Peter M.
Richardson
a,
Andrew J.
Parrott
b,
Olga
Semenova
a,
Alison
Nordon
b,
Simon B.
Duckett
*a and
Meghan E.
Halse
*a
aCentre for Hyperpolarisation in Magnetic Resonance, Department of Chemistry, University of York, UK. E-mail: meghan.halse@york.ac.uk; simon.duckett@york.ac.uk
bWestCHEM, Department of Pure and Applied Chemistry and CPACT, University of Strathclyde, Glasgow, UK
First published on 19th June 2018
Benchtop NMR spectrometers operating with low magnetic fields of 1–2 T at sub-ppm resolution show great promise as analytical platforms that can be used outside the traditional laboratory environment for industrial process monitoring. One current limitation that reduces the uptake of benchtop NMR is associated with the detection fields’ reduced sensitivity. Here we demonstrate how para-hydrogen (p-H2) based signal amplification by reversible exchange (SABRE), a simple to achieve hyperpolarization technique, enhances agent detectability within the environment of a benchtop (1 T) NMR spectrometer so that informative 1H and 13C NMR spectra can be readily recorded for low-concentration analytes. SABRE-derived 1H NMR signal enhancements of up to 17000-fold, corresponding to 1H polarization levels of P = 5.9%, were achieved for 26 mM pyridine in d4-methanol in a matter of seconds. Comparable enhancement levels can be achieved in both deuterated and protio solvents but now the SABRE-enhanced analyte signals dominate due to the comparatively weak thermally-polarized solvent response. The SABRE approach also enables the acquisition of 13C NMR spectra of analytes at natural isotopic abundance in a single scan as evidenced by hyperpolarized 13C NMR spectra of tens of millimolar concentrations of 4-methylpyridine. Now the associated signal enhancement factors are up to 45500 fold (P = 4.0%) and achieved in just 15 s. Integration of an automated SABRE polarization system with the benchtop NMR spectrometer framework produces renewable and reproducible NMR signal enhancements that can be exploited for the collection of multi-dimensional NMR spectra, exemplified here by a SABRE-enhanced 2D COSY NMR spectrum.
One of the main drawbacks of benchtop NMR is the low sensitivity. In this work we overcome this limitation through the use of hyperpolarization.13 Hyperpolarization is a general term that refers to the generation of a nuclear spin alignment that is significantly larger than that dictated by the Boltzmann distribution at thermal equilibrium.14,15 There are various approaches to hyperpolarization, each with different strengths and weaknesses. Some of the most popular methods include: dynamic nuclear polarization (DNP),16,17 spin-exchange optical pumping (SEOP),18 and para-hydrogen (p-H2) induced polarization (PHIP).19–26 In the context of low-field NMR spectroscopy, the ideal hyperpolarization method needs to be low-cost, fast, and relatively simple to implement so that the overall cost and portability advantages of compact NMR are not compromised.13 In this work, we focus on the PHIP approach where the source of hyperpolarization, p-H2, is relatively easy and cheap to access.27–30
Molecular hydrogen, H2, has two nuclear spin isomers: the triplet state, ortho-hydrogen, and the singlet state, p-H2. p-H2 has no net angular momentum and so is NMR silent. In order to exploit the spin order that is stored within p-H2 the symmetry of the p-H2 molecule must be broken, typically through a chemical reaction. In the original PHIP experiments of Bowers and Weitekamp (PASADENA and ALTADENA) visible hyperpolarization is generated through hydrogenation of an unsaturated precursor.27,31,32 This leads to dramatic NMR signal enhancements on the product molecule. However, the need to generate hyperpolarized molecules via hydrogenation significantly limits the scope of this approach to a relatively narrow range of chemicals with a suitable chemical precursor. In this work we focus on the signal amplification by reversible exchange (SABRE) method, which uses p-H2 as a source of hyperpolarization but does not require hydrogenation of the target molecule, providing renewable hyperpolarization over a timescale of tens of seconds with the addition of fresh p-H2.19,20
As illustrated in Fig. 1, SABRE is essentially a catalytic polarization transfer process. It works via the transient binding of the p-H2 and target substrate(s) to a transition-metal complex to form a J-coupling network that permits the transfer of spin order from the p-H2 to the nuclei on the substrate of interest.19 This transfer is achieved by carrying out the exchange reaction in a small magnetic field referred to as the polarization transfer field (PTF). For optimal polarization transfer in SABRE, the dominant J-coupling interaction in the active complex needs to be comparable to the chemical shift difference between the 1H derived from p-H2 and the target substrate nuclei.28,33 This corresponds to a PTF of tens of gauss for transfer to 1H,34 with much lower PTF values required for direct transfer to heteronuclei such as 15N and 13C.22
Fig. 1 An active SABRE catalyst reversibly binds both p-H2 and a substrate to allow polarization transfer from p-H2 to the substrate in a polarization transfer field (PTF) of a few tens of gauss. |
The conditions for efficient SABRE hyperpolarization are that the p-H2 and substrate reversibly bind to the active catalyst and the two p-H2-derived hydride nuclei couple differently to the bound substrate nuclei.28 The rates of exchange of the substrate and p-H2, along with the NMR relaxation times (T1) of the system and the propagating spin–spin couplings, dictate the maximum level of polarization that can be built up.35 It has been found that N-heterocycles such as pyridine are efficient SABRE targets when used with catalysts such as [IrCl(COD)(IMes)] (where COD = 1,5 cyclooctadiene and IMes = 1,3-bis (2,4,6-trimethyl-phenyl)-imidazolium).36 Recent work has shown that deuteration of the catalyst can result in 1H hyperpolarization levels of up to 50%.37 In addition, the scope of molecules amenable to hyperpolarization using the SABRE approach has recently been dramatically increased to include molecules with any functional group containing an exchangeable proton through the introduction of the SABRE-Relay mechanism.38 We focus here on the traditional SABRE method, but these developments will transfer directly to the novel SABRE-Relay approach to achieve a truly versatile platform to transform low-field analysis.
SABRE itself has already been shown to hyperpolarize a wide range of nuclei including 19F,3931P,40,41119Sn,4213C19,43–45 and 15N.22,33,46,47 For benchtop NMR, natural abundance 13C{1H} spectra are of particular interest because, unlike 1H spectra acquired at 43 MHz, which suffer from significant peak overlap and second-order coupling patterns due to limited chemical shift dispersion, 13C{1H} NMR spectra at 43 MHz and 400 MHz are often virtually identical in terms of observed spin-dilute peak patterns. However, due to the lower gyromagnetic ratio and very low natural abundance (1.109%) of 13C, the signal strength is very weak and so concentrated samples or isotopic labelling coupled with many transients (i.e. long experiment times) are often required.
13C SABRE hyperpolarization can also be achieved through direct transfer of polarization from the p-H2-derived hydrides of the active SABRE complex (Fig. 1) to the 13C nuclei of the substrate.33 In this case, polarization transfer fields of around 0.25 μT (or 2.5 mG), achieved through the use of mu-metal shields to exclude the Earth's magnetic field, are required to fulfil the necessary resonance condition.22,33 Due to the coupling between the 1H and 13C nuclei, it is also possible to hyperpolarize two-spin-order states involving 1H and 13C on the substrate. This type of hyperpolarization can be generated in PTFs similar to the 1H SABRE experiments (i.e. tens of G).48
1H SABRE hyperpolarization has been applied to compact NMR systems with detection field strengths ranging from tens of mT49–53 down to zero to ultra-low field.54–58 SABRE-hyperpolarized 13C NMR has also been demonstrated in a field of 5.75 mT, where very high NMR signal enhancements were observed (e.g. 30000000 in the 13C case); however no chemical shift information is available in this field regime.59 SABRE hyperpolarized 1H NMR spectra have also recently been shown for benchtop NMR spectrometers with fields of 1–2 T with sub-ppm chemical shift resolution using single-shot acquisition.13,60,61
In this work we explore the combination of SABRE hyperpolarization with a 1 T (43 MHz) benchtop NMR spectrometer using both manual and automated flow-based approaches. We compare the SABRE hyperpolarization levels that are achieved with high (9.4 T) and low-field (1 T) detection, explore the effect of protio solvents on the observed enhancements at low field, and demonstrate single-scan SABRE-hyperpolarized 1H and 13C benchtop NMR spectra. We explore the reproducibility of the renewable SABRE hyperpolarization using an automated flow and illustrate how this system can be used to achieve SABRE hyperpolarized 2D NMR spectroscopy.
In the second approach, p-H2 is bubbled through the sample at 4 bar in a PTF between 0 and 140 G produced by an electromagnet. Following a release of the p-H2 pressure, the solution is flowed into the benchtop NMR spectrometer under a pressure of nitrogen gas for signal detection. The automated flow system has been described previously for use with high-field NMR spectrometers48,62 and has been adapted here for use with a 1 T (43 MHz) NMR spectrometer (Spinsolve Carbon, Magritek). The full system consists of a hydrogen electrolysis cell (Peak Scientific), a p-H2 generator operating at 38 K to provide an estimated 92% p-H2 enrichment (Bruker), a mixing chamber within an electromagnet (a solenoid), and a glass cell inside the benchtop NMR spectrometer. The mixing chamber and flow cell are connected using fluorinated ethylene propylene (FEP) tubing. The glass flow cell was custom designed and built in house using a 30 mm section of a high quality NMR tube incorporated into the middle of a glass capillary (outer diameter of 4.3 mm and internal diameter of 2 mm). The lengths of the capillary were such that the NMR tube section was located in the detection region of the NMR instrument. The sample was transferred between the mixing chamber and the flow cell using a pneumatic control unit fed with nitrogen gas (5 bar). The pneumatic control unit (Bruker) allows for the return of the sample to the external mixing chamber after measurement. The flow cell has an exhaust to regulate the pressure during the transfer steps.
All 1D 1H NMR spectra were acquired following a single broadband 90° RF pulse. The 1D 13C NMR spectra were obtained either (a) following a single broadband 90° RF pulse without decoupling or (b) following an echo-based refocusing sequence and interleaved broadband 1H decoupling, see ESI† for details.27 The 2D COSY was acquired using a standard gradient selective COSY sequence that was modified to allow for repolarization by SABRE between each step of the experiment. For each mixing time, a single transient was acquired following: bubbling of p-H2 through the solution within the mixing chamber for 15 seconds, a 3 s delay for the release of the p-H2 pressure, 0.9 s for transfer of the sample to the spectrometer and a 0.1 s settling time inside the spectrometer. Following acquisition, the sample was returned to the mixing chamber. A delay of 10 s was included before each repolarization step to allow for full recovery of p-H2 pressure within the generator.
All samples used 5.2 mM [IrCl(COD)(IMes)] pre-catalyst loading with varying concentrations of substrate. The pre-catalyst and substrate were added to either 0.6 mL (for the manual approach) or 3 mL (for the automated approach) of methanol and mixed until fully dissolved. Both methanol-d4 (CD3OD) and protonated methanol (CH3OH) were used (Sigma Aldrich). In all cases, the active catalyst form [Ir(H)2(IMes)(sub)3]Cl (Fig. 1), is generated after reaction of [IrCl(COD)(IMes)] with H2 and the substrate.63,64 In the case of the manual shaking method the activation procedure was to add H2 to the degassed sample and shake vigorously; the gas was then evacuated and replaced with fresh H2 gas. This was repeated 6 times over 10 minutes before being left for an additional 5 minutes to allow for complete activation. In the case of the flow system a comparable activation procedure was used, however, in this case the sample was injected into to the mixing chamber and H2 bubbled through the solution for 15 s, again repeated 6 times over a period of 10 minutes.
Fig. 2A shows a comparison between 400 MHz 1H NMR spectra acquired without (top) and with (bottom) SABRE hyperpolarization for a sample containing 52 mM pyridine with 5.2 mM of catalyst in methanol-d4. As expected, the hyperpolarized spectrum contains six enhanced resonances, corresponding to the three distinct 1H resonances of pyridine in free solution (solid shapes), and of trans pyridine bound to the catalyst (hollow shapes). Fig. 2B presents the same comparison for detection using a 43 MHz benchtop NMR spectrometer. It can be readily observed that the signal to noise has vastly increased in the hyperpolarized spectrum (bottom) relative to the thermal NMR case (top). However, unlike for high-field detection, only the 1H signal for the ortho protons of pyridine can be fully resolved, due to the reduced chemical shift dispersion at 1 T. The observed SABRE efficiency in the two detection fields can be evaluated by calculating the SABRE enhancement factor (ε), which is the ratio of the integral of the hyperpolarized spectrum, to the integral of the thermal NMR spectrum (see ESI†) as shown in Fig. 2C.
The enhancement factors observed at low field (1 T) are much higher than those at high field (9.4 T). This is to be expected because of the lower Boltzmann polarization at 1 T. To directly compare efficiency, we report the polarization level, P, obtained by scaling the Boltzmann polarization at thermal equilibrium in the detection field by the observed enhancement factor, ε (see ESI†). Fig. 2D confirms that comparable levels of polarization are observed for both detection fields. The maximum enhancement values observed for the ortho proton resonances were 1610 fold (5.2%) and 17100 fold (5.9%) for the 9.4 T and 1 T detection fields, respectively. The higher polarization found for the benchtop measurement is attributed to the reduction in sample transfer time in the benchtop case, which reduces the loss of hyperpolarization due to NMR relaxation during transfer.
While pyridine is an attractive test substrate for SABRE experiments due to its high levels of polarization, the reduced chemical shift dispersion of the benchtop NMR spectrometer gives rise to complicated 1H NMR spectra, with significant peak overlap in the aromatic region (Fig. 2B). This issue can be avoided through the use of para-substituted pyridine derivatives, such as 4-methylpyridine, as illustrated by the 1H NMR spectra at 1 T in Fig. 3A. As with pyridine, large SABRE signal enhancements are obtained, with the additional benefit that all three SABRE hyperpolarized 1H resonances are resolved at 1 T.
Fig. 3B presents thermally polarized and hyperpolarized 1H NMR spectra of 4-methylpyridine in protonated methanol. Remarkably, hyperpolarized analyte characterization is again possible. This contrasts with high-field NMR, where protonated solvents lead to dynamic range problems and radiation damping68 that results in the need for solvent suppression techniques.69–71 In the lower magnetic field the hyperpolarized analyte actually yields a larger response than that of the solvent. As illustrated by the enhancement factors presented in Fig. 3C (deuterated solvent) and Fig. 3D (protonated solvent), the efficiency of the SABRE hyperpolarization is comparable in both solvents, with the best enhancements in the two solvents being 16000 fold (5.6%) and 14000 fold (4.9%), respectively. The small systematic reduction in enhancement in the protonated solvent case can be attributed to NMR relaxation, which is typically faster when using protio solvents. The ability to use protonated solvents for these measurements is a significant benefit for industrial applications as it removes the need for costly sample preparation steps prior to analysis.
Using an appropriately scaled thermally polarized spectrum as a reference, the 13C enhancement factor is estimated to be ε = 13700 (P = 1.2%) for the carbon in the para position (Cc). Larger signal enhancement factors are observed at lower analyte concentrations, where the effective catalyst loading is higher and so SABRE is more efficient. A maximum enhancement factor of 45500 fold (P = 4.01%) was observed for 26 mM of substrate with 5.2 mM of catalyst. See ESI† for full calculation and all enhancement values.
One of the major benefits of natural abundance 13C NMR is the relative simplicity of 13C{1H} spectra even in the low-field (1 T) regime. Fig. 4B shows a comparison of natural abundance 13C{1H} NMR spectra of neat 4-methylpyridine (top, thermally polarized, 256 scans in 5 h) and 156 mM 4-methylpyridine (bottom, SABRE hyperpolarized, 1 scan in 15 s). Prior to 1H decoupling, the anti-phase 13C signals were first refocused using an echo sequence (see ESI† for more details). The inset in Fig. 4B highlights the separation of the ortho and para13C resonances in the SABRE spectrum, despite a difference in chemical shift of <1 ppm. Note these peaks appear at different chemical shift values relative to the 13C{1H} spectrum of neat 4-methylpyridine due to solvent effects (see ESI† for more details). An interesting feature of the hyperpolarized 13C spectra is that the greatest SABRE enhancement is observed for the quaternary carbon (Fig. 4B, Cc, ε = 17000 and P = 1.5%); this can be attributed to its longer magnetic state lifetime, as there is no directly bound proton to drive relaxation. This is in contrast to standard 13C NMR, where these same factors make quaternary carbons the most difficult to detect.
Using high-field NMR detection, it has been shown previously that 2D SABRE hyperpolarized experiments such as 1H–1H COSY and 13C–1H HMQC62 and 2D DOSY72 are possible using the automated flow approach or using ultrafast single-shot methods with SABRE.73 Here we demonstrate that SABRE hyperpolarized 2D NMR can also be achieved using a benchtop NMR spectrometer for detection. Fig. 6 presents a SABRE hyperpolarized 2D gradient selective COSY spectrum of 52 mM of 4-methylpyridine acquired with 64 points in the indirect dimension, with re-hyperpolarization of the sample between each step (see ESI† for pulse sequence). The resulting COSY spectrum in Fig. 6 shows the expected peak patterns with the three resonances on the diagonal and as well as off-diagonal peaks indicative of the coupling between these resonances.
The example analytes used herein to illustrate the potential of this method were optimized to provide signal enhancements by factors of more than 104. We anticipate even with the more modest enhancements (on the order of 100–1000 fold) that could be routinely achieved for a broader range of analytes, the use of SABRE has the potential to open up exciting analytical applications by bridging the sensitivity gap between benchtop and high-field NMR spectroscopy. Furthermore, with the recent introduction of the SABRE-Relay method, the range of analytes amenable to hyperpolarization has been dramatically increased and now includes molecules with exchangeable protons such as amines, amides, alcohols, carboxylic acids, phosphates and carbonates.38 This approach has already been shown to work with 1H, 13C, 15N, 19F and 31P responses and a future truly versatile low field platform can now be envisaged.
By integrating an automated polarization system with the 1 T benchtop NMR spectrometer, we have shown how the continuous nature of the SABRE process can be exploited to repeatedly re-polarize the sample on a timescale of tens of seconds. The automated system allows for software control over SABRE parameters, namely: the bubbling time, transfer time, and polarization transfer field. This results in reproducible levels of total hyperpolarization as well as a highly consistent distribution of polarization within the target analyte. This reproducible SABRE hyperpolarization was exploited to carry-out a SABRE-enhanced 2D COSY NMR by transferring the sample back and forth to the polarizing unit for repolarization between each step of the experiment. The increased resolution which can be achieved from 13C and 2D experiments means that enhanced benchtop NMR now has the potential to be both highly specific and highly sensitive and therefore offers the ability to detect, identify, and quantify analytes at low concentration in complex mixtures (where many other components might be SABRE active in the 1H spectrum). Given the number of 2D sequences that have been shown to be amenable to this technique at high field (e.g. DOSY72 and HMBC62) and progress towards quantitative analysis with SABRE,74 we believe that this technology has the potential for rapid development into a versatile low cost technique for use in many analytical applications.