Issue 26, 2019

A portable oligonucleotide-based microfluidic device for the detection of VEGF165 in a three-step suspended-droplet mode

Abstract

Vascular endothelial growth factor (VEGF165), an important glycosylated protein from the VEGF family, is a type of signal protein highly associated with the development and progression of cancers. In this work, we designed a G-quadruplex-based aptasensing platform for the sensitive and selective detection of VEGF165 in aqueous solution and red blood cell solution. A long-lived phosphorescence iridium(III) complex (1) with promising photophysical properties and a large Stokes shift was chosen as a selective G-quadruplex probe. The platform could achieve a limit of detection (LOD) down to the picomolar level using a conventional fluorometer. Furthermore, we successfully applied the platform to a three-step suspended droplet (SD)-based microfluidic device for the monitoring of VEGF165. In contrast to the channel-based and digital microfluidic chips, SD-based chips allow easy introduction of liquid samples, valve-free manipulation of multiple reaction steps and flexible volume range. Importantly, polypropylene (PP), a hydrophobic and thermally stable material, was chosen as a substrate to fabricate the chip for the SD-based microfluidic device. The PP-based chip allows the combination of superhydrophobic force, gravity and surface tension for effective driving of the suspended droplet throughout the channel without reverse migration. After assembling all the major components, including a UV lamp, a rotatable chip holder, a filter and a camera into the portable device, we successfully demonstrated the applicability of the device to detect VEGF165 in aqueous solution with a LOD of 0.33 nM at a signal-to-noise ratio (S/N) of 3 and a linear range of 1–100 nM.

Graphical abstract: A portable oligonucleotide-based microfluidic device for the detection of VEGF165 in a three-step suspended-droplet mode

Supplementary files

Article information

Article type
Paper
Submitted
30 jan 2019
Accepted
26 mai 2019
First published
31 mai 2019

Dalton Trans., 2019,48, 9824-9830

A portable oligonucleotide-based microfluidic device for the detection of VEGF165 in a three-step suspended-droplet mode

C. Ko, H. Sun, K. Wu, C. Leung, K. Ren and D. Ma, Dalton Trans., 2019, 48, 9824 DOI: 10.1039/C9DT00427K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements