Issue 6, 2003

Analysis of individual platelet-derived microparticles, comparing flow cytometry and capillary electrophoresis with laser-induced fluorescence detection

Abstract

Platelet-derived microparticles (PMPs) formed by vesiculation during platelet activation seem to play a role in blood coagulation and in pathological disease states. Flow cytometry is currently the gold standard to characterize platelets and PMPs. Using this technique we distinguished between platelets and PMPs based on size and the presence of phosphatidyl serine (PS); PMPs were arbitrarily defined to be smaller than one micrometer and capable of forming a stable complex with fluorescently-labeled Annexin V, a protein that forms a calcium-dependent complex with PS. Further confirmation of PMP and platelet identity was done by use of fluorescently-labeled antibodies against CD41a, a glycoprotein found on the surface of both platelets and PMPs. In this report we also introduce the use of capillary electrophoresis with post-column laser-induced fluorescence detection (CE-LIF) for the analysis of fluorescently labeled platelets and PMPs. While both flow cytometry and CE-LIF can measure individual fluorescent events, only CE-LIF allowed us to calculate individual electrophoretic mobilities of activated platelets and PMPs that were then represented as distributions. A comparison between distributions suggests that PMPs have less negative mobilities. The fact that activated platelet preparations include PMPs partially obscure the interpretation of the data. While PMP and platelet number ml−1 determined by flow cytometry is lower than the same parameter determined by CE-LIF, signal-to-noise ratio was 20 fold better for flow cytometry than for CE-LIF. This is the first time that a direct comparison between these two techniques is reported.

Article information

Article type
Paper
Submitted
24 Jan 2003
Accepted
09 May 2003
First published
27 May 2003

Analyst, 2003,128, 581-588

Analysis of individual platelet-derived microparticles, comparing flow cytometry and capillary electrophoresis with laser-induced fluorescence detection

G. Xiong, O. Aras, A. Shet, N. S. Key and E. A. Arriaga, Analyst, 2003, 128, 581 DOI: 10.1039/B301035J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements