The GR/L-cys/Bi/SPE (reduced graphene oxide/L-cysteine/Bi/screen-printing electrode) was prepared to simultaneously determinate Cd2+ and Pb2+ ions by using SWASV (square wave anodic stripping voltammetry). The experimental conditions of the Cd2+ and Pb2+ determination including the deposition potential, the deposition time, pH and Bi3+ concentration were optimized. Well-defined peaks of Cd2+ and Pb2+ ions could be achieved. Under the optimized conditions, the calibration curves were obtained with linear ranges from 10 µg/L to 50 µg/L for Cd2+, and from 10 µg/L to 50 µg/L for Pb2+, with detection limit of 0.98 and 0.56 µg/L, respectively. Their fitting equations are I(µA)=0.13C(µg/L) +0.60 (R2=0.9976) for Cd2+ ion, and I(µA)=0.10C(µg/L)+0.51 (R2=0.9896) for Pb2+ ion, respectively. The GR/L-cys/Bi/SPE exhibits many merits with high selectivity, low detection limit, low fabrication costs and wide linear ranges. The GR/L-cys/Bi/SPE has great potential for the determination of Cd2+ and Pb2+ in real water samples.

Back to tab navigation

Supplementary files

Publication details

The article was received on 27 Jul 2018, accepted on 13 Sep 2018 and first published on 13 Sep 2018


Article type: Paper
DOI: 10.1039/C8AY01676C
Citation: Anal. Methods, 2018, Accepted Manuscript
  •   Request permissions

    Simultaneous determination of trace Cd2+ and Pb2+ using GR/L-cysteine/Bi modified screen-printing electrodes

    C. Li, X. Zhao and X. Han, Anal. Methods, 2018, Accepted Manuscript , DOI: 10.1039/C8AY01676C