Jump to main content
Jump to site search

Issue 11, 2010
Previous Article Next Article

On-chip three dimensional microcoils for MRI at the microscale

Author affiliations

Abstract

We present for the first time a fully MEMS-integrated technology to manufacture 3D geometrically perfect solenoidal microcoils for microscale MRI applications. We report 25 µm isotropic resolution MR images of a copper sulfate aqueous phantom. These images are acquired using microcoils with 5 windings of insulated 25 µm diameter Au wire and with quality factors as high as 46 at the operating frequency (400 MHz).

Graphical abstract: On-chip three dimensional microcoils for MRI at the microscale

Back to tab navigation

Article information


Submitted
13 Jan 2010
Accepted
26 Mar 2010
First published
21 Apr 2010

Lab Chip, 2010,10, 1387-1390
Article type
Communication

On-chip three dimensional microcoils for MRI at the microscale

V. Badilita, K. Kratt, N. Baxan, M. Mohmmadzadeh, T. Burger, H. Weber, D. v. Elverfeldt, J. Hennig, J. G. Korvink and U. Wallrabe, Lab Chip, 2010, 10, 1387