An Adaptive Weighted Polynomial Baseline Correction Method for Electrochemical Aptamer-based Sensor

Abstract

A background signal, or baseline, is typical a low frequency signal that is composited with the target signal and commonly occurs in electrochemical biosensing data. Square wave voltammetry (SWV) has been widely used to acquire data for electrochemical aptamer-based (EAB) biosensors. However, one challenge with SWV is that the true baseline is unable to be assessed, but can only be estimated. The background signal of SWV usually contains various features, such as levels, trends, and shapes. These features are usually uninformative and, if unaccounted for, they may confuse the results of the analysis. Consequently, standardizing the signal by correcting the baseline is an essential step processing electrochemical sensing results. In this research, we present an adaptive polynomial baseline correction method for the baseline correction of SWV data from real E-AB biosensors. This method can automatically identify the uninformative regions in the signal and provide a robust mathematical equation to estimate the baseline. Employing real world sensing data, We compared our method with other published methods and showed that our method performs more reliably within acceptable errors. We also used the baseline-corrected E-AB biosensing data to develop a statistical model for predicting the concentration of cocaine and THC in saliva samples and developed a friendly user interface that enables front-end users to analyse the data without code interaction. This work shows potential to facilitate data automation to detect specific analytes for point-of-care (POC) applications.

Supplementary files

Transparent peer review

To support increased transparency, we offer authors the option to publish the peer review history alongside their article.

View this article’s peer review history

Article information

Article type
Paper
Submitted
22 Dec 2025
Accepted
27 Feb 2026
First published
12 Mar 2026

Analyst, 2026, Accepted Manuscript

An Adaptive Weighted Polynomial Baseline Correction Method for Electrochemical Aptamer-based Sensor

Z. Wen, Y. Liu, O. Raymond, E. J. Itumoh, J. Stacey and J. Curran, Analyst, 2026, Accepted Manuscript , DOI: 10.1039/D5AN01347J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements