aBundesanstalt für Materialforschung und -prüfung (BAM), Richard-Willstätter-Str. 11, D-12489 Berlin, Germany. E-mail: Carlos.Abad@bam.de; Zengchao.You@bam.de
bDepartment of Chemistry, Humboldt Universität zu Berlin, Brook-Taylor-Str. 2, D-12489 Berlin, Germany
First published on 28th April 2023
We investigated the performance of nitrogen microwave inductively coupled atmospheric-pressure plasma mass spectrometry (MICAP-MS) under matrix effects and its applicability to trace element analysis in steels. Influences of different gas flows and ion optics on the matrix tolerance are studied, indicating that nebulizer gas flow has the most significant impact. Optimization of ion optics improves matrix tolerance for light elements due to the reduction of the inelastic collisional scattering effect. With optimized operating conditions, MICAP-MS achieves an internal standard intensity recovery of over 90% at an Fe concentration of 500 mg L−1. Even at an Fe concentration of 1 g L−1, the recovery remains above 80%. Three certified reference materials – non-alloy, low-alloy and high-alloy steel – were analyzed using MICAP-MS. The determined mass concentrations of the trace and minor components show metrological compatibility to the reference values. No significant differences are observed between the results obtained with aqueous and matrix-matched calibration, demonstrating the strong matrix tolerance of MICAP-MS, and its promising applicability to steel analysis.
Inductively coupled plasma mass spectrometry (ICP-MS) has been successfully applied for trace element analysis due to its capability to measure trace levels of most elements of the periodic table.10–12 In the last few decades, it has also been used for trace element analysis in steels.13–15 However, a noticeable limitation in these measurements is the large contribution of metals to the matrix effects,16,17 which may result in plasma suppression and long-term signal drift due to matrix deposition on the interface cones.18,19 Due to the excessive influx of positively charged matrix ions space charge effects can occur, leading to the defocusing of the ion beam to the mass spectrometer.20 Another limitation of the argon plasma sources is the interference resulting from Ar-related species. For example, 40Ar+ and 40Ar12C+ can interfere with the most abundant isotopes of calcium and chromium, which restrict the characterization of these elements. Besides, the high consumption of Ar gas demands a substantial budget for its operation.
The Ar-related limitations have driven continued interest in the development of alternative sources for ICP.21,22 Promising ion sources are microwave plasmas,23,24 especially the recently developed nitrogen-based microwave inductively coupled atmospheric-pressure plasma (MICAP),25–27 which utilizes a dielectric resonator in place of a load coil as in ICP, simplifying the instrument's electronics. By using nitrogen as a plasma gas, the typical Ar-related polyatomic interferents are not encountered. It has been demonstrated by several studies that MICAP-MS achieves comparable limits of detection (LOD) and sensitivity to ICP-MS for most of the elements, but costs significantly less for gas consumption.26,28,29 For practical applications, the investigation of matrix effects in MICAP is of great interest. In the 1980s, Urh et al. developed a microwave-induced plasma optical emission spectrometer (MIP-OES), and showed that matrix effects in MIPs are much more severe than in the Ar-ICP.30 About 30 years later, Thaler et al. implemented MICAP with OES in 2017, and demonstrated that matrix effects caused by sodium could suppress or enhance the analyte signal.31 Thereafter, Pelipasov et al. further investigated the matrix effects induced by elements with different ionization potentials in MICAP-OES. They indicated that with a decrease in the ionization potential of the matrix elements, matrix effects for both atomic and ionic lines increase.32 However, the influence of matrix effects in MICAP-MS has not been reported so far.
In the present work, we investigated the matrix tolerance of MICAP-MS and its applicability to quantification of minor components and trace elements in steels. The influences of gas flows and ion optics were studied. The recovery of the internal standards (IS) at different iron, nitric acid and hydrochloric acid concentrations was studied to evaluate the performance of MICAP-MS related to matrix tolerance without significantly compromising signal intensity. Non-alloy, low-alloy and high-alloy steels containing Al, V, Cr, Mn, Co, Cu, Ga, Nb, Mo, Pb, Sn, and Sb were analyzed after digestion with aqua regia, and the mass fractions of the selected elements were determined. Results obtained with aqueous and matrix-matched calibrations were compared and discussed.
High-purity deionized water with a resistivity of 18 MΩ cm obtained from a Milli-Q system (Merck Millipore, Germany) was used throughout the experiments. HNO3 (Merck AG, Germany), HCl (Merck AG, Germany), and HF (Merck AG, Germany) were used after purification by sub-boiling distillation in PFA containers. Calibration solutions were prepared from single-element standard solutions (Certipur®, Merck AG, Germany), and an ICP-MS IS solution (Analytik Jena GmbH, Germany) containing 6Li, 45Sc, 89Y, 115In, 159Tb, and 209Bi by dilution in 2% HNO3. In addition, matrix-matched calibration solutions were spiked with an Fe plasma standard solution (Specpure®, Alfa Aesar, United Kingdom).
For parameters optimization, blank and matrix solutions each containing 50 μg L−1 IS in 2% HNO3 were prepared. In addition, the matrix solutions contained 350 mg L−1 Fe, 90 mg L−1 Cr, and 40 mg L−1 Ni. All torch parameters (gas flow rates and sample depth) and ion optics parameters (voltages on lenses and mirrors) were tuned for high matrix tolerance and high sensitivity. For trace elements quantification, aqueous and matrix-matched calibrations were performed with six calibration levels. The calibration range and corresponding IS for each element can be found in Table S1.† Each calibration standard contained 50 μg L−1 IS in 2% HNO3. In addition, 500 mg L−1 Fe was added to each standard for matrix-matched calibration, whereby no further elements were added for aqueous calibration.
MICAP-MS | |
---|---|
Plasma power | 1500 W |
Nebulizer gas flow | 0.9 L min−1 N2 |
Auxiliary gas flow | 0.8 L min−1 N2 |
Plasma gas flow | 11 L min−1 N2 |
Sampling depth | 5 mm |
Sampling cone | Pt 1.1 mm |
Skimmer cone | Ni 0.5 mm |
We found potential memory effects in the tubing for certain elements (V, Co, Nb, W). After two washing steps with 10% HNO3, the signal returns to the blank level. Washing steps were performed after measurement of calibration standards and samples with high content of these elements.
Since the degree of the matrix effect is strongly dependent on the applied nebulizer gas flow in ICP-MS, influences of the nebulizer gas flow on the matrix tolerance in MICAP-MS were studied first. IS intensities were monitored under varying nebulizer gas flow rates from 0.45 L min−1 to 1.5 L min−1. Measurements were performed twice with blank and matrix solutions to give an insight into the variation of the matrix tolerance. The obtained results are shown in Fig. 1.
It can be observed that the signal intensities of the isotopes increased with a higher nebulizer gas flow rate, which was most likely due to the more efficient aerosol transport and ion sampling. However, a reduction could be observed for most of the isotopes (89Y, 115In, and 209Bi) when the gas flow rate exceeded 1.35 L min−1. This reduction might indicate that an excess aerosol load reduces the plasma temperature, ultimately leading to signal suppression.33 Furthermore, this observation agrees with the MICAP-MS study reported by Kuonen et al.,34 which indicates that higher nebulizer gas flow can also induce oxide formation in the plasma. 6Li did not demonstrate this intensity reduction (see Fig. 1a), but it showed much lower intensity compared to other isotopes at a low gas flow rate, which might have resulted from the stronger ion diffusion in the plasma. Signal suppression could be observed in the results obtained with the alloy matrix. A clear reduction of the matrix tolerance occurred when a nebulizer gas flow rate higher than 0.9 L min−1 was applied. This reduction indicates that the matrix effect tends to be more severe at a high nebulizer gas flow rate, which might have resulted from the increased matrix plasma load and the shorter residence time of the aerosol. This reduction was especially obvious for 6Li and 209Bi (Fig. 1a and f), which was possibly caused by the increased space charge effect for the lighter ions, and the slightly higher first IE of Bi (7.3 eV), respectively. It has also been demonstrated that the non-linear mass dependency on the matrix effect was minimized by optimizing matrix tolerance. At a nebulizer flow rate of 0.9 L min−1, the IS recovery in the matrix ranges from 87% to 92%. However, at a nebulizer flow rate of 1.35 L min−1 (optimized for high signal), not only does the recovery decrease, but the recovery range also becomes wider, spanning from 55% to 73%.
Apart from the nebulizer gas flow, influences of plasma gas flow, auxiliary gas flow, sheath gas flow, and sample depth on the matrix tolerance of MICAP-MS were also investigated. Fig. 2 shows the effects of varying conditions on 115In. It can be observed in Fig. 2a and S1† that the plasma gas flow did not significantly influence the plasma intensity and matrix tolerance, since it is the outer gas flow in the plasma torch. The auxiliary (intermediate) gas flow showed a similar tendency to the nebulizer gas flow, especially for 6Li and 209Bi (see Fig. S2†). However, its influence is less pronounced. Sheath gas flow is the nitrogen gas flow after nebulization, which surrounds the aerosol. Increasing the sheath gas flow rate and sampling depth could reduce the signal intensity (see Fig. 2c and d, S3 and S4†), possibly due to the reduced aerosol transport efficiency. Moreover, no improvement in the matrix tolerance could be observed after increasing these two parameters.
To investigate the influence of ion optics in MICAP-MS, the voltages applied to various ion optics were optimized and compared (see Fig. S5–14†). As examples, the results obtained with 6Li and 115In using different first lens and right mirror voltages are shown in Fig. 3. For 115In, the lens and mirror voltages had no significant influence on the matrix tolerance, whereas a noticeable improvement or deterioration effect could be observed for 6Li. A possible explanation is that due to the lighter ion mass, Li was more significantly affected by the inelastic collision with the N2 residual gas molecules behind the skimmer cone, reducing the number of ions transported to the mass analyzer. Therefore, optimizing ion optics could improve the ion transport efficiency and correspondingly reduce the influence of this matrix effect on light elements. Similar to ICP-MS, mass bias could also be observed in the ion optics of MICAP-MS, where the ions with elevated masses demonstrated shifted optimum voltages (see Fig. S15 and S16†).
When optimizing the parameters for plasma conditions and ion optics, the selection criterion was the average matrix tolerance for all six elements under investigation. In cases where matrix tolerance was not significantly affected, the criterion shifted to the average signal intensity. This approach represents a compromise. But the optimization of the average coincides, to a certain extent, with the optimization of element 115In, which is situated at the midpoint of the mass range.
Fig. 4 shows that despite the Fe-induced matrix effects, the recovery rates of all elements remained above 90% at an Fe concentration of 500 mg L−1. Even at an Fe concentration of 1 g L−1, the recovery rates remained above 80%, which demonstrates the strong matrix tolerance of MICAP-MS. Influences of the acid content on the matrix tolerance were also investigated by performing the measurements with HNO3 and HCl, whose concentrations range from 2% to 25%. It can be observed in Fig. S17† that high acid concentration leads to signal suppression for most of the elements in MICAP-MS. Both signal suppression and enhancement were observed for 6Li. This agrees with the theory described by Stewart and Olesik35,36 for Ar-based ICP-MS, which revealed that increasing the acid concentration could alter the aerosol evaporation, volumetric flux, and size distribution in the spray chamber. Consequently, the transport efficiency and the signal intensity could be reduced.
Fig. 4 Intensity recovery of 45Sc, 89Y, 115In, 159Tb, and 209Bi in 2% HNO3 with different Fe concentrations in relation to the IS solution with no matrix. |
It is noteworthy that we observed significant matrix deposition on the sample cone during our experiments. However, after an initial period of deposition, the signal stabilized, which can be attributed to the minimal matrix effect. A method focusing on higher signal intensity might reveal a more drifting signal as matrix deposition on the sample cone progresses. Upon cleaning the sample cone and allowing initial deposition to occur again, the signal reached a stable state again. Throughout the entire calibration and sample measurement campaign, no further cleaning was required.
(1) |
Elements | Reference | Aqueous calibration | Matrix-matched calibration | ||||||
---|---|---|---|---|---|---|---|---|---|
Mean | SM | Mean | SD | ζ-Score | Mean | SD | ζ-Score | ||
096-2 | Mn (%) | 1.320 | 0.012 | 1.330 | 0.016 | 0.5 | 1.335 | 0.016 | 0.8 |
Cr (%) | 0.0243 | 0.0009 | 0.0253 | 0.0001 | 1.1 | 0.0251 | 0.0001 | 0.9 | |
Mo (%) | 0.0020 | 0.0003 | 0.0020 | 0.0000 | 0.0 | 0.0019 | 0.0000 | −0.3 | |
Al (%) | 0.0460 | 0.0011 | 0.0456 | 0.0003 | −0.4 | 0.0469 | 0.0004 | 0.8 | |
Cu (%) | 0.0170 | 0.0005 | 0.0160 | 0.0005 | −1.4 | 0.0177 | 0.0004 | 1.1 | |
Nb (%) | 0.025 | 0.001 | 0.0252 | 0.0008 | 0.0 | 0.0240 | 0.0007 | −1.0 | |
179-2 | Mn (%) | 0.539 | 0.009 | 0.543 | 0.011 | 0.3 | 0.555 | 0.011 | 1.1 |
Cr (%) | 1.08 | 0.03 | 1.090 | 0.022 | 0.1 | 1.085 | 0.022 | 0.0 | |
Mo (%) | 0.070 | 0.006 | 0.0716 | 0.0009 | 0.3 | 0.0704 | 0.0009 | 0.1 | |
Cu (%) | 0.111 | 0.004 | 0.116 | 0.003 | 1.1 | 0.116 | 0.003 | 1.1 | |
V (%) | 0.188 | 0.007 | 0.206 | 0.005 | 2.1 | 0.200 | 0.005 | 1.4 | |
Ga (%) | 0.00129 | 0.00017 | 0.00128 | 0.00003 | 0.0 | 0.00100 | 0.00002 | −1.8 | |
Nb (%) | 0.00144 | 0.00018 | 0.00151 | 0.00002 | 0.4 | 0.00148 | 0.00002 | 0.2 | |
Sb (%) | 0.00175 | 0.00017 | 0.0020 | 0.0003 | 0.6 | 0.0025 | 0.0003 | 1.9 | |
286-1 | Mn (%) | 1.919 | 0.025 | 1.970 | 0.008 | 1.9 | 1.971 | 0.008 | 2.0 |
Mo (%) | 0.329 | 0.009 | 0.338 | 0.005 | 0.9 | 0.337 | 0.004 | 0.9 | |
Co (%) | 0.151 | 0.008 | 0.159 | 0.001 | 1.1 | 0.152 | 0.001 | 0.2 | |
Sb (%) | 0.0014 | 0.0004 | 0.0020 | 0.0002 | 1.3 | 0.0019 | 0.0002 | 1.1 | |
Sn (%) | 0.0084 | 0.0009 | 0.0092 | 0.0001 | 0.9 | 0.0093 | 0.0001 | 1.0 | |
Pb (%) | 0.00028 | 0.00013 | 0.00024 | 0.00001 | 0.3 | 0.00024 | 0.00001 | 0.3 |
The results obtained with aqueous and matrix-matched calibration for most elements were comparable and agreed well with the reference values. This indicates that no significant signal suppression or enhancement was observed during the measurements. However, the determined mass fraction of 51V in ECRM 179-2 was slightly higher than the reference value, which might have resulted from the polyatomic ion interferent 14N37Cl. Moreover, the LOD and LOQ determined with different calibration methods were in the same order of magnitude for most elements (see Table S3†). As the skimmer cone used is made of Ni, it was not quantified in the steel samples. Cr and W serve as the primary alloying elements in ECRM 286-1 and ECRM 179-2, respectively. The high contents of these elements are not in the working range of the method.
With optimized operating conditions, MICAP-MS showed high matrix tolerance, which achieved an intensity recovery of more than 90% at an Fe concentration of 500 mg L−1 and 80% at 1000 mg L−1. Like Ar-based ICP-MS,35,36 high acid concentration could result in intensity suppression in MICAP-MS, which might relate to changes in aerosol transport efficiency. The mass fractions of Al, Mn, Mo, Co, Cr, Mo, Cu, V, Ga, Nb, Sb, Sn, and Pb in the reference steel samples determined by MICAP-MS were comparable and compatible with the reference values. No significant difference was observed between the results obtained with aqueous and matrix-matched calibration. The LOD and LOQ determined with different calibration methods were also mostly in the same magnitude, which validates the excellent matrix tolerance of MICAP-MS. This study proves the applicability of MICAP-MS to element analysis in steels, with reduced matrix effect of Fe including primary alloying elements (Ni, Cr, Mn, W) with similar first ionization energy and atomic radii. The quantification covers a wide concentration range over five orders of magnitude from trace elements to minor components. Future research could expand on this work by exploring the matrix effect of easily ionizable elements such as Ca, Na, K, and Li in MICAP-MS, potentially broadening its applicability even further.