Fazzolari
Sandro
,
Hattendorf
Bodo
and
Günther
Detlef
*
ETH Zurich, Department of Chemistry and Applied Biosciences, Vladimir-Prelog-Weg 1, 8093 Zurich, Switzerland. E-mail: guenther@inorg.chem.ethz.ch
First published on 3rd December 2024
A fundamental study of four different sample introduction systems was carried out to evaluate the upper size limit of microplastics measured by inductively coupled plasma-time-of-flight-mass spectrometry (ICP-TOFMS). Three different, certified microplastic samples (PS, PMMA and PVC) within a size range of 3–20 µm in suspension were measured. In this study, no particles larger than 10 µm could be detected using pneumatic nebulization for sample introduction. However, we were able to extend the upper size limit to 20 µm by either using a falling-tube device or a vertical downwards-pointing ICP-TOFMS. Particle transport efficiencies could only be estimated and were within a range of 13% to 184%. The particle size was quantified by using dissolved citric acid (non-matrix matched) and agreed with reference values. The critical size values were 2.3 µm for PS, 2.4 µm for PMMA and 3.0 µm for PVC. Additionally, in the case of PVC, chlorine could also be detected and the critical size value was 3.9 µm based on the 35Cl+ ion signal.
To our knowledge there has been no report of introducing and quantifying particles of 10 µm or larger into the ICP-MS using a suspension-based approach. One of the main limitations of sp-ICPMS or single cell (sc)-ICPMS is the sample transport efficiency, which was found to decrease with increasing particle size.15,17,26–28 For example, even a so-called high efficiency cell introduction system, specially designed for single-cell analysis, showed that the transport efficiency decreases significantly with increasing cell size.28 Miyashita et al. were able to introduce 2–3 µm cells into the ICP with a transport efficiency close to 100%, which dropped to 86% for 6.4 µm cells.28 A similar observation was encountered if the high efficiency cell introduction system (HECIS) was used for sp-ICPMS.15,17 Laborda et al. reported a transport efficiency of 29.9% for 2.22 µm PS and 12.5% for 4.82 µm PS particles, respectively.15 Hendriks and Mitrano achieved a transport efficiency of 77% for 50 nm gold nanoparticles and 20% for 4 µm PS particles.17 As reported by Olesik and Bates,29 the commonly used spray chambers remove the majority of droplets >10 µm in size, which obviously also limits the particle sizes that can be transported into the ICP. In order to cope with the challenges encountered during the analyses of microparticles and cells, an ICP-TOFMS with a downward-pointing ICP had been developed.30 This arrangement allows the introduction of samples with a downward-pointing trajectory regardless of size and mass due to using low gas flows and gravity for sample transport. Initial studies have shown that 100% transport efficiency can be achieved even for droplets up to 93 µm in diameter.25,30,31
In this work, aqueous suspensions of 3–20 µm microparticles of different composition were analysed and size quantified using four different ICP-TOFMS configurations. All instrument configurations were studied with a special focus on particle transport capabilities. In addition, quantitative analysis was carried out to determine the size of the particles using non-matrix matched quantification. Finally, figures of merit for all different sample introduction approaches are summarized.
PS | PMMA | PVC | |
---|---|---|---|
Density ρP [g mL−1] | 1.05 | 1.18 | 1.38 |
Chemical formula of the repeat unit | (C8H8)n | (C5H8O2)n | (C2H3Cl)n |
Carbon concentration cc [g mL−1] | 0.97 | 0.71 | 0.53 |
Particles ranging in size from 3–20 µm (manufacturer values) were purchased from Lab261 (Palo Alto, CA, USA) in 5 mL suspensions with a solid content of 1% in deionized water containing 2 mM NaN3 as an anti-microbial agent and 0.1% Tween 20 as surfactant. 20 µm PVC was also purchased from Lab261 and additional samples (4 µm and 6 µm PVC from Lab261) were provided by Dr Thibaut Van Acker (Ghent University, Belgium). Prior to all experiments, the microplastic bead samples were inspected by optical microscopy (BX51, Olympus, Shinjuku, Japan) and images were taken using the software “ProgRes CapturePro 2.10.0.1”. For the analysis, microplastic suspensions were sonicated for 15 min and diluted in 2% HNO3 (subboiled in 18.2 MΩ ultrapure water) solution with a particle number concentration (PNC) range as listed in Table S1.† The samples were prepared gravimetrically in conditioned vials. Three aliquots had been prepared for each microparticle sample. Cesium (100 µg mL−1) or aluminium (1000 µg mL−1, both Inorganic Ventures, Christiansburg, VA, USA) were used as droplet tracers to count the number of droplets for setups A and B2 and to determine the aerosol transport efficiency in setups B1 and C.21
The carbon quantification was carried out via one-point calibration using citric acid monohydrate (Fisher Scientific, Loughborough, UK) in 2% HNO3 solution. The sensitivity for 12C+ was determined either from 500 mg L−1 or 1000 mg L−1 solutions, respectively, whereas a 5000 mg L−1 carbon content was used for determining the 13C+ sensitivity.
In Fig. 1, setups B1, B2 and C show the three other configurations used for the horizontal ICP-TOFMS system (icpTOF2R from Tofwerk AG, Thun, Switzerland). Setup B1 was based on pneumatic nebulization for particle introduction and using the online microdroplet calibration for quantification.21,23 The second approach used the same setup; however, the particles were introduced via the 50 µm MDG (Microdrop Technologies, Germany) into the falling-tube. The calibration solution was then introduced using the pneumatic nebulizer (Fig. 1B2). The third approach was identical to the second but the cyclonic spray chamber had been exchanged for a high efficiency cell introduction system (HECIS) (Elemental Scientific, Omaha, NE, USA) (Fig. 1C). A 500 mg L−1 C solution was also used for quantification.
The operating conditions for the microplastic measurements were optimized for maximum 12C+ sensitivity. The single-ion-signal histogram was collected for the compound Poisson critical value expression.33
The operating conditions for the downward-pointing ICP-TOFMS prototype are listed in Table 2.
Parameters | Conditions |
---|---|
a The 70 µm MDG was operated for the first two batches. b The last batch was introduced with the 50 µm MDG to study if a smaller nozzle-sized MDG can introduce particles up to 20 µm. The piezo voltage and the pulse width for both MDGs were adjusted using the control unit MD-E-3021 SN: 0154 from microdrop Technologies GmbH, Norderstedt, Germany | |
MDG (70 µm, MD-K-150-020, SN: 218)a | Piezo voltage: 74–75 V, pulse width: 24 µs |
MDG (50 µm, MD-K-150, SN: 217)b | Piezo voltage: 92 V, pulse width: 24 µs |
Dispensing frequency [Hz] | 50 |
He 1 [L min−1] | 0.58–0.77 |
He 2 [L min−1] | 0.52–0.67 |
Ar (sweep gas) [L min−1] | 0–0.09 |
Ar [L min−1] | 0.41–0.62 |
Ar auxiliary [L min−1] | 1.46 |
Ar plasma [L min−1] | 14.1–15.8 |
Plasma power [W] | 900 |
Einzel lens voltage [V] | 5 |
MCP voltage [V] | 2030–2100 |
Time resolution [ms] | 3 |
The measurement conditions for the horizontal icpTOF are listed in Table 3. All the particle measurements were carried out using a total analysis time of 182 s for the spray chamber systems and 600 s when using the falling-tube. The icpTOF was optimized to achieve high sensitivity for low mass elements (m/z < 80) and thus, Al was chosen as a droplet tracer.
Parameters | Conditions |
---|---|
MDG (50 µm, AD-KH-501-L6, S/N: 78) | Piezo voltage: 48 V, pulse width: 21 µs |
Dispensing frequency [Hz] | 50 |
Ar/He gas mixture [L min−1] | 0.75 (Ar) and 0.54 (He) |
Ar nebulizer [L min−1] | 0.75 |
Ar auxiliary [L min−1] | 0.8 |
Ar plasma [L min−1] | 16 |
Plasma power [W] | 1550 |
MCP voltage [V] | 2040 |
Time resolution [ms] | 3.02 |
The measurement conditions for the microparticle measurement using the HECIS (high efficiency cell introduction system designed for TOFWERK) spray chamber are listed in Table 4. The HECIS spray chamber was provided by Dr Kharmen Beham Billimoria (LGC, UK).
Parameters | Conditions |
---|---|
MDG (50 µm, AD-KH-501-L6, S/N: 78) | Piezo voltage: 48 V, pulse width: 21 µs |
Dispensing frequency [Hz] | 50 |
Ar/He gas mixture [L min−1] | 0.75 (Ar) and 0.4 (He) |
Ar 1 (nebulizer) [L min−1] | 0.31 |
Ar 2 [L min−1] | 0.22 |
Ar auxiliary [L min−1] | 0.8 |
Ar plasma [L min−1] | 16 |
Plasma power [W] | 1550 |
MCP voltage [V] | 2040 |
Time resolution [ms] | 2.9 |
![]() | (1) |
The microplastic particle volume VP (eqn (2)) was calculated by dividing the background-corrected particle ion signal IP (12C+ or 13C+, counts) by the sensitivity Si (counts per g) and the respective carbon concentration cc (Table 1).
![]() | (2) |
The transport efficiency was determined by dividing the calculated PNC (PNCcalc) by the final PNC (PNCfinal) as listed in Table S1.† The PNCcalc in these experiments was obtained viaeqn (3) where NP is the counted number of particle events, ND is the counted number of droplets (based on the 133Cs+ tracer signal) and VD is the mean volume of the monodisperse droplets.
![]() | (3) |
Samples | Reference diameter [µm] | Measured [µm] |
---|---|---|
3 µm PS | 3.10 ± 0.01 | 3.1 ± 0.1 |
4 µm PS | 4.20 ± 0.17 | 3.5 ± 0.5 |
6 µm PS | 6.05 ± 0.10 | 6.1 ± 0.1 |
8 µm PS | 8.12 ± 0.12 | 8.4 ± 0.4 |
10 µm PS | 9.98 ± 0.10 | 9.9 ± 0.1 |
20 µm PS | 23.33 ± 1.09 | 22.5 ± 4.0 |
5 µm PMMA | 5.2 ± 0.1 | 5.4 ± 0.5 |
6 µm PMMA | 6.2 ± 0.2 | 6.0 ± 0.1 |
8 µm PMMA | 7.52 ± 0.12 | 7.6 ± 0.1 |
10 µm PMMA | 10.22 ± 0.30 | 12.3 ± 3.0 |
20 µm PMMA | 20.03 ± 0.81 | 18.8 ± 2.6 |
4 µm PVC | 3.69 ± 0.18 | 3.5 ± 0.5 |
6 µm PVC | 5.59 ± 0.51 | 6.8 ± 1.5 |
20 µm PVC | 20.93 ± 0.89 | 22.0 ± 2.0 |
![]() | ||
Fig. 2 Transient signals for a period of 2 second from the 5 µm PMMA (A and D), 10 µm PMMA (B and E) and 20 µm PMMA (C and F) measurements with the downward-pointing ICP-TOFMS (A–C) and the setup B2 (D–F). The transient signals were split-event and background-corrected. For each plot, 50 droplets were counted. The dispensing rate was 50 Hz. Asterisks indicate the signals for double droplet events, exceeding a threshold of 5 * σ +µ. The standard deviation σ and the mean value µ were obtained from the Gaussian fits shown in Fig. 3. |
The histograms of the droplet signals from these measurements are shown in Fig. 3A–F. The number of droplets in Fig. 3C was higher than that in Fig. 3A and B because the lower transport efficiency for 20 µm particles required a longer measurement time (35 min instead of 10 min) to record at least 40 microparticle events.
![]() | ||
Fig. 3 Histogram of the droplet tracers (133Cs+ and 27Al+) from experiments shown in Fig. 2 of the 5 µm PMMA (A and D), 10 µm PMMA (B and E) and 20 µm PMMA (C and F) measurements. The histogram counts were normalized to the total number of droplets injected and fitted with a Gaussian function (red). The bin size was set to 10 counts. |
Fig. 4A–F show the corresponding particle-derived carbon ion signals of the 5 µm, 10 µm and 20 µm PMMA particles. The signal intensities generally showed a Gaussian distribution except for the 20 µm PMMA (Fig. 4C and F), where 49 microparticles were detected. Additionally, double events have been observed for the smaller particles and even triple events (see Fig. 4A, B and E). The occurrence of double and triple events was below 2% and 0.5%, respectively. The 13C+ signal intensity of the 20 µm particle was evaluated for carbon content quantification because the 12C+ ion signal exceeded the linear dynamic range of the ion detection system.
![]() | ||
Fig. 4 Histogram of the particle-derived carbon signals from the 5 µm (A and D), 10 µm (B and E) and 20 µm (C and F) PMMA collected from the experiment shown in Fig. 2A–F. The signal distribution was fitted with a Gaussian function. The 13C+ ion signal was collected only for the 20 µm particles due to exceeding the linear dynamic range of the 12C+ ion signal. The bin size was set to 20 counts. |
Another interesting observation that was made during these experiments was an apparent shift of the signals for droplets containing the microparticles. Especially when introducing droplets with 20 µm particles, we observed that their arrival in the ICP was distinctly offset from the expected time, based on the injection frequency. While “empty” droplets were typically observed to arrive within periods of 20 ms, the particle containing ones frequently appeared at irregular offsets of the expected time, while a gap occurred (e.g. at 79.8 s in Fig. 2C) in the droplet tracer signals. It seems that these particles arrived into the ICP at a different flight velocity than the empty droplets. The droplet and particle flight velocities as well as the drying efficiency of various droplet sizes in the downward-pointing ICP-TOFMS need to be further investigated.
Droplets with 20 µm microparticles generally had a lower droplet tracer signal than the 10 µm particles or the empty droplets (Fig. 5), indicating a matrix effect caused by the particles. This was not observed in experiments using the falling-tube setup B2 and seems to indicate that matrix effects were more pronounced for the low plasma power used here.
The mean particle sizes were determined by considering their stoichiometry and shape via the non-matrix matched calibration with all four instrumental configurations. Fig. 6 compares the measured diameters with the reference values provided by the manufacturer. 20 µm particles could in this study only be detected for the downward-pointing plasma and the falling tube setup. Using the cyclonic spray chamber setup, it was furthermore not even possible to introduce the 10 µm particles. Yet an accurate size determination of detected microplastic particles was generally possible with all four methods, indicating that the sensitivity of carbon is practically independent of the composition of the particle and even comparable to that of dissolved citric acid. The uncertainty of the size determination was typically below 15% except for the 20 µm particles when using the falling tube setup. While sizing of microplastics of 8 µm and below seems to be generally possible using pneumatic nebulization with a cyclonic spray chamber, larger particles seem to be not efficiently transported to the ICP. The HECIS can also provide transport of 10 µm particles but the largest sizes were only detected with a vertical arrangement and via microdroplet introduction. Critical sizes or size limits of detection were determined from the carbon sensitivities of the different configurations. Due to the differences in carbon concentration in the different materials their critical sizes vary. Using the downward-pointing ICP, the critical size value was 2.3 µm for PS, 2.4 µm for PMMA and 3.0 µm for PVC, respectively. The sensitivity using the horizontal icpTOF2R was higher by a factor of 1.6–2 in comparison to the downward-pointing ICP-TOFMS. However, the 12C+ background signal was also higher by a factor of 3.7. The critical sizes for the icpTOF2R were thus higher, namely 2.8, 3.2 and 3.8 µm for PS, PMMA and PVC respectively.
![]() | ||
Fig. 6 Comparison of the measured and reference sizes for the different microplastics and sample introduction setups. (A) Downwards ICP, (B1) horizontal ICP using the pneumatic nebulizer and cyclonic spray chamber, (B2) falling-tube and the (C) HECIS. Numerical values are provided in Table S2.† The plots were labeled the same as the four configurations in Fig. 1. |
Sample introduction system, transport efficiency [%] | ||||
---|---|---|---|---|
Sample | Downward-pointing | Horizontal ICP | ||
Cyclonic | HECIS | Falling-tube | ||
3 µm PS | 84 ± 57 | 14 ± 1 | 49 ± 02 | 34 ± 06 |
4 µm PS | 72 ± 74 | 4 ± 1 | 60 ± 15 | 25 ± 10 |
6 µm PS | 77 ± 40 | 1 ± 1 | 67 ± 12 | 60 ± 03 |
8 µm PS | 84 ± 13 | 0.3 | 26 ± 01 | 56 ± 06 |
10 µm PS | 90 ± 23 | 0 | 10 ± 03 | 123 ± 30 |
20 µm PS | 38 ± 27 | 0 | 0 | 22 ± 20 |
5 µm PMMA | 25 ± 05 | 2 ± 1 | 22 ± 12 | 57 ± 64 |
6 µm PMMA | 64 ± 19 | 0.3 ± 0.3 | 13 ± 05 | 109 ± 35 |
8 µm PMMA | 115 ± 18 | 0.2 | 4 ± 04 | 80 ± 20 |
10 µm PMMA | 85 ± 35 | 0 | 1 ± 01 | 184 ± 108 |
20 µm PMMA | 62 ± 51 | 0 | 0 | 132 ± 88 |
4 µm PVC | 29 ± 03 | 2 ± 1 | 2 ± 01 | 13 ± 05 |
6 µm PVC | 69 ± 33 | 0.4 ± 0.2 | 1 ± 01 | 14 ± 08 |
20 µm PVC | 43 ± 18 | 0 | 0 | 25 ± 15 |
We thus re-assessed the actual particle number concentrations for three dried aliquots of 10 µL of four selected particle sizes analysed using microscopy (Table S3†). Using these instead of the reference values in most cases reduced the spread of the number concentrations and improved the calculated transport efficiencies. There were however substantial inconsistencies between the results for the downwards ICP and falling tube setup and for individual sample batches. The transport efficiency for example for 8 µm PMMA still ranged from 63% to 241% for the falling tube setup (95–128% for downwards ICP) or between 18% and 140% for 10 µm PMMA.
Analysing dried droplets of five individual, freshly prepared batches of three microplastic samples of different size by microscopy again revealed a notable spread in particle number concentrations and a deviation of the mean to the reference values by up to a factor of four (Table S4†). This indicates that sample preparation has indeed a substantial impact on the particle number concentrations in the analysed suspension. It is not clear at this stage where these variations originate but it must be concluded that the uncertainty of the determined transport efficiencies is at the order of a factor of two or larger.
Due to the high uncertainties discussed before, it remains difficult to assess whether there is a significant difference in transport efficiencies obtained with the downwards ICP and the falling tube setup for the different sample materials and their sizes. The particle number concentrations determined, however, differ by less than a factor of two from the reference values in most cases, which is in general agreement with the uncertainties observed during sample preparation.
![]() | ||
Fig. 7 Transient signals of the 4 µm (A), 6 µm (B) and 20 µm (C) PVC particles showing coincident particle-derived carbon and chloride ion signals. |
The histograms of the particle-derived chlorine ion signals (35Cl+) are shown in Fig. 8 and fitted by an approximation to a Gaussian distribution. There were also larger signals observed for the 6 µm PVC outside of the Gaussian fit function which might come from larger PVC particles (see Fig. 8B). It was observed from the microscopic images that the 6 µm PVC sample also contained particles up to 9 µm.
![]() | ||
Fig. 8 Histogram of the particle-derived 35Cl+ ion signals for 4 µm (A), 6 µm (B) and 20 µm (C) PVC particles with a Gaussian fit function. |
Fig. 9 shows a linear correlation of the particle derived chlorine signal with the calculated mean particle volume which was derived from the carbon mass. The critical size value of PVC, based on the 35Cl+ sensitivity, was 3.9 µm.
Nonetheless, the experiment demonstrated that also chlorine can be detected in microplastics but only if the particle is sufficiently large like the 20 µm (and partly 6 µm) PVC in this case. The measurement indicated that the applicability of the ICP-TOFMS can be further extended to identify and distinguish chlorinated from non-chlorinated microplastics for particles above 6 µm diameter.
Yet, the determination of transport efficiencies for the samples analysed here remained challenging, because the particle number concentration in the analysed suspensions appeared to be affected by the sample preparation. A high variability was obtained for the particle detection events of individual batches of the same sample type and across the different materials and in the absence or presence of Tween 20 as surfactant. Control experiments by microscopy also revealed variations in particle number concentration by approximately a factor of four to the reference values. Therefore, the estimated transport efficiencies are considered to be of similar magnitude. For the spray-chamber based arrangements, the estimated transport efficiencies decreased with increasing particle sizes, which was not the case for the vertical particle introduction. The straight vertical trajectories of droplets and particles for the downward pointing ICP, however, allowed the detection of more particle events per measurement sequence for the largest particle size studied. This is considered to be an advantage for routine applications when analysing microplastic suspensions of unknown size distribution. In all cases where detection was successful, however, the particle sizes could be accurately quantified via the 12C+ or 13C+ ion signals using a standard of dissolved citric acid as the carbon source. Furthermore, the detection of chlorine was also possible for 4 µm PVC particles, which can be applied to distinguish chlorinated from non-chlorinated microplastics.
Particles of 20 µm diameter were reproducibly detected and size quantified by the vertical droplet and particle injection, suggesting that also larger objects can be analysed with these configurations.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4ja00323c |
This journal is © The Royal Society of Chemistry 2025 |