Inter-laboratory note. Laser ablation inductively coupled plasma mass spectrometric transient signal data acquisition and analyte concentration calculation

Abstract

Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) produces complex, time-dependent signals. These require significantly different treatment both during data acquisition and reduction from the more steady-state signals produced by solution sample introduction. This paper discusses, in detail, data acquisition and reduction considerations in LA-ICP-MS analysis. Optimum data acquisition parameters are suggested. Equations are derived for the calculation of sample concentrations and LOD when time-resolved data acquisition is employed, sensitivity calibration is obtained from reference materials with known analyte concentrations and naturally occurring internal standards are used to correct for the multiplicative correction factors of drift, matrix effects and the amount of material ablated and transported to the ICP.

Back to tab navigation

Article type: Paper
DOI: 10.1039/JA9961100899
J. Anal. At. Spectrom., 1996,11, 899-904

  •   Request permissions

    Inter-laboratory note. Laser ablation inductively coupled plasma mass spectrometric transient signal data acquisition and analyte concentration calculation

    H. P. Longerich, S. E. Jackson and D. Günther, J. Anal. At. Spectrom., 1996, 11, 899
    DOI: 10.1039/JA9961100899