High-precision analysis of Sr/Ca and Mg/Ca ratios in corals by laser ablationinductively coupled plasma optical emission spectrometry

Abstract

A method has been developed to determine high-precision Sr/Ca and Mg/Ca ratios in corals by laser ablation inductively coupled plasma optical emission spectrometry (LA-ICP-OES) using aqueous solution standard calibration. Simultaneous determination of the signals of the entire analytical wavelengths by ICP-OES and the high performance of the new type of LA system (Resonetics 193 nm ArF excimer laser-ablation system, RESOlution M-50) improve the precision for elemental ratios. Repeated measurements on a coral base synthesized working standard, BH-7, provide precisions of about 0.4% and 0.8% for the Sr/Ca and Mg/Ca ratios, respectively, which are better than the formerly reported precision by the LA-ICP-MS method, about 1%. Such precision is comparable to those obtained by the solution nebulization-ICP-OES (SN-ICP-OES) method, and is adequate for paleoclimate reconstruction. In addition, the LA-ICP-OES can provide results with much higher spatial/time resolution. Comparisons between the LA and SN methods were handled by measuring along the same track of a coral. The Sr/Ca results by these two methods agree quite well with each other. The LA-ICP-OES method is very promising for the analysis of element/Ca ratios in coral and other carbonates used in paleoclimate studies such as stalagmite. Systematic discrepancy, however, was observed in the Mg/Ca ratios, likely due to the existing state of magnesium in the coral skeleton.

Graphical abstract: High-precision analysis of Sr/Ca and Mg/Ca ratios in corals by laser ablation inductively coupled plasma optical emission spectrometry

Article information

Article type
Technical Note
Submitted
19 Jun 2009
Accepted
17 Sep 2009
First published
14 Oct 2009

J. Anal. At. Spectrom., 2010,25, 84-87

High-precision analysis of Sr/Ca and Mg/Ca ratios in corals by laser ablation inductively coupled plasma optical emission spectrometry

W. Deng, Y. Liu, G. Wei, X. Li, X. Tu, L. Xie, H. Zhang and W. Sun, J. Anal. At. Spectrom., 2010, 25, 84 DOI: 10.1039/B912112A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements