratory also presented a simpler, faster screening method.91 Using EDXRF spectrometry, total Se was determined in dietary supplements with an LOD of 4 μg g−1. The seleno-amino acids were also measured directly by organic MS combined with an ambient ionisation source (direct analysis in real time high resolution accurate mass–mass spectrometry; DART-HRAM-MS). Of concern to consumers and the health authorities, discrepancies between labelled ingredients and detected species were found.
Herbal medicines are widely used throughout the world. Milk thistle (Silybum marianum), is said to be used in the treatment of liver diseases of toxic and viral origin and also in Wilson’s disease to prevent chronic liver impairment. Suspecting that medicines derived from milk thistle have high concentrations of Cu, Jedlinszki and colleagues173 analysed ground plant material, crude milk thistle extract and products containing purified silymarin, using ICP-MS. The levels found ranged from 0.01 to 114.18 μg g−1 with the highest concentrations in preparations containing ground milk thistle fruits or the crude extract. Given that Wilson’s disease is a condition of Cu toxicity, the authors suggested that use of these milk thistle products should be avoided for these patients. The work of Santos et al.,76 using ETV-ICP-OES to measure concentrations of eight elements in powdered medicinal herbs was described in Section 4.3.
Several new approaches to the extraction and speciation of As in food samples were reported in the last year. The replacement of HNO3 for As extraction using sub-critical H2O–H2O2 was demonstrated by Maher et al.174 to completely separate As from a wide variety of rice samples and CRMs. This was carried out in PFA vessels with microwave heating to 200 °C and 25 bar pressure. These conditions converted AsIII to AsV which facilitated quantification by HPLC-ICP-MS because the peak retention time shifted past the void and DMA peak (Supelcosil LC-SCX cation exchange column; 250 × 4.6 mm, 5 μm). This method obviated the need for a DRC for total As determination as extracted C did not enhance ICP-MS signals. A non-chromatographic method for As speciation using electrolysis-HG, with a Cys-modified graphite electrode (GE) – previously shown to be capable of converting MeHg to Hg0 – was described as more sensitive than conventional HG by Yang et al.88 The modification of the polished and cleaned GE was carried out by initial immersion in H2SO4 (0.1 mol L−1) to create carboxylic acid surface groups using a 1.8 V potential followed by immersion in a solution of either GSH or cysteine for the formation of peptide bonds. With a current of 0.4 A, only AsIII is selectively reduced to AsH3 for detection by AFS on the GSH/GE which also enables determination of total As following prior reduction. Using the CYS/GE both AsIII and AsV are reduced at 0.6 A to give iAs while DMA is also detected at a higher current of 2.0 A; MMA was determined by difference. Calibration graphs were linear in the range 0.5–50 μg L−1 with RSDs (n = 6) which increased from 3.4% for AsIII (20 μg L−1) to 7.3% for DMA (20 μg L−1). The LODs were <0.25 μg L−1 for all species except DMA at 0.37 μg L−1.
Entomophagy as a sustainable source of dietary protein is witnessing a marked increase in interest. Latunde-Dada et al.179 assessed the nutritional mineral composition (Ca, Cu, Fe, Mg, Mn and Zn) of four ‘commonly-eaten’ insects, with reference to beef sirloin, using ICP-AES. Cricket and beef had higher Ca, Fe and Mn compared with grasshopper, meal- and buffalo worm although all insects had significantly higher Fe solubility.
