Issue 1, 2016

Use of Ga for mass bias correction for the accurate determination of copper isotope ratio in the NIST SRM 3114 Cu standard and geological samples by MC-ICPMS

Abstract

The absolute determination of the Cu isotope ratio in NIST SRM 3114 based on a regression mass bias correction model is performed for the first time with NIST SRM 944 Ga as the calibrant. A value of 0.4471 ± 0.0013 (2SD, n = 37) for the 65Cu/63Cu ratio was obtained with a value of +0.18 ± 0.04‰ (2SD, n = 5) for δ65Cu relative to NIST 976. The availability of the NIST SRM 3114 material, now with the absolute value of the 65Cu/63Cu ratio and a δ65Cu value relative to NIST 976 makes it suitable as a new candidate reference material for Cu isotope studies. In addition, a protocol is described for the accurate and precise determination of δ65Cu values of geological reference materials. Purification of Cu from the sample matrix was performed using the AG MP-1M Bio-Rad resin. The column recovery for geological samples was found to be 100 ± 2% (2SD, n = 15). A modified method of standard-sample bracketing with internal normalization for mass bias correction was employed by adding natural Ga to both the sample and the solution of NIST SRM 3114, which was used as the bracketing standard. An absolute value of 0.4470 ± 0.0013 (2SD, n = 37) for 65Cu/63Cu quantified in this study was used to calibrate the 69Ga/71Ga ratio in the two adjacent bracketing standards of SRM 3114, their average value of 69Ga/71Ga was then used to correct the 65Cu/63Cu ratio in the sample. Measured δ65Cu values of 0.18 ± 0.04‰ (2SD, n = 20), 0.13 ± 0.04‰ (2SD, n = 9), 0.08 ± 0.03‰ (2SD, n = 6), 0.01 ± 0.06 (2SD, n = 4) and 0.26 ± 0.04‰ (2SD, n = 7) were obtained for five geological reference materials of BCR-2, BHVO-2, AGV-2, BIR-1a, and GSP-2, respectively, in agreement with values obtained in previous studies.

Graphical abstract: Use of Ga for mass bias correction for the accurate determination of copper isotope ratio in the NIST SRM 3114 Cu standard and geological samples by MC-ICPMS

Article information

Article type
Paper
Submitted
24 दिसम्बर 2014
Accepted
06 मार्च 2015
First published
24 मार्च 2015

J. Anal. At. Spectrom., 2016,31, 280-287

Author version available

Use of Ga for mass bias correction for the accurate determination of copper isotope ratio in the NIST SRM 3114 Cu standard and geological samples by MC-ICPMS

Q. Hou, L. Zhou, S. Gao, T. Zhang, L. Feng and L. Yang, J. Anal. At. Spectrom., 2016, 31, 280 DOI: 10.1039/C4JA00488D

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements