Issue 4, 2025

Non-suppressed ion chromatography-tandem electrospray mass spectrometry using a short column for simultaneous analysis of dichloroacetic acid, trichloroacetic acid, and bromate: aqueous ammonia as the eluent additive

Abstract

Dichloroacetic acid (DCAA), trichloroacetic acid (TCAA), and bromate (BrO3) are disinfection byproducts (DBPs) formed during drinking water treatment and pose health risks. Rapid and reliable detection of these DBPs is essential for ensuring water safety. Non-suppressed ion chromatography (IC)-electrospray ionization mass spectrometry (IC-ESI-MS/MS) offers a promising approach for simultaneous analysis of organic haloacetic acids (HAAs) and inorganic oxyhalides, but previous methods using toxic methylamine can pose health risks. Herein, a new rapid non-suppressed IC-ESI-MS/MS method using a short AG18 guard column (13 μm, 2 mm × 50 mm) with aqueous ammonia as the alkaline modifier was developed for the simultaneous quantification of DCAA, TCAA, and BrO3. This method allows direct sample injection without pretreatment and achieves sub-μg L−1 detection limits within 5 minutes. Aqueous ammonia (0.4 M) and acetonitrile were used as mobile phases in a gradient elution mode, providing good linearity (R2 >0.995), low limits of quantification (0.48–1.13 μg L−1), and high recoveries (76.3–107.3%). Intra-day (n = 9) and inter-day (3 days, n = 9) precision were validated with relative standard deviations of 5.8–8.9% and 6.4–7.7%, respectively. The method also demonstrated limited matrix interference from common water parameters such as Cl, SO42−, HCO3/CO32−, and natural organic matter (NOM), making it suitable for real water analysis. The applicability of this method was confirmed by monitoring DBP formation during the chlorination of NOM, showing that two-step chlorine dosing strategies reduce DBP formation compared to single-stage dosing. This newly established method provides a robust and efficient tool for DBP monitoring in drinking water treatment processes.

Graphical abstract: Non-suppressed ion chromatography-tandem electrospray mass spectrometry using a short column for simultaneous analysis of dichloroacetic acid, trichloroacetic acid, and bromate: aqueous ammonia as the eluent additive

Supplementary files

Article information

Article type
Paper
Submitted
05 Nov 2024
Accepted
17 Dec 2024
First published
18 Dec 2024

Anal. Methods, 2025,17, 841-849

Non-suppressed ion chromatography-tandem electrospray mass spectrometry using a short column for simultaneous analysis of dichloroacetic acid, trichloroacetic acid, and bromate: aqueous ammonia as the eluent additive

M. Peng, J. Chen, S. Cheng, H. Zhang, L. Pu, E. Du, Q. Cheng, Y. Zuo and A. Li, Anal. Methods, 2025, 17, 841 DOI: 10.1039/D4AY02007C

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements