Accurate quantification of mercury in river water by isotope dilution MC-ICP-SFMS and ICP-QMS detection after cold vapour generation

Abstract

Precise and accurate determination of Hg in surface water is challenging due to volatilization and adsorption effects of Hg and low ng L−1 concentrations. The aim of our study was to examine the parameters, which affect the uncertainty of Hg concentrations determined by CV-ICP-MS, employing isotope dilution analysis for accurate quantification. We used two different types of mass spectrometers (MC-ICP-SFMS and ICP-QMS) and further evaluated two widely used isotope dilution approaches to achieve accurate data with low uncertainties for ultra trace level Hg determination in surface water. Obtained LODs were significantly below 1 ng L−1. Validated concentrations of Hg obtained in the new ground water reference material ERM-CA615 were in good agreement with the indicative value of 0.037 μg L−1 (95% confidence interval, k = 2). A reduction in the propagated uncertainty was evident with CV-MC-ICP-SFMS and also by undertaking a mass bias drift correction during the measurement session with certified isotope ratio values as currently accepted by the IUPAC. We could demonstrate that accurate Hg quantification can be achieved with either multicollector or quadrupole detection within a level of uncertainty of 27%/52% (uc, k = 2) down to 1 ng L−1 and better than 2%/5% (uc, k = 2) from 5 ng L−1. Taking the advantages of matrix separation/pre-concentration obtained from CV sample introduction, the method allows quantifying Hg in surface water within the data quality requirements regarding the Water Framework Directive.

Graphical abstract: Accurate quantification of mercury in river water by isotope dilution MC-ICP-SFMS and ICP-QMS detection after cold vapour generation

Article information

Article type
Technical Note
Submitted
01 Mar 2012
Accepted
03 Sep 2012
First published
25 Sep 2012

J. Anal. At. Spectrom., 2012,27, 1983-1991

Accurate quantification of mercury in river water by isotope dilution MC-ICP-SFMS and ICP-QMS detection after cold vapour generation

L. Fischer, M. Brunner, T. Prohaska and S. Hann, J. Anal. At. Spectrom., 2012, 27, 1983 DOI: 10.1039/C2JA30076A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements