Abstract

A model allowing simultaneous determination of the detector dead time and the mass discrimination factor in inductively coupled plasma mass spectrometry (ICP-MS), as well as the corresponding uncertainties, is presented and compared for three representative isotope systems, namely magnesium, indium and thallium. The advantages of using the model presented are firstly that both the detector dead time and the mass discrimination factor can be obtained simultaneously and secondly that the sampling time can be spent entirely on the isotopes of interest.

Article information

Article type
Inter-laboratory Note
Submitted
10 Feb 2000
First published
25 Apr 2000

J. Anal. At. Spectrom., 2000,15, 557-560

A model for calculating dead time and mass discrimination correction factors from inductively coupled plasma mass spectrometry calibration curves

P. K. Appelblad and D. C. Baxter, J. Anal. At. Spectrom., 2000, 15, 557 DOI: 10.1039/B001152P

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.