AAS | Atomic absorption spectrometry |
AB | Arsenobetaine |
AC | Arsenocholine |
ACN | Acetonitrile |
AE | Atomic emission |
AES | Atomic emission spectrometry |
AFS | Atomic fluorescence spectrometry |
APDC | Ammonium pyrrolidine dithiocarbamate |
ASU | Atomic Spectrometry Update |
CL | Confidence limits |
CPE | Cloud point extraction |
CRM | Certified reference material |
CS | Continuum source |
CSF | Cerebrospinal fluid |
CV | Cold vapour |
DART-HRAM-MS | Direct analysis in real time high resolution accurate mass–mass spectrometry |
DDTC | Diethyldithiocarbamate |
DLLME | Dispersive liquid–liquid microextraction |
DMA | Dimethylarsenic |
DNA | Deoxyribonucleic acid |
DPhT | Diphenyl tin |
DRC | Dynamic reaction cell |
Dw | Dry weight |
EBC | Exhaled breath condensate |
EDTA | Ethylenediamine tetraacetic acid |
EDXRF | Energy dispersive X-ray fluorescence |
EELS | Electron energy loss spectroscopy |
ERM | European RM |
ESI | Electrospray ionisation |
ETAAS | Electrothermal atomic absorption spectrometry |
EtHg | Ethyl mercury |
FAAS | Flame atomic absorption spectrometry |
FAES | Flame atomic emission spectrometry |
FDA | Food and Drug Administration |
FI | Flow injection |
Fs | Femtosecond |
FT-MS | Fourier transfer MS |
GD | Glow discharge |
GE | Graphite electrode |
GC | Gas chromatography |
GSH | Glutathione |
HCC | Hepatocellular carcinoma |
HDPE | High density polyethylene |
HG | Hydride generation |
HILIC | Hydrophilic interaction liquid chromatography |
HPLC | High performance liquid chromatography |
HR | High resolution |
IAEA | International Atomic Energy Authority |
iAs | Inorganic arsenic |
IBMK | Isobutyl methyl ketone |
IC | Ion chromatography |
ICP-AES | Inductively coupled plasma atomic emission spectrometry |
ICP-MS | Inductively coupled plasma mass spectrometry |
ICP-OES | Inductively coupled plasma optical emission spectrometry |
ID | Isotope dilution |
IEC | Ion exchange chromatography |
IL | Ionic liquid |
INAA | Instrumental neutron activation analysis |
IP | Ionisation potential |
ISE | Ion selective electrode |
iSe | Inorganic selenium |
JCGM | Joint Committee for Guides in Metrology |
KED | Kinetic energy discrimination |
LA | Laser ablation |
LC | Liquid chromatography |
LIBS | Laser induced breakdown spectroscopy |
LIF | Laser induced fluorescence |
LOD | Limit of detection |
LOQ | Limit of quantification |
MALDI | Matrix-assisted laser desorption ionisation |
MC | Multicollector |
MeHg | Methyl mercury |
MMA | Monomethylarsonic acid |
MRI | Magnetic resonance imaging |
MS | Mass spectrometry |
MSPD | Matrix solid phase dispersion |
MWCNT | Multi-wall carbon nanotubes |
NDL | Nutrient Data Laboratory |
NIST | National Institute of Standards and Technology |
NP | Nanoparticle |
PE | Polyethylene |
PEG | Polyethylene glycol |
PEO | Plasma electrolytic oxidation |
PFA | Perfluoroalkoxy alkane |
PhHg | Phenyl mercury |
PIGE | Particle-induced gamma ray emission |
PIXE | Particle-induced X-ray emission |
PT | Proficiency testing |
PTFE | Poly(tetrafluoroethylene) |
PVP | Polyvinylpyrrolidone |
Q-ICP-MS | Quadrupole-ICP-MS |
QQQ | Triple quadrupole |
REE | Rare earth element |
RM | Reference material |
RP | Reversed phase |
RSD | Relative standard deviation |
SEC | Size exclusion chromatography |
SeCys | Selenocysteine |
SF | Sector field |
SEC | Size exclusion chromatography |
SEM | Scanning electron microscopy |
SEM-EDX | SEM-energy dispersive X-ray spectrometry |
SeMet | Selenomethioneine |
SI | International System of units |
SIMS | Secondary ion MS |
SOD | Superoxide dismutase |
SPE | Solid phase extraction |
spICP-MS | Single particle ICP-MS |
SRM | Standard reference material |
SR-XRF | Synchrotron radiation X-ray fluorescence |
TBT | Tributyl tin |
TEM | Transmission electron microscopy |
TIMS | Thermal ionisation MS |
TMAH | Tetramethyl ammonium hydroxide |
TOF | Time-of-flight |
TPhT | Triphenyl tin |
TXRF | Total reflexion XRF |
UV | Ultra violet |
XANES | X-ray absorption near-edge structure |
XAS | X-ray absorption spectroscopy |
XRA | X-ray absorption |
XRF | X-ray fluorescence |
XRM | X-ray microscopy |
This journal is © The Royal Society of Chemistry 2018 |