Use of vitreous carbon as a working electrode for a coulometric redox titration

Abstract

Vitreous or glassy carbon has been examined as to its suitability for use as a working electrode in the coulometric generation of bromine. It has been found to compare well with platinum for this purpose, although an appreciable increase in the electrode potential is required to maintain a given current density. The theoretical current efficiency for the generation of bromine has been found to be better than 99·9 per cent. The theoretical value has been confirmed by the titration of 100 µequiv of arsenic(III) with an accuracy and precision each of 0·1 µequiv.

Article information

Article type
Paper

Analyst, 1972,97, 923-926

Use of vitreous carbon as a working electrode for a coulometric redox titration

V. J. Jennings, A. Dodson and A. M. Atkinson, Analyst, 1972, 97, 923 DOI: 10.1039/AN9729700923

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.