Águeda
Cañabate
a,
Esperanza
García-Ruiz
b,
Martín
Resano
b and
José-Luis
Todolí
*a
aDepartment of Analytical Chemistry, Nutrition and Food Sciences, University of Alicante, P. O. Box 99, 03080 Alicante, Spain. E-mail: jose.todoli@ua.es
bDepartment of Analytical Chemistry, Aragón Institute of Engineering Research (I3A), University of Zaragoza, Pedro Cerbuna 12, 50009, Zaragoza, Spain
First published on 21st November 2016
In this work, the performance of a high temperature torch Integrated Sample Introduction System (hTISIS) using inductively coupled plasma mass spectrometry (ICP-MS) for the multi-element analysis of whole blood is evaluated. Two different quadrupole-based spectrometers were tested and the optimization studies were performed with a reference whole blood material. Sample dilution factor and hTISIS temperature were taken as the variables. The sample was introduced following two different procedures: continuous aspiration and air-segmentation. The optimum performance of the system in terms of both analytical figures of merit and accuracy was obtained for a hTISIS temperature of 200 °C and for either 1:10 or 1:25 dilution factors depending on the ICP-MS device used. An improvement in terms of sensitivity and detection limits was obtained with the hTISIS as compared to the conventional cyclonic spray chamber. Thus, for 1:25 diluted blood samples, the hTISIS improved the sensitivity by a factor ranging 2.0–4.5. Moreover, at 200 °C, the limits of detection for hTISIS at 200 °C were 1.1 to 8.4 times lower than those for the cyclonic spray chamber. This analytical parameter was in the ng L−1 range for the detected elements. Matrix effects in turn became less severe as the hTISIS temperature increased (from room temperature to 80 °C). All these results were obtained without a severe degradation of the fundamental plasma parameters. In fact, under optimum conditions BaO+/Ba+ and Ba2+/Ba+ signal ratios were 1.2% and 2.0%, respectively. The developed method was applied for the analysis of low volumes (ca. 2.5 μL) of real blood samples after a minimally invasive collection by volumetric absorptive microsampling. The hTISIS is an easy-to-implement sample introduction system that can be employed for the routine analysis using ICP-MS.
Element determination in blood samples has been carried out via techniques such as atomic absorption spectroscopy,4–8 total reflection X-ray fluorescence (TXRF),5,9 and hydride generation atomic fluorescence spectroscopy (HG-AFS).10 However, due to the increasing demand for the determination of a number of elements in blood, alternative analytical methods are required.
Inductively coupled plasma optical emission spectroscopy (ICP-OES),5,11,12 and particularly, mass spectrometry (ICP-MS)5,13–17 offer a multi-element measurement capability, low detection limits,18 a wide dynamic range, and short analysis time.16 However, these techniques suffer from spectral as well as non-spectral interferences that may degrade the accuracy of the blood analyses.19
Interferences arising from the atomic and molecular ions generated in the plasma through the combination of argon and matrix constituents hinder ICP-MS quantification.17 In the case of blood analysis, the most severe specific spectral interferences are due to the presence of chloride, sodium, sulfur or carbon.17,19,20–23 These interferences can be minimized or even removed using a reaction and/or a collision cell.14,17,18,24,25
Non-spectral interferences can also be observed while carrying out the analysis of blood samples. Therefore, high concentrations of inorganic and organic solutes may cause changes in the aerosol generation, transport and/or properties of the plasma. Furthermore, dissolved salts may lead to a partial blockage of the injector torch or the ICP-MS cones.10 To overcome some of these problems, internal standardization or standard addition is classically employed.15
Moreover, matrix effects can also be minimized by lowering the sample consumption rate and using a total sample consumption system. This way, the matrix plasma load is decreased and the interferences associated with the sample introduction system may be removed. The latter assessment is based on the fact that with such system, the analyte transport efficiency is close to 100%, regardless of the sample nature.26,27 The so-called high temperature torch integrated sample introduction system (hTISIS) consists of a micronebulizer connected to a low inner volume single pass spray chamber.26–29 The main advantages of the hTISIS over a conventional spray chamber are as follows: (i) analyte transport efficiencies are virtually 100%; as a result, (ii) very low sample volumes (ca. 2–5 μL) are required to perform the analysis; thus, (iii) the solvent plasma load is also low; (iv) memory effects are practically negligible; and (v) virtually no waste is generated.30 Besides, the extent of the matrix effects caused by both organic and inorganic concomitants is less pronounced as compared to that of the conventional cyclonic spray chamber.28 To date, hTISIS has been applied for the accurate analysis of a few environmental, biological, and clinical reference materials.27,31,32
Thus, the goal of the present work was to test the suitability of the hTISIS, operated at a high temperature, as a system for the analysis of samples of a very complex nature, but of great importance (such as blood) by ICP-MS. Relevant parameters, such as the hTISIS chamber temperature and sample dilution factors, were optimized for the blood reference samples. The analytical figures of the merit were compared against those obtained with a conventional sample introduction system.
Single element stock solutions (CertiPUR, Merck) containing the required element at a 1000 mg L−1 concentration were used to prepare the standards (10 μg L−1 and 1 μg L−1). These standards also contained Ba at 10 μg L−1, whereas Rh at 25 μg L−1 was employed as the internal standard.
The reference materials analyzed were Seronorm Trace Elements Whole Blood L-2 and Seronorm Trace Elements Whole Blood L-3 (Sero, Billingstad, Norway; Reference 210205 and 210305, respectively). These lyophilized human whole blood materials were reconstituted in 5 mL Milli-Q water, according to the instructions provided in the certificate. Then, aliquots of the L-2 solution were diluted to 1:10, 1:25, 1:50, and 1:100 with a 1% HNO3 aqueous solution and L-3 was diluted to 1:10. Real samples were obtained from three different adult individuals.
The sampling procedure was performed via the so-called volumetric absorptive microsampling (VAMS) method. A puncture with a safety lancet was performed on the finger of several individuals and a small volume (10.5 μL) was retained on the absorptive tip of the VAMS microsampler (Mitra™, Neoteryx LLC, Torrance, USA). Afterwards, the samples were dried under ambient conditions for 4 hours followed by the extraction of the absorbed samples with 100 μL of 1% HNO3 aqueous solution using an ultrasonic bath (generator power: 50 W, and frequency: 50 Hz) for 1 hour to promote the extraction of the analytes. Finally, the analysis of the resulting solutions was performed by external calibration with a set of plain water standards. Anonymous samples were provided by the Hospital Universitario Miguel Servet, Zaragoza. The principles outlined in the declaration of Helsinki regarding all the experimental research involving humans or animals were followed. In addition, an informed consent was obtained from the participants involved.
Perkin-Elmer NexION 300X | Agilent 7700x | |
---|---|---|
Chamber temperature | RT-200 °C | 200 °C |
Nebulizer gas flow rate | 1.1 L min−1 | 0.5 L min−1/0.56 L min−1 HMI |
Sample uptake rate | 33 μL min−1 | |
RF power | 1600 W | 1600 W |
Plasma gas flow rate | 18 L min−1 | 15 L min−1 |
Auxiliary gas flow rate | 1.2 L min−1 | 0.90 L min−1 |
Cell gas flow rate (He) | 0.8 mL min−1 | 4.3 mL min−1 |
Moreover, an Agilent 7700x ICP-MS (Agilent, Santa Clara, USA) was also used. The instrument is equipped with a High Matrix Introduction (HMI) accessory that increases the tolerance to salty matrices as compared to that of the conventional ICP-MS instruments. The sample introduction system used with this instrument was hTISIS and a High Efficiency Nebulizer (HEN, Meinhard Glass Products).
Samples were supplied to the nebulizer both in continuous and discrete modes with the spectrometer peristaltic pump using flared end PVC-based tubing of 0.19 mm i.d. (Glass Expansion, Melbourne, Australia). For the segmented-flow injection mode, air was continuously aspirated and at a given time, 2.5–10 μL of either samples or standards were injected by adapting the nozzle to the flared end tubing. The solution was thus directed towards the nebulizer while avoiding the sample dispersion, as the carrier stream was simply air.
For the NexION 300X instrument used in the present study, effect of the hTISIS temperature was tested under continuous aspiration for four reference whole blood samples at different dilution factors (1:100, 1:50, 1:25, and 1:10). It was found that the most diluted samples provided a steep increase in signal with temperature. However, for the blood samples diluted at 1:10, the maximum sensitivity was achieved at 80 °C. This was likely due to the loss of nebulizer performance and ion transfer efficiency to the mass spectrometer. In fact, at 200 °C, a solid deposit was found at the nebulizer tip and the sampler cone.
The hTISIS is particularly designed to handle low liquid sample volumes. While working under segmented-flow injection mode, solid deposits observed at the interface were less severe than those in the continuous mode. Fig. 1 shows the effect of the temperature on the sensitivity under segmented-flow injection mode. Peak areas obtained with hTISIS were divided by those obtained with the conventional cyclonic spray chamber. The improvement factor in terms of sensitivity, contributed by the former device, increased with increase in temperature, however, depended on both the analyte atomic mass and the sample dilution factor. For instance, the relative area was higher for light (59Co, 63Cu, 66Zn) isotopes than that for heavy (110Cd and 208Pb) isotopes. This might suggest that the mass discrimination phenomena for light elements over heavy ones were less marked at high chamber temperatures than that at room temperature. It should be noted that for a conventional sample introduction setup, the presence of salty matrices induces space charge effects that decrease the sampling efficiency of light ions.35 Moreover, heavy ions preferentially remain in the center of the ion beam and are more efficiently transferred to the mass spectrometer through the interface. The higher peak area enhancement factor found for light elements when increasing the temperature suggests that they became efficiently focused under these operating conditions.
Note that as expected, the dilution factor played an extremely important role in terms of hTISIS performance when compared with that of the reference chamber. Fig. 2 plots the normalized area with respect to the cyclonic spray chamber for five different nuclides at the four evaluated dilution factors. In general terms, it was observed that despite the low injected sample volume (ca. 5 μL), the normalized area was the lowest for 1:10 dilutions. This effect was much more remarkable when the hTISIS was set at 200 °C than when it was set at room temperature. Obviously, the introduction of highly concentrated samples resulted detrimental in terms of hTISIS-ICP-MS signal production because of the abovementioned formation and growth of solid deposits at the interface, thus giving rise to a decrease in the ion transmission efficiency.
Detection limits in the blood reference samples (Table 2) were calculated according to the 3sb criterion (where sb is the standard deviation of a series of five consecutive blank measurements). In the present situation, sensitivity was calculated as the peak area divided by the analytical concentration of a reference material. It was observed that the hTISIS generally provided LODs lower than that provided by the cyclonic spray chamber when the diluted reference blood sample was analyzed. Table 2 also shows that the 7700x ICP-MS instrument provided LODs similar to those obtained with the NexION 300X system. Note that the injected sample volume was two times lower for the former spectrometer. Consequently, the absolute LODs would be lower while working with the 7700x system. These results could be assigned to the fact that, as abovementioned in the Experimental section, a HEN was employed for the 7700x system, whereas analyses with the NexION 300X spectrometer were carried out with a C-type pneumatic concentric nebulizer. As the HEN is a more efficient nebulizer, it was able to provide similar LODs as the C-type with a lower injected sample volume. It was observed that high LODs were obtained for 66Zn and the 7700x system. It was verified that the background signal for this isotope was actually high. This isotope suffers from polyatomic spectral interferences caused by the presence of sulfur and calcium as concomitants (i.e., 34S16O2, 32S34S, 33S2, 48Ca18O).36 Owing to the same reason, a decrease in the dilution factor (two last columns in Table 2) did not improve the limits of detection neither for 66Zn nor for 63Cu (likely interfered by 40Ar or 23Na+). Apparently, the collision cell operating conditions had to be further optimized to accurately determine this element using the abovementioned spectrometer. As this was not the goal of the present work, all the analyses were performed under the conditions summarized in Table 1.
Cyclonic segmenteda | hTISIS 200 °C segmenteda | hTISIS 200 °C segmentedb | hTISIS 200 °C segmentedc | |
---|---|---|---|---|
a 1:25 diluted sample (NexION 300X ICP-MS instrument). Injected volume: 10 μL. b 1:25 diluted sample (7700x ICP-MS instrument). Injected volume: 5 μL. c 1:10 diluted sample (7700x ICP-MS instrument). Injected volume: 5 μL. | ||||
59Co | 59 | 7 | 6 | 29 |
63Cu | 699 | 617 | 748 | 945 |
66Zn | 4357 | 3300 | 19000 | 21000 |
111Cd | 29 | 6 | 4 | 15 |
208Pb | 113 | 17 | 102 | 136 |
Interestingly, it was verified that an increase in the sample concentration led to higher LODs (see the two last columns in Table 2). In conclusion, when the NexION 300X ICP-MS was used, a 1:25 dilution was advisable to obtain the lowest LODs, whereas 1:10 was the dilution factor that was recommended while working with the 7700x system.
RSD (%) | Cyclonic | hTISIS | ||
---|---|---|---|---|
Continuous | Segmented-flow injection | Continuous | Segmented-flow injection | |
a 1:25 diluted blood sample; hTISIS temperature: 200 °C; injected sample volume in segmented-flow injection mode: 10 μL. | ||||
59Co | 4.2 | 16.2 | 6.4 | 5.0 |
63Cu | 0.5 | 5.2 | 4.2 | 5.2 |
66Zn | 0.6 | 10.1 | 2.7 | 8.0 |
111Cd | 1.5 | 9.0 | 3.5 | 8.9 |
208Pb | 1.2 | 7.5 | 4.0 | 4.9 |
To evaluate the magnitude of the matrix effects, signal obtained for the 1:25 diluted sample was divided by that measured for the 1:10 dilution sample. These were the two most concentrated samples and thus the interferences were expected to be more severe than those for the rest. The effect of the hTISIS temperature is considered in Fig. 4 for both continuous (Fig. 4a) and air-segmented (Fig. 4b) modes. In the absence of interferences, the ratios obtained should be equal to the ratio between the dilution factors, irrespective of the element and matrix considered. As calculated from the dilution factors considered and by assuming a 10% deviation as acceptable, the target value was set at 0.40 ± 0.04. To alleviate the solid deposit formation at the ICP-MS interface, an injected volume of 5 μL was used. When the sample was continuously delivered to the nebulizer at room temperature (Fig. 4a), the ratios were generally very close to 0.4. Furthermore, the increase in the chamber temperature to up to 60–80 °C also led to acceptable results. Regarding the results found for the air-segmented injection mode (Fig. 4b), matrix effects were observed (i.e., average ratios higher than 0.4) when the hTISIS was operated at room temperature. However, these effects became less severe as the hTISIS temperature increased. This observation was in agreement with the fact that the aerosol transport interferences were alleviated while working at 80 °C.
Table 4 summarizes the concentrations obtained with the cyclonic spray chamber and the hTISIS in both continuous and discrete modes. To quantify these elements, a series of plain water standards were employed. By considering the confidence intervals, it could be observed that with the cyclonic spray chamber, only copper and lead provided acceptable results in continuous mode, whereas in the segmented-flow injection mode, acceptable results were obtained for cobalt, copper, and zinc. With the hTISIS operated in continuous sample aspiration, it was noticed that the concentrations for cadmium, cobalt, and zinc were different with respect to the reference values. The appearance of matrix effects and/or the too high LODs could be the main reasons to account for the eventual deviations from the reference values. Finally, when the hTISIS was operated in a segmented-flow injection mode at 200 °C, concentrations similar to the reference ones were obtained for the five elements under study.
Element | Reference values | Cyclonic continuous | hTISIS continuous 200 °C | Cyclonic segmented | hTISIS segmented 200 °C |
---|---|---|---|---|---|
a NexION 300X; 1:25 sample dilution factor; confidence intervals were calculated as ±(t s)/(N)1/2 being N = 5. | |||||
Co | 5.8 ± 1.2 | 8.0 ± 0.7 | 1.9 ± 0.7 | 5.3 ± 0.9 | 5.5 ± 1.2 |
Cu | 1330 ± 270 | 1286 ± 8 | 1695 ± 115 | 1375 ± 74 | 1160 ± 82 |
Zn | 6500 ± 300 | 8775 ± 63 | 9085 ± 392 | 6137 ± 831 | 6509 ± 863 |
Cd | 5.8 ± 0.2 | 1.3 ± 0.4 | 8.8 ± 0.5 | 12 ± 3 | 7 ± 2 |
Pb | 310 ± 62 | 366 ± 3 | 324 ± 23 | 443 ± 19 | 376 ± 30 |
Among the different devices used for such protocols, volumetric absorptive microsampling (VAMS) is a new one that is particularly promising as it permits to collect a constant and known amount of blood in a very simple way.46,47 Furthermore, the collected sample volume is independent of the hematocrit level.48 It has been very recently demonstrated that when aiming at multi-element analysis, this methodology allows the recovery of the sample in a simple way by applying a single extraction step with 1% HNO3 without any further manipulation.49 However, the use of a conventional spray chamber and nebulizer requires a 100-fold dilution to be able to obtain the sample that is sufficient for the analysis. Using the hTISIS, in principle, a much lower value could be used with the subsequent improvement in the LODs, even for labs in which an ICP-MS/MS device (used in ref. 49) is not available.
As abovementioned, hTISIS was also coupled to a spectrometer containing an accessory for aerosol phase dilution (i.e., High Matrix Introduction, HMI, system). Compared with NexION, with 7700x spectrometer, it was possible to perform the quantification of the 1:10 diluted whole blood samples by lowering the injected sample volume down to 2.5 μL. This was also possible because in this case, a high efficiency micronebulizer (HEN) having a narrow capillary (80–100 mm id) and hence, a low dead volume was employed. Up to 11 elements were analyzed with this device. Again, aqueous standards were employed for calibration. It could be positively concluded that for all the evaluated elements, the obtained concentrations were in agreement with the reference ones (Table 5).
Isotope | Seronorm L-3. Reference values | Seronorm L-3. This work | Recovery (%) |
---|---|---|---|
9Be | 12.5 ± 2.5 | 13.6 ± 2.2 | 109 |
51V | 5.7 ± 1.1 | 5.9 ± 0.6 | 104 |
53Cr | 23.2 ± 4.7 | 31.0 ± 4.0 | 134 |
55Mn | 47.3 ± 9.5 | 40.0 ± 7.0 | 86 |
63Cu | 2470 ± 250 | 2890 ± 170 | 117 |
65Cu | 2470 ± 250 | 2930 ± 220 | 119 |
75As | 30.4 ± 7.3 | 28.9 ± 2.2 | 95 |
78Se | 272 ± 55 | 265 ± 28 | 97 |
111Cd | 12.1 ± 1.3 | 12.2 ± 2.5 | 100 |
121Sb | 22.4 ± 4.5 | 24.3 ± 1.0 | 109 |
205Tl | 34.1 ± 6.9 | 36.1 ± 1.8 | 106 |
206Pb | 447 ± 46 | 452 ± 17 | 101 |
208Pb | 447 ± 46 | 410 ± 17 | 92 |
According to the data in Table 5, it was observed that the method developed could be applied for the analysis of blood samples with a 1:10 dilution factor. Furthermore, quantitative data corresponding to 11 elements were acquired with a single 2.5 μL injection volume. This made the hTISIS especially suitable for the analysis of blood microsamples and fully compatible with the home-based collection protocols.
The developed method was finally applied for the analysis of three real blood samples (Table 6). The element quantification was performed in two different ways: (i) using aqueous standards and (ii) single standard calibration, where the reference sample acted as a matrix-matched standard. The results revealed an acceptable concordance between the concentrations obtained with both methods for 51V, 75As, 78Se, 121Sb, and 205Tl. Only for copper, the results significantly differed depending on the calibration method employed. This could be due to the differences between the matrices of a real blood sample and the reference one. In a different experiment, real samples were spiked and the recoveries were obtained by applying external calibration with plain water standards. According to the results, this parameter ranged from 96% (55Mn) to 109% (121Sb), thus giving an additional proof for the accuracy of the employed procedure.
Isotope | Real sample 1 | Real sample 2 | Real sample 3 | |||
---|---|---|---|---|---|---|
Calibration standards | Reference material | Aqueous | Reference material | Aqueous | Reference material | Aqueous |
a Injected volume: 2.5 μL. | ||||||
9Be | <2 | <2 | <2 | <2 | <2 | <2 |
51V | 1.5 ± 0.4 | 2.1 ± 0.5 | 1.5 ± 0.5 | 2.4 ± 0.8 | 1.6 ± 0.6 | 2.5 ± 0.8 |
53Cr | <5 | <5 | <5 | <5 | <5 | <5 |
55Mn | <31 | <31 | <31 | <31 | <31 | <31 |
63Cu | 1522 ± 93 | 2029 ± 116 | 1242 ± 54 | 1863 ± 150 | 713 ± 80 | 865 ± 134 |
65Cu | 1552 ± 95 | 2028 ± 112 | 1294 ± 94 | 1905 ± 198 | 714 ± 83 | 973 ± 163 |
75As | 2.3 ± 0.3 | 3.1 ± 0.3 | 8.4 ± 0.5 | 12.2 ± 1.2 | 3.1 ± 0.7 | 4.3 ± 0.8 |
78Se | 145 ± 20 | 195 ± 22 | 124 ± 18 | 188 ± 29 | 119 ± 22 | 155 ± 23 |
111Cd | <2 | <2 | <2 | <2 | <2 | <2 |
121Sb | 1.3 ± 0.3 | 1.7 ± 0.4 | 2.2 ± 0.3 | 3.1 ± 0.4 | 1.7 ± 0.5 | 2.1 ± 0.5 |
205Tl | 18.0 ± 3.8 | 23.5 ± 4.3 | 23.5 ± 3.5 | 33.1 ± 5.8 | 27.3 ± 4.2 | 35.2 ± 4.1 |
206Pb | <36 | <36 | <36 | <36 | <36 | <36 |
208Pb | <26 | <26 | <26 | <26 | <26 | <26 |
Finally, by comparing the element concentrations determined in the real blood samples with those of the reference samples (i.e., containing pathological levels of elements), it was concluded that all three individuals had metal levels below the pathological levels.
As a final step, potential contaminations from the VAMS tip were evaluated. In this case, the hydrophilic polyolefin tips were washed with 100 μL of the nitric acid solution and the element concentrations in the resulting extracts were determined by ICP-MS. The corresponding results are shown in Table 7. As previously stated in ref. 49, elements, such as vanadium, chromium, and aluminum, were present in the solid sorbent. Based on the data obtained in the present study it can be stated that although significant concentrations of seven elements were detected in the obtained solutions, it was possible to quantify all of them in the samples with an acceptable accuracy (see Table 5).
Isotope | Concentration |
---|---|
a Concentration after taking into account the 10-fold dilution. Confidence intervals were calculated as ±(t s)/(N)1/2 being N = 5. | |
51V | 5.9 ± 0.7 |
53Cr | 24.4 ± 1.6 |
55Mn | 89 ± 18 |
63Cu | 714 ± 349 |
65Cu | 759 ± 359 |
111Cd | 3.4 |
205Tl | 7.4 |
206Pb | 100 ± 12 |
208Pb | 98 ± 9 |