A low-cost voltammetric sensor based on multi-walled carbon nanotubes for highly sensitive and accurate determination of nanomolar levels of the anticancer drug Ribociclib in bulk and biological fluids†
Abstract
In this study, we present the development and comprehensive characterization of the first electrochemical sensor utilizing multi-walled carbon nanotubes (MWCNTs) for the sensitive and precise detection of Ribociclib (RIBO), an important anticancer drug. The sensor underwent systematic optimization, focusing on critical parameters such as pH, deposition potential, and cumulative time to enhance its electrocatalytic activity and expand the linear range for RIBO determination. The MWCNTs/GCE sensor exhibited excellent reproducibility and repeatability, ensuring reliable and consistent results. The applicability and feasibility of the sensor for real sample analysis were extensively evaluated by analyzing human serum, urine, and tablet samples using the standard addition method. The obtained percent recovery values demonstrated the sensor's exceptional accuracy and precision. Furthermore, interference studies revealed the sensor's remarkable selectivity, with minimal impact from common interfering substances. The developed sensor displayed a wide linear range of 0.01 μM to 5.0 μM, with a limit of detection (LOD) and limit of quantification (LOQ) calculated to be 0.69 nM and 2.31 nM, respectively, affirming its high sensitivity for detecting low RIBO concentrations. The MWCNTs/GCE sensor demonstrates substantial promise for diverse practical applications with its simplicity, cost-effectiveness, and excellent analytical performance.