Abstract

Diabetes is a common chronic metabolic disease with a wide range of clinical symptoms and consequences and one of the main causes of death. For the management of diabetes, painless and continuous interstitial fluid (ISF) glucose monitoring is ideal. Here, we demonstrate continuous diabetes monitoring using an integrated microneedle (MN) biosensor with an emergency alert system. MNs are a novel technique in the field of biomedical engineering because of their ability to analyze bioinformation with minimal invasion. In this work we developed a poly(methyl methacrylate) (PMMA) based MN glucose sensor. The device was produced by the 3D printing technique, microfabrication, electrodeposition, and enzyme immobilization step. The in vitro test for the glucose MN sensor showed a linear range from 1.5 to 14 mM with a sensitivity of 1.51 μA mM−1, limit of detection (LOD) of 0.35 mM and good selectivity. Highly repeatable sensing is observed with good reproducibility. The interference-free detection of glucose in the presence of physiologically relevant concentrations of ascorbic acid, uric acid, and mannose is demonstrated, along with the operational stability of the array. After resolving the biofouling consequences linked to on-body sensing, this MN platform would be appealing for minimally invasive electrochemical glucose monitoring. An alert is sent to confidants via email or SMS when the values are abnormal. The application is also able to display the recorded values in the form of a graph to help determine the state of health of the user over a period of time. It can be concluded that continuous monitoring and an emergency alert system are important for keeping an eye on diabetic patients and can send alert in case of an abnormal situation of the patient.

Graphical abstract: Early detection of hypo/hyperglycemia using a microneedle electrode array-based biosensor for glucose ultrasensitive monitoring in interstitial fluid

Article information

Article type
Paper
Submitted
26 Apr 2024
Accepted
05 Jul 2024
First published
10 Jul 2024

Lab Chip, 2024,24, 3958-3972

Early detection of hypo/hyperglycemia using a microneedle electrode array-based biosensor for glucose ultrasensitive monitoring in interstitial fluid

S. H. Tawakey, M. Mansour, A. Soltan and A. I. Salim, Lab Chip, 2024, 24, 3958 DOI: 10.1039/D4LC00365A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements