Jump to main content
Jump to site search

Issue 6, 2011
Previous Article Next Article

Radial sample preconcentration

Author affiliations

Abstract

In this paper we present a radial sample preconcentration strategy enabled by axisymmetric concentration polarization in a microfluidic chamber on a uniform nanoporous film. Sample analytes are focused into the centre, creating a concentrated plug that is injected vertically into the microfluidic analysis layer. No balancing pressure driven flows or tangential fields are required, and the process has essentially zero footprint on the analysis layer. An electrokinetic loading scheme enables repeat loading/concentration cycles, and a finned radial chamber geometry dampens instabilities and accommodates larger volumes. Modelling results indicate over 1800-fold concentration increases are possible in 10 s, for high mobility buffers and high applied field strength. At moderate field strength and buffer mobility, experiments demonstrate a 168-fold increase in concentration of FITC–BSA protein in 36 s.

Graphical abstract: Radial sample preconcentration

Back to tab navigation

Supplementary files

Publication details

The article was received on 24 Aug 2010, accepted on 06 Jan 2011 and first published on 11 Feb 2011


Article type: Paper
DOI: 10.1039/C0LC00326C
Citation: Lab Chip, 2011,11, 1102-1109
  •   Request permissions

    Radial sample preconcentration

    B. Scarff, C. Escobedo and D. Sinton, Lab Chip, 2011, 11, 1102
    DOI: 10.1039/C0LC00326C