A throughput model explaining non-linearity in discrete ion counters used in mass spectrometry

Abstract

A mathematical model is presented to calculate the expected throughput rate in a discrete ion counter with imposed non-extending dead time. The count loss mechanism consists of a combined effect of the imposed dead time, interfered with pulse pileup owing to the finite time resolution of the electronic pulses generated in the ion counter. This model may be applicable to mass spectrometers making use of discrete ion counters, thus providing a means to reproduce counting-rate dependency of observed atomic abundance ratios.

Graphical abstract: A throughput model explaining non-linearity in discrete ion counters used in mass spectrometry

Article information

Article type
Paper
Submitted
05 Mar 2025
Accepted
16 May 2025
First published
25 Jun 2025

J. Anal. At. Spectrom., 2025, Advance Article

A throughput model explaining non-linearity in discrete ion counters used in mass spectrometry

S. Pommé and S. F. Boulyga, J. Anal. At. Spectrom., 2025, Advance Article , DOI: 10.1039/D5JA00085H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements