UV-femtosecond and nanosecond laser ablation-ICP-MS: internal and external repeatability

Author affiliations

Abstract

The internal (precision within an ablation spot) and external repeatability (precision between ablation spots on the sample) were evaluated during repetitive 266 nm femtosecond and nanosecond laser ablation-inductively coupled plasma spectrometry. Femtosecond laser ablation sampling into the ICP-MS improved the precision by reducing systematic errors related to the particle size distribution and resultant spikes on the signal intensity. The data are reported in terms of measured concentrations for several elements in silica glass reference materials NIST 610-612.

Graphical abstract: UV-femtosecond and nanosecond laser ablation-ICP-MS: internal and external repeatability

Back to tab navigation

Publication details

The article was received on 07 Mar 2006, accepted on 29 Jun 2006 and first published on 11 Jul 2006


Article type: Technical Note
DOI: 10.1039/B603492F
J. Anal. At. Spectrom., 2006,21, 778-784

  •   Request permissions

    UV-femtosecond and nanosecond laser ablation-ICP-MS: internal and external repeatability

    J. González, S. H. Dundas, C. Yi Liu, X. Mao and R. E. Russo, J. Anal. At. Spectrom., 2006, 21, 778
    DOI: 10.1039/B603492F