Abstract

Different nebulizers (cross-flow, ultrasonic and two microconcentric nebulizers) were used for sample introduction of radioactive solutions into a quadrupole-based inductively coupled plasma mass spectrometer (ICP-QMS). The best sensitivity (from 420 to 850 MHz, which is about one order of magnitude higher in comparison with the cross-flow nebulizer) for long-lived radionuclides ( 226 Ra, 230 Th, 237 Np, 238 U and 241 Am) was observed using the ultrasonic nebulizer. However, using the ultrasonic nebulizer, a significantly higher sample size (26-fold) in comparison with the micronebulizers is required. Sample introduction by micronebulization with a small sample size in the low picogram range is the method of choice for the determination of long-lived radionuclides. The precision of determination of a 10 ng l –1 concentration was in the low-% range (and sub-% range) for all measurements using different nebulizer types. The detection limits for the determination of long-lived radionuclides in aqueous solutions applying the different nebulizers were 0.01-0.6 ng l –1 . The flow injection analysis approach was optimized for isotope dilution analysis of 232 Th (using 20 µl of 5 µg l –1 230 Th) by ICP-QMS. The isotopic abundance ratios of 230 Th- 232 Th isotope mixtures ( 230 Th/ 232 Th=0.01, 0.001 and 0.0001) were determined using a microconcentric nebulizer and 1 µg l –1 Th solutions with a relative external standard deviation of long-term stability measurements (over 20 h) of 0.17, 0.62 and 2.66%, respectively.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1999,14, 933-937

Determination of long-lived radionuclides by inductively coupled plasma quadrupole mass spectrometry using different nebulizers

J. Sabine Becker, R. S. Soma, K. L. Sutton, J. A. Caruso and H. Dietze, J. Anal. At. Spectrom., 1999, 14, 933 DOI: 10.1039/A809252D

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements