Issue 12, 2022

High-resolution continuum source graphite furnace molecular absorption spectrometry for the monitoring of Sr isotopes via SrF formation: a case study

Abstract

High-resolution continuum source graphite furnace molecular absorption spectrometry (HR CS GFMAS) can provide isotopic information under certain conditions, thus broadening its field of application. However, to date, only elements with two major stable isotopes have been monitored via this technique. In this work, the possibilities of HR CS GFMAS to determine isotope ratios of elements with more than two stable isotopes are evaluated for the first time. For this purpose, Sr was chosen as the analyte and SrF as the target species, so four different signals corresponding to four stable Sr isotopes (88Sr, 87Sr, 86Sr and 84Sr) should be distinguished. Nevertheless, due to the number of strontium isotopes, the shape of the peaks, and the resolution that the instrument exhibits in the spectral window, isotopic signals overlap, thus leading to potentially biased results. To circumvent this issue, a deconvolution protocol, consisting of measuring and correcting for the contribution of each isotope on the signals of the rest, was developed. These contributions were calculated as the signal ratio between the absorbance of the monoisotopic profile at the wavelengths where the maxima of other isotopes are expected and at its own maximum. Therefore, the interference can be simply subtracted from the net signal registered for the interfered isotope. The performance of this method was demonstrated for both naturally abundant and isotope-enriched Sr standards, paving the way for future applications in this field. Analysis of a real sample (tap water) spiked with a 84Sr solution is also demonstrated.

Graphical abstract: High-resolution continuum source graphite furnace molecular absorption spectrometry for the monitoring of Sr isotopes via SrF formation: a case study

Supplementary files

Article information

Article type
Paper
Submitted
15 जुलाई 2022
Accepted
29 सितम्बर 2022
First published
30 सितम्बर 2022

J. Anal. At. Spectrom., 2022,37, 2517-2528

High-resolution continuum source graphite furnace molecular absorption spectrometry for the monitoring of Sr isotopes via SrF formation: a case study

A. Bazo, R. Garde, E. Garcia-Ruiz, M. Aramendía, F. V. Nakadi and M. Resano, J. Anal. At. Spectrom., 2022, 37, 2517 DOI: 10.1039/D2JA00245K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements