Owen T.
Butler
*a,
Warren R. L.
Cairns
b,
Jennifer M.
Cook
c,
Christine M.
Davidson
d and
Regina
Mertz-Kraus
e
aHealth and Safety Executive, Harpur Hill, Buxton, UK SK17 9JN. E-mail: owen.butler@hsl.gsi.gov.uk
bCNR-IDPA, Universita Ca’ Foscari, 30123 Venezia, Italy
cBritish Geological Survey, Keyworth, Nottingham, UK NG12 5GG
dUniversity of Strathclyde, Cathedral Street, Glasgow, UK G1 1XL
eInstitut für Geowissenschaften, Johannes Gutenberg-Universität, Becher-Weg 21, 55099 Mainz, Germany
First published on 15th December 2017
This is the 33th annual review of the application of atomic spectrometry to the chemical analysis of environmental samples. This update refers to papers published approximately between August 2016 and June 2017 and continues the series of Atomic Spectrometry Updates (ASUs) in environmental analysis1 that should be read in conjunction with other related ASUs in the series, namely: clinical and biological materials, foods and beverages;2 advances in atomic spectrometry and related techniques;3 elemental speciation;4 X-ray spectrometry;5 and metals, chemicals and functional materials.6 In the field of air analysis, highlights within this review period included the fabrication of new air samplers using 3D printer technology, development of a portable aerosol concentrator unit based upon electrostatic precipitation and instrumental developments such as a prototype portable spark emission spectrometer to quantify metal particles in workplace air. The advent of ICP-MS/MS systems has enabled analysts to develop improved methods for the determination of PGEs and radioactive elements present in airborne particles. With such instruments, the capacity to eliminate or minimise many isobaric interferences now enables analysts to forego the use of many onerous sample clean-up procedures. Improvements in the capabilities of aerosol mass spectrometers were noted as were developments in other complimentary measurement techniques such as Raman. In the arena of water analysis there are growing concerns regarding engineered NPs e.g. Ag NPs, entering water courses resulting in the development and optimisation of new methods based upon FFF and sp-ICP-MS techniques to measure such inputs. Similar concerns exist for MRI contrasting agents e.g. Gd-based compounds and here improved methodologies that involve the use of sample preconcentration using chelating columns and ICP-MS analysis have been proposed. In the field of plant and soil analysis, similar to developments in the water sector, there has been increased interest in the measurement of NPs. Many comparisons of sample digestion or extraction methods have been reported but a key issue rarely addressed is transferability, i.e. whether methods preferred by one group of researchers using particular apparatus are also optimal in a different laboratory using different apparatus. New sample preconcentration methods continued to appear although – as in previous years – the CRMs selected for method validation often failed to reflect the nature of the intended sample(s). A noteworthy advance is the use of HR-CS-ETMAS for elemental analysis. Developments in LIBS included greater use of TEA CO2 lasers in place of Nd:YAG lasers and increased use of stand-off measurement. The past year has also seen a rise in proximal sensing using LIBS and pXRFS. In the field of geological analysis, the quest continues for well-characterised matrix-matched materials suitable for the calibration of elemental and, particularly, isotopic measurements by microanalytical techniques. Increasing interest in stable isotope analysis by SIMS is reflected by the number of matrix-matched RMs developed specifically for this technique. Much work continues on ways of improving isotope ratio measurements by ICP-MS and TIMS for a wide range of different isotope systems relevant to geochemical studies. High spatial resolution analysis by LIBS, LA-ICP-MS and SIMS to obtain data on chemical and isotopic variations in minerals and biogenic materials in two and three dimensions are the foundation for many new insights in geoscientific research. In XRFS and LIBS, the advantages and limitations of portable instrumentation continue to be major focus of activity.
New approaches for assessing occupational exposure to airborne pollutants are most interesting to read. A small wearable aerosol microconcentrator developed by NIOSH researchers16 employed corona discharge to ionise sampled particles which were subsequently precipitated onto an electrode. It was considered that this approach could potentially be superior to alternatives such as filtration, focused impaction using aerodynamic lens or spot collection using condensational growth. They envisaged that this unit could form the basis of a new generation of compact hand-held aerosol analysers offering real-time in situ measurements. Analysis of exhaled breath condensate (EBC) offers a potential for a non-invasive method for detecting inhaled NPs. A study17 of methodologies for the preservation of inhaled welding fume in EBC concluded that flash-freezing was preferable to storing samples at room temperature for maintaining the original particle morphology. Spanish researchers18 provided evidence of the direct adsorption of Hg in human hair after analysing samples from individuals who were occupationally exposed.
The Beta Attenuation Monitor is a widely used ambient air particulate sampler which provides PM2.5 mass concentration data in real time, often at hourly intervals. Particles are collected and deposited as a spot on a filter tape and the attenuation of a radiation beam passed through this spot is exponentially dependent on the mass of particles deposited. Once a reading is taken, the filter spool is wound on to present a fresh surface for the next sampling interval. Researchers based in the US19 investigated whether it was possible to recover and analyse filter spots to provide additional time-resolved chemical data. The glass-fibre filter tapes currently used contain elemental impurities which limit the successful determination of many metals of potential interest. Another limitation was the inability to determine elemental C, a marker widely used for monitoring diesel engine emissions. The authors suggested that replacement with a cleaner Teflon filter tape could well be beneficial. Coincidentally, a new sampler incorporating a XRFS unit is now commercially available.
German researchers20 evaluated techniques for sampling volatile As species (AsH3 and methylated species) present in fumarolic gases and emitted from volcanoes. They noted that previously reported sampling techniques, such as needle trap devices and cryotrapping, did not perform well due to either poor collection efficiency or artifact formation during storage. Tedlar™ gas bags performed well as samplers, although 60 to 90% losses of methylated species remained possible if test items were stored for extended periods prior to analysis. The dominant species found in a study of Italian fumaroles was AsH3 suggesting that microbially catalysed methylation was of minor importance.
Sampling atmospheric trace species can be challenging. An investigation21 into memory effects of gold-coated sorbent tubes, typically used for sampling trace levels of atmospheric Hg0, concluded that repetitive thermal cleaning of the sampling tubes to remove stubborn adsorbed traces of Hg and hydrocarbons resulted in a two-fold improvement in LOD. The AirCore sampler is a balloon-mounted air sampler which collects a continuous air profile during descent through the stratosphere. A new sub-sampler, dubbed the SAS (Stratospheric Air Sub-sampler) was developed22 to extract aliquots of collected air from this AirCore device, for subsequent determination of 14CO2 by IRMS, whilst maintaining vertical height profile information.
A suite of isotopic CH4reference gases in air25 was developed and certified with uncertainties of <1.5% and <0.2‰ for δ2H and δ13C. Aliquots of these reference gases were diluted to provide CH4 standards with typical atmospheric concentrations and branded as JRAS-M16 (Jena Reference Air Set-Methane 2016). Calibration of Hg monitoring instruments typically relies upon gas standards prepared from the saturated head-space above a pool of liquid Hg kept at a constant temperature. New SI-traceable results26 for the Hg0 concentration at saturation in air between 15 and 30 °C, are welcome. An innovative automated temperature-controlled sampling system, constructed for this purpose, enabled defined aliquots of Hg-enriched head-space air to be collected at different temperatures. These were collected into a liquid medium, mixed with defined volumes of an isotopically enriched Hg solution and presented for ID-ICP-MS analysis.
Bioaccessibility measurements are conducted to ascertain what fractions of metals can potentially be solubilised within humans once inhaled or injested. The SBET (simplified bioaccessibility extraction test) and the stomach phase of the UBM (unified bioaccessibility method), originally developed for contamination land studies procedures, have been miniaturised29 for air filter applications. This was geared towards the analysis of filters employed in the TEOM air sampler, widely used within the UK and other air monitoring networks. This air sampler measures PM10/2.5 mass in real-time but filters are often then discarded so their further use was welcome. Bioaccessibility measurements of As, Cd, Cr, Fe, Mn, Ni, Pb and Zn in small mg quantities of particulate matter were undertaken. Contamination issues were overcome by prewashing syringe filters in dilute acid. Rasmussen and coworkers30 investigated the effect of pH, particle size and crystal form on the dissolution behaviour of Zn, ZnO and TiO2 engineered-nanomaterials in simulated body-fluids. They concluded that experimental pH and temperature conditions should be reported in all future studies so as to facilitate the comparability of published data.
Radionuclide measurements often require laborious and complex sample manipulation, separation, clean-up and preconcentration steps prior to analysis. Following the Fukushima Dai-ichi nuclear power plant incident, a Japanese consortium developed31 an ICP-MS/MS procedure for the simultaneous determination of 135Cs, 137Cs, 239Pu and 240Pu isotopes in suspended particles. Radiocesium and Pu were sequentially preconcentrated using ammonium molydophosphate and ferric hydroxide precipitation and then separated from matrix elements using a two-stage ion-exchange chromatographic procedure. Limbeck and his team32 improved the manipulation of radioactive particles by upgrading and automating a fission track method. They incorporated a laser micro-dissection system to facilitate the isolation of these particles of interest and developed a harvester unit to enable the collection and transfer of identified particles to other micro-analytical techniques. This new capability enabled U isotopic signatures in nuclear materials to be determined within a turn-around of 12 days. Researchers at the University of Erlangen33 evaluated various methods for sampling carbonyls in air for subsequent 14C dating using AMS. Two of these procedures utilised the Strecker synthetic method to form amino acids from carbonyls using either sodium cyanide or trimethylsilylcyanide and the third semicarbazide to form crystalline carbazones. All resultant adducts were separated and purified using TLC. The performance of these procedures compared favourably against the more conventional carbonyl sampling approach employing derivatisation with 2,4-dinitrophenylhydrazine and separation of adducts using HPLC.
A wide-ranging review paper34 (112 references) both outlined the various approaches used to speciate or fractionate Mn in a diverse range of matrices such as air, animals, plants and waters and summarised a number of useful extraction procedures.
A portable system37 for the on-line determination of Hg speciation in flue and process gases consisted of two AAS detectors in parallel, each hyphenated to a gas sampling system which used impingers to trap Hg species. In the first system, total Hg was measured by reducing all Hg species in the sampled gas stream to Hg0 using SnCl2 as reductant. In the second system, sampled flue gas was initially passed through a KCl solution to remove HgII and thereby allowing only Hg0 to be detected. The HgII species concentration in the sampled flue gas was calculated from the difference between the two results. Usefully, calibration units, that produced Hg0 and HgII gas standards, were also developed.
On-line measurements are most useful when attempting to elucidate the kinetics of chemical processes. In one example,38 an SMPS was coupled to an ICP-AES system for the real-time measurement of the elemental composition of size-resolved NP aggregrates synthesised within a diffusion flame reactor. There was a correlation between precursor molar ratios and elemental ratio in magnetite and SiO2 particles in the size range 50–140 nm. In a second example,39 an automated ICP-AES system for the continuous leaching of air filter samples to assess the bioaccessibility of Al, Ba, Cu, Fe and Mn in PM10 particles was developed.
Work developing new prototype instruments for in situ measurements continues. Reseachers at the University of Miami deployed40 their LIF instrument, designed for in situ trace Hg measurement in ambient air, during a recent instrumental comparison exercise held in Reno. At a temporal resolution of ca. 5 minutes, their Hg0 measurements were within 10–25% of measurements obtained by two, more conventional, AFS instruments equipped with gold-trap concentrator units. Comparative oxidised Hg measurements were also performed using KCl-coated annular denuders to preconcentrate air samples. The resultant Hg0 was determined following a thermal reduction step. Although results agreed with those measured by AFS instruments also equipped with KCl-denuders, measurement uncertainties were large. This was attributed to the challenges of quantifying a measurand at the pg m−3 level. Researchers41 at NIOSH developed a field-portable microplasma instrument for the real-time determination of carbon nanomaterials in workplace air. Their approach involved the preconcentration of aerosols on a microelectrode tip (see also Section 1.2) for analysis by spark emission spectrometry (SES) at the characteristic C line of 247.856 nm. The absolute LOD was 1.6 ng, equivalent to ca. 240 ng m−3 for a nominal 7.5 L air sample. This study demonstrated that SES could form the basis of a new type of portable aerosol analyser for in situ elemental analysis in the workplace.
An intriguing but perhaps complicated and not very portable approach44 for the direct analysis of mercury in air involved coupling a gas-to-particle conversion unit to an ICP-MS instrument. The process involved sampling air containing traces of Hg0 which were converted to HgO and subsequently agglomerated with NH4NO3 particles. These agglomerates were then introduced into the ICP-MS instrument through a gas-exchange device in which the matrix air was replaced with Ar sufficient to sustain the plasma. The method LOD of 0.12 ng m−3 was considered sufficient to measure airborne Hg concentrations of ca. 2 ng m−3 with a high temporal resolution without recourse to sample preconcentration. The usefulness of isotopes for elucidating the biogeochemical cycle of mercury is attracting widespread interest. In a study from Beijing,45 the Hg isotopic composition was determined both for particles from 30 potential industrial source materials and for those collected on 23 PM2.5 filter samples. The results were influenced by seasonal variations in both mass-dependent fractionation (represented by the ratio 202Hg/198Hg, δ202Hg) and mass-independent fractionation of isotopes with odd and even mass numbers (represented by Δ199Hg and Δ202Hg). All filter samples were highly enriched in Hg and displayed wide ranges of both δ202Hg (−2.18 to 0.51‰) and Δ199Hg (−0.53 to 0.57‰) as well as small positive Δ202Hg (0.02 to 0.17‰). Such variations were attributed to all-year-round anthropogenic emission sources such as smelting and cement production, as well as seasonal variations dominated by coal combustion in winter and biomass burning in autumn. The more positive Δ199Hg values, noted in air samples collected in the spring and early summer, were attributed long-range Hg emissions that had undergone extensive photochemical reduction. In another China-based study,46 the fractionation of Hg isotopes in particles arising from coal combustion and seawater flue gas desulfurisation (SFGD) processes was examined. Researchers employed MC-ICP-MS with a precombustion step for the elegant liberation of Hg from solid matrices. It was demonstrated that whilst stack emissions were enriched with lighter Hg isotopes, the SFGD end-product was enriched with heavier Hg isotopes. The authors concluded that this information could be most useful in tracing Hg emissions from such power plants. An interesting and more theoretical paper47 described the first measurements of isotope fractionation during the oxidation of Hg0 by halogen atoms. The study used a laboratory gas chamber to simulate photochemical reactions that occur in the atmosphere. A more experimental study48 was of the applicability of Tl as IS for mass bias corrections in the determination of Hg isotopes by MC-ICP-MS.
The 240Pu/239Pu; 241Pu/239Pu and 242Pu/239Pu isotope ratios in mixed-oxide particles were measured49 by ICP-MS following sample dissolution and separation of Pu from matrix using UTEVA™ resins. The 238Pu/239Pu ratio was calculated from the 238Pu/(239Pu + 240Pu) activity ratio measured by α-spectrometry and the 240Pu/239Pu ratio measured by ICP-MS. A method50 for the continuous determination of 239Pu in ice cores coupled both a SF-ICP-MS instrument and a continuous-flow ion analyser to an ice core melter. Advantages of this approach included less sample preparation and reduced analysis time. A simple, sensitive and robust method51 employed GC-MC-ICP-MS to determine 32S/34S isotopes in trace organic sulfur compounds (OSCs) found in sour CH4 gas streams. The GC system was equipped with a thermal conductivity detector to determine H2S and a fast valve switching system to enable a precise chromatographic heart cut to be taken, thereby preventing saturation of the ICP-MS system. Any matrix effects arising from high concentrations of CH4 (up to 95% v/v) and H2S (up to 10% v/v) in gases produced by the pyrolysis of sulfur-rich kerogen were minimised. The accuracy of this approach was better than 0.3‰ for OSCs at a concentration of 25 pmol (1.4 ppm).
The scanning mobility particle sizer (SMPS) provides size distribution and concentration of nanoparticles in air with a temporal resolution of a few minutes or less. As reported in last year’s update,1 there is ongoing interest in the benefits of coupling SMPS to ICP-MS systems so that complementary elemental data can be gathered. Swiss-based researchers52 determined the Cu, K and Na concentrations in aerosol particles (13 to 340 nm) released by the combustion of wood samples that had been impregnated with different metal salts. Use of a rotating disk diluter allowed the controlled dilution of particulate-laden air in order not to extinguish the plasma. A widely-used instrument for collecting size-segregated airborne particles is the micro-orifice uniform deposit impactor (MOUDI). Laser ablation53 was key to providing a fast and sensitive method for the direct elemental analysis of MOUDI filter samples of urban air. Filter sections were ablated and elemental image maps, created with ImageJ software, converted to concentration values using results from the parallel ablation of a calibrant filter (NIST SRM 2783 (air particulate on filter media)). For a nominal 4.32 m3 air sample (30 L min−1 over a 24 h period) the LODs (3SD) ranged between 0.003 ng m−3 for V to 0.370 ng m−3 for Fe. Agreements were good (R2 = 0.94 to 0.99, slope = 0.83 to 1.52) between data obtained for As, Fe, Mn, Pb, V and Zn using this new LA approach and data generated using a more conventional acid leach procedure with analysis by ICP-MS. The poorer agreement noted for Cu, Ni and Sb which was attributed to their low concentrations in air and hence greater uncertainties in the resultant measurements. The determination of isotopic ratios of elements within single particles by the interrogation of transient signals from LA-ICP-MS analysis can be time consuming as integration parameters need to be set manually on a peak-by-peak basis. A new approach,54 which relied on a point-to-point data reduction process and applied robust statistical estimator methods, was claimed to be a simple but robust alternative that did not require supervision.
Two review papers55,56 (20 and 146 references) usefully described advances in the application of sp-ICP-MS for the characterisation of nanomaterials.
A combination63 of TOF and SF SIMS techniques was used for the determination of both elemental and isotopic composition of particles in nuclear forensic samples. Qualitative elemental scans obtained by TOF-SIMS were useful for assessing isobaric overlaps on U isotopes. The U isotope analysis was performed using SF-SIMS in a molecular shift mode. Measuring UO2+ species rather than U+ minimised the isobaric overlaps identified previously. A new algorithm64 enabled better identification of isotopically unique U-containing particles than possible with the image data processing techniques currently used in SIMS analysis. Images from test samples (composites of NIST SRMs) were successfully evaluated. Fusion65 of SIMS images with elemental maps from EDS facilitated, for the first time, the location of isotopically unique U-bearing particles in powdered samples. A simulant sample, consisting of NIST SRMs (U particles) dispersed in a river sediment (NIST SRM 8704 (Buffalo river sediment)), was used to evaluate the utility of this approach. The Cameca NanoSIMS 50 ion probe66 was capable of measuring accurately species-specific, stable N-isotope ratios in bulk NaNO3 and KNO3 powders deposited on a gold substrate. Typical within-run precision was ±1.3‰ and the accuracy for long-term measurements on an in-house NaNO3 standard was ±1.9‰ for a raster size of 5 × 5 μm2. The authors suggested that this approach would be useful for determining the isotopic composition in single μm-sized nitrate particles in order to understand better the fractionation processes that can occur when NO molecules react on the surfaces of sea-salt and dust particles.
A method67 for the simultaneous determination of both concentration and isotopic composition of U in individual micro-particles by ID-TIMS was validated by the successful analysis of NIST SRM U030, either as particles or as digests. Results were in good agreement with certified 235U/238U and 236U/238U values. An automated particle-screening software package,68 originally developed for SIMS analysis, was applied to highly precise TIMS data to identify specific U particles. The advantage of TIMS analysis was that molecular- and hydride-derived isobaric inferences were minimised.
Improved isotope ratio mass spectrometry instrumentation included the new automated system69 for the simultaneous determination of both δ2H and δ13C and δ13C and δ18O in atmospheric CH4 and CO2 gas samples, respectively. Air samples collected in flasks or cylinders were split into two streams and introduced simultaneously into two IRMS instruments. Analysis of 16 samples which included 4 reference and 1 QC air sample took 36 h. Long-term precisions, derived from the repeated analysis of a QC air sample since 2012, were: 0.04‰ (δ13C of CO2); 0.07‰ (δ18O of CO2); 0.11‰ (δ13C of CH4) and 1.0‰ (δ2H of CH4). Within a single day, the system exhibited a typical methane δ13C repeatability standard deviation (1σ) of 0.06‰. The high throughput and reliability of operation were important for coping with the large number of air samples anticipated from ongoing atmospheric green-house gas monitoring programmes. An improved HR-MC-IRMS method70 for measuring six singly- or doubly-substituted isotopic variants of N2O required only 10 μM of sample, took only 8–10 h per sample and offered a powerful new analytical capability for identifying the environmental sources and sinks for this important atmospheric molecule. A new instrumental setup71 combining LA nano-combustion-GC with IRMS enabled δ13C measurements to be performed on samples such as pollen that contained only ng quantities of C. Analysis of the IAEA CH-7 polyethylene standard had good accuracy and precision of <0.5‰ on sample masses as low as 42 ng C. This level of performance was sufficient to discriminate between pollen species.
A round-up of X-ray applications included the assessment75 of trace metals (20 elements) in Parisian air through the analysis of dust deposited on moss samples growing in cemeteries. Use of a slurry suspension-TXRFS methodology avoided a digestion step, reduced the risk of sample contamination and worked well with small sample quantities. Work focused on utilising Grimmia pulvinata, a moss species common in the 21 cemeteries studied and commonly found across Europe. The oxidation state of S in atmospheric particulate matter76 was investigated using a combination of NEXFS and XRF microscopy. Study of oxidation states S0 and SVI led to the finding that the source of the SVI species was either ammonium sulfate from the reaction of NH3 with SO2 or metal sulfates from industrial emissions. The S0 species most likely originated from primary emission sources such as bacteria or incomplete combustion. An assessment of worker exposure,77 conducted for the first time at a nuclear fuel fabrication plant, involved the sampling of U-containing particles using a cascade impactor and elemental characterisation using SEM-EDS. An eight-stage Marple impactor, with particle cutoff points of 21.3, 14.8, 9.8, 6.0, 3.5, 1.6, 0.9 and 0.5 μm, was used. Carbon tape was employed as an impaction substrate as its conductivity was necessary for SEM analysis. The extent of any bias in the EDS measurements was assessed by analysing standard test samples, prepared by depositing aliquots of NIST SRM U-010 (high purity U3O8 particles) onto this carbon substrate. A bias in U mass measurements of up to 15% was attributed to beam interactions with oxygen in this substrate material which resulted in an over-estimation of the oxygen content in the uranyl particles interrogated. The performance of a Rietveld XRD procedure78 for the quantification of crystalline phases in fly ash was optimised after careful assessment of various experimental conditions: internal standards (SiO2 or TiO2, spike dosages in the range 10–50% m m−1); incident X-rays (laboratory or synchrotron sources) and software (GSAS or TOPAS programmes). The procedure gave good reproducibility with arithmetic mean errors and the standard errors of identified main phases as low as ca. 1%.
Thermal-optical analysis for the determination of carbonaceous content of atmospheric particles is well established but differences in protocols have led to discrepancies in reported elemental C data. Researchers in Hong Kong81 used the same instrument to analyse 1398 filter samples by both the NIOSH TOT (thermal optical transmittance) and the IMPROVE (Interagency Monitoring of Protected Visual Protected Environments) TOR (thermal optical reflectance) protocols. About 80% of this discrepancy was attributable to a difference in the peak inert mode temperature employed and the remaining ca. 20% to a difference in the optical method (TOT vs. TOR). Fuller et al.82 compared light absorption of PM10 particles collected on four different filter substrates (Emfab™ filters (Teflon-coated glass-fibre filter), mixed cellulose ester filters, quartz fibre filters and Teflon filters) by both reflectrometry and transmissometry techniques. They established relationships between these absorption measurements and elemental C concentrations derived from thermal-optical analysis. There were strong linear relationships between reflectrometry and transmissometry data for each filter type but absorption characteristics differed. This illustrated the complexity of light scattering and the loading effects of particles collected on various filter media. The relationship between elemental carbon values and light absorption was non-linear regardless of filter type. Real-time optical black carbon instruments, used for regulatory measurements of particle matter emissions from aircraft turbine engines, require calibration against elemental C mass measurements derived from thermo-optical analysis. An international consortium83 evaluated the response of two black carbon instruments to surrogate soot particles generated using a mini-CAST soot generator. A key finding of this study was that, in addition to defining the elemental C content as currently undertaken, there is a need going forward to define the particle size distribution of any challenge soot used.
A feasibility study84 demonstrated that Raman microscopy could quantify <1 μg of respirable-size crystalline silica collected on an air filter sample. The LOD of ca. 0.2 μg for both quartz and cristobalite forms was ca. 10-fold lower that currently achievable using FTIR or XRD techniques. Signal attenuation due to matrix species such as haematite, as found in foundry dusts, remains an issue however and this new approach may well therefore be best suited for measuring low concentration of silica in air in those workplaces where other matrix species are not present to any significant extent. As the global production of carbon nanotubes (CNTs) increases, concerns grow about their safe handling during production and use. Raman spectroscopy85 unequivocally identified that airborne particles settled on a work surface contained CNTs. Furthermore, the technique could be used to ascertain whether the CNTs were synthesised by an electric-arc method or by a chemical vapour deposition method.
A round-up of instrumental developments and applications included a new IMS-TOF instrument86 for the determination of water-soluble organic C and organosulfate species in aerosol samples and a review paper87 (112 references) detailing the current status and application of IMS for the detection of chemical warfare agents. Cavity ring-down spectroscopy offers a growing potential for the ultra-sensitive and real-time detection of trace gas species, as exemplified by studies88 which measured NO2 at a concentration of 38 pptv and measured8914CO2 at <10 ppqv!
The most significant advances in analyte preconcentration for water analysis are summarised in Tables 1–3. It should be noted that the last entry in Table 1 (ref. 99) is for a passive sampler and the figures of merit are dependent on deployment time.
Analytes | Matrix | Substrate | Coating or modifier | Detector | LOD (in μL−1) unless stated otherwise | Validation | Reference |
---|---|---|---|---|---|---|---|
AlIII, CuII, FeIII, MnII, PbII, ZnII | Natural waters | Silica | Polyhexamethylene guanidine and 4,5-dihydroxy-1,3-benzenedisulfonic acid | ICP-AES | 0.02 (CuII, FeIII, MnII, PbII, ZnII) and 0.15 (Al) | Spike recovery | 105 |
Au | Environmental water | Magnetite | Graphene oxide | MP-AES | 0.005 | Spike recovery | 106 |
Bi | Sea, lake, river, stream and rain waters | Halloysite nanotubes | HR-CS-ETAAS | 0.005 | NIST SRM 1643e (trace elements in water) and NWRI CRM TMDA-54.5 (a high level fortified sample for trace elements) | 107 | |
CdII, CoII CuII, FeII, NiII, MnII, ZnII | Tap and mineral waters | Amberlite® XAD-16 | 2-(2-Benzothiazolylazo)-5-dimethyl aminobenzoic acid | FAAS | 0.71 (Cu) to 3.55 (Cd) | NIST SRM 1643e (trace elements in water) and spike recovery | 108 |
Cr, Cu, Zn | Water | Graphene oxide | Glycine | ED-XRFS | 0.07 (Zn) to 0.15 (Cr) ng mL−1 | Spike recovery | 109 |
Ce, Dy, Gd, La, Nd, Sm, Y | Well water | MWCNTs coated cellulose acetate membrane | 1-(2-Pyridylazo)-2-naphthol | ICP-AES | 0.05 (La) to 0.3 (Nd and Ce) | Spike recovery | 110 |
REEs, U | Lake, well and tap waters | Cotton roll | 1,2,5,8-Tetrahydroxyanthraquinone | ICP-AES | 0.002 (La) to 0.420 (U) | Spike recovery | 111 |
Sb (total), SbIII | Bottled water | Fe3O4 NPs coated with Ag0 | Sodium 2-mercaptoethane-sulphonate | HR-CS-ETAAS | 0.02 | NIST SRM 1640a (trace elements in water), NWRI CRM, CRM TMRain-04 (simulated rain water) | 112 |
Ag nanoparticles | Lake water | MWCNT passive sampler | ICP-MS | 45 nm (LOD for Ag not reported) | Spike recovery | 99 |
Analytes | Matrix | Method | Reagents | Detector | LOD in μg L−1 (unless stated otherwise) | Method validation | Reference |
---|---|---|---|---|---|---|---|
Ag, Cd, Ir, Os, Pd, Pt, Re, Ta | Water | DLLME | Chloroform and Aliquat®336 (tricaprylmethylammonium chloride) | ICP-MS | 0.04 (Pt) to 0.3 (Ag) | Spike recovery of inorganic ventures CMS-2 (certified precious metal standard) and comparison with SF-ICP-MS results | 113 |
Be, Cd, Cr, Hg, Pb, Tl | Natural and waste water | CPE | Aliquat®-336 (N-methyl-N,N,N-trioctylammonium chloride) and nonionic, Triton™ X-114 | CS-ETAAS | 0.01 (Be) to 0.3 (Hg) | NIST SRM 1640a (natural water), IRMM CRM BCR-713 (final effluent) and IRMM CRM BCR-715 (industrial effluent) and spike recovery | 114 |
Cd, Co, Cu, Ni, Pb, Zn | Water | CPE | 1,5-Di-phenylthiocarbazone (dithizone) and Triton™ X-114 | USN-ICP-AES | 0.009 (Cd) to 0.02 (Zn) | Spike recovery | 115 |
CrVI | Tap, waste and mineral waters, air, blood and urine | SFOD-LLME | Sodium dodecyl sulfate, 1,5-diphenylcarbazide, and 1-undecanol | ETAAS | 0.003 | NIST SRM 1640a (natural water) | 116 |
In | Water | DLLME | 1-(2-Pyridylazo)-2-naphthol, undecanol | ETAAS | 55.6 ng L−1 | Spike recovery | 117 |
Analytes | Matrix | Method | Substrate | Reagents | Detector | LOD in μg L−1 (unless stated otherwise) | Method validation | Reference |
---|---|---|---|---|---|---|---|---|
Ag | Seawater | SBME | Capillary hollow fibre stir bar | Aliquat® 336 in kerosene solution and dodecan-1-ol | ETAAS | 0.09 ng L−1 | Spike recovery and comparison with DLLME method | 118 |
As | Water | DME | MWCNT | Diethylenetriamine | HG-AAS | 3.05 ng L−1 | NRCC CRM SLRS-5 (river water) and NWRI CRM TM-28.3 (Lake Ontario water) | 119 |
Ni | Sea and river waters | Magnetic SLME | Graphene oxide | 1-Hexadecyl-3-methylimidazolium chloride | FAAS | 0.16 | Spike recovery | 120 |
All the single element speciation papers covered waters as well as other matrices. A review of antimony speciation in environmental samples102 (41 references) discussed how the redox chemistry of Sb plays an important role in the chromatographic separation of its species and how complex formation can cause artefacts during extraction procedures. The challenges and significance of Mn speciation in various matrices was reviewed34 (112 references) with a focus on the determination of the bioavailability or bioactivity of Mn in the environment. A review103 (65 references) of extraction methods for either CrIII or CrVI reflected the abundance of methods in the literature. The flow and miniaturisation systems used to automate the analyses were also covered.
Nanomaterials are now being exploited as stationary phases for speciation analysis such as the elemental speciation of environmental, food and biological samples covered in a 162-reference review.104
Methods for non-chromatographic elemental speciation abound. Those methods in which only one form of an analyte is preconcentrated and the concentration of another deduced by difference are included in Tables 1 to 3 This section covers methods in which more than one form of an analyte was speciated. In one such paper,122 AsIII and AsV were preconcentrated simultaneously on 40 mg of metallic iron NPs that had been coated with SiO2 and functionalised with polyethyleneimine. The As species were preconcentrated from 300 mL of sample with a recovery >80%. The NPs were separated magnetically from the sample and the analytes extracted with 4 mL of a solution containing 1% m/v thiourea and 5% v/v HCl prior to detection by HG-AFS. The LOD was 0.002 μg L−1 as As and the precision for each species was 1.95% RSD for AsIII and 2.55% RSD for AsV (n = 6). Mixed micelle DLLME123 was used to extract CrVI and CrIII sequentially from natural waters and effluents. Samples of 1–10 mL were adjusted to pH 2 and CrVI extracted with 1 mL of a 10% m/v solution of the cationic surfactant Aliquat-336 with 0.6 mL of trichloroethylene added as the dispersive solvent. The organic-rich upper layer was removed and dissolved in 0.2 mL of methanol containing 2% v/v HNO3. The pH of the remaining sample was readjusted to pH 2 if necessary and the CrIII extracted with 2 mL of a 10% m/v sodium dioctyl sulfosuccinate solution after the addition of trichloroethylene. The trichloroethylene-rich layer that formed was treated in the same way as the first extract. Both extracts were analysed directly by ETAAS. The LODs were 0.5 pg mL−1 for CrIII, 0.6 pg mL−1 for CrVI and 0.7 pg mL−1 for total Cr with a maximum RSD of 6% for 10 replicates within the linear range of 0.02 to 1 ng mL−1. The method was validated against the IRMM CRMs BCR 714 and 715 (industrial effluent) and the NIST SRM 1640a (natural water). The total Cr values and the sum of the concentrations for the two species agreed with the certified values for the total Cr concentration. A field method for trapping VIV and VV involved124 the use of Agilent Bond Elut (500 mg bed) strong anion-exchange cartridges preconditioned with 0.2 M Na2EDTA to trap VIV and VV from 10 mL filtered water samples. Subsequently VIV was eluted in the laboratory with 15 mL of a solution of 5 mM Na2EDTA, 2 mM tetrabutylammonium hydroxide and 4% v/v methanol at pH 4. The cartridge was dried by passing air through then Vv eluted with 15 mL of 80 mM dihydrogen ammonium phosphate. The V content of the eluents was determined by ICP-MS with a method LOD of 0.05 μg L−1 for V. In the absence of suitable speciation RMs, the total V values in the NIST SRM 1640a (natural water) and the NWRI CRM TMDA-54.2 (trace metal fortified water) were compared with those obtained using HPLC-ICP-MS. The species could be preserved on the column for up to 15 days. Comparison of results obtained by this in situ trapping method with those obtained for water samples taken and analysed back in the laboratory revealed that the vanadyl species was probably converted into a vanadate species during transport.
One of the most widely used methods for NP counting and sizing at environmentally relevant concentrations is the use of single particle counting ICP-MS. An attempt was made126 to formalise a systematic and reproducible approach for the determination of NPs in environmental samples, since they are more complex than mixtures of pure water and NPs. For a soil extract spiked with 30 nm Ag NPs at a concentration of 5.05 μg L−1, the mean particle size was measured as 68.2 nm, but diluting the sample to near the method LOD (0.05 μg L−1) resulted in a measured mean particle size of 37.4 nm. A further two-fold dilution was used to confirm the result. It has been suggested127 that dilution followed by analysis may not be suitable for seawater samples as the behaviour of NPs in dilute seawater is different to that in undiluted seawater. An alternative approach was the use of FI sample introduction with on-line dilution to lessen matrix problems as well as to reduce the time for which the sample remained diluted prior to analysis to <100 ms. Using a 20 μL sample loop, a carrier flow rate of 10–50 μL min−1 and a capillary inserted into the nebuliser, introduction of a sample at 0.8 mL min−1 and 5 mm from the tip of the nebuliser was possible. Matrix effects were reduced to the point where the intensity for NPs in seawater was up to 80% of that in deionised water. The size LOD of 18 nm in deionised water and 25 nm in seawater was sufficient to follow the dissolution dynamics of engineered NPs spiked into seawater at a concentration of 200 ng Ag L−1, the equivalent of 1.7–2.1 × 105 particles per mL for 60 nm Ag NPs. Another advantage of using FI-sp-ICP-MS128 was that the calculation steps were less complicated as known volumes were injected and so the peak area was directly proportional to the mass of NPs injected if the instrument had been previously calibrated against a NP standard. Therefore it was not necessary to maintain a constant sample uptake rate, so the flow fluctuations caused by the matrix were less important. The size LOD was 20 nm for an instrument dwell time of 5 ms and a 200 μs settle time. This was the same LOD as obtained using sp-ICP-MS.
Despite the flexibility of sp-ICP-MS, coupled techniques are still used, some even with detection in sp mode. The coupling of asymmetric flow FFF with sp-ICP-MS overcame129 some of the shortcomings of the two techniques. The limitations of sp-ICP-MS in the presence of a concentrated dissolved fraction were removed because the dissolved fraction was eliminated to waste in the asymmetric flow. Using a 1.5 mL min−1 channel flow rate, the FFF separated a mixture of Ag NPs and Ag–SiO2 by their hydrodynamic diameters within 30 minutes. A limitation of asymmetric flow FFF analysis is that the mass concentration is based upon the peak height but the particle number concentration is calculated by assuming all the NPs detected are spherical and of the same composition. This assumption breaks down when the sample content is a mix of coated and uncoated samples. With sp-ICP-MS as the detection method, each particle was an integration event so the particle number concentration was counted. From the difference in individual signal heights it was possible to distinguish between Ag and Ag–SiO2 NPs in deionised water with a background Ag+ concentration as high as 9.65 μg L−1. To determine TiO2 NPs in lake water,130 sedimentation field FFF combined with ICP-MS/MS was used to overcome the interferences that affect Ti determination by ICP-MS. The separation was optimised using two TiO2 standards (21 and 50 nm) at a flow rate of 0.5 mL min−1 and with a sedimentation field of 1000 rpm. The instrumental LOD for Ti+ was 10 ng L−1 using NH3/He as the collision/reaction gas at 2 mL min−1. The method LOD for Ti NPs when using sedimentation flow FFF-ICP-MS/MS was 6.8 μg L−1. This was sufficient to detect Ti NPs in lake water. Analysis of a 500 mL lake water sample before and after preconcentration by centrifugation at 10000 rpm for 2 h gave very similar size distributions (75–400 nm). Sunscreens or remobilised lake sediments were considered to be potential sources of the Ti NPs detected. Unfortunately, neither the mass of Ti found nor the particle count per mL was reported so comparisons with other methods was not possible.
The selective detection of CrVI in water used132in situ selective thermal separation and ETAAS analysis. The CrIII present in the sample was complexed with 2-thenoyltrifluoracetone and the resulting product vaporised at 400 °C during the ashing step. This left only CrVI present at the 1200 °C atomisation step. The CrIII concentration was determined by difference between the total Cr and CrVI values. The LOD was 0.046 μg L−1 for CrVI and the analytical precision 3.1% RSD (n = 5) at a concentration of 1 μg L−1. The method was validated with the Chinese CRMs GBW(E)080403a and GBW(E)080404b (chromium in simulated water) and results for bottled water samples compared with those obtained by IC-ICP-MS analysis.
A photochemical vapour generation method using ICP-MS as detector134 was developed for the determination of Os in natural waters. When the sample was acidified to 5% v/v HNO3, volatile OsO4 was easily produced under UV irradiation. To improve sensitivity without increasing the blank, the irradiation time was increased by using a stopped-flow FI system. The optimum irradiation time of 150 s had the added advantage that the irradiation photo-reduced the sample matrix and heated the sample to 90 °C, thereby improving the separation of OsO4 from the sample in the GLS. Memory effects from residual OsO4 were almost eliminated by using a 5% v/v NH4OH solution as the carrier. This was a big improvement over the use of 5% HNO3 as carrier when the signal from the preceding sample dropped by only 80% in the wash out time of 60 s. The LOD was 0.8 pg mL−1, the precision 3.2% RSD (n = 5) at a spike concentration of 1 ng mL−1 and the recoveries 97% (river water) to 109% (seawater). The swift and quantitative liberation135 of Hg from water samples (from a water purification plant and with concentrations of up to 100 μgL−1) in a GLS was made possible by the addition of 10% v/v formic acid and use of a short 10 s UV irradiation pulse. An AFS system was then used to determine Hg with analytical LOD of 0.003 μg L−1. The LOD for Hg with ICP-AES detection was reduced136 to 90 pg mL−1 by adding 15% v/v formic acid to the sample and irradiating with an 11 W UV-C lamp prior to pneumatic nebulisation. All the Hg was converted to Hg0 in a spray chamber held at 40 °C and no negative effects were seen for the simultaneous detection of the non-volatile analytes (Cu, Fe and Mn). The method was validated with the IRMM CRMs ERMCA713 (waste water) and ERCCC580 (estuarine sediment).
