Zuzana
Gajdosechova
* and
Joshua
Marleau-Gillette
National Research Council Canada, Metrology, 1200 Montreal Rd, Ottawa, ON K1A 0R6, Canada. E-mail: Zuzana.Gajdosechova@nrc-cnrc.gc.ca
First published on 19th April 2024
The popularity of vaping cannabis-derived concentrates has grown rapidly since its legalization in several countries around the world. There is some limited evidence that metal-bearing nanoparticles (NPs) may be present in cannabis vape liquids. However, owing to the complexity of the cannabis vape liquids matrix, data on particle number concentration and particle size distribution are not available. Single particle inductively coupled plasma mass spectrometry (ICP-MS) is a technique widely used to obtain this information for samples in aqueous suspension; however, cannabis vape liquids are not water soluble. Additionally, this technique has not been well tested on naturally formed polydisperse nanoparticles, such as those found in cannabis vape liquids. An approach based on single particle analysis in organic mode ICP-MS is proposed for assessing particle number concentration and particle size distribution in cannabis vape liquids. Three approaches for determining transport efficiency were assessed, and particle sedimentation under various conditions was studied. Finally, cannabis vape liquids from nine cartridges were analyzed for several metal-bearing particles (Fe, Ni, Al, Co, Cu, Cr, V, Zn, Pb, and Sn). All samples contained a substantial number of NPs, with three samples having a number of particles one to two orders of magnitude higher than the rest of the studied samples, and the particle size distribution was, in most instances, below 150 nm for all metals.
Although regulations in countries where the consumption of cannabis products is legalized require testing for arsenic (As), cadmium (Cd), mercury (Hg), and lead (Pb), recent publications have found significant levels of other metals in tested cannabis vape liquids.15–18 Specifically, in several tested cannabis vape liquids, cobalt (Co) and vanadium (V) were found almost 10 times, chromium (Cr) and copper (Cu) 100 times, and nickel (Ni) 1000 times over the permitted levels of elemental impurities in inhaled products established by the European Pharmacopoeia.15 The studies reported large relative standard deviations of replicate measurements, which could be caused by several factors. Cannabis vape liquids have high viscosity; thus, poor sample homogeneity can have a significant impact on the reproducibility of the analysis. Kubachka et al.16 investigated several homogenization strategies and reported a decrease in the relative standard deviation between replicates when samples were heated at 120 °C for 30–60 min and continuously stirred with a stirrer bar rather than pipette tip, a comparative stirring method. Currently, cannabis vape liquids are sold pre-filled in single-use cartridges or disposable pens, generally containing between 0.5 to 1 g of product. Similar to studies of nicotine vape liquids, there is evidence that metal contaminants are leached from the metal parts of vaping cartridges.15 Therefore, depending on the quality of the metal components, contamination may vary significantly between samples of the same production lot. Such observations were recently reported where identical cannabis vape liquids obtained from two vaping cartridges were visually different and contained significantly different amounts of metals.15 Therefore, combining mass fractions obtained from identical cannabis vape liquids, but originating from two vape cartridges, could also have a significant contribution to high relative standard deviations. Another major contributor to low reproducibility are metal particles, the presence of which was recently identified by scanning electron microscopy and laser ablation inductively coupled plasma mass spectrometry.15 Naturally formed metal particles generally vary in size, and particles of uneven size, although homogeneously distributed within the sample matrix, will significantly impact the reproducibility of the measurement.
Metal NPs have been shown to be transported into the aerosol generated by nicotine vape devices,19,20 and as the mechanism of inhalation of cannabis vape liquids is identical, it is very likely that users of cannabis vape liquids are also exposed to metal NPs. The exposure levels may, however, differ due to difference in the vape liquid matrices, the designs, and operation conditions of vape devices, which may have an impact on the transport efficiency of the metal particles. Inductively coupled plasma mass spectrometry operated in a single particle mode (SP-ICP-MS) is often a method of choice for analysis of metal NPs. It provides information about the elemental composition of the NPs, the particle number concentration (PNC), particle size distribution (PSD), and the mass fraction of dissolved analytes. However, in contrast to nicotine vape liquids, cannabis vape liquids are not water soluble and instead require dilution into an organic solvent. Although ICP-MS is perfectly suitable for the analysis of organic solvents, there is only a handful of publications discussing measurements of NPs in organic matrices. The earliest analyses found their applications in the petrochemical industry and included investigations of naturally formed Fe, Mo,21 and HgS22 NPs in petroleum hydrocarbons. An additional study by Nelson et al.23 looked at the stability and accuracy of the detection of engineered Fe3O4 NPs dispersed in crude oil. A more recent study focused on sizing TiO2 NPs in sunscreen products using a microdroplet injector, which minimized the organic solvent load in the plasma and increased the transport efficiency of the sample.24 The general instrument requirements for the analysis of NPs in organic matrices include the addition of O2 into plasma to convert the carbon-based matrix into CO2, eliminating the deposition of soot on the interface cones. As a result of O2 addition, the plasma temperature increases, requiring the substitution of nickel interface-cones with platinum cones. The optimization of the introduction system (i.e., the liquid flow rate and nebulizer gas flow rate) plays a crucial role as it determines the amount of aerosol reaching the plasma and, therefore, the ionization and atomization characteristics of the plasma. Both the sample and nebulizer flow rates are generally decreased compared to aqueous analysis, resulting in a lower amount of solvent being introduced into the plasma and increasing the plasma residence time of the NP. Increased plasma residence time is particularly important for the analysis of NPs to ensure complete ionization and, hence, accurate size determination of large NPs. As suggested by others, the plasma residence time of particles can be further increased by adjusting the plasma gas flow rate, the power for plasma generation, inner diameter of the injector and sampling position in the plasma.25
The aim of this study was to characterize metal particles present in widely available cannabis vape liquids purchased from the Canadian legal market. In the process, an investigation was conducted into the suitability of various methods (dynamic mass flow and size method using ligand transfer, and sequential dilution of the reference material (RM)) for the determination of NP transport efficiency, a crucial parameter for accurate sizing of NPs. Additionally, we assessed and compared the sedimentation of several reference materials in organic solvent and highlighted the possible implications this may have on PNC. Finally, the optimized method was applied to the analysis of nine cannabis vape liquids and python-based open-source SPCal software was used for data interpretation of NPs containing ten different metals.
Sample ID | Packaging date | Mass (g) | Δ9-THC | Total THC | CBD | Total CBD | Additional information |
---|---|---|---|---|---|---|---|
1 | 31-Mar-2023 | 1 | 825 | 825 | 2 | 2 | Glass housing and ceramic core 510 thread cartridge |
2 | 29-Aug-2023 | 1 | 766 | 809 | <10 | <10 | 510 thread cartridge |
3 | 03-Jul-2023 | 1.2 | 770.0 | 770.0 | 12.5 | 12.5 | 510 thread cartridge |
4 | 18-Sep-2023 | 1 | 763.5 | 763.5 | 0.0 | 0.0 | Terpinolene 5.8%; ocimene 1.5%; limonene 0.9%, 510 thread cartridge with metal components coated in black paint |
5 | 03-Jul-2023 | 1 | 320.548 | 723.117 | 0 | <6 | Total terpenes 9.077%: myrcene, pinene, limonene, 510 thread cartridge with metal components coated in white paint |
6 | 05-Jun-2023 | 1 | 850 | 850 | <6.0 | <6.0 | Pax cartridge |
7 | 07-Jun-2023 | 1 | 810 | 810 | 3.4 | 3.4 | 510 thread cartridge |
8 | 28-Aug-2023 | 1.2 | 833.3 | 833.3 | <6.0 | <6.0 | 6% terpenes: terpinolene, myrcene, ocimene, limonene, caryophyllene, 510 thread cartridge |
9 | 08-Oct-2023 | 1 | 866 | 900 | <2.5 | <2.5 | 510 thread cartridge with glass inner channel |
10 | 11-Aug-2023 | 0.5 | 858 | 858 | 2.5 | 2.5 | Disposable vape pen |
Another portion of the citrate-stabilized 60 nm Ag NPs reference material was transferred into toluene through a ligand and solvent transfer reaction, as described elsewhere.26 Briefly, 1 mL of Ag NPs was added to 1 mL of 10 mM DDT in trichloromethane, followed by the addition of approximately 5 mL of acetone to form a single-phase solution. Subsequently, the solution was shaken for 30 s and centrifuged at 12000g for 2 min to sediment DDT-capped NPs. The sedimented NPs were washed twice with ethanol/trichloromethane (5:1 v/v) before being dispersed in toluene by 30 s of shaking and 5 min of sonication.
Tune mode | NH3 | O2 |
---|---|---|
Scan mode | MS/MS | |
RF power (W) | 1500 | |
Sampling depth (mm) | 10 | |
Nebulizer gas (L min−1) | 0.8 | 0.6 |
Option gas: O2 (L min−1) | 0.3 (30%) | |
Spray chamber temp (°C) | 5 | |
He flow (mL min−1) | 1.0 | — |
NH3 flow (mL min−1) | 3.0 (30%) | — |
O2 flow (mL min−1) | — | 0.38 (25%) |
Octopole bias (V) | −7.5 | −3.0 |
Axial acceleration (V) | 2.00 | 2.0 |
Energy discrimination (V) | −7.2 | −8.0 |
Quantitation was performed using a multi-level external calibration method, and transport efficiency (also called nebulization efficiency) was compared among two methods. Firstly, using particle size method dilution of citrate-stabilized Ag NPs into PremiSolv via PGME or from the size of DDT-capped Ag NPs. Secondly, by using dynamic mass flow (DMF) described elsewhere.28 Briefly, a container with PremiSolv was placed on an analytical balance with both the sample and waste tube immersed in the container and allowed to equilibrate for 30 min. The mass change (i.e., the mass flow of sample reaching the plasma) was monitored for 60 min, and the change in the mass was recorded every 10 min. The recorded weights of the mass reaching the plasma were plotted against the time at which the weight was recorded (slope 1). To determine the mass flow of the sample uptake, the waste tubing was removed from the container and allowed to equilibrate for 30 min. The sample uptake was monitored for 15 min, and the mass change was recorded every 5 min. The recorded weights of the sample uptake were plotted against the time at which the weight was recorded (slope 2). The transport efficiency (η) was calculated using eqn (1):
(1) |
A ligand exchange reaction was performed to determine if the currently available NP RM in a water-based matrix could be used for organic solvent analysis. Following the ligand transfer reaction, DDT-capped 60 nm Ag NPs were dispersed in toluene and examined by transmission electron microscopy, which demonstrated the absence of aggregation and particles of a spherical shape without any substantial changes to the particle size (data not shown). The measured particle size by SP-ICP-MS was 59 ± 2 nm (n = 7), which was in good agreement with the size provided by the manufacturer (59 ± 6 nm). The ligand exchange reaction was not a quantitative reaction, as a visible amount of Ag coated the walls of the reaction container, so transport efficiency using this material could be calculated using only the particle size and not the particle frequency method. In comparison with the decreased polarity dilution transfer of NPs, the transport efficiency using DDT-capped Ag NPs was, on average, 0.3% lower and thus could be considered a suitable option for preparation of RMs in organic solvent to calculate transport efficiency.
Lastly, the use of citrate-stabilized 60 nm Ag NPs sequentially diluted via PGME into organic solvent resulted in a transport efficiency of around 3.0%, when particle size was used for the calculation. However, when particle frequency was used, the transport efficiency was significantly lower, around 1.8%, resulting in an incorrect size determination of the tested RMs. When using particle frequency for the determination of transport efficiency, the calculation was based on the accurate concentration of aspirated NPs in the solution. Clearly, the particle frequency method underestimated the transport efficiency. The transport efficiency using particle frequency method is expressed as a ratio between the number of observed particle events and the number of introduced particles in the suspension of known PNC. The accuracy of this method can be hindered by particles loss along the sample introduction pathway which could happened during addition of O2 gas. Oxygen is being added in the elbow connector of the spray chamber which may introduce turbulence to the sample flow and hence disperse some NPs which eventually don't reach the plasma.
Although the addition of 10 mM DDT in combination with sonication and vortexing appeared to be an ideal condition for the analysis, the addition of 10 mM DDT into the cannabis vape liquid resulted in phase separation. Cannabis vape liquids contain a large number of organic complexes that may be sensitive to the polarity of the matrix. Therefore, the cannabis vape liquids were diluted in the organic solvent without DDT. The diluted sample was placed into the autosampler, and measurement of Co and Sn NPs were recorded every 5 min for a duration of 25 min, which was the total time required for the analysis of one sample. As seen in Fig. 1d., the sedimentation trend varied significantly between the two studied analytes. Cobalt NPs exhibited sedimentation over the studied time interval, whereas Sn NPs displayed a random change in the detected particle number.
All samples were analyzed for NPs containing ten elements (Fe, Ni, Al, V, Cu, Co, Zn, Cr, Pb, and Sn), which were found to be present in most of the samples. Wash samples containing only the diluent (organic solvent) were measured between each cannabis sample to monitor NP carry-over. Most of the wash samples contained, on average, ≤20 NPs of selected analytes, with the exception of Cu, which had an average count of 77 NPs, Fe with an average count of 233 NPs, and Al with an average count of 245 NPs. Unfortunately, Fe and Al NPs seemed to be present in the diluent, as a similar average count of these NPs was detected throughout the entire work with this solvent.
It should be noted that some samples contained a large number of metal particles whose signal distribution appeared as if these samples contained several distinct sizes of NPs. Fig. 2a shows the signal distribution of Cu in sample 8, which was diluted 13-fold. The signal distribution could be interpreted as corresponding to NPs of several distinct sizes; however, after additional dilution (100-fold), the signal distribution changed to one typically observed for samples containing polydisperse NPs (Fig. 2b). However, the detected particle number did not correspond to the dilution factor applied. The number of particles detected in the 13-fold diluted sample 8 was calculated as 8133, but after 100-fold dilution, the number of particles decreased only slightly to 7388. Clearly, not all particles have been counted in the 13-fold diluted sample.
Fig. 2 Signal distribution of Cu detected in sample 8 (a) after a 13-fold dilution, and (b) after an additional 100-fold dilution. |
In an ideal system, the particle number should decrease linearly with an increased dilution factor. However, in polydisperse samples, such as samples with naturally formed NPs, this non-linear decrease in PNC may be caused by several factors. Firstly, with high PNC, there is a very large probability of multiple particle events, which will result in the underestimation of the PNC and overestimation of the PSD. Thus, when the sample is being diluted, the number of multiple particle events is reduced, and hence the PNC doesn't follow a linear decrease. Secondly, with the introduction of sample dilution, the mass fraction of dissolved analyte is diluted, resulting in a lower particle detection threshold. Consequently, smaller particles become detectable and increase the PNC. Therefore, several dilution factors should be tested when analyzing polydisperse NP samples as it was previously suggested by others.33,34
NPs present in cannabis vape liquids are naturally formed, most likely as a result of corrosion or thermal expansion and cooling-induced degradation, such as chipping or cracking of the metal components. Due to the size heterogeneity of the NPs and the possible presence of dissolved ionic fractions of the studied analytes, assigning the signal produced by NPs and dissolved analytes was quite challenging. As seen in Fig. 3, there was a distinct drop in the frequency of signal distribution between dissolved and particulate Ag when the RM was analyzed, known as the particle detection threshold (Fig. 3a), which was absent in the real samples with a heterogeneous size distribution (Fig. 3b).
Although both samples contained a similar number of NPs (2837 for Ag RM and 2487 for Sn in the cannabis vape liquids), assigning the particle detection threshold in the cannabis samples was challenging based on visual examination of the signal distribution chart. The calculation of the particle detection threshold in the software provided by the instrument manufacturer was determined by the approximation of the ionic signal portion as an exponential function, as described in detail elsewhere.35 Briefly, the signal distribution plot was constructed by integration of the raw signals, which were then counted within each period. Applying an exponential function to the data points in the signal distribution plot, plurality of approximate curves approximating the ionic signal portion was calculated. Then, the coefficient of determination (R2) was calculated for each approximation curve, and the one with the maximum correlation was selected. Data points corresponding to the curve with the maximum correlation were selected as the particle detection threshold.
However, because the raw data are treated as a chromatographic spectrum a continuous uniform baseline has been applied, and only the signal above the baseline was integrated. This approach introduced several inaccuracies. The criterion for the baseline selection is not transparent and because the baseline is continuous over the entire acquisition time, corrections for possible signal fluctuation are not possible. More importantly, application of the uniform baseline results in the reduction of peak area and possible negative bias in counting of the small particles ultimately leading to larger size distribution and detection of fewer particles. Therefore, the raw data were processed using an open-source software SPCal, applying iterative automatic threshold detection method. Depending on the signal intensity, the software automatically switches between Poisson filter (<50 counts) and Gaussian filter (>50 counts).36 The critical value (LC), defined as signal intensity that is significantly higher than the background and limit of detection (LD) were first calculated based on selected expression for paired observations.37 The signal above the LD was then background-corrected and integrated to generate a single particle distribution plot. Additionally, to reduce the signal intensity contribution from dissolved ions in the sample and thus reduce the LD, samples were sequentially diluted. The raw data in this study were processed using both Currie's expression and Formula C, and significant differences were found in the output results. When applying Currie's expression (Fig. 4a), a higher PNC was obtained (3191 particles) with a lower mean PSD (34 nm) compared to Formula C (Fig. 4b), which counted 696 particles with a mean PSD of 53 nm. Upon closer examination of the raw signal, it was observed that Formula C filtered out a larger number of data points belonging to the dissolved analyte. As a result of its higher filtration efficiency, i.e., a more conservative approach, the PSD seemed to be fully distinguished from the dissolved background. On the other hand, the PSD was incomplete, limited by the LOD, when Currie's expression was applied. Based on these observations, data were processed using Formula C are reported.
Fig. 4 Comparison of particles size distribution plots obtained by data processing using Carrie's expression (a) which counted 3191 particles and Formula C (b) which counted 696 particles. |
As shown in Table 3, all studied samples contained metal particles. Samples 5, 8, and 9 had the highest number of particles, generally one or two orders of magnitude higher than the rest of the studied samples. Samples 8, 9, and 10 belonged to the same producer but were contained in different designs of vape devices. Samples 8 and 9 were in a vape cartridge, which was mounted by a user on a reusable battery, whereas sample 10 was packaged in a disposable pen. Additionally, the inner channel of the vape cartridge of sample 8 was coated with black paint, while the inner channel of sample 9 was made of glass. Sample 5 was produced by a different manufacturer and was contained in a cartridge of a similar design to most of the studied samples. The inner channel of the cartridge was coated in white paint. Similarly, sample 4 had the inner channel of the cartridge coated in black paint, and contrary to samples 5 and 8, the PNC was within the range of the majority studied samples. Thus, without deeper investigation into the composition of the cartridge parts, it was not feasible to explain why samples 5, 8, and 9 contained a significantly higher PNC of all studied metals. Comparison of PNC between different samples can be somewhat misleading, because the amount of dissolved analyte has a direct impact on the LOD, which affects the PNC; thus, samples with considerably different LODs may not be directly comparable.
Analyte | Sample ID | Mean size (nm) | LOD (nm) | NPs/device | Analyte | Sample ID | Mean size (nm) | LOD (nm) | NPs/device |
---|---|---|---|---|---|---|---|---|---|
a NQ – not quantified due <100 detected particles per minute of acquisition time. | |||||||||
Fe | 1 | 71 | 30 | 1.52 × 105 | Co | 1 | NQ | NQ | ND |
2 | NQ | NQ | NQ | 2 | NQ | NQ | ND | ||
3 | 103 | 53 | 6.32 × 105 | 3 | 66 | 28 | 5.32 × 105 | ||
4 | 75 | 32 | 2.96 × 105 | 4 | NQ | NQ | ND | ||
5 | 63 | 29 | 1.30 × 108 | 5 | 65 | 29 | 1.44 × 106 | ||
7 | 122 | 67 | 1.98 × 106 | 7 | NQ | NQ | ND | ||
8 | 60 | 27 | 8.37 × 107 | 8 | 83 | 40 | 3.86 × 106 | ||
9 | 68 | 28 | 1.07 × 108 | 9 | 77 | 30 | 6.20 × 106 | ||
10 | 96 | 45 | 3.60 × 107 | 10 | 63 | 36 | 2.76 × 107 | ||
Ni | 1 | NQ | NQ | NQ | Zn | 1 | NQ | NQ | ND |
2 | NQ | NQ | NQ | 2 | NQ | NQ | ND | ||
3 | 110 | 46 | 6.24 × 105 | 3 | 141 | 68 | 2.29 × 106 | ||
4 | NQ | NQ | NQ | 4 | NQ | NQ | ND | ||
5 | 109 | 43 | 3.80 × 106 | 5 | 161 | 81 | 3.55 × 107 | ||
7 | 107 | 43 | 2.28 × 105 | 7 | 179 | 87 | 2.28 × 105 | ||
8 | 98 | 43 | 8.69 × 107 | 8 | 121 | 61 | 9.03 × 107 | ||
9 | 120 | 51 | 7.61 × 106 | 9 | 181 | 84 | 5.22 × 106 | ||
10 | 104 | 42 | 2.19 × 106 | 10 | 127 | 68 | 1.24 × 107 | ||
Al | 1 | 120 | 47 | 1.12 × 105 | Cr | 1 | NQ | NQ | ND |
2 | 149 | 67 | 3.17 × 105 | 2 | 87 | 44 | 4.52 × 105 | ||
3 | 177 | 110 | 1.40 × 106 | 3 | 72 | 29 | 3.91 × 105 | ||
4 | 135 | 49 | 2.36 × 105 | 4 | NQ | NQ | ND | ||
5 | 76 | 39 | 1.14 × 108 | 5 | 69 | 30 | 1.50 × 106 | ||
7 | 154 | 81 | 9.34 × 106 | 7 | 69 | 30 | 6.41 × 105 | ||
8 | 54 | 26 | 1.74 × 108 | 8 | 76 | 33 | 6.42 × 106 | ||
9 | 90 | 44 | 8.59 × 107 | 9 | 75 | 30 | 3.82 × 106 | ||
10 | 98 | 44 | 5.51 × 106 | 10 | 73 | 29 | 5.00 × 105 | ||
V | 1 | NQ | NQ | NQ | Pb | 1 | NQ | NQ | NQ |
2 | 99 | 67 | 5.81 × 106 | 2 | NQ | NQ | NQ | ||
3 | NQ | NQ | NQ | 3 | NQ | NQ | NQ | ||
4 | NQ | NQ | NQ | 4 | NQ | NQ | NQ | ||
5 | 54 | 29 | 5.49 × 105 | 5 | NQ | NQ | NQ | ||
7 | NQ | NQ | NQ | 7 | NQ | NQ | NQ | ||
8 | 59 | 34 | 6.61 × 105 | 8 | 67 | 33 | 9.35 × 105 | ||
9 | 53 | 28 | 2.37 × 105 | 9 | 75 | 42 | 7.88 × 105 | ||
10 | NQ | NQ | NQ | 10 | ND | ND | ND | ||
Cu | 1 | NQ | NQ | NQ | Sn | 1 | NQ | NQ | NQ |
2 | NQ | NQ | NQ | 2 | NQ | NQ | NQ | ||
3 | 86 | 37 | 1.82 × 106 | 3 | 97 | 39 | 4.79 × 105 | ||
4 | 74 | 31 | 7.37 × 104 | 4 | NQ | NQ | NQ | ||
5 | 95 | 35 | 7.09 × 106 | 5 | 110 | 38 | 8.07 × 105 | ||
7 | 103 | 41 | 2.57 × 105 | 7 | NQ | NQ | NQ | ||
8 | 207 | 94 | 5.84 × 108 | 8 | 145 | 66 | 1.20 × 107 | ||
9 | 81 | 34 | 5.57 × 106 | 9 | 110 | 42 | 2.08 × 107 | ||
10 | 93 | 45 | 8.92 × 106 | 10 | 98 | 41 | 4.31 × 105 |
Vanadium is rarely reported in studies of vaping devices; however, small quantities of V were previously measured in cannabis vape liquids,15 and V NPs were detected in some of the studied samples (Table 3). The PNC below the level of quantitation was measured in majority of samples analysed for Pb particles. Samples in which <100 particles were measured within a minute of acquisition time were excluded from reporting, as 100 particle events is the minimum number concentration limit of quantitation under ideal counting conditions.37 As mentioned previously, Al- and Fe-containing NPs were detected in the diluent; however, the PNC in the samples was >1500 particle events, whereas the blanks systematically contained <250 NPs. Therefore, Al and Fe PNCs reported in Table 3 were corrected for the average PNCs measured in the blanks; however, the PSD may be skewed by the PSD found in the blanks.
The PSD varied considerably between analytes. While the mean size of particles containing V, Co, Cr, and Pb was <100 nm in all samples, the mean size of particles containing Zn was >120 nm in all instances. The mean sizes of particles containing other metals were mostly distributed in the range of 60 to 150 nm. Similar mean sizes of metal particles were previously reported in the aerosols from nicotine vape devices,19 where also a considerable size variation between aerosol generated from different nicotine devices was observed. However, as discussed previously, the presence of dissolved analytes in the samples may have led to an overestimation of particle size, and thus, the reported PSD and PNC in this study should be taken as an informational value only. The presence of dissolved analytes in both the sample and diluent matrix have had a direct impact on the LOD of the measured particles. For some metals such as Co, the LOD range across studied samples has shown relatively narrow variability, ranging from 28 to 44 nm. In contrast, the LOD for Al has exhibited a wider range, varying between 26 and 110 nm. Additionally, the calculated particle size is influenced by the assumption that the particles are spherical and possess a density equivalent to that of a bulk metal, which may not hold true for naturally formed composite particles.
The analyzed cannabis vape liquids were found to contain a large number of NPs with various metals. Three analyzed samples exhibited a particle count one or two orders of magnitude higher than the rest of the studied samples, and the particle size distribution for most metals was predominantly below 150 nm. The presence of dissolved analytes and polydisperse particles in the studied samples was evident and may have influenced the reported PNC in a negatively biased manner and the PSD in a positively biased manner. Consequently, it is advisable to consider the reported values as semi-quantitative. Moreover, although a significant number of metal-bearing particles were detected in cannabis vape liquids, it does not guarantee that all these particles can be transported into the aerosol, and further studies should be conducted to investigate the presence of NPs in aerosols.
This journal is © The Royal Society of Chemistry 2024 |