Issue 9, 1996

Simple convenient method for measuring injection volumes in flow injection and sequential injection analysis

Abstract

A simple, convenient method for determining injection volumes in flow injection and sequential injection analysis is presented. It is based on comparing the dilution of an injected dye with a standard calibration curve. The proposed colour method gave the volume of the whole injection device more accurately than methods where the inner dimensions of the injection device are not precisely known. The colour method is within the 95% confidence level with an RSD of 0.8%.

Article information

Article type
Paper

Anal. Commun., 1996,33, 339-341

Simple convenient method for measuring injection volumes in flow injection and sequential injection analysis

J. F. van Staden and D. Malan, Anal. Commun., 1996, 33, 339 DOI: 10.1039/AC9963300339

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.