The problems associated with B analysis in water have been rediscovered.138 Preservation of samples and immediate analysis (within 24 h) after sampling were essential. In the ICP-AES analysis, the Fe and Cr emission lines close to the main B emission line could present problems. Filtration would reduce the measured B concentrations at low actual concentrations but increase them at high actual concentrations. When appropriate care was taken, results within 5% of the assigned values of proficiency scheme samples could be obtained. Strangely, no mention was made of known B memory effects within sample introduction systems, or of any attempts to reduce them.
A commercially available system with FI and preconcentration on an IDA column resin was used off-line to prepare seawater samples for REE analysis.139 Using ID-ICP-MS analysis, REEs were successfully determined in only 11 mL of seawater with a procedural long-term error of ≤3.9% 2σ RSD (n = 16) for all REEs except Ce and Gd. Results for measured REE concentrations in the samples from an international intercalibration (Geotraces) were within 7% (2σ RSD) of the consensus values except for Ce (71%), Gd (14%) and Lu (12%). The REE concentrations in Brazilian rivers were determined140 in filtered samples, with and without preconcentration by SPE with bis-(2-ethyl-hexyl)-phosphate. This preconcentration approach was unable to retain Gd complexes, suspected to be present in discharges from hospitals that use MRI contrasting agents. Such complexes only dissociate slowly in acidic water to produce the free ion which is retained by the phosphate ester. Strong correlations with parameters indicative of industrial effluents and recycled water (B, NH4OH and DOC) seemed to confirm the finding that this Gd source was indeed anthropogenic in origin. A anthropogenic input was also observed in seawater samples from San Francisco Bay.141 The Gd levels in the southern part of the Bay had risen more than 13-fold from 8.27 pmol kg−1 in 1993 to 112 pmol kg−1 in 2013. It was suggested that preconcentration with the NOBIAS PA-1® 20 resin followed by SF-ICP-MS detection could be used to monitor increases in the environmental concentrations of Gd and other exotic trace elements in waste streams from hospitals and technological industries.
The quantification and fractionation of fine particulate carbon in stream water was achieved by coupling asymmetric flow FFF with SF and quadrupole ICP-MS.142 Using an optical carbon detector as a reference method, recoveries were 122 ± 22% and the LOD 0.3 mg L−1. The latter was poorer than that (0.04 mg L−1) achievable with a dedicated carbon detector. The particulate carbon concentrations in a fresh water sample and a bovine serum albumin candidate RM were determined. If FFF-quadrupole-ICP-MS were used with FI sample introduction, recoveries could be improved to 90–113%.
Accurate isotope ratio measurements by ICP-MS typically require laborious preconcentration and matrix separation steps before analysis. To overcome these problems, IC-MC-ICP-MS was investigated for Mg143 and Sr144 isotope ratio analyses. Using an IonPac CS16 cation-exchange column (5 × 250 mm, 5 μm particle size) with corresponding guard column and suppressor unit at a flow rate of 1 mL min−1, the Sr was isolated from the matrix and determined within 14 minutes with a precision that was worse than that of traditional off-line methods but within the uncertainty of the certified value of the NIST SRM 987 (strontium carbonate isotopic standard). For Mg, a narrower (3 mm id) version of the same column was used so that the flow rate (0.36 mL min−1) was compatible with the desolvation system. A good precision of ±0.15‰ (2SD) on 26Mg/24Mg isotope ratios in matrix-rich natural waters was achieved with an analysis time of 18 minutes per replicate, comparable to the precision obtainable with off-line matrix separation methods. The use of DGT samplers, typically used to obtain time-weighted average concentration in bodies of water, was tested145 under controlled conditions in a clean room to ascertain whether accurate Pb and Zn isotope ratios could be obtained. Use of Chelex®-100 resin as the accumulation layer introduced no fractionation of Pb isotopes but fractionation of Zn isotopes did occur. A correction factor was applied to account for the exposure time and thickness of the diffusion layer. Photographic black and white film was used146 as a DGT analogue for in situ sampling of S in sediment pore water in order to obtain a 2D (22 × 3.5 mm) map of the δ34S distribution. Photographic film disturbed the sediments less and had a 103 superior binding affinity for S than typically used DGT binding matrices. Interrogating the film with a LA-MC-ICP-MS instrument allowed a high spatial resolution (102 μm) map of δ34S in the pore waters to be obtained with a reasonable accuracy (<0.9‰, 2σ) and external precision (<0.8‰, 2σ) across an environmentally relevant concentration range (0.01–20 mM).
Typically, trace elements in water need to be preconcentrated onto or into a solid matrix before LIBS analysis. One method148 used a magnesium alloy to trap Cd, Cr, Cu, and Pb into a thin surface layer. The linear working range was 0–7 μg mL−1 for a mixed element standard (except for Cu which gave a curved response at all concentrations) and the LODs were between 0.025 (Cr) and 0.420 (Cd) μg mL−1. Alternatively,149 the target trace elements could be incorporated into a poly(vinyl alcohol) (PVA) membrane. The PVA solution was cast into a Petri dish with 10 mL of the sample or standard so that the elements were incorporated directly into a 0.34 mm thick polymer film once dried. The LODs were 1 ng mL−1 (Ag) to 5 μg mL−1 (Cd) when using a series of “differently linear” calibration curves. A sample preparation time of 90 minutes for the curing and drying of the polymer is a limitation on the utility of this method for the moment and work on improvements continues.
At the other end of the portability scale, a sensitive TXRFS method was developed152 at the Shanghai synchrotron facility. Using a 100 s read time, the LODs were 0.03 (Zn) to 0.11 (Cr) pg for a 10 μL sample deposited as a 100 μm diameter spot size on a conditioned Si wafer. Results for waters spiked with 100 and 500 μg L−1 Co, Cr, Cu, Mn, and Ni agreed within 10% of results obtained by ICP-MS.
Two review articles concerning the determination of radionuclides were of interest. The first154 (109 references) focused on the fission products 135Cs, 137Cs and 90Sr, with particular reference to analyses using AMS, ICP-MS and TIMS. The second155 (155 references) provided a comprehensive overview of the status of radiochemical analysis in Nordic countries.
Other notable reviews discussed: the use of elemental and molecular mass spectrometry in rhizosphere research156 (118 references); techniques and strategies for quantifying and mapping chemical elements in single cells157 (103 references); methods for separation and characterisation of environmental NPs and sub-μm particles158 (97 references); and progress and challenges in the application of (predominantly atomic-spectrometry-based) analytical techniques in archaeological soil analysis159 (81 references).
Articles featuring the development of analytical procedures specifically for use in the certification of new CRMs included a detailed description165 of an ID-ICP-MS procedure that contributed Cd, Cu, Hg, Pb and Zn mass fraction values for IAEA 458 (marine sediment). The evaluation166 of CV-AAS and ICP-MS methods included application to four candidate INCT CRMs, one of which was a sediment.
Several improved methods for radionuclides were reported. Researchers in Japan172 recommended CaF2/LaF3 co-precipitation followed by sequential application of three commercial resins – TEVA, UTEVA, DGA – for removal of interfering elements in their method for determination of Pu in soils and sediments by SF-ICP-MS. Members of the same team developed a method173 for ultratrace determination of 241Am in soil and made a detailed study174 of TRU and DGA resins for interference removal. Other authors175 reported an improved version of their automated anion-exchange procedure for sequential separation of La, Pb, Pu, Th and U and its application to tree ring samples. A novel multi-step procedure176 to study the distribution of 99mTc in seaweed by ICP-MS separated the total content into eight fractions. The majority of Tc was found as TcO4− or in water-, acid- and alkali-soluble forms (15%, 29%, 14% and 25% of the total content, respectively) with the remainder bound to pigment (3%), algin (11%) and cellulose (3%). No Tc bound to water-soluble protein was detected. Extraction procedures specifically designed for use with ICP-MS/MS included a simple method177 for purification of soil digest prior to measurement of U and 236U/238U ratios and a procedure178 for ultratrace determination of 135Cs/137Cs in environmental samples at global fallout levels. The latter procedure may have the potential to rival analysis using TIMS.
Losses due to venting during microwave-assisted extraction with hot mineral acids did not always affect179 the accuracy of concentrations measured by ICP-AES or ICP-MS, even for relatively volatile analytes such as As. The authors suggested that condensation promoted by the temperature gradient between the bottom and top of the (tall) digestion vessels aided analyte recovery. A MAE with aqua regia followed by addition of 0.1 M NH4SCN prevented the loss of I both during mineralisation and in the spray chamber when analysing biota (including seaweed) by ICP-MS.180 A hotplate digestion181 with HCl (2 h at 150 °C) avoided formation of insoluble precipitates that led to inaccuracies when MAE with HCl–HNO3–HF was applied to carbonate-rich sediment prior to analysis by ICP-MS.
The feasibility of microwave-induced combustion to digest soil with high inorganic matter content for determination of halogens was demonstrated182 for the first time. A 100 mg sample was mixed with 400 mg of microcrystalline cellulose to aid combustion, placed in a quartz cell filled with 20 bar of O2 and ignited. The liberated halogens were trapped in 6 mL of 100 mM NH4OH solution. Following dilution, Cl and F were measured by IC and Br and I by ICP-MS. Spike recoveries were 94–103%.
An ultrasound assisted extraction183 for measurement of I in food by ICP-AES involved sonication of 0.1 g samples in 10 mL of deionised water for 5 minutes and gave a recovery of 98% for analysis of NIST SRM 1573a (tomato leaves). A UAE for soil184 using a deep eutectic solvent (choline chloride/oxalic acid in a 1:2 molar ratio) was optimised by two-level full factorial design using NIST SRM 2709 (San Joaquin soil). When sonication of 0.1 g test samples for 10 minutes in 10 mL solvent was followed by a 5 minute period of further heating in a block digestor at 120 °C, recoveries by ICP-AES were similar to certified and indicative values for As, Cd, Mo, Sb, Se and V. Sonication in a mixture185 of 5 mM 1-butyl-3-methyl-imidazolium tetrafluoroborate ionic liquid, 0.005 M EDTA and 0.1 M NaHCO3 for 7 minutes gave performance similar to that obtained with 6 h conventional shaking in 0.05 M EDTA solution for determination of available Cd, Cu, Ni, Pb and Zn in sediment by FAAS. In another sediment study186 sonication in 4 M HCl and 0.75% (m/v) thiourea for 20 minutes were optimal for extraction of Hg, giving results close to target values for analysis of IRMM CRM BCR 320 R (channel sediment) and NIST SRM 2709a (San Joaquin soil) by CV-AAS.
Comparisons of extraction methods for soil and sediments included a study187 of the effects of soil properties on the efficiency of extraction of roxarsone and its metabolites by different reagents for determination by HPLC-ICP-MS. Soil pH, CEC and Fe content all affected the extraction of As species but satisfactory performance was obtained with 9:1 (v/v) 0.1 M NaH2PO4 + 0.1 M H3PO4. To improve understanding of total P content in agricultural runoff, a comparison188 was made of acid persulfate, USGS and alkaline persulfate methods for measuring P in suspended sediment by ICP-AES. Use of the acid persulfate method was recommended. Dilute HNO3 (0.75–1.0 M) was189 more efficient than chelating agents for extraction of 20 elements from NRCC CRM MESS-3 (marine sediment). However, whether it could “provide a better assessment of potentially bioavailable metals” is open to debate since the true bioavailable concentration was unknown. A valuable study190 used ID to compare the selectivity of 0.43 M HNO3, 0.43 M CH3COOH, 0.05 M Na2H2EDTA and 1 M CaCl2 for extraction of labile Cd, Cu, Ni, Pb and Zn from four soils with varying characteristics. Whilst CH3COOH and CaCl2 failed to recover the labile pool quantitatively, HNO3 and EDTA solubilised additional, non-labile analyte. It was suggested that a lower concentration of EDTA might represent the optimal extractant for estimating the labile PTE fraction in soils.
Comparisons of methods for extraction of plants included a study191 on CRMs – amongst them NIST SRM 1571 (orchard leaves) – that found MAE in 4 mL of HNO3 + 2 mL of H2O2 gave best recoveries for the determination of Cu, Fe and Zn by FAAS. The extraction of As species was studied192 using NRCCRM GBW 10049 (green onion) with determination by HPLC-ICP-MS. Three common extraction methods (shaking at 90 °C in a water bath, UAE and MAE) and three extractants (water, ammonium phosphate and HNO3) were assessed by summing the concentrations of As species released and comparing the sum with the certified total As concentration. Best recovery (94%) was obtained by MAE in 1% HNO3. A method193 for measuring As, Cd, Cu, Ni, Pb and Zn in artichoke involved extraction of 0.5 g sample in 10 mL 25% HNO3 at room temperature with passage of ozone through the extractant, followed by centrifugation to isolate the supernatant for analysis by AAS. Recoveries were similar to those obtained using UAE, those for NIST SRM 1570a (spinach leaves) were 98–101%. The same SRM featured in a comparison194 of dry ashing, hotplate digestion (5 + 1 HNO3:H2O2) and MAE digestion (7 + 1 HNO3:H2O2). All of the procedures gave results close to target values but the microwave procedure was the most accurate. Dry ashing at 480 °C followed by dissolution of the ash in HNO3 gave recoveries superior to those obtained using hot acid digestion in 1:2 (v/v) HNO3 + H2O2 for the determination of REEs in plants by ICP-MS.195 Hotplate digestion196 with mixtures of HCl and HNO3 was however preferred to dry ashing for extraction of Cu and Fe from orange peel and leaves for determination by FAAS. Four digestion methods197 were compared for the determination of 19 elements in samples of wood by ICP-AES and CV-AAS: MAE according to method EN 15290; wet digestion using USEPA Method 3050B; and dry ashing (again according to EN 15290), both with and without a subsequent LiBO2 fusion. Through analysis of IRMM CRM BCR 100 (beech leaves) and BCR 62 (olive leaves) it was shown that wet digestion was the most fit for purpose, giving results close to certified values. Wet digestion also allowed a relatively large sample mass (5 g) to be taken in order to have analyte concentrations above LOD values.
Studies involving DGT and related devices for estimating the available fraction of analytes in soil and sediments included a comparison198 of high resolution (precipitated zirconia and ferrihydrite) and conventional (ferrihydrite slurry) gels for measurement and mapping of W in soils and sediments by LA-ICP-MS. All gave comparable W concentrations but the high resolution approach was better at identifying hotspots. A novel sampling device that incorporated a strong basic anion-exchange resin was developed and tested by a research group in Austria for the ICP-MS determination199 and MC-ICP-MS isotopic analysis200 of labile sulfate in soil. A further innovation201 from the same group was the combination of DGT and a planar optode for pH sensing in the same device. Detailed images of pH and trace element distribution in soil around a plant root were obtained using LA-ICP-MS for elemental information and a ratiometric method to map the pH-sensitive luminophore, 2′,7′-dichloro-5(6)-N-octadecyl-carboxamidofluorescein. Black-and-white photographic film146 was used to obtain a high-resolution 2D map of δ34S in sediment pore waters by LA-MC-ICP-MS. The advantages of this substrate over those commonly used in DGT devices are that it is readily available and requires minimal preparation before deployment.
In bioaccessibility testing, a dynamic version of the unified bioaccessibility test202 was successfully developed and coupled with ICP-AES for on-line determination of Cu, Fe and Mn in soybeans.
An eight-step sequential extraction protocol for fractionation of V in soil203 was proposed to overcome poor recoveries obtained using previous methods. A five-step procedure for the extraction of As from soils and sediments204 was updated and shown, by spike recovery tests, to preserve AsIII and AsV speciation during extraction. A variant205 on the ultrasound assisted version of the BCR protocol mentioned in last year’s update1 was shown to be applicable to Cd, Ni and Pb.
Analyte(s) | Matrix | Coprecipitating agent | Detector | LOD (μg L−1) | CRMs or other validation | Reference |
---|---|---|---|---|---|---|
Ag | Plant, water | 1,6-Diamino-4-(4-chlorophenyl)-2-oxo-1,2-dihydropyridine-3,5-dicarbonitrile | FAAS | 1.60 | NKK CRMs 916 and 920 (Al alloy) | 206 |
Cd, Pb | Tea, tobacco, water | 3-(4-tert-Butylphenyl)-5-phenyl-4-(2-hydroxy-5-methoxybenzylamino)-4H-1,2,4-triazole | FAAS | 2.0 for Pb, 0.2 for Cd | HPS CRM SA-C (sandy soil C) | 207 |
Analyte(s) | Sample matrix | Method | Reagent(s) | Detector | LOD (μg L−1) (unless otherwise stated) | CRMs or other validation | Reference |
---|---|---|---|---|---|---|---|
Ag, Cd | Nuts, vegetables | UA-CPE | 3,7-Diamino-2,8-dimethyl-5-phenyl-phenaziniumchlorid, poly(ethyleneglycol-mono-p-nonylphenylether) | FAAS | 0.02 for Ag, 0.9 for Cd | NIST SRM 1573a (tomato leaves) and SRM 1643e (trace elements in waters) | 208 |
As | Sediment, soil, water | MIL-AALLME | Magnetic ionic liquid 1-butyl-3-methylimidazolium tetrachloroferrate | ETAAS | 0.029 for AsIII | NRCCRM GBW 080231 (seawater), 08605 (water), 07402 (soil), 070008 (soil), 07309 (sediment) | 209 |
Cd | Sediment, water | UA-DLLME | 5-(4-Dimethylaminobenzylidene)rhodanine | FAAS | 0.23 | AR APS 1071 (drinking water), RTC 016 (freshwater sediment 3), NIST SRM 1566b (oyster tissue), SRM 1573a (tomato leaves) | 210 |
Cd | Herbs | SFODME | 1-(2-Pyridylazo)-2-naphthol | ETAAS | 0.0052 | NIST SRM 1573a (tomato leaves) | 211 |
Cd, Cu | Tea, water | DLLME | N,N′-Bis(2-hydroxy-5-bromo-benzyzl)1,2-diaminopropane, carbon tetrachloride | FAAS | 0.69 for Cd, 2.0 for Cu | NRCC SLRS-5 (river water), NIST SRM 1573a (tomato leaves) | 212 |
Co | Spinach, water | DLLME | Thiocyanate, cetylpyridinium chloride, acetone, carbon tetrachloride | ETAAS | 0.02 | NWRI CRM-TMDW (drinking water) | 213 |
Cr | Soil | LLME | Methyltrioctylammonium chloride, xylene | FAAS, ETAAS | 2 mg kg−1 (FAAS), 25 μg kg−1 (ETAAS) | RTC CRM041-30 G chromium VI-soil | 214 |
Hg | Coal, rock, soil | LLME | Exchangeable water–polar organic solvent system | CV-AAS | 0.0005 | NIST SRMs 1633b (coal fly ash), 2711 (Montana soil), NRCCRM GBW 07309 (stream sediment) | 215 |
Mn, Sn, Zn | Food of plant origin | UA-CPE | 2-(3,4-Dihydroxyphenyl)-3,5,7-trihydroxy-4H-chromen-4-one, cetyltrimethylammonium bromide | FAAS | 0.34 for Mn, 0.70 for Sn, 0.15 for Zn | NIST SRMs 1548a (typical diet), 2385 (slurried spinach) | 216 |
Mo | Various foodstuffs | VA-SFODME | 4-Amino-3-hydroxy-1-naphthalenesulfonic acid, 1-undecanol | ETAAS | 0.010 | NIST SRM 1568a (rice flour), 577b (bovine liver), NRCCRM GBW 07605 (tea) | 217 |
Ni | Alloy, effluent, soil | CPE | N,N′-Bis(salicylaldehyde) ethylenediamine | FAAS | 0.15 | Spike recovery | 218 |
Pb | Food, sediment, soil, water | VA-DLLME | 2-Hydroxypyridine-3-carboxylic acid, carbon tetrachloride | FAAS | 8.3 | NWRI TMDA 64.2 (water), TMDA 53.3 (water), SPS-WW2 (wastewater), NRCCRM GBW 07425 (soil), IRMM BCR146R (sewage sludge), HR-1 (harbour sediment) and NCS DC 73349 (bush branches and leaves) | 219 |
Pd | Food, soil, water | DLLME | 2-Mercaptobenzimidazole, chloroform | FAAS | 8.0 | GEOSTATS GPP-10 (ore) | 220 |
REEs | Moss, plant, water | LLE | 3-Ethylaminobut-2-enoic acid phenylamide | ICP-MS | 0.0002–0.0037 | NCS DC 73348 (bush branches and leaves) | 221 |
Analyte(s) | Matrix | Substrate | Substrate coating/modifying agent or analyte complexing agent | Detector | LOD (μg L−1) | CRMs or other validation | Reference |
---|---|---|---|---|---|---|---|
As species | Tea leaves and infusions | TiO2 nanotubes | SFODME + SPE | ETV-ICP-MS | 0.000046 for AsIII, 0.000072 for Asv | NRCCRM GBW 07605 (tea leaves) | 222 |
As, Cd, Cr, Pb | Rice | Dispersion of MWCNTs | ICP-MS | 5.0 for As, 0.6 for Cd, 10 for Cr, 2.1 for Pb | NRCCRM rice CRMs GBW10010, GBW10043, GBW10044, GBW10045 | 223 | |
Au, Pd | Anodic slime, electronic waste, gold ore, soil, water | Amberlite XAD-1180 | N-(4-Methylphenyl)-2-{224acetyl}hydrazine | FAAS | 0.29 for Au, 0.19 for Pd | RTC CRM SA-C (sandy soil C) | 224 |
Au, Pd, Pt | Road sediment | Activated carbon | 3,4-Dihydroxycinnamic acid | ICP-AES | 0.12 for Au, 0.18 for Pd, 0.32 for Pt | NRCCRM GBW 07291 (PGE for geochemical analysis) | 225 |
Cd | Chocolate, cigarettes, herb, water | MWCNTs | Polyvinylpyridine | TS-FF-AAS | 0.036 | NRCC DORM-3 (fish protein) | 226 |
Cd | Cigarettes, coriander, lettuce, sediment | Amberlite XAD-2 | 3,4-Dihydroxybenzoic acid | FAAS | 0.49 | NIST SRM 1646-1 (estuarine sediment) | 227 |
Cd, Pb | Water | Amberlite XAD-1180 | 2-Mercaptobenzothiazole | FAAS | 0.35 for Cd, 5.0 for Pb | NIST RM 8704 (Buffalo river sediment) | 228 |
Cd, Pb, Zn | Various foodstuffs, water | Polyhydroxybutyrate diethanol amine polymer | DDC | FAAS | 3.47 for Pb, 1.39 for Cd, 0.43 for Zn | NIST SRM 1515 (apple leaves), IAEA 336 (lichen), NRCCRM GBW 07605 (tea) | 229 |
Cd, Pb, Zn | Vegetables, water | Amino-functionalised hollow core mesoporous shell silica spheres | ICP-MS | 0.004 for Cd, 0.021 for Pb, 0.007 for Zn | NIST RM 8433 (corn bran) | 230 | |
Cd, Zn | Fertiliser, water | Amberlite XAD-2 | 2-(5-Bromo-2-pyridylazo)-5-diethylaminophenol | FAAS | 0.63 for Cd, 0.38 for Zn | NIST SRM 1573a (tomato leaves) | 231 |
Cu | Fish, tomato leaves, water | Amberlite XAD-4 | N-p-Anisidine-3,5-di-tert-butylsalicylaldimine | HR-CS-FAAS | 0.56 | NIST SRM 1573a (tomato leaves), IAEA 407 (fish tissue) | 232 |
Cu | Walnuts, water | Amberlite CG-120 | SQT-FAAS | 0.23 | NIST SRM 1570a (spinach leaves) | 233 | |
Cu | Plants, water | Ion-imprinted polymer | 1-(2-Pyridylazo)-2-naphthol | ICP-AES | 0.29 | Spike recovery | 234 |
Cu, Fe, Mn | Blood, hair, honey, spinach, water | Graphene oxide | 1-Phenyl-3-(2-thienylmethyl)thiourea | FAAS | 0.06 for Cu, 0.37 for Fe, 0.49 for Mn | NWRI TMRAIN-95 (spiked rainwater), NRCC SLRS-4 (river water), NIST SRM 1640a (spring water) | 235 |
Cu, Ni, Zn | Water | Ion-imprinted polymer gel | ICP-AES | 0.54 for Cu, 0.62 for Ni, 0.58 for Zn | NRCCRM GBW 08301 (river sediment) | 236 | |
Cu, Pb | Spice, street dust, water | Carbon nanospheres | Ionic liquid | FAAS | 0.30 for Cu, 1.8 for Pb | LGC 6016 (estuarine water), NIST RM 8704 (Buffalo river sediment) and IRMM BCR 482 (lichen) | 237 |
Ga, In, Tl | Liquid crystal displays, sediment, water | Amino silica gel | 2-Hydroxy-5-(2-hydroxybenzylideneamino)benzoic acid | FAAS | 4.1 for Ga, 1.5 for In, 1.2 for Tl | Spike recovery | 238 |
Hg | Water | Dispersion of TiO2 NP | CV-AAS | 0.004 | NRCC DOLT-2 (dogfish liver), IAEA 085 (human hair), NIST SRM 2709 (San Joaquin soil), SRM 2711 (Montana soil) and SRM 2704 (Buffalo river sediment) | 239 | |
Mn | Walnut | Amberlite CG-120 | SQT-FAAS | 0.22 | NIST SRM 1570a (spinach leaves) | 240 | |
Pb | Apple, lentil, street dust, vegetables, water | NiCo2O4 microspheres | FAAS | 5.9 | IRMM BCR 482 (lichen), NWRI TMDA 70 (lake water), NIST RM 8704 (Buffalo river sediment) | 241 | |
REEs | Water | Carbon nanofibres | 1-Phenyl-3-methyl-4-benzoyl-5-pyrazone | ICP-MS | 1.5 for Ce, 0.9 for Eu, 0.5 for Gd, 0.3 for Yb | NCS GBW 07602 (bush branch and leaves) | 242 |
REEs | Blood, water | Gold NPs on poly (glycidyl methacrylate-ethylene dimethacrylate) | Mercaptosuccinic acid | ICP-MS | 0.00016 for Tb, 0.00085 for Gd | NRCCRM GBW 07301 (stream sediment) | 243 |
An interesting trend is the increasing use of molecular species as the basis for elemental analysis by HR-CS-ETMAS, either alone or together with HR-CS-ETAAS. In a method250 for direct analysis, soils (up to 0.25 mg) were pyrolysed at 800 °C, and then Cd determined at an atomisation temperature of 1700 °C, and Cr and Fe (as atomic species) and Al (as AlH) at 2600 °C. Results were in agreement with certified values for NIST SRM 2710a (Montana I soil) according to a t-test at 95% CI. Two SS-HR-CS-ETMAS methods for determination of F, one for analysis of solid samples of baby food251 and the other for 1% (w/v) slurries of flour,252 gave recoveries within the uncertainty of the certified value for the NCS CRM DC73349 (bush branches and leaves). In both methods F was determined at 606.440 nm as the CaF molecular species, formed following addition of 10 μL of a 4 mg mL−1 Ca solution to the sample in the furnace. Even isotopic information has been obtained with this approach. A study253 successfully distinguished Ca79Br and Ca81Br on the basis of their electronic transitions and then used solid sampling, platform vaporisation and Pd-NP matrix modifier to determine Br concentration in NIST SRM 1573a (tomato leaves). The poor recoveries (<8% of the certified value) obtained with conventional calibration were attributed to Cl interference. However, when ID was used for calibration the result (1390 ± 70 μg g−1, n = 5) was close to the certified value of 1300 μg g−1.
Turner et al.254 studied interference from Fe in the determination of B in highly weathered tropical forest soils by ICP-AES. They recommended use of the less sensitive 208.957 nm line when measuring B in Mehlich-III extracts and soil digests.
Interest has continued in alternative optical emission sources for atomic spectrometry. A novel flowing liquid anode APGD gave255 superior performance – an average 20-fold improvement in LODs – than the more frequently studied flowing liquid cathode APGD for determination of Ag, Cd, Hg, Pb, Tl and Zn. Spike recoveries from tea leaves ranged from 98.9% for Hg to 102% for Ag. In a prototype microanalytical system,256 10 μL of soil digest was vaporised on a Rh coil and the vapour introduced into a CCP for measurement of Ag, As, Cd, Cu, Hg, Pb, Sb, Sn and Zn using a low-resolution microspectrometer. Results and LODs were similar to those obtained with ICP-AES but the new system had the advantages of markedly lower capital and running costs.
The first room temperature chelate vapour generation method for determination of Zn involved258 continuous on-line mixing of sample (acidified to 0.05 M HNO3) and 0.4% (m/v) Na-DDC solutions followed by immediate sparging of the volatile zinc chelate from solution, using a stream of Ar gas, for determination by AFS.
Two on-line automated FI-HG-AFS methods were optimised by a research group in Brazil. The first259 was used for the sequential determination of As, Sb and Se in peanuts and the second260 for the determination and speciation of Sb in soils.
Three types of pellet were compared264 for determination of 30 elements in soil and sediment by LA-ICP-MS: original sample; sample mixed with boric acid; and sample mixed with hexane, dried at 50 °C, then mixed with DCM and dried again. Treatment with organic solvents was recommended as results for IAES CRM Soil-5 were within 10% of target values.
Several studies focused on trace element mapping in plants by LA-ICP-MS. Imaging of As, Cd, Cr, Cu, H, Mn, Ni, Pb and Zn in individual grains of rye and wheat265 was facilitated by use of a novel though time-consuming serial polishing strategy to expose layers of the sample for analysis. The new open source software package LA-iMageS266 was used for rapid processing and visualisation of LA-ICP-MS data as part of a method267 for imaging Cd, Cu, Fe and Mn in sunflower seeds that had been cultivated in soil spiked with different concentrations of Cd. A combination268 of LA-ICP-MS and TOF-SIMS imaging revealed that Gd and Y were primarily located at the epidermis in roots of Zea Mays. It was suggested that formation of precipitates under physiological conditions limited uptake and transport of the REEs.
In an interesting article,523 an atmospheric pressure pulsed rf discharge was proposed as a means to ablate soils and other non-conducting solid for analysis by ICP-MS. Linear calibration graphs were based on analysis of a series of NRCCRM soil CRMs. Although quantitative analysis of unknowns was not demonstrated, the approach shows promise as an inexpensive alternative to LA-ICP-MS.
Methods for the measurement of isotope ratios by MC-ICP-MS included a procedure28 for Cr that used a single AG1-X8 column to remove matrix elements and achieved an analyte recovery of >93% from CRM digests. The method was suitable for 53Cr/54Cr determination in lichens and mosses and has potential for studying environmental contamination arising from stainless steel manufacture. A single column of Bio-Rad AG-MP-1M was used269 to isolate Cd (recovery >95%) for measurement of 114Cd/110Cd ratio in plants.
Interference reduction in ICP-MS/MS has received attention. Machado et al.270 successfully determined Pt and Rh in sediments as 195Pt16O+ and 103Rh16O+ by introducing 0.3 mL min−1 O2 into the reaction cell. At the same time, 112Cd+, 106Pd+ and 120Sn+ were determined as their atomic ions. Virgilio et al.271 used 0.5 mL min−1 O2 to measure 75As16O+ and 51V16O+, along with 52Cr+, 60Ni+, 112Cd+ and 208Pb+, in NIST SRM 1573a (tomato leaves). Recoveries were 85–113%. Researchers in China272 recommended use of O2 reaction gas for determination of As and Se in plant CRMs because results obtained were more consistent with certified values than when H2 or He were used in collision mode. Beryllium was preferred to Li and Rh as IS for the determination of B in plants using ICP-MS/MS with O2 reaction gas.273 Recoveries from NIST SRMs 1515 (apple leaves) and 1573a (tomato leaves) were 90–108%.
Analytical methods featuring chromatographic separation coupled with ICP-MS included a HPLC-ICP-MS method274 for Sb speciation in plants, which was applied to ryegrass grown hydroponically under controlled conditions. Uptake of trimethylantimony by roots and subsequent translocation to shoots was demonstrated conclusively for the first time. A powerful combination of elemental and molecular MS techniques275 including HILIC-ICP-MS and SEC-ICP-MS was used to identify more than 50 metal complexes in xylem and embryo sac liquid of pea plants. Also reported were HPLC-ICP-MS methods for the speciation of mercury276 and tin277 in sediments and for the determination of AsIII and AsV in iron-rich precipitates originating from acid mine drainage.278 A modified CE-ICP-MS interface279 improved sensitivity 2–5 fold for As speciation analysis in Solanum lyratum extracts. An interesting study142 demonstrated the feasibility of determining C along with other analytes (Al, Fe, Mn, P and Si) in both synthetic and natural fine particulate organic matter in water and soil extracts by asymmetric flow FFF-ICP-MS.
The novel coupling of an ETV unit and tungsten coil trap with ICP-MS was used to determine Cd and Zn in wheat.280 Up to 20 mg of solid sample was ashed in a sampling boat and then heated in a porous carbon vaporiser to release the analytes for capture on the coil. Results for corn, rice and wheat CRMs were close to certified values. LODs were 0.01 μg kg−1 for Cd and 0.1 μg kg−1 for Zn.
The combination of multi-element analysis by sp-ICP-TOF-MS with advanced chemometrics represented281 a potentially significant advance for discriminating between engineered NPs and naturally occurring nanoscale particulates in soil.
Interest continued in the development of LIBS methods for soil analysis. Two research groups used soil as a test substrate to demonstrate new approaches for analysis of particulate matter. The first284 employed double-sided foam tape to trap particles from a stream of argon gas, while the second285 mixed 2 mg of sample with 2 mg of Si grease and presented it to a TEA CO2 laser as a thin layer smeared on a Ni subtarget. Labutin et al.286 showed that generation of synthetic spectra by thermodynamic modelling could aid determination of REEs in rocks and soils by highlighting potential spectral overlaps in advance. A research team in China287 optimised a LIBS-LIF method to eliminate spectral interferences in the determination of Pb in soil altogether. Pressed pellets of soil CRMs were irradiated with a Nd:YAG laser in the normal way then, after an optimal delay time of 14 μs, a second laser at 283.31 nm (resonant with the E0 → E2 transition of the Pb atom) was fired through the sample plasma. This enhanced the Pb signal markedly, increased sensitivity and improved the LOD from 24 to 0.6 μg g−1. Another interesting article288 showed that the 235U isotopic abundance could be estimated through application of appropriate decomposition algorithms to LIBS spectra of mixtures containing NIST SRM 2710a (Montana I soil) and up to 8.6% (w/w) 235U-enriched U3O8. Applications in nuclear forensics were envisaged.
The application of LIBS to the analysis of plants included the successful determination289 of Pb in soft biological materials with high water content, specifically carrots and mango leaf, for which use of a ps rather than a ns laser was necessary to increase the power density delivered to the sample. A method290 for the determination of Ba, Fe, Mg, Mn and Sr in powdered samples of blue-green algae could be performed in air at atmospheric pressure using a TEA CO2 laser and non-gated detection. Workers in Beijing showed that LIBS could be used to profile the Ca content in endosperm, embryo and epidermis of a single grain of corn under laboratory conditions291 and that elemental maps could be obtained from living plants in the field using their portable LipsImag system.292 Both 2D and 3D maps of chlorpyrifos on leaves of sprayed maize plants were generated by a multi-variable regression model based on the emission lines of Cl and P.
The combination of multivariate chemometric methods and LIBS has been used to classify soils. One procedure293 was based on LS-SVM and the emission lines of Al, Ca, Fe, K, Mg, Na and Si. Another294 (in Chinese with English abstract) used PCA and Al, Ca, Fe, Mg, Si and Ti. Tobacco mosaic virus was detected295 in a procedure involving PLS-DA.
Remote detection296 of Cd, Cr, Ni and Pb in NIST SRMs 2710a (Montana I soil) and 2586 (trace elements in soil – containing Pb from paint) at a distance of up to 6 m was achieved by stand-off LIBS following careful optimisation of measurement parameters including gate width, gate delay and laser pulse energy. Other workers297 demonstrated remote LIBS detection of Al, Ca and Mg in soil from a distance of 15 m during commissioning of a truck-based mobile laser spectroscopy laboratory. Approaches demonstrated under laboratory conditions, but with clear potential for eventual deployment in stand-off mode, included a procedure298 to discriminate between weeds and crops based on the use of plasma temperature and a method299 to detect drought stress in plants based on relative concentrations of Ca, Fe, K and Na.
Improvements in pXRFS were achieved by using either soil CRMs301 or solid mixtures302 of organic (plant) and inorganic (soil and sediment) CRMs as calibrants. Use of pXRFS analysis of till for geochemical exploration was investigated by Sarala303 who compared three instruments – two handheld and one truck-mounted system – for analysis of samples from Finland, and Hall et al.304 who used a single handheld instrument for analysis of archived samples from sites close to mineral deposits in Canada. Although the concentrations of some pathfinder elements were below the LODs, both groups concluded that pXRFS was fit-for-purpose. A useful method305 for measuring P, an indicator of human activities, in archaeological soils was also reported. Dao306 showed that normalising the intensity of the P-Kα line to that of the Ag-Kα line was an effective way to compensate for variations in plant moisture content in the determination of P in maize leaves. This meant that results could be expressed on a dry weight basis.
Proximal sensing using pXRFS continued to be of interest. A method307 for in situ estimation of pH in frozen and permafrost-affected soils proved successful in field trials at five sites in northern Alaska. Models308 based on pXRFS and VisNIR DRS data could be used to predict total C and N contents in organic-rich soils containing up to ca. 50% C. Combining information from MIR, pXRF and VisNIR spectroscopies through the use of a model averaging technique was recommended309 for best prediction of soil geochemical properties based on a study of 322 samples from the Irish National Soil Database Archive.
The preparation of thin specimens in a TXRFS method310 for the determination of Br in soil involved suspension of 20 mg soil in 1% Triton X-110, addition of Rh as IS and homogenisation. Finally, a 10 μL aliquot was dried onto a quartz sample holder. The LOD was 3.7 mg kg−1 and the recoveries for analysis of 12 soil and sediment CRMs were 89–133%.
A TXRFS method for the analysis of plants311 was developed by the determination of Ca, K, Mn, and S in leaves of Arabidopsis thaliana following digestion of 0.3 mg samples in 150 μL HNO3. More recently, it was demonstrated312 that sample digestion was unnecessary prior to analysis of vegetal foodstuffs by TXRFS. In a procedure similar to that already cited in this update for determination of Br in soil, the sample (20 mg) was suspended in deionised water and Ga added as IS before drying onto a quartz holder. Methods for trace element mapping in wheat grains313 and carrots314 by μXRFS were described. Although the resolution of benchtop instruments is poorer than that achievable with synchrotron-based methods, the advantages in terms of cost and accessibility are clear.
There are few certified values available for Au and PGEs in ferromanganese nodule RMs, despite the fact that Mn- and Fe-rich materials are of great economic interest. In a study315 on the ferromanganese nodule RM USGS NOD-A-1, Au, Ir, Pd, Pt and Ru were preconcentrated using anion-exchange following acid digestion and their concentrations determined by ICP-MS. The suitability of the Fe- and Mn-rich natural nodule RMs GSJ JMn-1, USGS NOD-P-1 and NOD-A-1, and BGR FeMn-1 for trace element determinations in microanalytical research was assessed by analysing pressed powder pellets by LA-ICP-MS using fs and ns LA systems.316 The pellets exhibited heterogeneities on the μm-scale for several elements at spot sizes of <40 μm. In contrast, the synthetic USGS RM FeMnOx-1, was shown to be homogeneous for spot sizes 8–10 μm and so gave low RSD values of <10% for all elements measured, confirming its suitability as a future calibration material.
Mass fractions of Br, Cl, F and I in USGS geological glass RMs BHVO-2G, BIR-1G, BRC-2G, GSD-1G and GSE-1G, and NIST SRMs 610 and 612 (trace elements in glass) were determined by a variety of techniques.317 The relatively high halogen contents of glasses GSD-1G and GSE-1G made them well suited for the calibration of Br, Cl and I in microanalytical studies of silicate glasses, and melt and fluid inclusions by EPMA and LA-ICP-MS.
Molybdenum is an element of increasing interest in magmatic and hydrothermal geology, in planetary sciences and for tracking environmental pollution. The Mo mass fractions and isotopic composition of 24 international geological RMs encompassing a wide range of compositions (i.e. coral, igneous and sedimentary rocks, manganese nodules, ferromanganese crusts and molybdenite) were determined by ID MC-ICP-MS using a 97Mo–100Mo double-spike.318 The Mo isotopic composition of USGS RMs BCR-2 (basalt), SCo-1 (shale) and SGR-1b (shale) were reported for the first time in a study which used MC-ICP-MS analysis after purification of Mo by ion-exchange chromatography.319
Zinc isotopes are important tracers of hydrothermal systems relevant in ore genesis studies, identification of sources of contamination in soils and aerosols, monitoring palaeo-pH and studies of the origin and evolution of the solar system. Chen et al.320 reported high-precision Zn isotope data (better than ±0.06‰ δ66Zn, 2SD) for 17 rock RMs with a range of compositions, including the first published Zn isotope data for USGS RMs GSP-1 (grandodiorite), GSP-2 (grandodiorite) and RGM-1 (rhyolite). Determination of δ66Zn values (n = 98) in NIST SRM 683 (metal nuggets) by MC-ICP-MS proved the homogeneity of this RM and so demonstrated its suitability as an isotopic CRM for Zn. Araujo et al.163 presented the first Zn isotope ratio measurements for environmental NIST SRMs 2709 (soil) and 1646a (estuarine sediment) as well as δ66Zn data for USGS rock RMs AGV-2 (andesite), BCR-2 (basalt) and BHVO-2 (basalt).
There is a trend to publish isotope ratio data for some of the more popular USGS RMs originally characterised for their trace element compositions. Thus, the determination of the Ca isotope compositions of 16 USGS RMs extended321 the range of geological materials analysed and provided the first Ca isotope data for three sedimentary RMs and one carbonatite RM. Kimura et al.322 observed heterogeneities in B and Li isotope ratios for USGS GSD-1G (synthetic basalt glass) when ablating spot sizes of <100 μm by LA-ICP-MS. The first 230Th/232Th values for CRPG RM BE-N (nephelinite) and USGS RM BIR-1 (Icelandic basalt) were published in a study323 of Th isotopic compositions in silicate RMs using a two-step isolation protocol and MC-ICP-MS measurements. The authors also reported 230Th/232Th values for USGS RMs BHVO-2 (basalt) and AGV-2 (andesite).
Well-characterised reference solutions are a cornerstone of reliable and traceable measurements. The isotopic composition of two new Cu isotope standard solutions developed by Yuan et al.324 was determined in three different laboratories using MC-ICP-MS. The National Institute of Metrology, China325 prepared and certified one natural and 10 synthetic Se solutions with a wide range of Se isotope ratios, for calibration and quality control of chemical and MS procedures. A new standard solution326 for the Zn isotope community, intended as a replacement for the current reference standard solution JMC-Lyon Zn, was calibrated against existing Zn reference solutions.
New microanalytical RMs included MUL-ZnS-1, a sulfide RM327 prepared by sintering a powdered (Zn,Fe)S matrix doped with several elements (Ag, As, Cd, Co, Cr, Cu, Ga, Ge, In, Mn, Mo, Ni, Pb, Sb, Se, Sn and Tl). This RM was intended primarily for the matrix-matched calibration of LA-ICP-MS measurements of sphalerites but could also be applied to other sulfides. Chew et al.328 proposed a protocol for developing natural apatite mineral RMs. Gem quality crystals of Durango apatite were analysed by LA-ICP-MS and solution ICP-MS and potential interferences examined in detail. The authors claimed that this protocol could easily be modified to characterise other natural minerals as potential microanalytical RMs. Cruz-Uribe et al.329 showed that the ca.100 μm wide rims of quartz grains from the Bishop Tuff quartz (California, USA) had homogeneous Ti mass fractions of ca. 41 μg g−1 and that they could be used to calibrate Ti-in-quartz thermometry measurements by LA-ICP-MS. Not only did this provide a matrix-matched RM, but it was possible to use 48Ti, the most abundant Ti isotope, because there was no interference of 48Ca on 48Ti due to the absence of Ca in quartz. For the quantification of Br abundances in natural materials, a set of 21 Br-doped glasses330 were prepared from natural volcanic rocks and characterised by INAA. These were used to compare the figures of merit of three micro-analytical techniques (LA-ICP-MS, SIMS and SRXRFS): SRXRF gave the best accuracy and precision (10% for Br contents ≥10 μg g−1 Br) with an LOD of <1 μg g−1 Br, with SIMS not far behind. In contrast, the precision and accuracy for LA-ICP-MS were relatively poor (ca. 20% for 100 s to 1000 s μg g−1 Br). Roberts et al.331 characterised a calcite RM (WC-1, 254.4 ± 6.4 Ma, marine calcite cement, Guadalupe Mountains, Texas) for the LA-ICP-MS U–Pb dating of carbonates and advocated its use as the principal RM for U–Pb normalisation. This RM could be used reliably despite its variable common Pb content, because its isotopic composition fell on a mixing line between initial and radiogenic Pb.
Most available zircon RMs are younger than 1000 Ma so a careful evaluation of a potential zircon RM for U–Pb geochronology of the Archean is of particular note.332 Uranium–Pb data obtained on the Stillwater AN2 zircon (Montana, USA) by ID-TIMS, CA-TIMS and LA-ICP-MS yielded indistinguishable concordia and a weighted mean 207Pb/206Pb age of 2710 Ma. The LA-ICP-MS results demonstrated homogeneity within grain and across multiple grains of Stillwater AN2 zircon on the scale of tens to hundreds of μm. This new RM plugs an important gap in U–Pb geochronology. Nasdala et al.333 advocated zircon M127-A as a new zircon RM for SIMS U–Pb geochronology as well as isotopic analysis of Hf, O and, potentially, Li. A mean concordia age of 524.37 Ma was determined from TIMS analyses made in several laboratories. Zircon M127-A was homogeneous in terms of its chemical and isotopic composition (δ18O 8.26 ± 0.06‰ (2 s), 176Hf/177Hf 0.282396 ± 0.000004 (2 s), and δ7Li −0.6 ± 0.9‰ (2 s)) and had a remarkably concordant U–Pb isotopic system. This made it particularly suitable for studies that aim to use the same sample mounts for the determination of O (SIMS), Li (SIMS) or Hf isotopes (LA-ICP-MS). A gem-quality zircon megacryst334 from Sri Lanka, named LGC-1, analysed extensively by a variety of techniques, was homogeneous with respect to its He, Th and U distributions and so was considered as a suitable grain for in situ (U–Th)/He dating. However, the authors emphasised the need for developing more such RMs because of the limited material available. This would also be true for most natural mineral RMs.
Increasing interest in in situ measurement of sulfur isotope ratios in a variety of minerals prompted the development of a suite of S isotope RMs (pyrite, chalcopyrite, pyrrhotite and pentlandite) relevant to magmatic, magmatic-hydrothermal and hydrothermal ore systems.335 The new sulfide RMs were characterised for their chemical (EPMA) and isotopic (SIMS) compositions and provided appropriate matrix-matched RMs for the samples under investigation. The first scapolite RM (CB1) for in situ S isotope analysis using SIMS was characterised using LA-ICP-MS, EPMA, SIMS and bulk fluorination gas source IRMS.336 Results of four SIMS analytical sessions demonstrated that the megacryst was isotopically homogeneous (reproducibility of δ34S values was 0.40‰ (2SD), n = 244) and that the crystallographic orientation did not affect the analysis. The δ34S values in a chalcopyrite glass RM, YN411-m, intended for in situ S isotope ratio measurements, were determined by GS-MS and solution MC-ICP-MS.337 The homogeneity was evaluated by replicate analyses by fs LA-MC-ICP-MS, which gave an external precision of 0.28‰. It was concluded that, for the analysis of a natural chalcopyrite prepared as a pressed powder pellet, YN411-m was better suited as a bracketing RM than a non-matrix-matched pressed pellet made from RM IAEA-S-1 (silver sulfide).
In an excellent publication,338new protocols for the determination of U–(Th–)Pb ages, as defined by the LA-ICP-MS geochronology international community, included a workflow defining the appropriate propagation of uncertainties. This is not only important for reporting U–Pb data but other analytical data as well. New radiogenic ratios presented for commonly-used zircon RMs, based on ID-TIMS measurements, were considered more suitable for LA-ICP-MS calibrations than an age-equivalent ratio calculated from a published age, which can introduce inaccuracies if only one age (207Pb/206Pb, 206Pb/238U, 207Pb/235U or 208Pb/232Th date, or the concordia age) was assumed to be the correct age of the reference zircon. Recommendations on how to present data and new data reporting standards were provided with the aim of improving the peer-review process when such data are submitted for publication.
Accurate information on isotope abundances and atomic weights, which can vary in natural materials because of physical and chemical processes, is of great importance for mass spectrometric analysis. Therefore, an IUPAC Technical Report339 providing isotopic abundances, delta values and atomic weights with the upper and lower boundaries for 12 elements (B, Br, C, Cl, H, Li, Mg, N, O, S, Si, and Tl) is welcome.
Preconcentration and separation protocols can be required in the determination of trace elements by ICP-MS when the sample matrix might otherwise contaminate or physically damage the instrument. Strnad et al.343 reported two relatively simple methods for the extraction of Au from complex matrices using a mixture of HBr, HClO4, HI and aqua regia. These extracts could be analysed by ICP-MS without any further preconcentration or separation. Although the procedures were developed primarily for the determination of Au in electronic waste, they were also applicable to the analysis of geological materials. The accuracy achieved for nine geological RMs was better than ±10% in most cases, with a precision of <5% RSD (n = 5). Possible pitfalls such as nugget effects, spectral interferences and memory effects were evaluated. No significant interferences were observed on 197Au. Dubenskiy et al.344 tested the efficiencies of six new hypercrosslinked polystyrene sorbents for the simultaneous quantitative extraction of Au, Pd, Pt and Ru as their chlorocomplexes from rock digests following NiS fire assay pretreatment. The sorbant Stirosorb-514 in TBA, with 1 M HCl in ethanol as the eluent, was the most effective reversible sorption system for the preconcentration of PGEs prior to ICP-MS measurement.
Several recent studies discussed new or improved chromatographic separation schemes for isotope ratio measurements. Li et al.345 based a rapid tandem-column separation procedure for the separation of Hf, Nd, Pb and Sr from silicate matrices on the enhanced elemental selectivity permitted by the Sr Spec and TODGA Spec resins. The procedure offered a significant improvement in separation efficiency and a reduction in cross-contamination compared to conventional four column protocols. It took ca. 8 h to process a batch of 25 samples, including column setup, cleaning and preconditioning. The robustness of the method was verified by determining the isotopic compositions of Nd, Pb and Sr (by TIMS) and of Hf (by MC-ICP-MS) in eight silicate rock RMs with a wide range of matrices. A SPE procedure346 was developed for the preconcentration of 28 elements (Ba, C, Co, Cr, Cu, Li, Ni, Rb, REEs, Sr and Zn) present at very low concentrations in magnetite and pyrite prior to their determination by HR-ICP-MS. Following sample decomposition in a mixture of 6 M HCl and 8 M HNO3, two groups of elements were separated from the iron matrix using anion-exchange resin AG1-X8 and TRUspec ion-exchange resin. Recoveries were 100 ± 5% and the procedural blanks low. The CCRMP RM FER-2 (iron formation) was used for validation. A three-step ion-exchange method347 for the separation and purification of B from clay sediments was published in Chinese. In the first step, cation-exchange resin AG 50W-X8 was used to remove Al, Fe and the REEs, to prevent the formation of insoluble hydroxides that can absorb B in highly alkaline solutions. In subsequent steps, B was separated on Amberlite IRA 743 and purified on a mixed anion- and cation-exchange column. Recoveries were >90% with no obvious fractionation of B isotopes. To perform very precise determinations of Nd isotope ratios in geological samples, Kagami and Yokoyama348 developed a three-step column chemistry procedure involving cation-exchange chromatography for separating major elements from REEs, oxidative extraction chromatography to remove Ce and final purification of Nd using Eichrom Ln Spec resin. High recoveries (>91%) and effective separation of Nd from Ce and Sm (Cd/Nd <1.2 × 10−5 and Sm/Nd <5.2 × 10−6) were achieved. Small extraction chromatographic columns (alone or in tandem) were used349 to separate 11 elements involved in six radiogenic isotope systems from a rock digest prior to isotope ratio measurements by MC-ICP-MS or TIMS. Commercially available extraction resins, together with a resin prepared in-house for the separation of Rb, provided typical recoveries of 85–95% with procedural blanks of 10–100 pg. A simplified scheme for the separation of Hf, Nd, Pb and Sr only was also proposed.
Alkali fusion used for the dissolution of single zircon grains for (U–Th–Sm)/He thermochronometry had350 several advantages over previously used methods. Not only was it twice as fast but it avoided the use of HF, thereby making it unnecessary to remove the grains from the metal microvials, as usually required to avoid contamination by co-dissolution of the vial material. A mixture of B(OH)3 and LiBO2 was added to the sample in a Pt crucible and fused at 990 °C for 2 h. After dissolution in HNO3, Sm, Th and U were measured by ICP-MS. The (U–Th–Sm)/He ages of two reference zircons were in good agreement with reported ages and the precision was comparable to that of traditional dissolution procedures.
Chemical abrasion of zircons is a procedure which involves sequential annealing and leaching, and is commonly used to remove parts of the zircon grain that have experienced loss of Pb, thereby improving the precision and accuracy of U–Pb dating. Following a detailed evaluation,351 optimal conditions for CA were annealing at 900 °C for 48 h and leaching in a Parr digestion vessel at 190 °C in conc. HF for 15 h. The application of CA to U–Pb dating of zircons by SIMS rather than by LA-ICP-MS is relatively uncommon. In a recent SIMS study352 of the application of CA to zircons of late Eocene age, CA-treated grains yielded older 206Pb/238U ages than untreated grains. These differences in measured ages did not correlate however with trace element concentrations. Chemical abrasion appeared to improve the accuracy of the age determinations while the precision remained the same. Although the treatment could potentially have affected the secondary ion sputtering rates and so result in biased ages, as observed for LA-ICP-MS analysis, ion microprobe pit depths were in fact identical for both CA and non-CA reference zircons so no loss had occurred.
Fusion techniques are often used to produce homogeneous glass beads for bulk rock analysis by LA-ICP-MS. Zhang et al.353 advocated a novel flux-free fusion technique in which a high-energy IR laser was used to prepare glasses from pressed powder pellets. Although this method apparently worked well for felsic and zircon-bearing rock powders rocks, unfortunately the authors compared their LA-ICP-MS data for glasses with results from XRFS analysis of the original pressed powder pellets but not with glasses prepared by other fusion techniques such as Ir-strip fusion.354,355 The reported precision and accuracy of LA-ICP-MS measurements on the glass beads were better than those obtained from pressed powder pellets but were not as good as the figures of merit achievable by other flux-free methods. In addition, there remained issues of a longer analysis time due to the need to prepare pressed pellets, loss of volatile elements and a much larger sample mass requirement. However, if no Ir-strip heating device is available, this technique would provide suitable material for whole rock analysis by LA-ICP-MS, thus avoiding the need for time-consuming sample dissolution. The technique of Bao et al.356 involving a flux-free fusion in boron nitride vessels followed by rapid acid digestion was especially applicable to high-SiO2 rocks containing refractory minerals such as zircon and rutile. The optimum conditions were heating for 1 minute only at 1400 or 1600 °C for extrusive and intrusive rocks, respectively. Fusion was followed by a rapid HF–HNO3 digestion of the glasses in capped Savilex screwtop beakers on a hotplate. This procedure was eight times faster than methods using high-pressure PTFE digestion bombs. Element concentrations determined by ICP-MS for ten RMs with SiO2 contents between 44.6 and 73 wt% were in good agreement (<±10%) with reference values for most elements, including volatile elements such as Cs and Pb. Another fusion method357 (paper in Chinese), specifically for the preparation of sulfide materials prior to the determination of major and trace elements by LA-ICP-MS and XRFS, employed HNO3 to pre-oxidise the samples to avoid damaging the Pt crucibles. The addition of GeO2 prevented breakage of the glass beads, which were homogeneous (<5.6% (1 RSD) by XRFS and <3% (1 RSD) by LA-ICP-MS) in their major element composition.
An alternative approach to the preparation of samples for bulk rock analysis by LA-ICP-MS is the preparation of nanoparticulate pressed powder pellets. To evaluate this methodology, six rock RM powders were further pulverised by wet milling and pressed into pellets using microcrystalline cellulose as a binder.358 Analysis of the pellets by LA-ICP-MS using a 193 nm excimer laser gave a measurement repeatability as low as 0.5–2% (1 s, n = 6) using a spot size of 120 μm, demonstrating the pellets were homogeneous at this scale. Blank contributions were negligible for most elements and polyatomic interferences caused by the organic binder were insignificant. One advantage of this method was the ability to measure elements such as As, B, Bi, Li, Sb and Tl, which are not always accessible with other methods of sample preparation.
An excellent review375 (44 references) of the determination of Th by mass spectrometric techniques detailed the advantages and pitfalls associated with the measurement of Th isotope ratios by a range of techniques, including TIMS, ICP-MS and SIMS. Problems of contributions to the blank and the poor ionisation efficiency of Th were discussed and the need for CRMs with certified 230Th/232Th isotope ratios was highlighted.
A brief history376 of the development of laser ablation ICP-MS was presented in an issue of Elements (vol 12, number 5, 2016) dedicated to a wide range of applications of this technique to the Earth sciences. Other reviews of interest in this issue included articles on elemental analysis of igneous systems377 (30 references), advances in isotope ratio determinations378 (29 references) and the role of ICP-MS in palaeoclimate research379 (31 references). An overview (150 references) of the principles of LA-ICP-MS analysis of solid inorganic materials by Shazzo and Karpov380 emphasised the problems arising from elemental fractionation. A paper381 in Czech (104 references) focused on the principles of LA-ICP-MS instrumentation and its application to geochemistry, parameters influencing the ablation process, quantification strategies, limitations of the technique and the use of RMs. Bruand et al.382 (53 references) highlighted how the trace element and isotopic composition of apatite can provide important insights into the petrogenesis and provenance of igneous rocks. Recent advances383 (169 references) in the application of LA-ICP-MS to magmatic-hydrothermal ores to determine the trace element composition of minerals included appropriate methodology, RMs, new mineral matrices, analytical uncertainties and future applications.
The application of LA-ICP-MS to the analysis of fluid inclusions to derive information on fluid–rock interactions and ore deposit formation, as well as tracking evolution of fluids along fluid pathways, was summarised by Wagner et al.384 (30 references). To obtain absolute elemental concentrations in fluid inclusions, microanalytical measurements (by LA-ICP-MS) are usually interpreted in relation to microthermometric observations. Steele-MacInnis et al.385 (83 references) assessed the use of such data to obtain robust results for the composition of low- to intermediate-salinity chlorine-dominated inclusions.
The use of decrepitation to release the contents of fluid inclusions for analysis was first developed more than three decades ago and before the advent of LA-ICP-MS. In a reappraisal of the technique, Piperov et al.387 devised a system in which a simple ‘decrepitator’ was inserted between the nebuliser and the plasma torch of an ICP-AES instrument. The performance of this system was compared with conventional crush-leach or decrepitation leach methods of preparation, as well as analysis by LA-ICP-MS. The decrepitation ICP-AES technique was able to provide relatively rapid preliminary information at low cost and was particularly useful for sample materials that were not amenable to analysis by LA-ICP-MS.
Bachari et al.388 examined the effects of spectral interferences on the ICP-AES determination of U concentrations in phosphate ore. Results from three different acid digestion procedures confirmed that the acid used was one of the most important influences on the choice of emission line for measurement. Spectral interferences affecting the accurate determination389 of Rh in molybdenum and copper concentrates by ICP-AES were investigated using two different ICP-AES systems. Quantitative databases for the type and magnitude of spectral interferences in the presence of Al, Ca, Cu, Fe, Mg, Mo and Ti were obtained for both a radial viewing 40.68 MHz ICP with a high resolution spectrometer and an axial viewing 27.12 MHz ICP with a medium resolution spectrometer.
A very large suite of LIBS spectra of rock samples392 – 1356 spectra from 452 geological samples – was used to compare the accuracy of different models for prediction of elemental compositions. In general, the accuracy of the predicted values for Al, Ca, Co, Cr, Fe, K, Mg, Mn, Na, Ni, Si, Ti and Zn was much better for multivariate analysis using PLS than for univariate analysis, especially when based on single emission lines. This conclusion was not surprising given the wide range of geological matrices involved. The ChemCam team393 used a ‘sub-model’ method with PLSR to improve the accuracy of quantitative LIBS analysis of geological materials on the Mars Science Laboratory rover. The method was conceptually simple but did require a substantial number of RMs to provide the training set of data. Dyar et al.394 presented an excellent compilation of methods for baseline removal from LIBS spectra. Results from different approaches demonstrated the complexity of this issue and indicated that sophisticated types of multivariate analysis, with optimisation of adjustable parameters, were always necessary to improve prediction accuracy. Aguilera and Aragón395 proposed a new approach to quantitative analysis by LIBS, designated as CSigma-LIBS, where conventional calibration for each element using a range of RMs was replaced with a procedure based on CSigma graphs. These graphs were constructed to include reference lines of an emitting species with well-known transition probabilities, together with lines for the elements of interest, both in the same ionisation state. Two glasses of known composition, prepared from mixtures of pure compounds, were used for calibration. This technique was applied to the determination of mass fractions of Al, Ca, Fe, Mg, Mn, Si, Ti and V of geological RMs prepared as fused glass beads. The average precision was 3.4% for concentrations >0.1 wt%, with good agreement with the certified values for the four rock RMs analysed. The main drawbacks were the need to prepare fused glass beads, which removed the versatility and speed of the LIBS technique, and the dilution involved, thereby preventing the determination of trace elements.
There is increasing interest in high resolution geochemical characterisation of materials by LIBS. Lefebvre et al.396 demonstrated that LIBS was capable of providing semi-quantitative geochemical signatures of multi-layered, sub-mm thick coatings on a rock surface. The test sample was a basaltic rock with sub-mm Ca–Mg–Fe–Si rich mineral coatings. The authors considered the method suitable for fine-scale 3D LIBS analysis of coatings on Martian rocks.
An innovative approach397 of double-pulse LIBS coupled with a petrographic, optical microscope was used to analyse thin sections of iron meteorite samples directly in air rather than in a controlled atmosphere. Qualitative and quantitative analysis by calibration-free LIBS and one-point calibration LIBS methods, respectively, provided results in good agreement with data from ICP-MS and SEM analyses. Double-pulse LIBS,398 combined with multivariate statistics, was also used to generate a map of the spatial distribution of U in sandstone-hosted U ores. The lateral resolution was limited to 100 μm. Unfortunately, there was no comparison of the LIBS results with other in situ techniques such as EPMA or LA-ICP-MS.
A well-documented study399 on the application of handheld LIBS instrumentation to geological materials included an excellent review of the operational principles of LIBS and portable instrumentation in general, and its possible application to geochemical fingerprinting and provenance determination. Samples including carbonates, silicates, oxide minerals and native gold were analysed using a handheld LIBS instrument under simulated field conditions. Connors et al.400 evaluated the performance of one specific commercially-available handheld LIBS instrument for the analysis of geological materials. All the samples were tested after pressing into thin cakes with a 6-ton press and data reported using traditional peak ratio calibrations. This research indicated a significant potential for qualitative analysis and elemental mapping. Because of its high sensitivity, LIBS has the capability to measure elements such as Be and Li which cannot be measured by pXRFS. However, more work was necessary to implement PCA and PLS model generation software for fully quantitative analysis.
Spectral interferences have an impact on the accuracy and precision of trace element and isotope ratio measurements by ICP-MS. A detailed study405 investigated the formation of doubly charged alkaline earth-argon ions (MAr2+, e.g.86Sr40Ar2+) in the ICP, which can potentially interfere with isotopes of analyte elements, e.g.63Cu+. The relative abundances of the various MAr2+, M2+ and MAr+ ions agreed with assumed properties of the ICP and the ion extraction process in ICP-MS. An exception was the relatively low abundance of MgAr2+, which was assumed to result from collision-induced dissociation during the ion extraction process. The same authors406 looked at the abundance of doubly charged polyatomic Ar interferences in ICP-MS spectra and how they affect ultratrace analyses and isotope ratio measurements. Changing ICP parameters, such as raising the carrier gas flow or lowering gas temperatures, increased the abundance of the doubly charged molecular ions significantly.
The high spatial resolution of LA-ICP-MS analysis, one of the major advantages of this technique since it permits elemental and isotopic mapping and depth profiling, was exploited408 to resolve chemical variations in daily growth bands of Miocene and modern giant clams. A rotating rectangular aperture was used to maximise the spatial resolution of line scans across the sample surface. Data were acquired using laser spots of 4 × 50 μm at a scan speed of 1.5 μm s−1 and a 15 Hz repetition rate. No signal smoothing device was used. Six masses were monitored. Results obtained by LA-ICP-MS at a spatial resolution of 4 μm in this study were comparable to data obtained in other laboratories by nanoSIMS at a resolution of 2 μm. In another application of in situ multi-element 2D imaging by LA-ICP-MS, single pyrite grains409 were mapped for nine elements to obtain information on the redox conditions existing during pyrite formation. Raimondo et al.410 presented a technique for the routine acquisition of 2D trace element maps of garnet minerals at high spatial resolution. They also developed an extension to the XMapTools software package for rapid processing of the LA-ICP-MS data in order to visualise and interpret compositional zoning patterns. The elemental maps obtained were used to identify the mechanisms controlling geochemical mobility in garnet. The distributions of Au, Hg, Se and Te in reduction spheroids (cm-scale pale spheroids in an otherwise red rock) from continental red bed successions, measured by LA-ICP-MS, revealed411 increased concentrations of Te towards the core. This showed that Te is mobile and can be concentrated in such low temperature sedimentary environments, probably controlled by a process such as microbiological activity. Petrelli et al.412 examined the potential of a quadrupole LA-ICP-MS system for high resolution trace element analysis of volcanic glasses from ash particles using laser beam sizes of 5 to 15 μm. They concluded that, for most geological applications, a measurement precision of 15% for concentrations >1.7 μg g−1 and an accuracy of <15%, obtained using 42Ca as the IS and with an 8 μm spot size, was acceptable.
Warburton et al.413 tested the reliability of depth profiling to obtain high resolution trace element data for otolith microchemistry. The technique could yield accurate data if important factors such as tailing effects, elemental fractionation and spot size were considered carefully. A combination of SIMS U–Pb and elemental LA-ICP-MS depth profiling techniques for acquiring maps of unpolished zircon grains provided414 information on zircon recrystallisation, rim development and mineral–fluid interactions.
Matrix | Sample treatment | Technique | Analysis and figures of merit | Reference |
---|---|---|---|---|
Geological RMs (silicates, carbonates, barites) | Two methods of preparation: (a) for MC-ICP-MS, addition of 130Ba–135Ba double spike before cation-exchange procedure on Dowex 50W-X8. Total procedure blanks <0.9 ng; (b) for TIMS, digests mixed with 132Ba–136Ba enriched double spike, cation-exchange chromatography on AG 50-X8 resin | MC-ICP-MS, TIMS | Ba isotope data for 12 geological RMs by MC-ICP-MS and TIMS, reported relative to NIST SRM 310a (Ba solution). For both techniques, intermediate measurement precision (2SD) better than ±0.054‰ in δ137Ba/134Ba | 454 |
Chromites | High pressure/temperature acid digestion using an UltraCLAVE system and a one pass, single column matrix separation on AG 1-X8 anion-exchange resin | MC-ICP-MS | Measurements using 9 FCs in dynamic mode. Reproducibility over several months was 0.11‰ (2SD). NIST SRM 979 (Cr solution) used as δ-zero standard | 28 |
Silicate rock RMs | Li-free glass, prepared from SiO2, MgO, Al2O3, CaO, FeO, TiO2 and Na2O powders, used as a wash sample. Eight RMs and wash sample mounted in epoxy resin for ablation with 193 nm excimer laser | LA-ICP-MS | In situ measurements optimised for signal intensity and isotope ratio stability. SSB approach using USGS glass RM BCR-2G (basalt). Li-free glass ablated as a wash sample to reduce Li memory effect between each sample | 455 |
Iron meteorites and chondrites | Silicates digested in HF–HNO3 and Fe meteorites in HCl. Mo purified using two-stage anion-exchange chromatography on AG 1-X8 resin. Total analytical blank 0.9 ± 0.3 ng, yield 50–65% | NTIMS | Re double filaments with La(NO3)2 added as an electron emitter. Mo isotopes measured as MoO3− on 9 FCs in a single line static method. In situ oxide correction made using 100Mo18O16O2− and 92Mo17O16O2− measured on FCs equipped with 1012 Ω resistors | 456 |
Rock RMs and standard solutions | 100Mo–97Mo double-spike added before stepwise digestion methods according to lithology. Purification by anion/cation-exchange resin double-column procedure using AG 1-X8 (100–200 mesh) and AG 50W-X8 (200–400 mesh) | MC-ICP-MS | SSB and 100Mo–97Mo double-spike methods of correcting for mass fractionation compared. Results consistent, within error, but double spike method preferred for samples with <0.5 μg g−1 Mo. Analytical precision better than ±0.04‰ amu−1 for 101 to 103 ng g−1 Mo | 457 |
Rock RMs and synthetic Nd RMs | Rock powders digested with HF–HNO3; final solution in HCl. 4 column separation of Nd using: (a) AG 50W-X8 200–400 mesh resin (b) Ln-spec 50–100 mm resin, (c) Ln-spec 25–50 mm resin, (d) AG 50W-X8 200–400 mesh resin. Nd recovery was >99% | MC-ICP-MS | Nd isotopes measured in static low-resolution mode with FCs with 1011 Ω resistors; 140Ce and 147Sm monitored using FCs with 1012 Ω resistors to correct for isobaric interferences. SBB approach using GSJ RM JNdi-1 (Nd oxide) | 458 |
Iron meteorites | After leaching in HCl, samples dissolved in HNO3–HCl. Pd purified in a two column anion-exchange procedure using AG 1-X8 resin, with an intermediate evaporation step, to remove Ru, before loading onto the second anion-exchange column. Total yield from both columns was generally >50%. Procedural blanks were <1.2 ng g−1 Pd | MC-ICP-MS | All Pd isotopes measured simultaneously on FCs fitted with 1011 Ω resistors, while also monitoring 101Ru and 111Cd on FCs with 1012 Ω resistors. Instrumental mass fractionation correction via SSB. External reproducibility (2SD) was 1.29 for ε102Pd, 0.22 for ε104Pd, 0.11 for ε106Pd and 0.27 for ε110Pd | 459 |
Iron meteorites and chondritic metal | Dissolution in reverse aqua regia (3 + 1 HNO3 and HCl). After Os extraction using solvent extraction with CCl4, Re purified in 2 column procedure using cation-exchange resin AG 50W-X8 followed by anion-exchange on AG 1-X8 resin. Final solutions spiked with NIST SRM 3163 (W standard) | MC-ICP-MS | All Re isotopic ratios measured on an instrument equipped with 9 FCs and 1011 Ω resistors. Mass fractionation corrected by normalising to 186W/184W = 0.927672 using the exponential law. Precision of δ187Re values was ±0.02‰ (2SE) | 460 |
Chromitites, Os–Ir–Ru alloy grains and meteorites | Chromitites and Os–Ir–Ru alloys digested by the Carius tube technique, meteorite digested with HCl. Purification using a two-step column chemistry with AG 50W-X8 cation-exchange resin and AG 1-X8 anion-exchange resin. Ru further purified by microdistillation | NTIMS | Samples loaded onto single Pt filaments and coated with activator solution of NaOH–Ba(OH)2. Ru isotopes measured as trioxides (RuO3−) in a Static method on 7 FCs. Corrections for possible interferences from Ba, Mo, Pd and Zr were not necessary. External reproducibility for 100Ru/101Ru ±6.4 ppm (2SD) | 461 |
Geological RMs, chromitites | 98Ru–101Ru double spike added prior to sample digestion in Carius tubes. Purification of Ru by cation-chromatography on AG 50W-X8 resin followed by distillation of Ru as volatile oxides. Total recovery of Ru from column chemistry was >95%. Total procedural blank was 41 ± 17 pg Ru (1SD) | MC-ICP-MS | Addition of N2 to the sample gases minimised oxide formation. Simultaneous collection in a single cycle using FCs connected to amplifers with 1011 Ω resistors for Ru isotopes and 1012 Ω resistors for Mo and Pd isotopes. External reproducibility of 0.05‰ for δ102Ru/99Ru (2SD) | 462 |
Solid RMs | Isotopic RMs IAEA-S-1 (solid Ag2S), IAEA-S-2 (solid Ag2S) and IAEA-S-4 (pure S) dissolved by microwave-assisted digestion in HNO3 and H2O2 | SF-ICP-MS, Q-ICP-MS, ICP-MS/MS, MC-ICP-MS | Evaluation of 34S/32S measurements on different single collector ICP-MS instruments and by MC-ICP-MS using different instrumental isotopic fractionation corrections. Combined uncertainties 0.3–0.5% for Q-ICP-MS, 0.08% for SF-ICP-MS, 0.26% for ICP-MS/MS and 0.02% for MC-ICP-MS in edge mass resolution mode | 463 |
Pyrite | Pyrite and silver sulfide RMs were embedded separately in epoxy resin on glass slides and polished to obtain flat surfaces | LA-MC-ICP-MS | Study of S isotope fractionation behaviour in pyrite during LA with 193 nm ArF excimer system. SEM and TEM used to characterise debris formed during LA. SSB approach with matrix-matched pyrite RM recommended with high raster velocity to reduce fractionation | 464 |
Igneous and sedimentary rocks | Silicate rocks decomposed with HF–HClO4–HNO3, dolomitic sample leached with 6 M HCl. Separation of Sm from other REEs achieved by four-step cation-exchange column chromatography procedure using AG 50W-X8 resin. Sm yield >97%, total chemical blank of Sm <20 pg | TIMS | Comparison of unspiked and samples spiked with 150Sm–154Sm. Samples loaded on Re triple filaments together with 0.5 μL of 2 M H3PO4. Isobaric interferences of Nd+ on 144Sm, 148Sm, and 150Sm corrected using the Nd natural isotope ratios. Interference of Gd+ on 152Sm and 154Sm was negligible | 465 |
Cassiterite | Cassiterite powders mixed with KCN and reduced in a muffle furnace at 900 °C. One half of the Sn beads produced was dissolved in HCl, diluted and spiked with Sb. The other half was embedded in epoxy resin for LA | MC-ICP-MS, fs LA-MC-ICP-MS | Comparison of Sn isotope ratios measured by solution and fs LA MC-ICP-MS. In-house RMs defined for SSB, with a pure Sn rod used as the bracketing standard for LA and an Sb solution for mass bias correction | 466 |
Standard solutions | Ta standard solution prepared from high purity Ta metal purchased from the Ames laboratory, USA. Ta metal and commercially-available standard solutions were used throughout this study | MC-ICP-MS | Comparison of data from two different collector setups, involving FCs with 1010 Ω, 1011 Ω, 1012 Ω, and 1013 Ω resistors in different combinations. External mass bias correction using Yb spiked into Ta solutions. Isobaric interferences corrections based on 180Hf/178Hf = 1.2863 and 180W/183W = 0.008304. External reproducibility was ±4ε for 180Ta/181Ta (2SD) | 467 |
Ca–Al-rich inclusions in Allende CV3 chondrite | Synthetic glass RMs were prepared from Ca–Mg–Al–Si oxides to mimic composition of Ti–Al rich pyroxene commonly found in Ca–Al-rich inclusions in chrondites. Glass then powdered and doped with varying amounts of pure oxide powders of Ca, Cr and V, before re-melting and quenching to produce glasses for correction of potential isobaric interferences and matrix effects. These were mounted with an undoped glass in epoxy and polished | LA-MC-ICP-MS | Measurements by MC-ICP-MS in medium resolution in multi-dynamic mode on FCs in two peak jumping steps. Correction of instrumental and natural mass-dependent fractionation using SSB and internal normalisation to 49Ti/47Ti = 0.749766. Isobaric interferences from Ca, Cr and V corrected using the synthetic glasses. External reproducibility (2SD) was ±0.4, ±0.5 and ±1.8 for ε46Ti, ε48Ti and ε50Ti, respectively | 468 |
V-rich minerals and basalt RMs | (a) For in situ analyses, samples mounted in epoxy in separate holding rings. (b) For solution MC-ICP-MS, various digestion schemes depending on starting material. After drying, all samples were treated with HCl. Three column purification to remove, stepwise, Fe, Ti + Zr + Hf, and then the remaining matrix elements | fs LA-MC-ICP-MS, MC-ICP-MS | In situ analyses of stable V isotopes using a 194 nm fs LA system. Solution of Fe standard IRMM-014 aspirated into ablation aerosol for instrumental mass bias monitoring by measuring 57Fe/56Fe; δ51V values determined relative to V metal via SSB. LA and solution data agreed well with each other, typically within 0.1‰ units. Average precision of 0.1‰ (2SD) for δ51V for solution MC-ICP-MS and <0.33‰ (2SD) by fs LA-MC-ICP-MS | 469 |
W solutions and iron meteorites | Commercial W standard solutions measured without any pretreatment. Method said to be applicable to analysis of iron meteorites but no details given | NTIMS | Solutions loaded onto single Re filaments with activator solution containing 5 μg of La and Gd in 5% HNO3. Oxide corrections made using 86W16O218O− and 187Re16O218O− measured on 2 FCs equipped with 1012 Ω resistors. Long-term reproducibilities (2SD) for 182W/184W and 183W/184W better than 5.7 ppm and 6.6 ppm, respectively | 470 |
Metal alloys | NIST SRM 129c (Fe–Ni steel) digested in HNO3–HCl, dried and fluxed in concentrated HCl, dried and redesolved in HCl–HF. Purification of W two-column procedure using AG 1-X8 anion-exchange resin. Total chemical yields 60–80%, full procedural blank 0.26 ± 0.15 ng | MC-ICP-MS | 178Hf, 181Ta and 188Os monitored to correct isobaric interferences on 180W, 184W and 186W. Instrumental mass bias corrected via SSB with NIST SRM 3163 (W solution). External reproducibility (2SD) for ε180W was ±0.49 and ±0.48 using 186W/184W and 186W/183W, respectively, better by a factor of ca. 2.4 compared to previous studies. External precisions (2SD) for other W isotope ratios <±0.08 | 471 |
Essential reading for anyone new to geological applications of LA-ICP-MS is an issue of Elements (volume 12, number 5, 2016, see also Section 4.3.1) which consists of a series of review articles, including one378 (29 references) on the state-of-the-art in isotope ratio determinations by LA-MC-ICP-MS, written by some of the leading practitioners in the field. They highlighted developments in instrumentation and methodology that provided rapid and cost-effective isotopic analysis with considerable flexibility compared to other microanalytical techniques. The ability to acquire isotope maps and depth profiles was seen to be a particular advantage in geoscientific research such as the elucidation of zonation patterns. Continuing challenges such as matrix effects, uncertainty assessment, suitability of RMs and interferences were considered, as well as new hardware advances, such as ICP instruments coupled to TOF analysers.
The application of ICP-MS/MS to the LA analysis of geological materials is particularly advantageous for in situ Rb–Sr and K–Ca dating. Hogmalm et al.415 evaluated N2O and SF6 as reaction gases in preference to O2 for separating 87Sr from 87Rb in the reaction cell. The main reaction products were SrO+ with N2O and SrF+ with SF6; Rb does not react with either gas. Sensitivity was improved by a factor of ca. 10 with N2O and ca. 8 for SF6 compared to when O2 was used. The same precisions for isotopic ratios and Rb–Sr ages could be obtained with spot sizes of 50 and 80 μm. Furthermore, when SF6 was mixed with H2, it was possible to measure 40Ca+ free from 40K+ and 40Ar+ interferences, thereby enabling simultaneous K–Ca and Rb–Sr dating of high-K and low-Ca minerals, such as micas. Ohata et al.416 investigated the potential of ICP-MS/MS for reducing or eliminating polyatomic 16O18O1H and 36Ar1H interferences on 35Cl and 37Cl, respectively, during Cl isotope ratio measurements. The Cl isotope ratios were an order of magnitude more precise when measured using a mixture of H2 and O2 as reaction gas than when no reaction gas was used. In the concentration range 1–10 mg kg−1, the sensitivity and precision (0.2–0.5% RSD) obtained using H2 as reaction gas were similar to those using SF-ICP-MS.
Double-spike methods, commonly utilised for the measurement of radiogenic isotope systems by TIMS for many years, are now being more widely adopted by the MC-ICP-MS community. The double-spike technique normally requires the measurement of four isotopes of the element of interest, but Coath et al.417 demonstrated that it was possible to apply double-spiking to elements with three isotopes, e.g. K, Mg and Si. Precisions and accuracies of around 0.01‰ were obtained for Mg and Si isotopic abundances measured by MC-ICP-MS but, in contrast, the low abundance of 40K and the presence of the 40Ar isobar made K ill-suited to this technique. A new protocol418 used a 180W–183W double spike for the MC-ICP-MS determination of the stable W isotope composition of geological RM, steel, iron meteorite and chondrite samples. An important aspect of this procedure was the efficient chemical separation of Hf from W involving a three-stage anion exchange chromatography prior to measurement to avoid isobaric interferences from highly abundant 180Hf (ca. 35%) on the low abundance 180W (ca. 0.12%). Isobaric dilution analysis (IBDA) was proposed as a novel calibration tool for long-lived radionuclides.419 In contrast to IDA, the spike is not an isotope of the target element but of another element that shares an isobaric isotope and has similar chemical properties, thus potentially providing a cheaper alternative to conventional IDA. The principle was demonstrated by analysis of a mixture of Dy and Gd standard solutions by single collector SF-ICP-MS. These elements have isobaric isotopes at m/z 160. The resulting ratio of m/z 157 to m/z 160 differed by only 0.01‰ from the same ratio determined by IDA. The concept was subsequently applied to the determination of 99Tc in a contrast agent.
The importance of accurate correction of mass-dependent fractionation during Pb isotope measurements was highlighted420 in a MC-ICP-MS study of lavas, from islands in French Polynesia, previously characterised by TIMS for their Hf–Nd–Pb–Sr signatures. The Pb isotope ratios obtained by MC-ICP-MS were significantly different from TIMS literature data and had a much smaller spread. This discrepancy resulted from an improved mass fractionation correction procedure applied in MC-ICP-MS analysis, in which a Tl-doped solution was added to all the sample solutions in a SSB approach. Regional variations in the Pb isotope data seen in previous TIMS studies disappeared when the new MC-ICP-MS data were plotted, thus necessitating a revised geological interpretation of the origin of the lavas. Using silicate glasses as samples, Dai et al.421 employed the LA split-stream (LASS) technique with MC-ICP-MS and quadrupole ICP-MS analyses to obtain in situ major and trace element concentration data and Pb isotope ratio data simultaneously from a single laser spot. Operation of an excimer ArF LA system at a repetition rate of 15–20 Hz and a spot size of 160 μm gave precisions of 0.17%, 0.074%, 6.7%, 3.0% and 4.3% for 208Pb/206Pb, 207Pb/206Pb, 208Pb/204Pb, 207Pb/204Pb and 206Pb/204Pb, respectively, for samples with Pb contents >20 μg g−1. A gas-exchange device used422 in the measurement of Pb isotopes in ore samples by MC-ICP-MS removed Hg vapour from sample aerosol particles produced by LA of mercury-rich sulfides. The device reduced the interference of 204Hg on 204Pb without affecting the Pb signal intensity, thereby improving the within-run precision, external reproducibility and accuracy of the measurements. As noted by other researchers, matrix effects occurred when silicate RMs were used for external standardisation in the analysis of sulfide samples so matrix-matched calibration was necessary. The USGS RM MASS-1 (synthetic polymetal sulfide) and a natural sphalerite Sph-HYLM were demonstrated to be suitable for this purpose.
In the field of LA-ICP-MS U–Pb zircon geochronology, an interesting contribution423 demonstrated the capabilities of a detector system equipped with 16 FCs, 4 EMs and two Daly detectors for the simultaneous detection of 206Pb and 207Pb. The Daly detectors featured better long-term gain stability compared to a conventional EM during 10 h runs monitoring 135Ba/138Ba and 136Ba/138Ba. The wide dynamic range of the Daly detectors (100 to 107 cps) overlapped by several orders of magnitude that of a FC (104 to 1010 cps), making cross-calibration between the detectors possible. The 207Pb/206Pb ratio for the zircon 91500 RM was measured on the Daly detectors with an analytical precision of ca. 5% (n = 15) and a relative deviation from the literature value of −0.04%. This instrument configuration was particularly advantageous for the analysis of young zircons and small sample masses. In another LA-MC-ICP-MS zircon study,424 a multiple ion counting system equipped with six full-size EMs was used to detect simultaneously the ion signals from 202Hg, 204(Hg + Pb), 206Pb, 207Pb, 208Pb and 238U and so achieve high-precision U–Pb isotope ratio measurements at improved spatial resolution. The U–Pb age of 336.3 ± 1.9 Ma obtained for the Plešovice zircon (Czech Republic) using a LA spot size of 25 μm and ablation duration of 2.5 s was more precise than that of 340.3 ± 3.5 Ma achieved by single collector SF-ICP-MS under the same ablation conditions. To take full advantage of the instrumental setup, 1 s ablations were adopted. These produced craters with depths of <1 μm and diameters of 25 μm so, unsurprisingly, the resultant U–Pb age of 339.5 ± 6.7 Ma for Plešovice was significantly less precise but still comparable to the age obtained using conventional ablation with single collector ICP-MS. Marillo-Sialer et al.425 considered radiation-induced structural defects, trace element composition, colour and crystallographic orientation as possible causes of variation in ablation behaviour between different zircon matrices that might bias U–Pb ages as determined by LA-MC-ICP-MS. Crystallographic orientation and colour had a negligible effect compared to the combined influences of radiation damage and cation substitution in some zircon samples. Thermal annealing at temperatures >1000 °C resulted in much more uniform ablation behaviour between different zircon matrices, thereby improving accuracy of U–Pb age determinations significantly.
Guillong et al.426 applied a new 230Th/238U disequilibrium method to obtain age determinations of young zircons (10–300 ka) using single collector LA-SF-ICP-MS. Special attention was paid to the correction of polyatomic oxide interferences on m/z 228 and 230, to abundance sensitivity correction for the tail of the intense 232Th signal on 230Th and to corrections for mass bias and the relative sensitivity of Th and U. A robust procedure for the determination of the LODs, especially that for 230Th, was applied. The method was validated by analysing well-characterised zircons RMs (91500, AusZ-7, Temora2, GJ-1, OD-3, FCT and Plešovice) and samples previously analysed by SIMS. The Th–U ages derived by LA-SF-ICP-MS and their precision (ca. 1%) were comparable with SIMS data for these zircons.
The application of uranium–Pb geochronology to other accessory minerals, such as apatite, is also possible. When Thompson et al.427 used LA with quadrupole ICP-MS to determine the U–Pb ages of several apatite RMs, there was a systematic bias of up to 3% compared to data obtained by IDMS. This bias was attributed to elemental fractionation at the ablation site. Because accurate U–Pb age determinations in apatite are compromised by variable amounts of common Pb incorporated into the crystal during growth, Wohlgemuth-Ueberasser et al.428 used LA-ICP-MS to determine the Pb isotopic composition in a coexisting mineral in order to estimate the initial Pb isotopic composition of apatite. The resulting age information was deemed to be accurate and precise using plagioclase for the common Pb corrections, rather than K-feldspar which is normally used for this purpose. The U–Pb dating of hematite (α-Fe2O3) can provide valuable information on the formation of ore deposits, but relies on the availability of matrix-matched calibration materials for LA-ICP-MS analysis. To address this problem,429 a U–Pb solution in 2% HNO3 with known isotopic ratios was aspirated and mixed in the nebuliser with the aerosol produced by ablating a pure synthetic hematite sample. Apart from dispensing with the need for a isotopically homogeneous natural hematite RM, this approach meant that the concentration and isotope ratios of the solution could be adjusted to match those of the unknown samples, making the method efficient for dating hematite samples containing low (ca.10 μg g−1) to high (>1 wt%) U concentrations. An empirical study430 of allanite U–Pb geochronology using 193 nm LA-ICP-MS examined the use of NIST SRM 610 (trace elements in glass) for calibration to overcome the lack of homogeneous matrix-matched materials. Although such an approach has been attempted before with mixed results, optimised laser operating conditions of 25 s ablations at 1 J cm−2, 3 Hz and a 45 μm crater produced negligible time-dependent Pb/U and Pb/Th fractionation. The results of this study were promoted as a guide for the LA-ICP-MS community towards matrix-independent U–Th–Pb geochronology for a wide variety of major and accessory minerals.
Molybdenum isotope variations in magmatic rocks were the focus of a detailed review431 (229 references), which included an overview of separation techniques using anionic-exchange resins and analytical methods mainly involving MC-ICP-MS combined with double-spike methods. Willbold et al.432 developed a new low-volume, single-pass chromatographic technique for the separation of Mo from silicate and Fe-rich metal matrices prior to MC-ICP-MS analysis. The procedure, which used a combination of a strong anionic-exchange resin with a loading solution of a mixture of HCl and ascorbic acid, was said to be ideal for treating silicate samples with low Mo contents, such as basalts and andesites. An improved methodology433 for the calibration of Mo isotope ratio measurements by MC-ICP-MS employed NIST SRM 987 (Sr isotope standard) to correct for instrumental mass discrimination using a regression model. The novelty of the method was the use of controlled changes in the magnitude of instrumental mass discrimination, achieved by small incremental changes in plasma rf power, to define the relationship between isotope ratios of the analyte (Mo) and the IS (Sr).
The influence of LA parameters on the isotope fractionation of B and Li was explored322 using 193 nm ns and 266 nm fs laser systems coupled to a MC-ICP-MS instrument and crater diameters of 30–200 μm. Higher repetition rates and larger spot sizes increased the fractionation of lighter isotopes by up to 8‰ for δ11B and δ7Li. This was primarily caused by mass loading effects in the ICP, although thermal fractionation at the LA site was also a contributory factor. Thermal effects are usually smaller with fs laser system but, in this study, the shape of the fs laser craters was not very reproducible, resulting in a larger bias compared to the ns LA results. Reproducible ablation of basalts and reference glasses could be achieved by maintaining a constant spot size, a repetition rate of 5 Hz and an energy density of 20 J cm−2. A novel correction for ablation volume resulted in repeatability and laboratory bias better than 1‰ for both δ11B and δ7Li.
The influence of different inorganic acids on B isotope measurements by MC-ICP-MS was studied434 using B solutions containing HCl, HNO3 and HF at concentrations up to 0.2 M. In general, the acids enhanced the B signal intensities and reduced the isotopic mass bias compared to the same B concentration in water. The use of HNO3 was recommended since it yielded similar signal intensities as HCl but did not require rigorous acidity matching to avoid mass bias and was less hazardous than HF, which gave the highest signal intensities. Previous studies have observed differences in MC-ICP-MS and NTIMS measurements of δ11B in marine carbonates. This was investigated435 using a suite of 18 samples, comprising inorganic calcite grown under controlled pH conditions and planktonic foraminifera of Quaternary age. The NTIMS results were 0.5–2.5‰ higher than those from MC-ICP-MS but the cause for this offset remained unclear and could not be attributed to varying B/Ca ratio in NTIMS or the increased sample processing necessary for MC-ICP-MS. However, measurements of inorganic calcites precipitated from parent solutions of varying pH confirmed that both techniques showed the same relationship of pH to calcite δ11B within uncertainty, indicating that relative δ11B changes generated from NTIMS and MC-ICP-MS measurements were comparable.
An innovative analytical method144 involving on-line separation of Sr from the sample matrix by IC prior to MC-ICP-MS analysis provided high throughput, robust and sensitive on-line measurements of87Sr/86Sr ratios with minimum contamination in complex natural samples including carbonate and silicate rocks. The precision of 0.003% (2σ) was only 2–3 times worse than could be obtained by labour- and time-intensive off-line TIMS or MC-ICP-MS techniques. A thorough investigation436 of potential Ca- and P-based interferences, which could have a significant influence on Sr isotope ratio measurements in bioapatite and calcium carbonate matrices by LA-ICP-MS, underlined the complex nature of such analysis. Correction of instrumental isotopic fractionation was identified as the biggest challenge. Accurate Sr isotope ratios could be obtained provided adequate corrections for interferences and instrumental isotopic fractionation were incorporated into the analytical protocol and consideration was given to the resulting expanded uncertainties, which might limit the use of LA-MC-ICP-MS depending on the particular application. Another contribution437 on 87Sr/86Sr ratios measurements by LA-MC-ICP-MS focused on the analysis of fossil teeth. The major polyatomic interference at mass 87 was 40Ar31P16O rather than 40Ca31P16O and could significantly bias results towards higher values compared to TIMS analyses, especially for samples with low Sr mass fractions. Tuning the instrument to low oxide production rates reduced the offset between TIMS and LA-MC-ICP-MS 87Sr/86Sr ratios from 0.00266 obtained without special tuning to an absolute mean difference of 0.00003 (n = 21).
A method438 for high precision measurements of Os isotopes by MC-ICP-MS and NTIMS, incorporating a 188Os–190Os double spike, permitted simultaneous measurement of 187Os/188Os and 186Os/188Os ratios as well as the stable Os isotope composition expressed as δ190Os. Relative to the IAG DROsS Os isotope RM, the internal precision was 0.01–0.02‰ for δ190Os and <40 ppm for 187Os/188Os and 186Os/188Os for both techniques. This study reported the first stable Os isotope compositions for four geological RMs: Durham University in-house RM GP-13 (lherzolite) and CRPG RMs UB-N (serpentinite), CHR-Bkg (chromitite) and CHR-Pt+ (chromitite). The Os ratios were the same within the analytical uncertainty and so provided a preliminary δ190Os value of +0.130 ± 0.032‰ (2SD; n = 4) for the Earth’s upper mantle. Ultra-fine sulfide powder pellets and fused beads, used as calibrants for in situ measurements of Os isotope ratios in sulfides by LA-MC-ICP-MS, were prepared439 from CNRCG RMs GBW07268 (chalcopyrite) and GBW07267 (pyrite) doped with standard solutions of Ir, Os, and Rh. The homogeneity of the powder pellets for 187Os/188Os and 187Re/185Re was assessed by line scans and spot analyses. Greater phase separation was detected in the chalcopyrite fused beads than in the pyrite ones. Using the pressed pellets as matrix-matched RMs and applying a modified method to calculate mass fractionation coefficients, data for 187Os/188Os in sulfides were within 0.5% of the reference values for samples with 187Re/188Os <0.6.
To date, very few measurements have been made of the isotopic composition of Ga in geological samples. Zhang et al.164 developed a three-column ion-exchange chromatography method for the determination of Ga isotopes by MC-ICP-MS in a range of geological materials. One anion-exchange AG MP-1 M column and two cation-exchange AG 50W-X8 columns were employed in the purification scheme. The SSB technique corrected for instrument mass bias using Cu as the IS. The bracketing standard was a 200 ng g−1 solution of NIST SRM 994 (Ga metal) and the IS was prepared from NIST SRM 3114 (copper solution). The long-term reproducibility for δ71Ga was 0.04‰ (2SD). Interestingly, Kato et al.440 described a similar three-step ion-exchange chromatography protocol to separate Ga from rock samples prior to the determination of their Ga isotopic composition by MC-ICP-MS. Initially, matrix elements were removed on anion-exchange resin AG MP1, then Fe was separated from Ga on AG 50W-X12 and finally Ba was separated using a similar elution protocol to step 1 but with much smaller volumes. Analysis of a range of mafic rocks from geological settings yielded almost identical δ71Ga values of 0.00 ± 0.06 (2SD) relative to a commercial ICP Ga standard solution. This was interpreted as the best estimate for the Ga isotopic composition of the bulk silicate Earth. External reproducibility was 0.05‰ based on replicate measurements of basalts USGS BHVO-2 and BCR-2. However, in this study, all the δ71Ga data were calculated relative to a commercial ICP Ga standard solution and not NIST SRM 994. Consequently, the δ71Ga values for USGS RMs BCR-2 and BHVO-2 (basalts) differed greatly from those reported by Zhang et al.164 A two-stage purification method for the determination of Ga isotope ratios by MC-ICP-MS was based441 on columns containing AG 1-X4 and Ln-spec resins. The reproducibility of the MC-ICP-MS analyses was better than 0.05‰ (2SD). Some of the measurements were performed in the IPGP laboratories in Paris where the protocol of Kato et al.440 was developed, so similarities in some of the methodology is not surprising. Although it is likely that the large spread of −0.05‰ to +0.74‰ in δ71Ga values for BHVO-2 reported from these three studies can be explained by differences in correction strategies and the RMs employed, there is insufficient information to be certain. Unfortunately, none of the publications offered information on how well the 71Ga/69Ga ratios measured for the bracketing RMs agreed with certified or published data. What is clear is the Ga isotope community needs to address the issue of an internationally recognised delta zero for this system with some urgency!
Particle size distributions in aerosols produced during fs and ns LA were found to have a significant influence on the accuracy of stable Fe isotope ratio measurements by LA-MC-ICP-MS.442 The ablation characteristics of the ns laser (193 nm) varied according to the substrate and laser fluence and resulted in variable particle morphology, particle size distributions and Fe isotope fractionations, with possible non-stoichiometric sampling of different natural Fe minerals, including magnetite (Fe3O4), siderite (FeCO3), pyrrhotite (Fe4S5), and pyrite (FeS2). In contrast, fs LA (266 nm) produced more consistent ablation as these dependencies were minimised. It was concluded that fs LA was superior to ns LA because of the more uniform particle size distributions and stoichiometric sampling. The advantages and disadvantages of cold plasma conditions443 for Fe isotopic analysis by MC-ICP-MS were evaluated under wet and dry aerosol conditions and different instrument cone configurations. When standard interface cones were in place, changing from a hot to a cold plasma reduced Fe instrumental mass discrimination but with a high-efficiency jet interface mass discrimination was about the same under both plasma conditions. A cold plasma combined with pneumatic nebulisation (wet plasma) suppressed the spectral interferences 40Ar16O+ and 40Ar14N+ but not 40Ar16O1H+. Analysis of several geological RMs, including USGS PCC-1 (peridotite) and BHVO-2G (basalt) and CRPG IF-G (iron formation), demonstrated that accurate 56Fe/54Fe measurements could be achieved in low resolution mode using a cold plasma and standard interface.
In an investigation of factors affecting instrumental mass bias when measuring Si stable isotopes by MC-ICP-MS under dry plasma conditions, Oelze et al.444 discovered that the addition of Mg was very beneficial. Not only did the addition of Mg reduce variations in mass bias but it also increased Si ion intensities. In addition, variations in Mg and Si concentrations and HCl molarity between samples and the bracketing standards were tolerated to a much greater degree without affecting the instrumental mass bias. Two approaches for the simultaneous in situ determination of the elemental composition and stable Si isotope ratios of geological samples were evaluated.445 The first used LASS with MC-ICP-MS and quadrupole ICP-MS detection and the second one employed direct observation of the emission profile of an MC-ICP-MS plasma. A fs LA system provided the sample aerosol to both setups. International geological RMs were used to evaluate both methods, which provided accurate δ30Si data with an uncertainty of better than 0.23‰ (2SD). The LASS method was considered to be superior for the measurement of element concentrations down to the μg g−1 level because of its higher sensitivity and the ability to determine trace as well as major element concentrations.
Discrepancies in δ26Mg data obtained by different laboratories for the San Carlos olivine prompted a thorough investigation of the homogeneity of this material.446 Analytical artefacts (e.g. levels of matrix removal, low Mg yields of column chemistry, mismatch of Mg concentrations and acid molarity) and changes in δ26Mg values of standard solutions during long-term storage were responsible for the differences rather than any heterogeneity in the Mg isotopic composition of the olivine. The new δ26Mg data were homogeneous to within 0.03‰ and therefore consistent with the narrow range of δ26Mg values of mantle olivine worldwide. Thus it was confirmed that San Carlos olivine is a suitable RM for Mg isotope analysis.
The deterioration of FCs449 caused by strong ion beams is an issue when making high-precision isotope ratio measurements by TIMS. A comparison of the long-term fluctuations and reproducibility of Nd isotope ratios measurements acquired over 8 months in three different modes (static, dynamic and multistatic) revealed that static methods yielded high-precision data only when all the FCs were relatively new. Data from the multistatic method were more sensitive to deterioration of the FCs than those from the dynamic method, which therefore was most effective in minimising FC degradation. Li et al.450 developed a method for Cr isotope ratio measurements based on an improved two-column chemical separation procedure and addition of a highly sensitive Nb2O5 emitter to enhance the Cr ionisation efficiency. The 10-fold increase in sensitivity provided by the Nb2O5 emitter allowed the amount of sample on the filament to be reduced from 200 ng to 5 ng. In addition, the use of W filaments instead of the more commonly used Re filaments reduced costs significantly. The isotopic fractionation of Ca during column chemistry followed an exponential law.451 The more Ca that was eluted, the lower the δ44Ca/40Ca value of the eluate became. Instrumental fractionation deviated from an exponential law when mixing of sample reservoirs on the filaments occurred. A model for checking the behaviour and degree of instrumental fractionation was proposed as a quick check on the TIMS data quality. Fukami et al.452 reported an optimised analytical protocol for precise isotope ratio measurements of sub-ng amounts of Pb by total evaporation TIMS. Parameters examined included different ionisation activators, quantity of activator, width of the sample on the filament, various filament heating regimes and data reduction strategies. The method was suitable for use with a 204Pb–207Pb double spike, which is an advantage for laboratories that have difficulty in sourcing 202Pb–205Pb spike solutions.
Vollstaedt et al.453 investigated the potential and limitations of Se isotope ratio measurement with state-of-the-art TIMS instrumentation in static mode. The NTIMS analysis provided better precision than HG-ICP-MS with fewer isobaric interferences, which can be a big drawback in the measurement of Se by ICP-MS. However, Se memory effects accumulated during analysis and could not be removed sufficiently by cleaning techniques. This led to the degradation of precision for δ80Se/78Se over time. The authors concluded that if these memory problems could be solved, the accuracy and precision of NTIMS analysis of Se isotope ratios would be superior to those by HG-ICP-MS.
Additional contributions to high-precision isotope ratio determinations by TIMS are included in Table 7.
The relative sensitivity factors (RSFs) of REEs determined in the nanoSIMS analysis of zircon, apatite and NIST SRM 610 (trace elements in glass) varied475 considerably so it was concluded that absolute concentrations of REEs could only be determined using calibrations based on matrix-matched RMs. Steele et al.476 highlighted the difficulty in making accurate Cr/Mn measurements in carbonate minerals by SIMS because of matrix effects. The Cr/Mn RSFs for a range of carbonate phases implanted with Cr varied systematically with the chemical composition of the different carbonate minerals. An empirical correction scheme was therefore proposed.
The increasing importance of stable isotope analysis by SIMS is reflected in the number of newly-developed isotope RMs for SIMS analysis. Such RMs are vital because matrix-matched calibrations are essential for this technique. In addition to zircon M127-A333 (see Section 4.1), gem-quality apatites were evaluated as potential oxygen RMs.477 A suite of carbonate RMs was developed for the calibration of oxygen and carbon isotope measurements.478,479 Two quartz samples, UNIL-Q1 (Torres del Paine Intrusion, Chile) and BGI-Q1 (Shandong province, China), previously characterised as RMs for trace element analysis by Audétat et al.,480 were proposed481 as suitable RMs for SIMS analysis. The SIMS measurement precisions were <0.41‰ (2SD) for UNIL-Q1 and <0.48‰ (2SD) for BGI-Q1. Reference δ18O values were determined by CO2 laser fluorination. Monazite RMs are scarce. Three monazites,482 from Brazil, Madagascar and Namibia, were considered to be potential RMs for O isotope analysis because of their homogeneous chemical and oxygen isotope composition. Uncertainties for the SIMS measurements were in the range of 0.3 to 0.4‰ (2SD); δ18O values by laser fluorination were also available. To widen the applicability of SIMS oxygen isotope analysis to solid solution minerals, the homogeneity of two new pyroxene RMs (augite NRM-AG-1 and enstatite NRM-EN-2) was evaluated.483 As these materials showed no bias caused by crystallographic orientation and were homogeneous at a scale of 20 μm, they were considered to be suitable RMs for SIMS δ18O analysis.
Instrumental mass fractionation (IMF) during the measurement of stable isotopes by SIMS occurs at several stages of the analysis and is strongly influenced by the major element composition of the target material. To correct for these matrix effects, Fabrega et al.484 developed a predictive surface response methodology, based on multivariate statistics, and applied it to the measurement of δ18O in plagioclase, K-feldspar and quartz. The great advantage of SIMS for research into carbon sequestration is its ability to analyse very small sample volumes. For example, in situ δ13C and δ18O measurements could be made485 on single carbonate crystallites as small as 3–10 μm in width with sub per-mil (‰) precision and accuracy. The study focused on characterising the mineralogy and isotopic fingerprinting of underground rock formations prior to CO2-injection and made predictions regarding the δ13C values of carbonate mineral cements that may form in response to long-term storage of CO2. The oxygen isotope composition of coal and kerogen samples486 was determined by SIMS on organic matter residues isolated from a quartz matrix. The procedure included a method of filtering out δ18O data from the oxygen-bearing micro-inclusions of various mineral phases present in the impure OM. In order to constrain the influence of δ37Cl values derived from primary magmas, olivine-hosted melt inclusions were targeted for Cl isotope analyses. Data from different volcanic arc settings were obtained487 using an experimental setup designed specifically for the analysis of glasses containing >900 μg g−1 Cl. Two FCs were used instead of an EM to improve the measurement precision; total uncertainties were 0.4‰ (2SD). Observed variations in instrumental mass fractionation were mainly caused by the presence of Al2O3, K2O and SiO2.
Sharpe and Fayek488 demonstrated that the mass bias for U–Pb isotope ratio measurements in uraninite increased with increasing Pb content. A three-point calibration curve, based on uraninite RMs with different Pb concentrations, facilitated the accurate determination of U–Pb ratios. Alternatively, mass bias correction could be achieved by using a uraninite RM with a Pb content similar to that of the unknown samples. Bastnaesite is a major economic REE mineral in which the incorporation of Th is favoured relative to that of U. A comparison of ages gained by U–Pb dating of zircons and Th–Pb dating of bastnaesite provided489 information on intrusion of the host magma and the mineralisation ages of REE deposits in the Himalayan Mianning-Dechang REE belt. Monazite RW-1 was characterised490 as a RM for SIMS U–Pb and Th–Pb isotope analyses.
A novel SIMS scanning ion-imaging technique was developed491 to investigate μm-scale U–Pb systematics in a lunar zircon with a complex internal structure. The spatial resolution was improved by a factor of four compared to traditional SIMS analyses, facilitating the analysis of features of the zircon that could not be accessed by traditional SIMS methodology.
A novel approach for measuring 14C, using a combination of LA sampling with accelerator mass spectrometry (AMS), provided flexible and time-resolved acquisition of 14C profiles in carbonates.493 The considerable advantages over conventional gas measurements offered by the LA-AMS technique included the speed of analysis (because no time-consuming sample preparation was required), low sample consumption and higher spatial resolution. Depending on the measurement mode, typical precisions of 1–5% combined with a spatial resolution of 100 μm were obtained for speleothems and a variety of terrestrial and marine carbonates.
A new resonance ionisation mass spectrometer,494 known as the Chicago instrument for laser ionisation (CHILI), was developed specifically to measure isotopic ratios in small samples such as interstellar dust at high spatial resolution and high sensitivity. This RIMS instrument was capable of determining all the stable isotopes of Fe and Ni simultaneously in presolar SiC grains when a special timing scheme for the ionisation lasers was used to separate the Fe and Ni isotopes in the TOF spectrum. The lateral resolution was 0.8 μm: the ultimate goal was to achieve 10 nm resolution with 30–40% useful yield.
Two studies applied isotope ratio mass spectrometry (IRMS) to the determination of clumped isotopes in carbonates. Schauer et al.495 showed that the choice of 17O correction strategy could have a large effect on the accuracy of Δ47 values. Differences as large as 0.1‰ obtained in an inter-laboratory comparison were thought to have their origins in the 17O correction strategies applied. Wang et al.496 systematically evaluated the effectiveness of a Porapak™ Q absorbent trap (PQT) for purifying CO2. Comparison of the Δ47 values obtained from different experiments showed that a setup with CO2 passing through the PQT driven by liquid N2 and high vacuum pumping gave the most promising results, because it saved time without compromising the extraction efficiency. An elemental analyser (EA) equipped with purge-and-trap technology and interfaced to an IRMS instrument was used to perform high-precision S isotope measurements in bioapatite.497 These determinations are normally limited by the low S content of such samples but results from this study demonstrated the capacity of this analytical methodology to discriminate between marine, freshwater and terrestrial environments based on δ34S values from modern and fossil vertebrate bioapatite containing as little as 0.14–1.19 wt% S. This methodology has potential for characterising the living environment of an extinct vertebrate purely from its fossilised bones and teeth. The redox conditions for the extraction of N from an organic-rich RM prior to measurement by EA IRMS were investigated as these were known to affect the accuracy and precision of δ15N data.498 Injection of more than 20 mL O2 into the reaction tube resulted in improved combustion and better N-yields than when CuO or V2O5 was added.
The ultimate aim of Hoegg et al.499 in combining a liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasma ion source with an Orbitrap mass analyser was to produce an IRMS instrument capable of being deployed in the field. Their current system had minimal mass bias for 235U/238U, better than that obtained with beam-type mass analysers, and a dynamic range of 4 orders of magnitude. The precision obtained with this system500 was ca. 0.1% RSD for U isotope ratio measurements but poorer (1–3% RSD) for low abundance species.
Younis et al.504 used only two RMs to establish WDXRFS calibrations for elements of interest. The dilution factor in a fusion bead technique was varied to generate a wide range of concentrations from each RM. Calibration lines for major elements, including Al, Ca, Fe, Mg, Mn, Na, Si and Ti, were constructed from fused beads which had Li2B4O7 flux: sample ratios of 2 to 54. As a result, RMs with matrices similar to those of the samples to be analysed could be selected. This approach was validated by analysing several USGS geological RMs with different matrices.
Elemental variations in drill cores can be assessed by the fast and nondestructive technique of XRFS core-scanning. Six GSJ geological RMs were analysed to determine the influence of exposure times and choice of X-ray tube.505 Increasing the counting time had only a minor effect on the accuracy of measurements. Thus, caution was required when interpreting data for elements present at low concentrations or low sensitivities, e.g. Al, even when the counting times were long and scanning counts reasonably high. Based on linear regression lines and correlation coefficients of 17 elements with different exposure times and X-ray tubes, the use of a Mo X-ray tube rather than a Cr one was recommended for all XRFS-scanning experiments. Interstitial water can give rise to serious matrix effects during XRFS core-scanning so a new polynomial correction for water content was implemented506 for two existing calibration methods. This removed the influence of changes in water content in long sediment sequences.
Combining optical imaging and XRFS analysis had507 the aim of determining whether images of rock surfaces could be used to predict the elements present in a sample. The ultimate goal would be the remote characterisation of inaccessible underground mine voids. The machine learning algorithm developed was able to predict the elemental composition with a high degree of accuracy. However, the success of this approach was likely to depend on the suite of minerals present and the ability of expert geologists to identify minerals through visual inspection.
Portable XRFS instruments are valuable for informing rapid decision-making. The fast determination of the concentration of Cu and contaminants in Cu ore slurry allowed512 the slurry composition to be modified, if necessary, before entering a high-capacity smelter, thereby improving the smelting and refining process. Quantitative high-density data for drilled igneous rocks were obtained513 on-board a ship during an IODP expedition in order to improve the selection of samples for further analysis and provide key information for drilling operations. A combination of pXRFS and benchtop SEM was used514 in the field to characterise the igneous stratigraphy in a drill core and its relationship to Cu–Pd mineralisation. The geochemical data were used to inform the sampling strategy and increase the efficiency of mineral exploration. Mineralogical and geochemical characterisation of drill fines from a borehole by a combined pXRFS and portable XRD approach enabled geologists to identify different lithologies during drilling and modify their strategy as necessary.515 In this context, a comparison516 of pXRFS measurements on unprepared cores and their equivalent powders showed that data quality was not compromised if an average of multiple measurements of different points on the intact sample were taken. Between three and seven measurements were recommended for a significant improvement in precision.
Various algorithms and software packages for handling isotopic data have been made available to the scientific community, including ET_Redux,517 a free open-source software for U–Pb geochronology using LA-ICP-MS data. An excellent description was provided of the principles of LA-ICP-MS data handling, statistical analysis of the data and uncertainty propagation, as well as the underlying algorithms. The new software was developed as a part of a collaborative effort to aid the compilation, comparison and archiving of data from different laboratories. A novel modelling procedure,518 coded in R, has been made available for the analysis of U–Pb datasets that contain discordant analyses, often the case when analysing detrital zircons. A new R-based software package,519 IsotopeHf, was developed for handling data from Lu–Hf ID analyses by MC-ICP-MS.
For elemental analysis, an R script program520 called TERMITE provided very fast data reduction of trace element measurements obtained by the LA-ICP-MS of samples such as carbonates with a homogeneous matrix. The underlying calculations were described in detail. A Matlab-based program,521 AERYN, improved the reliability of element distribution maps acquired by techniques such as EPMA and LA-ICP-MS. A new Matlab algorithm522 was developed to quantify K, Th and U from natural gamma radiation spectra generated on-board ships. The performance of this algorithm was demonstrated by comparing results with those obtained using shore-based ICP-MS, ICP-AES and WDXRFS techniques.
2D | Two-dimensional |
3D | Three-dimensional |
AAS | Atomic absorption spectrometry |
AES | Atomic emission spectrometry |
AFS | Atomic fluorescence spectrometry |
AMS | Accelerator mass spectrometry |
APDG | Atmospheric pressure glow discharge |
AR | Alpha Resources Inc. (USA) |
ASU | Atomic Spectrometry Update |
BCR | Community Bureau of Reference (of the European Community) now IRMM |
C18 | Octadecyl bonded silica |
CA | Chemical abrasion |
CCP | Capacitively coupled plasma |
CCRMP | Canadian Certified Reference Materials Project |
CE | Capillary electrophoresis |
CNRCG | China National Research Centre for Geoanalysis |
CPE | Cloud point extraction |
CRDS | Cavity ring-down spectroscopy |
CRM | Certified reference material |
CRPG | Centre de Recherches Pétrographiques et Géochimiques (France) |
CS | Continuum source |
CV | Cold vapour |
DA | Discriminant analysis |
DCM | Dichlormethane |
DDC | Diethyldithiocarbamate |
DGT | Diffusion gradient in thin films |
DLLME | Dispersive liquid liquid microextraction |
DMA | Dimethylarsenic acid |
DME | Dispersive microextraction |
DOC | Dissolved organic carbon |
DRC | Dynamic reaction cell |
DRS | Diffuse reflectance spectrometry |
EDS | Energy dispersive X-ray spectroscopy |
EDTA | Ethyldiaminetetraacetic acid |
EDXRFS | Energy dispersive X-ray fluorescence spectrometry |
EM | Electron multiplier |
EN | European Committee for Standardisation |
EPA | Environmental Protection Agency (USA) |
EPMA | Electron probe microanalysis |
ERM | European Reference Material |
ETAAS | Electrothermal atomic absorption spectrometry |
ETMAS | Electrothermal molecular absorption spectrometry |
ETV | Electrothermal vaporisation |
FAAS | Flame atomic absorption spectrometry |
FC | Faraday cup |
FFF | Field flow fractionation |
FI | Flow injection |
fs | Femtosecond |
FTIR | Fourier transform infrared |
GBW | CRMs of the National Research Centre for Certified Reference Materials (China) |
GC | Gas chromatography |
GD | Glow discharge |
GEOSTATS | Geostats Pty Ltd (Australia) |
GLS | Gas liquid separator |
GSB | CRMs of the Institute for Environmental Reference Materials (of Ministry of Environmental Protection, China) |
GSJ | Geological Survey of Japan |
GS-MS | Gas source mass spectrometry |
HG | Hydride generation |
HILIC | Hydrophobic interaction liquid chromatography |
HPLC | High performance liquid chromatography |
HPS | High Purity Standards (USA) |
HR | High resolution |
IAEA | International Atomic Energy Agency |
IAG | International Association of Geoanalysts |
IC | Ion chromatography |
ICP | Inductively coupled plasma |
ICP-AES | Inductively coupled plasma atomic emission spectrometry |
ICP-MS | Inductively coupled plasma mass spectrometry |
ICP-MS/MS | Inductively coupled plasma mass spectrometry with a quadrupole-cell-quadrupole design |
id | Internal diameter |
ID | Isotope dilution |
IDA | Isotope dilution analysis |
IGGE | Institute of Geochemical and Geochemical Exploration (China) |
INAA | Instrumental neutron activation analysis |
INCT | Institute of Nuclear Chemistry and Technology (Poland) |
IMS | Ion mobility spectrometry |
IODP | International Ocean Discovery Program |
IPGP | Institute de Physique du Globe de Paris (France) |
IRMM | Institute for Reference Materials and Measurements |
IRMS | Isotope ratio mass spectrometry |
IS | Internal standard |
IUPAC | International Union of Pure and Applied Chemistry |
LA | Laser ablation |
LASS | Laser ablation split stream |
LGC | Laboratory of the Government Chemist (UK) |
LIBS | Laser induced breakdown spectroscopy |
LIF | Laser induced fluorescence |
LLE | Liquid liquid extraction |
LLME | Liquid liquid microextraction |
LOD | Limit of detection |
LOQ | Limit of quantification |
MAE | Microwave assisted extraction |
MC | Multicollector |
MIL | Magnetic ion liquid |
MIP | Microwave induced plasma |
MIR | Mid infra red |
MMA | Monomethylarsenic acid |
MP | Microwave plasma |
MRI | Magnetic resonance imaging |
MS | Mass spectrometry |
MWCNT | Multiwalled carbon nanotube |
m/z | Mass to charge ratio |
NBL | New Brunswick Laboratories (USA) |
NCS | National Analysis Centre for Iron and Steel (China) |
Nd:YAG | Neodymium doped:yttrium aluminum garnet |
NEXAFS | Near edge X-ray absorption fine structure |
NIOSH | National Institute of Occupational Safety and Health |
NIR | Near infrared |
NIST | National Institute of Standards and Technology (USA) |
NKK | Nippon Keikinzoku Kogyo (Japan) |
NMIA | National Measurement Institute of Australia |
NP | Nanoparticle |
NRCC | National Research Council (of Canada) |
NRCCRM | National Research Centre for Certified Reference Materials (China) |
ns | Nanosecond |
NTIMS | Negative thermal ionisation mass spectrometry |
NWRI | National Water Research Institute (Canada) |
PGE | Platinum group element |
PLS | Partial least squares |
PLSR | Partial least squares regression |
PM2.5 | Particulate matter (with an aerodynamic diameter of up to 2.5 μm) |
PM10 | Particulate matter (with an aerodynamic diameter of up to 10 μm) |
ppm | Parts per million |
PTE | Potentially toxic element |
PTFE | Poly(tetrafluoroethylene) |
PVG | Photochemical vapour generation |
pXRFS | Portable X-ray fluorescence spectrometry |
QC | Quality control |
REE | Rare earth element |
RIMS | Resonance ionisation mass spectrometry |
RM | Reference material |
rpm | Revolutions per minute |
RSD | Relative standard deviation |
RTC | RT Corporation (now Sigma-Aldrich RTC) (USA) |
SBME | Stir bar microextraction |
SD | Standard deviation |
SEC | Size exclusion chromatography |
SEM | Scanning electron microscopy |
SF | Sector field |
SFODME | Solidification of a floating organic droplet microextraction |
SI | Système international (d’unités) (International System of Units) |
SIMS | Secondary ion mass spectrometry |
SMPS | Scanning mobility particle sizer |
sp | Single particle |
SPE | Solid phase extraction |
SQT | Slotted quartz tube |
SPS | Spectrapure Standards (Norway) |
SR | Synchrotron radiation |
SRM | Standard reference material (of NIST) |
SSB | Sample-standard bracketing |
SVM | Support vector machine |
SXRFS | Synchrotron X-ray fluorescence spectrometry |
TBA | Tributylamine |
TEA | Transverse excited atmospheric |
TEM | Transmission electron microscopy |
TEOM | Tapered element oscillating microbalance |
TIMS | Thermal ionisation mass spectrometry |
TLC | Thin layer chromatography |
TOF | Time of flight |
TXRFS | Total reflection X-ray fluorescence spectrometry |
UAE | Ultrasonic extraction |
USGS | United States Geological Survey |
USN | Ultrasonic nebulisation |
UV | Ultra violet |
VALLME | Vortex assisted liquid liquid microextraction |
VIS | Visible |
WDXRFS | Wavelength dispersive X-ray fluorescence spectrometry |
XRD | X-ray diffraction |
XRF | X-ray fluorescence |
XRFS | X-ray fluorescence spectrometry |
This journal is © The Royal Society of Chemistry 2018 |