Issue 9, 2022

In situ measurement of Sm–Nd isotopic ratios in geological materials with Nd < 100 μg g−1 by LA-MC-ICP-MS

Abstract

In this study, we developed a method for measuring Sm–Nd isotopic ratios in samples with low Nd contents (<100 μg g−1) by LA-MC-ICP-MS. The use of jet sample and X skimmer cones, with the addition of a small water vapour flow (∼5 mL min−1) significantly improved the instrumental sensitivity by a factor of 1.9–3.6, and has little effect on Nd instrumental mass bias. The accuracy and precision of our analytical method is demonstrated by analysis of eight geological reference glasses (Nd = 21.4–60.9 μg g−1). The obtained Sm–Nd isotope data are consistent with their recommended values. The measurement of a clinopyroxene megacryst (JLM-1) indicates that matrix effects on 147Sm/144Nd calibration between clinopyroxene and the measured geological reference glasses are insignificant. The inverse correlation between the analytical precision and the Nd signal intensity shows that the 143Nd/144Nd internal precision is <120 ppm (2s) for a 146Nd signal >0.15 V, whereas the 143Nd/144Nd internal precision is <80 ppm (2s) for a 146Nd signal >0.3 V. The addition of water vapour did not suppress the matrix effects for 147Sm/144Nd calibration. The Sm–Nd isotopic compositions of the Panzhihua clinopyroxene measured with our method indicate that the Panzhihua layered intrusion behaved as a closed magmatic system that experienced limited crustal contamination. The method presented in this study is a rapid and accurate technique for in situ analysis of Sm–Nd isotopes of low-Nd samples, such as volcanic glass, clinopyroxene and hornblende.

Graphical abstract: In situ measurement of Sm–Nd isotopic ratios in geological materials with Nd < 100 μg g−1 by LA-MC-ICP-MS

Supplementary files

Article information

Article type
Technical Note
Submitted
26 मई 2022
Accepted
20 जुलाई 2022
First published
27 जुलाई 2022

J. Anal. At. Spectrom., 2022,37, 1776-1786

In situ measurement of Sm–Nd isotopic ratios in geological materials with Nd < 100 μg g−1 by LA-MC-ICP-MS

L. Zhang, F. Yang, L. Hong, Y. Zhang, J. Soldner, Y. Zhang and Z. Ren, J. Anal. At. Spectrom., 2022, 37, 1776 DOI: 10.1039/D2JA00181K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements