Method for isotope ratio drift correction by internal amplifier signal synchronization in MC-ICPMS transient signals

Author affiliations

Abstract

The measurement of isotope ratios by Multi Collection Inductively Coupled Plasma Mass Spectrometry (MC-ICPMS) in transient signals often shows a drift of the isotope signal ratio during signal acquisition. This “isotopic drift” is generally related to the small and distinct time lags between the responses of the amplifiers involved in the Faraday detector configuration. In this work, a method of synchronization of transient isotope signals for a duration of a few tens of seconds is proposed in order to: (1) quantify the time lags between the amplifiers using the ratios of the raw isotope signals and (2) correct the isotope ratio drifts. The method was successfully tested on lead isotope ratio measurements obtained from two different multi-collector mass spectrometers and setups (flow injection with direct injection and gas chromatography). This approach offered a precise determination of the time lag between the different amplifier systems and an effective correction of the isotope ratio drift. The performances of the methods traditionally used for isotope ratio calculation of transient signals were also compared before and after isotope ratio drift correction.

Graphical abstract: Method for isotope ratio drift correction by internal amplifier signal synchronization in MC-ICPMS transient signals

Back to tab navigation

Supplementary files

Publication details

The article was received on 08 Apr 2014, accepted on 23 May 2014 and first published on 27 May 2014


Article type: Paper
DOI: 10.1039/C4JA00118D
Author version
available:
Download author version (PDF)
J. Anal. At. Spectrom., 2014,29, 1607-1617

  •   Request permissions

    Method for isotope ratio drift correction by internal amplifier signal synchronization in MC-ICPMS transient signals

    A. Gourgiotis, S. Bérail, P. Louvat, H. Isnard, J. Moureau, A. Nonell, G. Manhès, J. Birck, J. Gaillardet, C. Pécheyran, F. Chartier and O. F. X. Donard, J. Anal. At. Spectrom., 2014, 29, 1607
    DOI: 10.1039/C4JA00118D

Search articles by author

Spotlight

Advertisements