Comment on: “Slope ratio calibration for analysis of plant leaves by laser-induced breakdown spectroscopy” by Lidiane C. Nunes, Fabio R. P. Rocha and Francisco J. Krug, JAAS, 2019, 34, 2314

Author affiliations

Abstract

In the paper “Slope ratio calibration for analysis of plant leaves by laser-induced breakdown spectroscopy”, recently published on JAAS, the authors present an allegedly new calibration method for LIBS analysis, called Slope Ratio Calibration (SRC). In this short comment we will demonstrate that this method is completely equivalent to Single-Point Calibration (SPC), when a fair comparison between the two methods is done.

Back to tab navigation

Article information


Submitted
11 Nov 2019
Accepted
15 May 2020
First published
22 May 2020

J. Anal. At. Spectrom., 2020, Advance Article
Article type
Comment

Comment on: “Slope ratio calibration for analysis of plant leaves by laser-induced breakdown spectroscopy” by Lidiane C. Nunes, Fabio R. P. Rocha and Francisco J. Krug, JAAS, 2019, 34, 2314

V. Palleschi, J. Anal. At. Spectrom., 2020, Advance Article , DOI: 10.1039/C9JA00381A